Camera-based Measurement of Respiratory Rates
Respiratory rate as an important predictor for adverse events, but still a neglected vital sign.
Precise and simple measurement of the respiratory rate in volunteers by a new electronic camera device in different settings.
Study Overview
Status
Conditions
Intervention / Treatment
Detailed Description
Background of the study The respiratory rate (RR) is an important vital sign to be monitored in the emergency room, as it is a sensitive predictor of risk and critical illness . An increased RR is a sensitive and independent marker of increased mortality in patients with community acquired pneumonia and in unselected patients presenting to the emergency department (ED): This is why the RR is a component in many risk scores for prediction of adverse outcomes, such as the 'Confusion/Urea/Respiratory rate/Blood pressure/-Age 65 or older' (CURB-65) score, the Modified Early Warning Score (MEWS), Trauma Score, Revised Trauma Score, APACHE II (Acute Physiology and Chronic Health Evaluation) and PRISM (Paediatric Risk of Mortality ). Additionally, it is part of the Systemic Inflammatory Response Syndrome (SIRS) Sepsis-definition and definitions for weaning a patient from a ventilator (RSBI, Rapid Shallow Breathing Index, ). In some cases, measurement of the oxygen saturation is not sufficient as tachypnoea may not always have a connection to hypoxia . Hence, RR plays an important role in triage of patients presenting to the emergency room. Especially tachypnoea brings the need for more detailed monitoring and investigation. Tachypnea/Bradypnoea can be seen as a red flag and may indicate a big variety of diseases like heart failure, pneumonia, metabolic disturbances or pulmonary embolism. Recent studies could demonstrate that also trends of RR are reliable predictors of clinical outcome. Especially an increase in tachypnoea during hospitalisation is associated with high mortality A careful and exact measurement is essential. In emergency situations the RR is one component of the Emergency Severity Index (ESI) algorithm and a mandatory vital sign measured at decision point "D". However, its measurement is often omitted due to time constraints . Moreover documented pathologic RR in ER scenarios indicating hypoxia are often neglected leading to fatal events and contributing to deaths caused by medical errors.
According to recommendations of the World Health Organization (WHO) the RR should be measured by counting the breaths over a period of 1 minute by counting how many times the chest rises. This should be done when the patient is at rest. This appears impractical due to time consumption and workload for nursing staff in an emergency department. Studies have shown that this fact is a main reason why RR is often not recorded and critical situations might be missed. A prospective study by Lovett et al. demonstrated that the measurement of RR by nursing staff and even by impedance plethysmography as a common used device have a low sensitivity in detecting brady- or tachypnoea. The vital sign measurements may also differ according to observer variability. The approach to use mobile devices for RR measurement has shown encouraging results to improve efficiency in comparison to traditional methods though the field of using mobile devices is still in an experimental stadium.
The investigators aim to compare the results given by the camera-based measurements to the results given by capnography. The goal of the survey is to analyse the accuracy of measurements performed by the software.
The investigators will recruit healthy volunteers without any history of pulmonary diseases for an analysis of RR monitoring. Upon agreement the investigators will measure the RR camera-based by a device and simultaneously count the RR by capnography as a reference method. The data will be collected through a standardised protocol and archived in an Access(R) database. Different settings (e.g. supine vs sitting) will be assessed. Results will be depicted in a Bland-Altmann plot. In a first step the investigators plan to do a feasibility study, after that, as a second step, they will continue to measure the RR in different positions and situations to simulate a situation similar to emergency conditions.
Study Type
Enrollment (Actual)
Phase
- Not Applicable
Contacts and Locations
Study Locations
-
-
Basel-Stadt
-
Basel, Basel-Stadt, Switzerland, 4031
- Intensive Care Unit, University Hospital Basel
-
-
Participation Criteria
Eligibility Criteria
Ages Eligible for Study
Accepts Healthy Volunteers
Genders Eligible for Study
Description
Inclusion Criteria:
- Age 18 and older
- healthy male volunteers
- BMI 19-25
Exclusion Criteria:
- pulmonary diseases
Study Plan
How is the study designed?
