TES HT100 临床结果 - TES HT100 脑电图仪在轻度头部创伤或疑似神经系统症状患者中用于脑损伤筛查的临床验证(TESCO)

The goal of this clinical study is to evaluate how accurately and safely the TES HT100 device can identify intracranial abnormalities in adult emergency department patients with mild head trauma or neurological symptoms that are not related to trauma and are not immediately life-threatening. The main questions it aims to answer are:

  • How well does TES HT100 identify patients with intracranial abnormalities compared with head CT, the reference standard?
  • Is TES HT100 safe to use in this patient population?
  • Could TES HT100 help improve patient triage and use of hospital resources in the emergency setting?

Researchers will compare the result of TES HT100 with the head CT report.

Participants will:

  • Be enrolled if they are adults presenting to the emergency department with mild head trauma or selected neurological symptoms
  • Undergo standard clinical assessment and head CT according to usual care
  • Have an additional TES HT100 examination lasting about 5 minutes, performed by trained personnel
  • Be assessed for device performance and any adverse events related to use of the device

研究概览

详细说明

This is a prospective, multicenter, open-label clinical investigation evaluating the TES HT100 as a non-invasive adjunctive screening device for detection of intracranial abnormalities in adult Emergency Department patients. The study is being conducted in three centers in Italy and Poland.

TES HT100 is intended for non-invasive detection of endocerebral lesions. The device used in this investigation is TES HT model 100, equipped with software version 1.3 or later and firmware version 1.4 or later, in the same configuration planned for commercialization. The system uses non-ionizing, ultra-low-power electromagnetic waves in the 500-6500 MHz frequency range and provides a binary result indicating the presence or absence of intracranial abnormality. The examination is brief, with an overall duration of about 5 minutes.

The study is designed to evaluate the diagnostic performance and safety of TES HT100 when used in addition to standard emergency care. Participants undergo routine clinical assessment and brain CT according to usual clinical practice, and the TES HT100 examination is performed as an add-on procedure. The reference standard for performance evaluation is the brain CT report. TES HT100 is not intended to replace standard clinical assessment or CT imaging.

To reduce interpretation bias, the TES HT100 examination is performed by trained personnel who do not have access to the CT report at the time of testing, and the radiologist interpreting the CT scan is blinded to the TES HT100 result. TES HT100 examinations are performed within 24 hours before or after CT, according to site workflow.

The investigation plans to enroll approximately 850 participants overall across two clinical cohorts. The study duration is planned to be 30 months. Three TES HT100 devices, one per participating center, will be used under the same qualification and maintenance procedures.

In addition to diagnostic validation against CT, the study includes safety monitoring during device use and an exploratory assessment of the potential clinical utility of TES HT100 in emergency department triage and resource management. The overall objective is to generate clinical evidence on the performance and safety of TES HT100 as a rapid, non-invasive screening support tool in the emergency setting.

研究类型

介入性

注册 (估计的)

850

阶段

  • 不适用

联系人和位置

本节提供了进行研究的人员的详细联系信息,以及有关进行该研究的地点的信息。

学习联系方式

学习地点

    • CE
      • Castel Volturno、CE、意大利、81030
        • 招聘中
        • Pineta Grande Hospital
        • 接触:
        • 副研究员:
          • Angelo Localzo, Registered Nurse
        • 副研究员:
          • Esaù Russo, Registered Nurse
        • 首席研究员:
          • Adriana Modestina Conte, MD
        • 副研究员:
          • Antonia Sorbo, MD - Radiologist
      • Gdansk、波兰、80-952
        • 尚未招聘
        • Emergency Department, University Clinical Centre, Medical University of Gdańsk
        • 接触:
        • 首席研究员:
          • Mariusz Siemiński, prof. dr hab. n. med.
      • Katowice、波兰、40-635
        • 招聘中
        • Emergency Department, Leszek Giec Upper Silesian Medical Centre of the Medical University of Silesia in Katowice
        • 接触:
        • 首席研究员:
          • Mateusz Majewski, dr n. med.

参与标准

研究人员寻找符合特定描述的人,称为资格标准。这些标准的一些例子是一个人的一般健康状况或先前的治疗。

资格标准

适合学习的年龄

18年 及以上 (成人、年长者)

接受健康志愿者

不

有资格学习的性别

全部

描述

Inclusion Criteria:

  • Age greater than 18 years
  • Provision of informed consent
  • Presentation to the Emergency Department with mild head injury, or
  • Presentation to the Emergency Department for non-traumatic neurological symptoms and not in a critical condition, including:
  • syncope
  • spatial disorientation and/or loss of balance
  • speech disturbance and/or facial hemi-paresis
  • visual alterations
  • memory loss and/or confusional state

Exclusion Criteria:

  • Age less than 18 years
  • Pregnant women
  • Metal plates or metal prostheses installed in the skull cap
  • Severe head trauma and/or extensive lacerated-contused areas
  • Critical condition
  • Visible skull fracture
  • Failure or inability to provide signed informed consent

学习计划

本节提供研究计划的详细信息,包括研究的设计方式和研究的衡量标准。

研究是如何设计的?

