Post-Market Clinical Study of the PanOptix Intraocular Lens (IOL) in a Chinese Population
Post-Market Clinical Study of the AcrySof® IQ PanOptix® Presbyopia Correcting Intraocular Lens in a Chinese Population
Study Overview
Status
Status
Conditions
Conditions
Intervention / Treatment
Intervention / Treatment
Detailed Description
Study Type
Study Type
Enrollment (Actual)
Enrollment
Phase
Phase
- Not Applicable
Contacts and Locations
Study Locations
-
-
-
Beijing, China, 100730
- Beijing Tongren Hospital Capital Medical University
-
Guandong, China, 510060
- Zhongshan Ophthalmic Center, Sun Yat-sen University
-
Hangzhou, China, 310009
- the Second Affiliated Hospital of Zhejiang University School of Medicine
-
Peking, China, 100044
- Peking University People's Hospital
-
Shandong, China, 250002
- Ophthalmic Hospital Affiliated to Shandong University of Traditional Chinese Medicine
-
Shanghai, China, 200031
- Eye & ENT Hospital of Fudan University
-
Shanxi, China, 710068
- Shanxi Eye Hospital
-
Sichuan, China, 610041
- West China Hospital of Sichuan University
-
Tianjin, China, 300384
- Tianjin Medical University Eye Hospital
-
-
Participation Criteria
Eligibility Criteria
Eligibility Criteria
Ages Eligible for Study
Accepts Healthy Volunteers
Description
Key Inclusion Criteria:
- Able to comprehend and sign an informed consent form;
- Able to complete all study visits required in the protocol;
- Chinese; diagnosed with cataracts in both eyes;
- Planned bilateral cataract removal by routine phacoemulsification;
- Pre-operative regular corneal astigmatism of less than 1.0 diopter (D);
- Pre-operative best corrected distance visual acuity (BCDVA) worse than or equal to 0.3 LogMAR in each eye;
- Potential postoperative best corrected distance visual acuity (BCDVA) of 0.3 logMAR or better in both eyes based on the investigator expert medical opinion.
- Other protocol-specified inclusion criteria may apply.
Key Exclusion Criteria:
- Glaucoma, diabetic retinopathy, retinitis pigmentosa and any pathologic changes associated with the optic nerve;
- Clinically significant corneal diseases;
- Clinically significant/severe dry eye that would affect study measurements based on the investigator's expert medical opinion;
- Previous intraocular or corneal surgery;
- Pregnancy or lactation during study or planning to be pregnant/lactating;
- Systemic medications that, in the opinion of the Investigator, may confound the outcome or increase the risk to the subject;
- Other planned ocular surgical procedures;
- Patients who desire monovision correction.
- Other protocol-specified exclusion criteria may apply.
Study Plan
How is the study designed?
Design Details
- Primary Purpose: Treatment
- Allocation: N/A
- Interventional Model: Single Group Assignment
- Masking: None (Open Label)
Number of Arms
Arms and Interventions
Participant Group / ArmParticipant Group / Arm |
Intervention / TreatmentIntervention / Treatment |
|---|---|
|
Experimental: PanOptix IOL
AcrySof IQ PanOptix Presbyopia Correcting IOL implanted in the capsular bag in the posterior chamber of the eye during cataract surgery
|
UV and blue-light filtering foldable multifocal IOL with a +2.17 diopter (D) addition power of intermediate visual acuity and a +3.25 D addition power of near visual acuity.
This device is approved in China.
Other Names:
Cataract extraction by phacoemulsification, followed by implantation of the AcrySof IQ PanOptix Presbyopia Correcting IOL
|
What is the study measuring?
Primary Outcome Measures
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Percentage of Eyes With Best Corrected Distance Visual Acuity (BCDVA) Equal to or Better Than 0.3 LogMAR - Monocular
Time Frame: Month 6 (180-210 days post second eye surgery, 187-238 days post first eye surgery)
|
Visual acuity (VA) was assessed monocularly (eyes individually) with best refractive correction in place under photopic (well-lit) conditions at a distance of 5 meters using Chinese Standard Logarithmic VA charts.
LogMAR typically ranges from -0.3 (20/10 vision on the Snellen chart) to 1 (20/200 vision).
Lower scores indicate better vision.
