Clinical Trial Comparing Devices Used for Cardioversion of Atrial Fibrillation
A Pragmatic Randomized Clinical Trial Comparing Rectilinear Biphasic Waveform and Biphasic Truncated Exponential Waveform Shocks for Cardioversion of Atrial Fibrillation (Zoll vs Lifepak)
Study Overview
Status
Status
Conditions
Conditions
Intervention / Treatment
Intervention / Treatment
Detailed Description
Study Type
Study Type
Enrollment (Actual)
Enrollment
Phase
Phase
- Not Applicable
Contacts and Locations
Study Locations
-
-
Virginia
-
Falls Church, Virginia, United States, 22042
- Inova Health System
-
-
Participation Criteria
Eligibility Criteria
Eligibility Criteria
Ages Eligible for Study
- Adult
- Older Adult
Accepts Healthy Volunteers
Description
Inclusion Criteria:
- Age ≥ 18 years
- In AF on presentation with plan for DCCV in either the Electrophysiology lab or DCCV procedure area
Exclusion Criteria:
- Participants who are receiving any other investigational agents.
- Parents/guardians or participants who, in the opinion of the Investigator, may be non-compliant with study schedules or procedures.
- Patients with arrhythmias other than AF
- Patients with hemodynamically unstable AF
- Patients with untreated hyperthyroidism
- Patients with known or suspected pregnancy
- Patients without sufficient anticoagulation or a transesophageal echocardiogram or computed tomography scan documenting the absence of intracardiac thrombi
Study Plan
How is the study designed?
Design Details
- Primary Purpose: Treatment
- Allocation: Randomized
- Interventional Model: Crossover Assignment
- Masking: None (Open Label)
Number of Arms
Arms and Interventions
Participant Group / ArmParticipant Group / Arm |
Intervention / TreatmentIntervention / Treatment |
|---|---|
|
Active Comparator: Output synchronized to 200J using Zoll Rectilinear Biphasic Waveform (RBW) defibrillator
Participants receiving Direct Current Cardioversion (DCCV) using full output Rectilinear Biphasic Waveform (RBW) shocks from a Zoll defibrillator (200 Joules) up to 2 shocks.
|
Patients with AF will be randomized to receive a full output synchronized 200J shock using a RBW waveform from a Zoll defibrillator
|
|
Active Comparator: Output synchronized to 360J Lifepak Biphasic Truncated Exponential Waveform (BTE) defibrillator
Participants receiving Direct Current Cardioversion (DCCV) using full output Biphasic Truncated Exponential Waveform (BTE) shocks from a Physiocontrol/Medtronic Lifepak defibrillator (360 Joules) up to 2 shocks.
|
Patients with AF will be randomized to receive a full output synchronized 360J shock using a BTE waveform from a Lifepak defibrillator
|
|
Active Comparator: Zoll defibrillator waveform used after unsuccessful second full output shock.
Participants who received first shock via Zoll that do not have success after a second full output shock will be crossed over to the Lifepak defibrillator waveform and receive up to 2 full output shocks with a minimum of 1 minute between each shock to ascertain the outcome of the immediately prior shock.
|
Patients with AF will be randomized to receive a full output synchronized 360J shock using a BTE waveform from a Lifepak defibrillator
|
|
Active Comparator: Lifepak defibrillator waveform used after unsuccessful second full output shock.
Participants who received first shock via Lifepak that do not have success after a second full output shock will be crossed over to the Zoll defibrillator waveform and receive up to 2 full output shocks with a minimum of 1 minute between each shock to ascertain the outcome of the immediately prior shock.
|
Patients with AF will be randomized to receive a full output synchronized 200J shock using a RBW waveform from a Zoll defibrillator
|
What is the study measuring?
Primary Outcome Measures
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
The Number of Participants in Normal Sinus Rhythm (NSR) by ECG After the First DCCV Shock
Time Frame: At least 1 minute after shock delivery
|
The frequency of NSR by ECG one minute after a single full output DCCV shock is delivered by either a 200J RBW waveform shock from a Zoll Series R defibrillator or 360J BTE waveform shock from a Lifepak 15 defibrillator
|
At least 1 minute after shock delivery
|
Secondary Outcome Measures
Secondary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
The Number of Participants in Normal Sinus Rhythm (NSR) by ECG After the First or Second DCCV Shock
Time Frame: At least 1 minute after shock delivery
|
The frequency of NSR by ECG one minute after a first or second full output DCCV shock is delivered using either a 200J RBW waveform shock from a Zoll Series R defibrillator or 360J BTE waveform shock from a Lifepak 15 defibrillator
|
At least 1 minute after shock delivery
|
|
The Frequency of NSR One Minute After the First or Second Full Output DCCV Shock is Delivered by a Zoll Series R RBW Waveform Defibrillator and a Lifepak Series 15/20 BTE Waveform Defibrillator After Crossing Over From the Alternative Defibrillator.
Time Frame: At least 1 minute after shock delivery
|
The frequency of NSR one minute after the first or second full output DCCV shock is delivered by a Zoll Series R RBW waveform defibrillator and a Lifepak Series 15/20 BTE waveform defibrillator after crossing over from the alternative defibrillator.
|
At least 1 minute after shock delivery
|
Collaborators and Investigators
Sponsor
Sponsor
Investigators
Investigators
- Principal Investigator: Brett Atwater, MD, Inova Health Care Services
Study record dates
Study Major Dates
Study Start (Actual)
Study Start
Primary Completion (Actual)
Primary Completion
Study Completion (Actual)
Study Completion
Study Registration Dates
First Submitted
First Submitted
First Submitted That Met QC Criteria
First Submitted That Met QC Criteria
First Posted (Actual)
First Posted
Study Record Updates
Last Update Posted (Actual)
Last Update Posted
Last Update Submitted That Met QC Criteria
Last Update Submitted That Met QC Criteria
Last Verified
Last Verified
More Information
Terms related to this study
Additional Relevant MeSH Terms
Other Study ID Numbers
Other Study ID Numbers
- U23-04-5042
Drug and device information, study documents
Studies a U.S. FDA-regulated drug product
Studies a U.S. FDA-regulated device product
product manufactured in and exported from the U.S.
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.