- ICH GCP
- Registre américain des essais cliniques
- Essai clinique NCT06190808
Surveillance continue de la glycémie dans l'unité de soins intensifs
Surveillance continue de la glycémie en unité de soins intensifs : validation et mise en œuvre
The aim of the proposed clinical trial is to study the use of the Dexcom G7 continuous glucose monitor (CGM) within the intensive care unit (ICU) environment. The overall goal is to validate the accuracy and practicality of the Dexcom G7 when used to assess glycemia in critically ill patients.
Aperçu de l'étude
Statut
Les conditions
Intervention / Traitement
Description détaillée
The specific goals are 1) to work with intensive care unit (ICU) staff, in a respiratory ICU setting, to determine how best to place and maintain Dexcom G7 continuous glucose monitors (CGMs) on patients and how to place recording devices and chargers in that environment, 2) download CGM readings to the Dexcom Dashboard in an analyzable format, and 3) to compare simultaneous CGM readings with glucose measurements obtained in the course of routine care as recorded in the electronic medical record. The primary analytical goal is the correlation of CGM glucose measurements with those obtained in the course of routine care. A secondary goal is to determine the frequency of device malfunctions that occur either as a result of device failure (e.g. it falls off) or adverse events (e.g. removal due to skin irritation).
Study subjects will be patients admitted to the 6-ICU UMass Memorial Medical Center (UMMMC) University Campus medical intensive care unit for any reason who have known diabetes or who are hyperglycemic (blood glucose concentration ≥200 mg/dL) with diabetes that was previously undiagnosed or unrecognized. All enrolled volunteers will have a Dexcom G7 CGM place on the skin according to manufacturer's specifications. The devices will remain in place for 10 days or until discharge from the ICU
Type d'étude
Inscription (Réel)
Phase
- N'est pas applicable
Contacts et emplacements
Lieux d'étude
-
-
Massachusetts
-
Worcester, Massachusetts, États-Unis, 01605
- UMass Memorial Medical Center
-
-
Critères de participation
Critère d'éligibilité
Âges éligibles pour étudier
Accepte les volontaires sains
Sexes éligibles pour l'étude
La description
Inclusion Criteria:
Campus medical intensive care unit for any reason who have known diabetes or who are hyperglycemic (blood glucose concentration ≥200 mg/dL) with diabetes that was previously undiagnosed or unrecognized.
- Persons with either type 1 or type 2 diabetes.
- Persons of either sex.
- Persons of any age over 18.
- Persons able to give informed consent or cognitively impaired adults whose legally authorized representative (LAR) can give informed consent.
- English or Spanish speaking adults.
Exclusion Criteria:
- Persons unable to provide informed consent.
- Cognitively impaired persons whose LAR declines to provide informed consent.
- Patients receiving hydroxyurea, which interferes with sensor function.
- Patients receiving high dose acetaminophen (> 4 gm/day), which interferes with sensor function.
- Anasarca affecting preferred areas for device placement (arm, upper buttocks, or thigh), which may interfere with sensor accuracy.
- Severe cachexia with absence of subcutaneous fat at preferred sensor placement sites, which might interfere with sensor accuracy.
- Persons who are pregnant.
- Prisoners.
Plan d'étude
Comment l'étude est-elle conçue ?
Détails de conception
- Objectif principal: Autre
- Répartition: N / A
- Modèle interventionnel: Affectation à un seul groupe
- Masquage: Aucun (étiquette ouverte)
Armes et Interventions
Groupe de participants / Bras |
Intervention / Traitement |
|---|---|
|
Expérimental: Single Arm
All members of the arm will have a CGM place and glucose concentration downloaded. |
Continuous glucose monitor |
Que mesure l'étude ?
Principaux critères de jugement
Mesure des résultats |
Description de la mesure |
Délai |
|---|---|---|
|
Mean Absolute Relative Difference (MARD).
Délai: 6 months
|
Concordance of glucose concentration results obtained by standard test methods with those obtained nearly simultaneously by a continuous glucose monitoring device was assessed by Mean Absolute Relative Difference (MARD). The MARD is the average percentage difference between the continuous glucose monitor readings measured over 6 months and glucose concentration results obtained by near simultaneous (+/- 5 minutes) standard test methods was used to gauge the accuracy of the CGM in the ICU setting. Typical MARD values reported for CGMs in the outpatient setting average approximately 8%. Typical MARD values reported for CGMs other than the one used in this study n the ICU setting average approximately 12%. |
6 months
|
|
Clarke Error Grid
Délai: 6 months
|
Concordance of glucose concentration results obtained by standard test methods with those obtained nearly simultaneously by a continuous glucose monitoring device was assessed by Clarke Error Grid. The Clarke Error Grid plots of reference glucose values obtained by point-of-care fingerstick or laboratory determination (x-axis) against CGM values (y-axis). The grid is divided into 5 zones (A-E) that indicate clinical implications of CGM glucose measurements when compared to a "gold standard" Region A: Clinically accurate. Region B: Benign measurement error. Region C: Overcorrection error region CGM values suggest unnecessary treatment. Region D: Serious hypo- or hyperglycemia detection error region. Region E: Erroneous treatment region. CGM fails to detect potentially dangerous hypo- or hyperglycemia Ideally, all values are in Regions A and B. We report the percentage of correlated values that fall in each region. |
6 months
|
Mesures de résultats secondaires
Mesure des résultats |
Description de la mesure |
Délai |
|---|---|---|
|
Unplanned Event Frequency
Délai: 6 months
|
Number of occurrences per patient of device failure or or unforeseen device removal |
6 months
|
Collaborateurs et enquêteurs
Parrainer
Les enquêteurs
- Chercheur principal: John P Mordes, MD, UMass Chan Medical School
Dates d'enregistrement des études
Dates principales de l'étude
Début de l'étude (Réel)
Achèvement primaire (Réel)
Achèvement de l'étude (Réel)
Dates d'inscription aux études
Première soumission
Première soumission répondant aux critères de contrôle qualité
Première publication (Réel)
Mises à jour des dossiers d'étude
Dernière mise à jour publiée (Réel)
Dernière mise à jour soumise répondant aux critères de contrôle qualité
Dernière vérification
Plus d'information
Termes liés à cette étude
Mots clés
Termes MeSH pertinents supplémentaires
Autres numéros d'identification d'étude
- STUDY00001138
Plan pour les données individuelles des participants (IPD)
Prévoyez-vous de partager les données individuelles des participants (DPI) ?
Description du régime IPD
Study design and deidentified aggregate glucose concentration data will be shared. No individual patient data will be shared.
Délai de partage IPD
One year
Critères d'accès au partage IPD
Publication
Type d'informations de prise en charge du partage d'IPD
- PROTOCOLE D'ÉTUDE
- SÈVE
- CIF
Informations sur les médicaments et les dispositifs, documents d'étude
Étudie un produit pharmaceutique réglementé par la FDA américaine
Étudie un produit d'appareil réglementé par la FDA américaine
produit fabriqué et exporté des États-Unis.
Ces informations ont été extraites directement du site Web clinicaltrials.gov sans aucune modification. Si vous avez des demandes de modification, de suppression ou de mise à jour des détails de votre étude, veuillez contacter register@clinicaltrials.gov. Dès qu'un changement est mis en œuvre sur clinicaltrials.gov, il sera également mis à jour automatiquement sur notre site Web .