- ICH GCP
- Registro de ensaios clínicos dos EUA
- Ensaio Clínico NCT06190808
Monitorização Contínua da Glicose na Unidade de Terapia Intensiva
Monitorização Contínua da Glicose na Unidade de Terapia Intensiva: Validação e Implementação
The aim of the proposed clinical trial is to study the use of the Dexcom G7 continuous glucose monitor (CGM) within the intensive care unit (ICU) environment. The overall goal is to validate the accuracy and practicality of the Dexcom G7 when used to assess glycemia in critically ill patients.
Visão geral do estudo
Status
Condições
Intervenção / Tratamento
Descrição detalhada
The specific goals are 1) to work with intensive care unit (ICU) staff, in a respiratory ICU setting, to determine how best to place and maintain Dexcom G7 continuous glucose monitors (CGMs) on patients and how to place recording devices and chargers in that environment, 2) download CGM readings to the Dexcom Dashboard in an analyzable format, and 3) to compare simultaneous CGM readings with glucose measurements obtained in the course of routine care as recorded in the electronic medical record. The primary analytical goal is the correlation of CGM glucose measurements with those obtained in the course of routine care. A secondary goal is to determine the frequency of device malfunctions that occur either as a result of device failure (e.g. it falls off) or adverse events (e.g. removal due to skin irritation).
Study subjects will be patients admitted to the 6-ICU UMass Memorial Medical Center (UMMMC) University Campus medical intensive care unit for any reason who have known diabetes or who are hyperglycemic (blood glucose concentration ≥200 mg/dL) with diabetes that was previously undiagnosed or unrecognized. All enrolled volunteers will have a Dexcom G7 CGM place on the skin according to manufacturer's specifications. The devices will remain in place for 10 days or until discharge from the ICU
Tipo de estudo
Inscrição (Real)
Estágio
- Não aplicável
Contactos e Locais
Locais de estudo
-
-
Massachusetts
-
Worcester, Massachusetts, Estados Unidos, 01605
- UMass Memorial Medical Center
-
-
Critérios de participação
Critérios de elegibilidade
Idades elegíveis para estudo
Aceita Voluntários Saudáveis
Gêneros Elegíveis para o Estudo
Descrição
Inclusion Criteria:
Campus medical intensive care unit for any reason who have known diabetes or who are hyperglycemic (blood glucose concentration ≥200 mg/dL) with diabetes that was previously undiagnosed or unrecognized.
- Persons with either type 1 or type 2 diabetes.
- Persons of either sex.
- Persons of any age over 18.
- Persons able to give informed consent or cognitively impaired adults whose legally authorized representative (LAR) can give informed consent.
- English or Spanish speaking adults.
Exclusion Criteria:
- Persons unable to provide informed consent.
- Cognitively impaired persons whose LAR declines to provide informed consent.
- Patients receiving hydroxyurea, which interferes with sensor function.
- Patients receiving high dose acetaminophen (> 4 gm/day), which interferes with sensor function.
- Anasarca affecting preferred areas for device placement (arm, upper buttocks, or thigh), which may interfere with sensor accuracy.
- Severe cachexia with absence of subcutaneous fat at preferred sensor placement sites, which might interfere with sensor accuracy.
- Persons who are pregnant.
- Prisoners.
Plano de estudo
Como o estudo é projetado?
Detalhes do projeto
- Finalidade Principal: Outro
- Alocação: N / D
- Modelo Intervencional: Atribuição de grupo único
- Mascaramento: Nenhum (rótulo aberto)
Armas e Intervenções
Grupo de Participantes / Braço |
Intervenção / Tratamento |
|---|---|
|
Experimental: Single Arm
All members of the arm will have a CGM place and glucose concentration downloaded. |
Continuous glucose monitor |
O que o estudo está medindo?
Medidas de resultados primários
Medida de resultado |
Descrição da medida |
Prazo |
|---|---|---|
|
Mean Absolute Relative Difference (MARD).
Prazo: 6 months
|
Concordance of glucose concentration results obtained by standard test methods with those obtained nearly simultaneously by a continuous glucose monitoring device was assessed by Mean Absolute Relative Difference (MARD). The MARD is the average percentage difference between the continuous glucose monitor readings measured over 6 months and glucose concentration results obtained by near simultaneous (+/- 5 minutes) standard test methods was used to gauge the accuracy of the CGM in the ICU setting. Typical MARD values reported for CGMs in the outpatient setting average approximately 8%. Typical MARD values reported for CGMs other than the one used in this study n the ICU setting average approximately 12%. |
6 months
|
|
Clarke Error Grid
Prazo: 6 months
|
Concordance of glucose concentration results obtained by standard test methods with those obtained nearly simultaneously by a continuous glucose monitoring device was assessed by Clarke Error Grid. The Clarke Error Grid plots of reference glucose values obtained by point-of-care fingerstick or laboratory determination (x-axis) against CGM values (y-axis). The grid is divided into 5 zones (A-E) that indicate clinical implications of CGM glucose measurements when compared to a "gold standard" Region A: Clinically accurate. Region B: Benign measurement error. Region C: Overcorrection error region CGM values suggest unnecessary treatment. Region D: Serious hypo- or hyperglycemia detection error region. Region E: Erroneous treatment region. CGM fails to detect potentially dangerous hypo- or hyperglycemia Ideally, all values are in Regions A and B. We report the percentage of correlated values that fall in each region. |
6 months
|
Medidas de resultados secundários
Medida de resultado |
Descrição da medida |
Prazo |
|---|---|---|
|
Unplanned Event Frequency
Prazo: 6 months
|
Number of occurrences per patient of device failure or or unforeseen device removal |
6 months
|
Colaboradores e Investigadores
Patrocinador
Investigadores
- Investigador principal: John P Mordes, MD, UMass Chan Medical School
Datas de registro do estudo
Datas Principais do Estudo
Início do estudo (Real)
Conclusão Primária (Real)
Conclusão do estudo (Real)
Datas de inscrição no estudo
Enviado pela primeira vez
Enviado pela primeira vez que atendeu aos critérios de CQ
Primeira postagem (Real)
Atualizações de registro de estudo
Última Atualização Postada (Real)
Última atualização enviada que atendeu aos critérios de controle de qualidade
Última verificação
Mais Informações
Termos relacionados a este estudo
Palavras-chave
Termos MeSH relevantes adicionais
Outros números de identificação do estudo
- STUDY00001138
Plano para dados de participantes individuais (IPD)
Planeja compartilhar dados de participantes individuais (IPD)?
Descrição do plano IPD
Study design and deidentified aggregate glucose concentration data will be shared. No individual patient data will be shared.
Prazo de Compartilhamento de IPD
One year
Critérios de acesso de compartilhamento IPD
Publication
Tipo de informação de suporte de compartilhamento de IPD
- PROTOCOLO DE ESTUDO
- SEIVA
- CIF
Informações sobre medicamentos e dispositivos, documentos de estudo
Estuda um medicamento regulamentado pela FDA dos EUA
Estuda um produto de dispositivo regulamentado pela FDA dos EUA
produto fabricado e exportado dos EUA
Essas informações foram obtidas diretamente do site clinicaltrials.gov sem nenhuma alteração. Se você tiver alguma solicitação para alterar, remover ou atualizar os detalhes do seu estudo, entre em contato com register@clinicaltrials.gov. Assim que uma alteração for implementada em clinicaltrials.gov, ela também será atualizada automaticamente em nosso site .