Prevention of Severe Hypoglycaemia With Hypo-Safe Hypoglycaemia Alarm Device

March 7, 2014 updated by: UNEEG Medical A/S

Hypoglycaemia is the most common acute complication in insulin-treated diabetes. The fear of hypoglycaemia discourages diabetic subjects from the attempt to maintain tight glycemic control, which in turn leads to increased diabetes related morbidity and mortality.

Neuroglycopenic hypoglycaemia in insulin-treated diabetic patients is associated with characteristic changes in EEG with a decrease in alpha activity and an increase in delta and theta activity. We have recently demonstrated that hypoglycaemia-associated EEG-changes can be recorded from subcutaneously placed electrodes using an automated mathematical algorithm based on non-linear spectral analysis. Experimental findings hold promises that an alarm, given at the time of EEG-changes, can help the patients to avoid severe hypoglycaemia by ingestion of carbohydrate.

This is the first larger scale trial testing the clinical applicability of a hypoglycaemia-alarm based on real-time analysis of EEG-signals.

Study Overview

Status

Withdrawn

Conditions

Detailed Description

The near-normalization of glycemic control has become an established treatment goal in diabetes in order to reduce the risk of late complications such as nephropathy, neuropathy, retinopathy and cardiovascular disease (1). However, the frequency of insulin-induced hypoglycaemia increases several-fold during intensified insulin therapy (2;3) and hypoglycaemia is the most common acute complication in insulin-treated diabetes. The fear of hypoglycaemia discourages diabetic subjects from the attempt to maintain tight glycemic control, which in turn leads to increased diabetes related morbidity and mortality (4;5). Symptoms of hypoglycaemia can be classified as autonomic (warning) symptoms caused by the release of catecholamines and neuroglycopenic symptoms caused by the lack of glucose supply to the brain. Symptoms of hypoglycaemia may be compromised at night-time (nocturnal asymptomatic hypoglycaemia) due to impaired glucose counterregulatory response by adrenaline and glucagon. Some 25% of patients with type 1 diabetes suffer from unawareness in various degrees increasing with long diabetes duration and tight glycemic control (4;6;7).

Several studies have evaluated the potential use of continuous glucose monitoring system (CGMS) as hypoglycaemia alarms but so far failed to show reduction in the frequency of severe hypoglycaemia (8;9). Although the technology is continuously being improved it is still associated with a number of problems (10). The technique is rather imprecise, particularly in the lower range of glucose measurements, and only about 33% of hypoglycemic events were detected in a larger clinical trial (11). The accuracy of the reading is reduced when rapid changes in blood glucose occur (12). There is a significant and variable delay from the change in blood glucose to the change in the interstitial compartment ranging from 4 - 10 minutes (13) and the catheters are rather costly and must be replaced every 72 hours.

The EEG signal reflects the functional state and metabolism of the brain. The brain is almost totally dependent on a continuous supply of glucose, and when the glucose level is lower than the metabolic requirements of the brain, its function deteriorates. Neuroglycopenic hypoglycaemia in insulin-treated diabetic patients is associated with characteristic changes in EEG with a decrease in alpha activity and an increase in delta and theta activity (14-17). These changes are clearly seen at blood glucose ~2.0mmol/l (14;15) preceding the development of severe cognitive dysfunction (18). We have recently demonstrated that hypoglycaemia-associated EEG-changes can be recorded from subcutaneously placed electrodes using an automated mathematical algorithm based on non-linear spectral analysis and that EEG-changes above a predefined threshold can be demonstrated more than 10 minutes before development of severe hypoglycaemia in the majority of the patients (19). We found a very low rate of false alarms and no adverse reactions related to implantation of the electrodes. We have subsequently performed a number of studies with real-time alarms type 1 diabetes patients exposed to insulin induced hypoglycaemia. In these studies the patients were instructed to ingest carbohydrates when he/she heard the alarm sound. In three out of four cases the patients were able to do so, while a fourth patient did not spontaneously ingest the meal although, he was fully conscious and not clinically affect by the hypoglycaemia. These findings hold promises that an alarm, given at the time of EEG-changes, can help the patients to avoid severe hypoglycaemia by ingestion of carbohydrate.