No hypothesis testing was pre-specified for this endpoint.
|
Month 6 (180-210 days post second eye surgery, 187-238 days post first eye surgery)
|
|
Percentage of Eyes With Best Corrected Distance Visual Acuity (BCDVA) Equal to or Better Than 0.3 LogMAR - Binocular
Time Frame: Month 6 (180-210 days post second eye surgery, 187-238 days post first eye surgery)
|
VA was assessed binocularly (eyes together) with best refractive correction in place under photopic conditions at a distance of 5 meters using Chinese Standard Logarithmic VA charts.
LogMAR typically ranges from -0.3 (20/10 vision on the Snellen chart) to 1 (20/200 vision).
Lower scores indicate better vision.
No hypothesis testing was pre-specified for this endpoint.
|
Month 6 (180-210 days post second eye surgery, 187-238 days post first eye surgery)
|
|
Mean Best Corrected Distance Visual Acuity (BCDVA) - Monocular
Time Frame: Month 6 (180-210 days post second eye surgery, 187-238 days post first eye surgery)
|
Visual acuity (VA) was assessed monocularly (eyes individually) with best refractive correction in place under photopic (well-lit) conditions at a distance of 5 meters using Chinese Standard Logarithmic VA charts.
LogMAR typically ranges from -0.3 (20/10 vision on the Snellen chart) to 1 (20/200 vision).
Lower scores indicate better vision.
No hypothesis testing was pre-specified for this endpoint.
|
Month 6 (180-210 days post second eye surgery, 187-238 days post first eye surgery)
|
|
Mean Best Corrected Distance Visual Acuity (BCDVA) - Binocular
Time Frame: Month 6 (180-210 days post second eye surgery, 187-238 days post first eye surgery)
|
VA was assessed binocularly (eyes together) with best refractive correction in place under photopic conditions at a distance of 5 meters using Chinese Standard Logarithmic VA charts.
LogMAR typically ranges from -0.3 (20/10 vision on the Snellen chart) to 1 (20/200 vision).
Lower scores indicate better vision.
No hypothesis testing was pre-specified for this endpoint.
|
Month 6 (180-210 days post second eye surgery, 187-238 days post first eye surgery)
|
|
Mean Distance Corrected Intermediate Visual Acuity (DCIVA) - Monocular
Time Frame: Month 6 (180-210 days post second eye surgery, 187-238 days post first eye surgery)
|
VA was assessed monocularly (eyes individually) with distance correction in place (plus or minus power) under photopic conditions at a distance of 60 centimeters using Chinese Standard Logarithmic VA charts.
LogMAR typically ranges from -0.3 (20/10 vision on the Snellen chart) to 1 (20/200 vision).
Lower scores indicate better vision.
No hypothesis testing was pre-specified for this endpoint.
|
Month 6 (180-210 days post second eye surgery, 187-238 days post first eye surgery)
|
|
Mean Distance Corrected Intermediate Visual Acuity (DCIVA) - Binocular
Time Frame: Month 6 (180-210 days post second eye surgery, 187-238 days post first eye surgery)
|
VA was assessed binocularly (eyes together) with distance correction in place (plus or minus power) under photopic conditions at a distance of 60 centimeters using Chinese Standard Logarithmic VA charts.
LogMAR typically ranges from -0.3 (20/10 vision on the Snellen chart) to 1 (20/200 vision).
Lower scores indicate better vision.
No hypothesis testing was pre-specified for this endpoint.
|
Month 6 (180-210 days post second eye surgery, 187-238 days post first eye surgery)
|
|
Mean Distance Corrected Near Visual Acuity (DCNVA) - Monocular
Time Frame: Month 6 (180-210 days post second eye surgery, 187-238 days post first eye surgery)
|
VA was assessed monocularly (eyes individually) with distance correction in place (plus or minus power) under photopic conditions at a distance of 40 centimeters using Chinese Standard Logarithmic VA charts.
LogMAR typically ranges from -0.3 (20/10 vision on the Snellen chart) to 1 (20/200 vision).
Lower scores indicate better vision.
No hypothesis testing was pre-specified for this endpoint.
|
Month 6 (180-210 days post second eye surgery, 187-238 days post first eye surgery)
|
|
Mean Distance Corrected Near Visual Acuity (DCNVA) - Binocular
Time Frame: Month 6 (180-210 days post second eye surgery, 187-238 days post first eye surgery)
|
VA was assessed binocularly (eyes together) with distance correction in place (plus or minus power) under photopic conditions at a distance of 40 centimeters using Chinese Standard Logarithmic VA charts.
LogMAR typically ranges from -0.3 (20/10 vision on the Snellen chart) to 1 (20/200 vision).
Lower scores indicate better vision.
