Multi-target Tracking in Prostate Radiotherapy Using MLC and KIM (KOALA)
Phase I Feasibility Study of Accounting for Relative Motion of Multiple Targets in Prostate Cancer Radiotherapy Using Realtime Multi-leaf Collimator Adaptation and Kilovoltage (kV) Intrafraction Monitoring
Study Overview
Status
Status
Conditions
Conditions
Intervention / Treatment
Intervention / Treatment
Detailed Description
This study will assess the feasibility of implementing real-time multi-leaf collimator (MLC) tracking to account for the relative motion of the moving prostate tumour target and the static pelvic nodal target for high-risk prostate cancer patients. The capability of tracking for the relative motion of multiple targets will ensure that all the treatment targets receive correct dose as prescribed by the doctor and minimising side effects to the critical organs.
During radiation treatment, the prostate position will be monitored in real time using the KIM technology. The nodal target will be imaged before and after each treatment to evaluate the nodal treatment margin. The MLC tracking is implemented by recalculating the radiation beam shape fit for the moved prostate and static nodal targets and sending the adjusted MLC leaf positions to the treatment delivery system. The actually delivered dose to the patient will be calculated after the treatment and compared to the dose without MLC tracking to assess the treatment efficacy.
Study Type
Study Type
Phase
Phase
- Not Applicable
Contacts and Locations
Study Locations
-
-
New South Wales
-
St Leonards, New South Wales, Australia, 2065
- Royal North Shore Hospital
-
-
Participation Criteria
Eligibility Criteria
Eligibility Criteria
Ages Eligible for Study
Accepts Healthy Volunteers
Description
Inclusion Criteria:
- Patients undergoing definitive external beam radiotherapy at Northern Sydney Cancer Centre
- Patients histologically proven prostate adenocarcinoma
- PSA obtained within 3 months prior to enrolment
- Pelvic lymph nodes are included for treatment for patients at stage T1c-T3b with Gleason 8-10 or prostate-specific antigen (PSA) > 20 ng/ml or pelvic lymph node positivity on conventional imaging or prostate-specific membrane antigen (PSMA) scans.
- Patient must be able to have gold fiducial markers placed in the prostate
- Eastern Cooperative Oncology Group (ECOG) performance status 0-2
- Ability to understand and the willingness to sign a written informed consent document.
- Prostate dimension that allows leaf span with tracking margin of ±8mm
Exclusion Criteria:
- Patients with artificial Hip(s), lumbar spinal surgical rods or other large metallic pelvic implants
- Patient's dimensions >40cm as measured at the level of the prostate
- Patients with overlapping implanted gold fiducials in x-ray imaging
Study Plan
How is the study designed?
Design Details
- Primary Purpose: Treatment
- Allocation: Non-Randomized
- Interventional Model: Single Group Assignment
- Masking: None (Open Label)
Number of Arms
Arms and Interventions
Participant Group / ArmParticipant Group / Arm |
Intervention / TreatmentIntervention / Treatment |
|---|---|
|
Experimental: Stage 1: Optimise nodal treatment margin
During Stage 1: Optimise nodal treatment margin, Combined real-time use of 'Multi-Leaf Collimator Adaptation' and 'kV Intrafraction Monitoring' will be used to reshape the radiation beam in real-time.
The nodal target position stability will be evaluated by the cone beam computed tomography (CBCT) imaging before and after each treatment session for the first 10 patients.
|
This intervention uses the simultaneous combined use of two technologies: 'Multi-Leaf Collimator Adaptation' and 'kV Intrafraction Monitoring'.
kV Intrafraction Monitoring (KIM) measures the motion of tissues targeted for radiotherapy in real time using kV imaging.The multi-leaf collimator (MLC) reshapes the radiation beam to maximise dose to the target tissue and minimise dose to the surrounding healthy tissue.
Combining KIM and MLC allows the shaped radiation beam to follow the moving target (the prostate) while remaining fixed on a stationary target (lymph nodes) that are also being treated at the same time.
Other Names:
|
|
Experimental: Stage 2: Use treatment margin
During Stage 2: Use treatment margin, Combined real-time use of 'Multi-Leaf Collimator Adaptation' and 'kV Intrafraction Monitoring' will be used to reshape the radiation beam in real-time.
At the same time, multi-leaf collimator (MLC) tracking will be used to reshape the radiation beam in real-time using the margin size determined in Stage 1.
|
This intervention uses the simultaneous combined use of two technologies: 'Multi-Leaf Collimator Adaptation' and 'kV Intrafraction Monitoring'.
kV Intrafraction Monitoring (KIM) measures the motion of tissues targeted for radiotherapy in real time using kV imaging.The multi-leaf collimator (MLC) reshapes the radiation beam to maximise dose to the target tissue and minimise dose to the surrounding healthy tissue.
Combining KIM and MLC allows the shaped radiation beam to follow the moving target (the prostate) while remaining fixed on a stationary target (lymph nodes) that are also being treated at the same time.
Other Names:
|
What is the study measuring?