设计细节

  • 主要用途:诊断
  • 分配:不适用
  • 介入模型:单组作业
  • 屏蔽:无(打开标签)

武器和干预

参与者组/臂
干预/治疗
实验性的:TES HT100 Examination

All enrolled participants receive standard emergency department evaluation, including head CT according to usual clinical practice, and undergo an additional TES HT100 examination as an add-on procedure. TES HT100 is evaluated against the head CT report used as the reference standard. The study includes adults with mild head injury and adults with neurological symptoms not related to head trauma and not in a critical or life-threatening condition.

A non-invasive, add-on diagnostic examination performed with the TES HT100 device to assess the presence or absence of intracranial abnormality in adult emergency department patients. The device uses ultra-low-power, non-ionizing electromagnetic waves in the 500-6500 MHz range and provides a binary on/off result. The examination lasts about 5 minutes, does not require direct contact with the patient's head, and is performed by trained personnel in addition to standard clinical evaluation and head CT.

研究衡量的是什么?

主要结果指标

结果测量
措施说明
大体时间
Diagnostic accuracy of TES HT100 for detection of intracranial abnormalities
大体时间:At the index emergency department visit, with TES HT100 performed within 24 hours before or after head CT

Diagnostic performance of TES HT100 compared with the head CT report as the reference standard, assessed by ROC-AUC and by sensitivity and specificity for identification of intracranial abnormalities.

At the index emergency department visit, with TES HT100 performed within 24 hours before or after head CT
Rate of device-related adverse events
大体时间:From TES HT100 examination through study participation completion, an average of 1 year

Incidence and type of adverse events related to the use of TES HT100 during the study.

From TES HT100 examination through study participation completion, an average of 1 year

次要结果测量

结果测量
措施说明
大体时间
Operator time associated with TES HT100-supported screening
大体时间:Day 1

Descriptive evaluation of operator time associated with use of TES HT100 during emergency department screening, based on the number of TES HT100 examinations performed per participant and the approximate duration of each examination (about 5 minutes), as described in the protocol.

Day 1
Procedures associated with TES HT100-supported screening
大体时间:Day 1

Descriptive evaluation of workflow-related procedures associated with use of TES HT100 during emergency department screening, as part of the exploratory assessment of potential clinical benefit described in the protocol. This may include patient handling or organizational procedures required during the diagnostic pathway, such as patient transfer to a different bed or room or need for additional staff support.

Day 1
Costs associated with TES HT100-supported screening
大体时间:Day 1

Descriptive evaluation of resource use and related cost considerations associated with use of TES HT100 during emergency department screening, as part of the exploratory assessment of potential clinical benefit described in the protocol.

Day 1
Diagnostic performance of TES HT100 in predefined patient subgroups
大体时间:At the index emergency department visit, with TES HT100 performed within 24 hours before or after head CT

Analysis of TES HT100 diagnostic performance, based on sensitivity and specificity versus head CT, in predefined subgroups, including younger patients, older patients, patients with or without prior injury or brain surgery, and patients with or without neurodegenerative disease.

At the index emergency department visit, with TES HT100 performed within 24 hours before or after head CT

合作者和调查者

在这里您可以找到参与这项研究的人员和组织。

赞助

调查人员

  • 研究主任:Fabrizio Alfieri、B&B srl

出版物和有用的链接

负责输入研究信息的人员自愿提供这些出版物。这些可能与研究有关。

一般刊物

研究记录日期

这些日期跟踪向 ClinicalTrials.gov 提交研究记录和摘要结果的进度。研究记录和报告的结果由国家医学图书馆 (NLM) 审查,以确保它们在发布到公共网站之前符合特定的质量控制标准。

研究主要日期

学习开始 (实际的)

2024年12月4日

初级完成 (估计的)

2026年12月30日

研究完成 (估计的)

2027年1月30日

研究注册日期

首次提交

2026年3月14日

首先提交符合 QC 标准的

2026年3月18日

首次发布 (实际的)

2026年3月24日

研究记录更新

最后更新发布 (实际的)

2026年9月24日

上次提交的符合 QC 标准的更新

2026年9月21日

最后验证

2026年9月1日

更多信息

与本研究相关的术语

计划个人参与者数据 (IPD)

计划共享个人参与者数据 (IPD)?

不

IPD 计划说明

De-identified participant-level data may be shared upon reasonable request, subject to ethics and institutional approval where applicable and execution of a data sharing agreement. Public sharing is not planned due to confidentiality and consent restrictions.

药物和器械信息、研究文件

研究美国 FDA 监管的药品

不

研究美国 FDA 监管的设备产品

不

此信息直接从 clinicaltrials.gov 网站检索,没有任何更改。如果您有任何更改、删除或更新研究详细信息的请求,请联系 register@clinicaltrials.gov. clinicaltrials.gov 上实施更改,我们的网站上也会自动更新.

订阅