No hypothesis testing was pre-specified for this endpoint.
|
Month 6 (180-210 days post second eye surgery, 187-238 days post first eye surgery)
|
|
Number of Ocular Treatment Emergent Adverse Events
Time Frame: Month 12 (360-390 days post second eye surgery, 367-418 days post first eye surgery)
|
An adverse event was defined as any adverse medical event that occurred during the clinical trial whether or not it was related to the investigational medical device or test procedure.
No hypothesis testing was pre-specified for this endpoint.
|
Month 12 (360-390 days post second eye surgery, 367-418 days post first eye surgery)
|
|
Number of Non-Ocular Treatment Emergent Adverse Events
Time Frame: Month 12 (360-390 days post second eye surgery, 367-418 days post first eye surgery)
|
An adverse event was defined as any adverse medical event that occurred during the clinical trial whether or not it was related to the investigational medical device or test procedure.
No hypothesis testing was pre-specified for this endpoint.
|
Month 12 (360-390 days post second eye surgery, 367-418 days post first eye surgery)
|
|
Number of Secondary Surgical Interventions
Time Frame: Month 12 (360-390 days post second eye surgery, 367-418 days post first eye surgery)
|
A secondary surgical intervention (SSI) was defined as a surgical procedure occurring after primary implantation of the IOL.
Posterior capsulotomies were excluded from the definition of secondary surgical interventions.
No hypothesis testing was pre-specified for this endpoint.
|
Month 12 (360-390 days post second eye surgery, 367-418 days post first eye surgery)
|
|
Number of Participants Reporting a Severe Visual Disturbance (QUVID)
Time Frame: Month 6 (180-210 days post second eye surgery, 187-238 days post first eye surgery); Month 12 (360-390 days post second eye surgery, 367-418 days post first eye surgery)
|
The Questionnaire for Visual Disturbances (QUVID) is a patient-reported outcomes questionnaire that collects responses about 7 vision-related experiences.
For each experience, the participant was asked, "In the Past 7 Days, How Severe Was Your Worst Experience" and responded on a 5-point scale, where 0-4 scale, where 0=none, 1=a little, 2=mild, 3=moderate, and 4=severe.
No hypothesis testing was pre-specified for this endpoint.
|
Month 6 (180-210 days post second eye surgery, 187-238 days post first eye surgery); Month 12 (360-390 days post second eye surgery, 367-418 days post first eye surgery)
|
|
Number of Participants Reporting a Most Bothersome Visual Disturbance (QUVID)
Time Frame: Month 6 (180-210 days post second eye surgery, 187-238 days post first eye surgery); Month 12 (360-390 days post second eye surgery, 367-418 days post first eye surgery)
|
The Questionnaire for Visual Disturbances (QUVID) is a patient-reported outcomes questionnaire that collects responses about 7 vision-related experiences.
For each experience, the participant was asked, "In the Past 7 Days, How Much Were You Bothered With" and responded on a 5-point scale, where 0=not bothered at all, 1=bothered a little, 2=bothered somewhat, 3=bothered quite a bit, and 4=bothered very much (most bothersome).
No hypothesis testing was pre-specified for this endpoint.
|
Month 6 (180-210 days post second eye surgery, 187-238 days post first eye surgery); Month 12 (360-390 days post second eye surgery, 367-418 days post first eye surgery)
|
Collaborators and Investigators
Sponsor
Sponsor
Investigators
Investigators
- Study Director: Sr. Clinical Project Manager, Surgical, Alcon (China) Ophthalmic Product Co., Ltd.
Study record dates
Study Major Dates
Study Start (Actual)
Study Start
Primary Completion (Actual)
Primary Completion
Study Completion (Actual)
Study Completion
Study Registration Dates
First Submitted
First Submitted
First Submitted That Met QC Criteria
First Submitted That Met QC Criteria
First Posted (Actual)
First Posted
Study Record Updates
Last Update Posted (Estimated)
Last Update Posted
Last Update Submitted That Met QC Criteria
Last Update Submitted That Met QC Criteria
Last Verified
Last Verified
More Information
Terms related to this study
Keywords
Additional Relevant MeSH Terms
Other Study ID Numbers
Other Study ID Numbers
- ILH297-C004
Plan for Individual participant data (IPD)
Plan to Share Individual Participant Data (IPD)?
Drug and device information, study documents
Studies a U.S. FDA-regulated drug product
Studies a U.S. FDA-regulated device product
product manufactured in and exported from the U.S.
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.