Design Details
- Primary Purpose: Diagnostic
- Allocation: N/A
- Interventional Model: Single Group Assignment
- Masking: None (Open Label)
Arms and Interventions
Participant Group / Arm |
Intervention / Treatment |
|---|---|
|
Other: Volunteers
Monitoring of respiratory rates via camera-based System Monitoring of respiratory rate by Philips®Vital Sign Device - Camera-based system (Prototype) and capnography simultaneously.
|
Respiratory rates monitored via camera-based system and capnography simultaneously The "Philips Vital Sign Device" will record chest movements by camera.
The program will use an algorithm to convert the chest movements into the respiratory rate
|
What is the study measuring?
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Absolute difference between respiratory rates at different breathing spectra measured via camera
Time Frame: study day 1
|
The study compares respiratory rates obtained by 3 methods. 1. Camera based measurement, 2. with capnography and 3. with the WHO criterion standard of counting breaths for 1 Minute (18 measurements in whole). The unit of respiratory rate is breaths per minute (e.g. 18/min) To test the accuracy of the Philips Software investigators will compare the camera-based measurements with the measurements done by capnography (as a gold standard). The statistical results will be depicted in a Bland-Altman-Plot |
study day 1
|
|
Absolute difference between respiratory rates at different breathing spectra measured via capnography
Time Frame: study day 1
|
The study compares respiratory rates obtained by 3 methods. 1. Camera based measurement, 2. with capnography and 3. with the WHO criterion standard of counting breaths for 1 Minute (18 measurements in whole). The unit of respiratory rate is breaths per minute (e.g. 18/min) To test the accuracy of the Philips Software investigators will compare the camera-based measurements with the measurements done by capnography (as a gold standard). The statistical results will be depicted in a Bland-Altman-Plot |
study day 1
|
|
Absolute difference between respiratory rates at different breathing spectra measured via WHO standard (counting breaths for one minute)
Time Frame: study day 1
|
The study compares respiratory rates obtained by 3 methods. 1. Camera based measurement, 2. with capnography and 3. with the WHO criterion standard of counting breaths for 1 Minute (18 measurements in whole). The unit of respiratory rate is breaths per minute (e.g. 18/min) To test the accuracy of the Philips Software investigators will compare the camera-based measurements with the measurements done by capnography (as a gold standard). The statistical results will be depicted in a Bland-Altman-Plot |
study day 1
|
Collaborators and Investigators
Investigators
- Principal Investigator: Christian Nickel, MD, University Hospital, Basel, Switzerland
Publications and helpful links
General Publications
- Hodgetts TJ, Kenward G, Vlachonikolis IG, Payne S, Castle N. The identification of risk factors for cardiac arrest and formulation of activation criteria to alert a medical emergency team. Resuscitation. 2002 Aug;54(2):125-31. doi: 10.1016/s0300-9572(02)00100-4.
- Strauss R, Ewig S, Richter K, Konig T, Heller G, Bauer TT. The prognostic significance of respiratory rate in patients with pneumonia: a retrospective analysis of data from 705,928 hospitalized patients in Germany from 2010-2012. Dtsch Arztebl Int. 2014 Jul 21;111(29-30):503-8, i-v. doi: 10.3238/arztebl.2014.0503.
- Pollack MM, Ruttimann UE, Getson PR. Pediatric risk of mortality (PRISM) score. Crit Care Med. 1988 Nov;16(11):1110-6. doi: 10.1097/00003246-198811000-00006.
- Bone RC, Sprung CL, Sibbald WJ. Definitions for sepsis and organ failure. Crit Care Med. 1992 Jun;20(6):724-6. doi: 10.1097/00003246-199206000-00002. No abstract available.
- Meade M, Guyatt G, Cook D, Griffith L, Sinuff T, Kergl C, Mancebo J, Esteban A, Epstein S. Predicting success in weaning from mechanical ventilation. Chest. 2001 Dec;120(6 Suppl):400S-24S. doi: 10.1378/chest.120.6_suppl.400s.
- Mower WR, Sachs C, Nicklin EL, Safa P, Baraff LJ. A comparison of pulse oximetry and respiratory rate in patient screening. Respir Med. 1996 Nov;90(10):593-9. doi: 10.1016/s0954-6111(96)90017-7.
- Subbe CP, Kruger M, Rutherford P, Gemmel L. Validation of a modified Early Warning Score in medical admissions. QJM. 2001 Oct;94(10):521-6. doi: 10.1093/qjmed/94.10.521.
- Goldhill DR, White SA, Sumner A. Physiological values and procedures in the 24 h before ICU admission from the ward. Anaesthesia. 1999 Jun;54(6):529-34. doi: 10.1046/j.1365-2044.1999.00837.x.
