Accuracy of Non-Invasive Blood Pressure Measurements at the Arm, Calf , and Finger (PANIC)
Prospective, Multi-center Evaluation of the Accuracy of Non-invasive Measurement of Blood Pressure Using an Arm, Calf and Finger Cuff.
Study Overview
Status
Status
Conditions
Conditions
Intervention / Treatment
Intervention / Treatment
Detailed Description
Study Type
Study Type
Enrollment (Actual)
Enrollment
Phase
Phase
- Not Applicable
Contacts and Locations
Study Locations
-
-
-
Nantes, France, 44093
- CHU de Nantes
-
Orléans, France
- CHR Orléans
-
Tours, France, 37044
- CHRU de Tours
-
-
Participation Criteria
Eligibility Criteria
Eligibility Criteria
Ages Eligible for Study
Accepts Healthy Volunteers
Genders Eligible for Study
Description
Inclusion Criteria:
- Patient hospitalised in one of the three recruiting intensive care units
- Patient already carrying an indwelling arterial catheter
- Stable BP over a 5-min period: no variation of mean BP above 10% and no change in vasopressor therapy
Exclusion Criteria:
- Brachial circumference > 42 cm.
- Any local fracture, wound, ischemic injury, infection, amputation, thrombophlebitis, that could prevent inflating a cuff around the arm or a finger
- Need of emergency therapy not compatible with the BP measurements schedule
- Asymmetry of mean BP between right and left arm (> 5 mmHg)
- Pregnancy
- Age < 18 years
- Patient under curatorship or guardianship
Study Plan
How is the study designed?
Design Details
- Primary Purpose: Diagnostic
- Allocation: N/A
- Interventional Model: Single Group Assignment
- Masking: None (Open Label)
Number of Arms
Arms and Interventions
Participant Group / ArmParticipant Group / Arm |
Intervention / TreatmentIntervention / Treatment |
|---|---|
|
Other: Combined non-invasive and invasive BP measurements
Patients will all undergo measurement of BP through 3 different techniques over a 30-min period: continuous noninvasive BP measurement (with the finger cuff and Clearsight™ device), repeated intermittent oscillometric NIBP measurements with a cuff placed around a calf or an arm, and continuous invasive BP measurement (through an indwelling arterial catheter).
|
Continuous (finger cuff) noninvasive and invasive BP measurements, over a 30-min period.
Pairs of noninvasive/invasive BP recorded each minute).
In the same time, 3 pairs of oscillometric NIBP (cuff placed around a calf), and invasive BP measurements, at 30-sec intervals.
The cuff is then placed around an arm and triplets of BP (arm cuff, finger cuff, intra-arterial catheter) will be taken over a 20-min period, at 3-min intervals and every time the invasive and/or the continuous finger cuff measurements cross the threshold of 65 mmHg for the mean BP.
If clinically indicated, a cardiovascular intervention that can be either vascular volume expansion or initiation of vasoactive drug therapy, can be administered during this period.
Finally, the cuff placed around the arm will be once again placed around the calf to take 3 pairs of oscillometric NIBP measurement with cuff placed around a calf, and simultaneous invasive BP measurement, at 30-second intervals.
|
What is the study measuring?
Primary Outcome Measures
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Discriminative power of the combined BP measurements with finger cuff and arm cuff to detect intra-arterial mean BP below 65 mmHg
Time Frame: 30 minutes
|
Area under the receiver operating characteristics curve of the combined BP measurements with finger cuff and arm cuff to detect intra-arterial mean BP below 65 mmHg
|
30 minutes
|
Secondary Outcome Measures
Secondary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Discriminative power of the combined BP measurements with finger cuff and arm cuff to detect intra-arterial systolic hypertension
Time Frame: 30 minutes
|
Area under the receiver operating characteristics curve of the combined BP measurements with finger cuff and arm cuff to detect systolic intra-arterial above 140 mmHg
|
30 minutes
|
|
Discriminative power of the combined BP measurements with finger cuff and arm cuff to detect intra-arterial diastolic hypertension
Time Frame: 30 minutes
|
Area under the receiver operating characteristics curve of the combined BP measurements with finger cuff and arm cuff to detect diastolic intra-arterial above 90 mmHg
|
30 minutes
|
|
Accuracy of continuous finger cuff measurements tested against the intra-arterial reference
Time Frame: 30 minutes
|
Accuracy of continuous finger cuff BP measurements tested against the intra-arterial reference, according to the ISO standard
|
30 minutes
|
|
Accuracy of intermittent, oscillometric arm cuff BP measurements tested against the intra-arterial reference
Time Frame: 30 minutes
|
Accuracy of intermittent, oscillometric arm cuff BP measurements tested against the intra-arterial reference, according to the ISO standard
|
30 minutes
|
|
Accuracy of intermittent, oscillometric calf cuff measurements tested against the intra-arterial reference
Time Frame: 30 minutes
|
Accuracy of intermittent, oscillometric calf cuff BP measurements tested against the intra-arterial reference, according to the ISO standard
|
30 minutes
|
Collaborators and Investigators
Sponsor
Sponsor
Investigators
Investigators
- Principal Investigator: Gregoire MULLER, Dr, CHR Orléans
Study record dates
Study Major Dates
Study Start (Actual)
Study Start
Primary Completion (Actual)
Primary Completion
Study Completion (Actual)
Study Completion
Study Registration Dates
First Submitted
First Submitted
First Submitted That Met QC Criteria
First Submitted That Met QC Criteria
First Posted (Actual)
First Posted
Study Record Updates
Last Update Posted (Actual)
Last Update Posted
Last Update Submitted That Met QC Criteria
Last Update Submitted That Met QC Criteria
Last Verified
Last Verified
More Information
Terms related to this study
Keywords
Additional Relevant MeSH Terms
Other Study ID Numbers
Other Study ID Numbers
- CHRO 2019-09
Drug and device information, study documents
Studies a U.S. FDA-regulated drug product
Studies a U.S. FDA-regulated device product
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.