Primary Outcome Measures
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Software or mechanical failure
Time Frame: The treatment period (2-9 weeks)
|
The percentage of fractions delivered without software or mechanical failure
|
The treatment period (2-9 weeks)
|
Secondary Outcome Measures
Secondary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Geometric accuracy
Time Frame: The treatment period (2-9 weeks)
|
The geometric accuracy of the beam shape, determined by comparing the ideal beam shape with the actual beam shape.
|
The treatment period (2-9 weeks)
|
|
Prostate motion trajectory
Time Frame: Treatment period (2-9 weeks)
|
Prostate motion trajectory measured by KIM.
|
Treatment period (2-9 weeks)
|
|
Dosimetric accuracy
Time Frame: Treatment period (2-9 weeks)
|
The estimated dose distributions will be compared to the original plan using the dose reconstruction method18 based on the prostate motion trajectory and the logged MLC positions (beam shapes).
|
Treatment period (2-9 weeks)
|
|
Acute toxicity
Time Frame: Treatment period (2-9 weeks) plus 3 months
|
Toxicity during treatment
|
Treatment period (2-9 weeks) plus 3 months
|
Collaborators and Investigators
Sponsor
Sponsor
Investigators
Investigators
- Principal Investigator: Thomas Eade, MBBS, RANZCR, Director of Research, Senior Staff Specialist
Study record dates
Study Major Dates
Study Start (Actual)
Study Start
Primary Completion (Actual)
Primary Completion
Study Completion (Actual)
Study Completion
Study Registration Dates
First Submitted
First Submitted
First Submitted That Met QC Criteria
First Submitted That Met QC Criteria
First Posted (Estimated)
First Posted
Study Record Updates
Last Update Posted (Actual)
Last Update Posted
Last Update Submitted That Met QC Criteria
Last Update Submitted That Met QC Criteria
Last Verified
Last Verified
More Information
Terms related to this study
Keywords
Additional Relevant MeSH Terms
Other Study ID Numbers
Other Study ID Numbers
- Multi-target tracking
Plan for Individual participant data (IPD)
Plan to Share Individual Participant Data (IPD)?
IPD Plan Description
Drug and device information, study documents
Studies a U.S. FDA-regulated drug product
Studies a U.S. FDA-regulated device product
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.
Clinical Trials on Prostate Cancer
-
NCT07156045RecruitingProstate Cancer Castration-resistant Prostate Cancer
-
NCT03880422RecruitingObesity | Overweight | Cancer Survivor | Prostate Adenocarcinoma | Stage I Prostate Cancer | Stage II Prostate Cancer | Stage III Prostate Cancer | Stage IV Prostate Cancer | Stage IIA Prostate Cancer | Stage IIB Prostate Cancer
-
NCT03477864WithdrawnStage III Prostate Cancer | Stage IV Prostate Cancer | Stage IVA Prostate Cancer | Stage IVB Prostate Cancer | Stage IIIA Prostate Cancer | Stage IIIB Prostate Cancer | Stage IIIC Prostate Cancer
-
NCT01469338TerminatedDiarrhea | Recurrent Prostate Cancer | Hormone-resistant Prostate Cancer | Stage I Prostate Cancer | Stage III Prostate Cancer | Stage IV Prostate Cancer | Stage IIA Prostate Cancer | Stage IIB Prostate Cancer
-
NCT02144649CompletedStage I Prostate Cancer | Stage III Prostate Cancer | Stage IV Prostate Cancer | Stage IIA Prostate Cancer | Stage IIB Prostate Cancer
-
NCT01882985CompletedRecurrent Prostate Cancer | Stage I Prostate Cancer | Stage III Prostate Cancer | Adenocarcinoma of the Prostate | Stage IIA Prostate Cancer | Stage IIB Prostate Cancer
-
NCT07298239RecruitingProstate Cancer Castration-resistant Prostate Cancer
-
NCT04457245TerminatedRandomized Trial of PSMA PET Scan Before Definitive Radiation Therapy for Prostate Cancer (PSMA-dRT)Stage II Prostate Cancer AJCC v8 | Stage IIIA Prostate Cancer AJCC v8 | Stage IIIB Prostate Cancer AJCC v8 | Stage IIC Prostate Cancer AJCC v8 | Stage III Prostate Cancer AJCC v8 | Stage IIIC Prostate Cancer AJCC v8 | Stage IIA Prostate Cancer AJCC v8 | Stage IIB Prostate Cancer AJCC v8 | Stage I Prostate Cancer American Joint Committee on Cancer (AJCC) v8
-
NCT00121238CompletedRecurrent Prostate Cancer | Stage I Prostate Cancer | Stage III Prostate Cancer | Stage IIA Prostate Cancer | Stage IIB Prostate Cancer
-
NCT07593079Not yet recruitingRecurrent Prostate Cancer | Prostate Cancer | Metastatic Prostate Cancer | Prostate Cancer Recurrent | Prostate Cancer Metastatic
Clinical Trials on Combined real-time use of 'Multi-Leaf Collimator Adaptation' and 'kV Intrafraction Monitoring'
-
NCT05987293Active, not recruitingTracheal Tube-related Tongue Injury