- Kellett J, Murray A, Woodworth S, Huang W. Trends in weighted vital signs and the clinical course of 44,531 acutely ill medical patients while in hospital. Acute Med. 2015;14(1):3-9.
- Grossmann FF, Nickel CH, Christ M, Schneider K, Spirig R, Bingisser R. Transporting clinical tools to new settings: cultural adaptation and validation of the Emergency Severity Index in German. Ann Emerg Med. 2011 Mar;57(3):257-64. doi: 10.1016/j.annemergmed.2010.07.021. Epub 2010 Oct 16.
- Venkatesh AK, Curley D, Chang Y, Liu SW. Communication of Vital Signs at Emergency Department Handoff: Opportunities for Improvement. Ann Emerg Med. 2015 Aug;66(2):125-30. doi: 10.1016/j.annemergmed.2015.02.025. Epub 2015 Mar 21.
- Hogan J. Why don't nurses monitor the respiratory rates of patients? Br J Nurs. 2006 May 11-24;15(9):489-92. doi: 10.12968/bjon.2006.15.9.21087.
- Lovett PB, Buchwald JM, Sturmann K, Bijur P. The vexatious vital: neither clinical measurements by nurses nor an electronic monitor provides accurate measurements of respiratory rate in triage. Ann Emerg Med. 2005 Jan;45(1):68-76. doi: 10.1016/j.annemergmed.2004.06.016.
- Edmonds ZV, Mower WR, Lovato LM, Lomeli R. The reliability of vital sign measurements. Ann Emerg Med. 2002 Mar;39(3):233-7. doi: 10.1067/mem.2002.122017.
- Karlen W, Gan H, Chiu M, Dunsmuir D, Zhou G, Dumont GA, Ansermino JM. Improving the accuracy and efficiency of respiratory rate measurements in children using mobile devices. PLoS One. 2014 Jun 11;9(6):e99266. doi: 10.1371/journal.pone.0099266. eCollection 2014. Erratum In: PLoS One. 2015;10(2):e0118260.
- Folke M, Cernerud L, Ekstrom M, Hok B. Critical review of non-invasive respiratory monitoring in medical care. Med Biol Eng Comput. 2003 Jul;41(4):377-83. doi: 10.1007/BF02348078.
Study record dates
Study Major Dates
Study Start
Primary Completion (Actual)
Study Completion (Actual)
Study Registration Dates
First Submitted
First Submitted That Met QC Criteria
First Posted (Estimate)
Study Record Updates
Last Update Posted (Estimate)
Last Update Submitted That Met QC Criteria
Last Verified
More Information
Terms related to this study
Keywords
Additional Relevant MeSH Terms
Other Study ID Numbers
- EKNZ 2015-306
Plan for Individual participant data (IPD)
Plan to Share Individual Participant Data (IPD)?
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.
Clinical Trials on Tachypnea
-
New York Hospital QueensCompleted
-
Zekai Tahir Burak Women's Health Research and Education...UnknownTransient Tachypnea of NewbornTurkey
-
Kanuni Sultan Suleyman Training and Research HospitalCompletedRespiratory Distress Syndrome | Intrauterine Growth Retardation | Transient Tachypnea of NewbornTurkey
-
Ain Shams UniversityUnknownCesarean Delivery Affecting Fetus | Transient Tachypnea NewbornEgypt
-
Benha UniversityCompletedTransient Tachypnea of Newborn and Fetal AcidosisEgypt
-
Massachusetts General HospitalCompletedThermal Imaging, Neonatal Pneumonia, TachypneaUnited States
-
Khyber Teaching HospitalCompleted
-
Makassed General HospitalCompleted
-
Kanuni Sultan Suleyman Training and Research HospitalCompletedRespiratory Distress Syndrome | Transient Tachypnea of Newborn
-
Medical University of WarsawJagiellonian University; University of Ottawa; Pomeranian Medical University... and other collaboratorsActive, not recruitingRespiratory Failure | TTN | PPHNPoland
Clinical Trials on Philips®Vital Sign Device - Camera-based system (Prototype)
-
EarlySense Ltd.The Chaim Sheba Medical CenterWithdrawn
-
Darma Inc.Virginia Commonwealth UniversityCompletedVital Sign MonitoringUnited States