For clinical applicability the device should be able to distinguish hypoglycaemia-induced EEG changes from noise, artefacts and physiological variations in the EEG including the low-frequency waves seen during sleep, with high sensitivity and specificity using a mathematical algorithm that classifies the EEG in real-time. There should be a "time-window" between hypoglycaemia-induced EEG changes and severe cognitive impairment. The device should be fully compatible with normal everyday activities. Thus, the device should be small, fully biocompatible and implantable, and the monitoring and processing unit should be small and have sufficient battery power.

This is the first larger scale trial testing the clinical applicability of a hypoglycaemia-alarm based on real-time analysis of EEG-signals.

Study Type

Interventional

Phase

  • Not Applicable

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Locations

      • Hillerød, Denmark, 3400
        • Hillerød Sygehus

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

16 years to 68 years (Adult, Older Adult)

Accepts Healthy Volunteers

No

Genders Eligible for Study

All

Description

Inclusion Criteria:

  • Male and female patient with type 1 diabetes for at least one year
  • Age 18-70 years
  • Impaired awareness of hypoglycaemia as defined by

    • A score of ≥4 on the Gold-scale (Appendix 8) or
    • Two or more occasions of severe hypoglycaemia (need of help from third person) within the past 12 month
  • Multiple injection insulin therapy or continuous insulin injection therapy
  • For female participants: Not pregnant and, if child bearing potential, usage of reliable anti-contraceptive method during the study period

Exclusion Criteria:

  • Severe cardiac disease

    • History of myocardial infarction
    • Cardiac arrhythmia
  • Previous stroke or cerebral haemorrhage and any other structural cerebral disease
  • Active cancer or cancer diagnosis within the past five years
  • Uremia defined as s-creatinine above 3 times upper reference value
  • Liver disease defined as s-ALAT above 3 times upper reference interval
  • Inability to understand the informed consent
  • Epilepsy
  • Use of antiepileptic drugs for any purposes
  • Clinical important hearing impairment
  • Use of active implantable medical device including

    • Pacemaker and ICD-unit
    • Cochlear implant
  • Use of following drugs

    • Chemotherapeutic drugs of any kind
    • Methotrexate
    • Third generation antipsychotic drugs (aripiprazole, quetiapine, clozapine, ziprasidone, paliperidone, risperidone, sertindole, amisulpride, olanzapine
  • Abuse of alcohol (defined as consumption of > 250g alcohol (in Danish: 21 "genstande") per week or abuse of any other neuroactive substances
  • Infection at the site of device-implantation
  • Any hemorrhagic disease
  • Diving (snorkel diving allowed) or parachute jumping
  • Patients that are judged incapable of understanding the patient information or who are not capable of carrying through the investigation

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Primary Purpose: Prevention
  • Allocation: Randomized
  • Interventional Model: Parallel Assignment
  • Masking: None (Open Label)

Arms and Interventions

Participant Group / Arm
Intervention / Treatment
Experimental: Hyposafe Hypoglycemia alarm device
EEG based hypoglycemia detection
An EEG based hypoglycemia alarm is tested against standard glucose measurement on the risk of severe hypoglycemia
Active Comparator: Regular glucose control
Regular glucose control group
An EEG based hypoglycemia alarm is tested against standard glucose measurement on the risk of severe hypoglycemia

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Time Frame
The frequency of severe hypoglycaemia event
Time Frame: six months
six months

Secondary Outcome Measures

Outcome Measure
Time Frame
The frequency of clinical hypoglycaemia (sensation of hypoglycaemia and BG<3.5mmol/l), biochemical hypoglycaemia (BG<3.5mmol/l), and nocturnal hypoglycaemia (waking up with a sensation of hypoglycaemia and BG<3.5mmol/l)
Time Frame: six months
six months

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Investigators

  • Principal Investigator: Birger Thorsteinsson, MD., Medical Department, Hillerød Sygehus

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start

November 1, 2009

Primary Completion (Anticipated)

January 1, 2010

Study Completion (Anticipated)

March 1, 2011

Study Registration Dates

First Submitted

November 18, 2009

First Submitted That Met QC Criteria

August 9, 2010

First Posted (Estimate)

August 10, 2010

Study Record Updates

Last Update Posted (Estimate)

March 10, 2014

Last Update Submitted That Met QC Criteria

March 7, 2014

Last Verified

August 1, 2010

More Information

Terms related to this study

Other Study ID Numbers

  • Hyposafe-hypo-02

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.

Clinical Trials on Type 1 Diabetes

Clinical Trials on Hyposafe Hypoglycaemia alarm device

3
Subscribe