Clinical Performance of Masimo INVSENSOR00040

May 15, 2024 updated by: Masimo Corporation

This study is designed to compare the accuracy of a noninvasive measurement of oxygen saturation and pulse rate to reference values obtained by a laboratory blood gas analyzer and a standard of care EKG monitor.

Study procedures follow ISO-80601-2-61:2011 standard requirements for basic safety and essential performance of pulse oximeter equipment. Arterial blood samples will be collected from subjects while undergoing a controlled desaturation procedure wherein the concentration of oxygen inhaled is slowly reduced until the subject's arterial oxygen concentration is approximately 70%.

Study Overview

Status

Completed

Conditions

Intervention / Treatment

Study Type

Interventional

Enrollment (Actual)

90

Phase

  • Not Applicable

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Locations

    • California
      • Irvine, California, United States, 92618
        • Masimo Corporation

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

  • Adult

Accepts Healthy Volunteers

Yes

Description

Inclusion Criteria:

  • Subject is 18 to 50 years of age.
  • Subject weighs a minimum of 110 lbs.
  • Subject has a hemoglobin value ≥ 11 g/dL.
  • Subject's baseline heart rate is ≥ 45 bpm and ≤ 85 bpm.
  • Subject's CO value is ≤ 3.0% FCOHb.
  • Subject's blood pressure: Systolic BP ≤ 140 mmHg and ≥ 90 mmHg, Diastolic BP ≤ 90 mmHg and ≥ 50 mmHg, and if systolic BP is lower than 100 mmHg and/or diastolic BP is lower than 60 mmHg, subject passes an orthostatic blood pressure test.
  • Subject is able to read and communicate in English and understands the study and the risks involved.

Exclusion Criteria:

  • Subject whose skin is not intact and/or has tattoos in the area of device placement (e.g., wrist).*
  • Subject is pregnant.
  • Subject has a BMI > 35.
  • Subject has a history of fainting (vasovagal syncope), blacking out or losing consciousness during or after a blood draw, or has a fear of blood draws.
  • Subject has open wounds, inflamed tattoos or piercings, and/or has any visible healing wounds that a medical professional determines may place them at an increased risk for participation.*
  • Subject has known drug or alcohol abuse.
  • Subject uses recreational drugs.*
  • Subject experiences frequent or severe headaches and/or migraine headaches, migraine auras, altitude sickness, and/or headaches accompanied by visual changes or sensitivity to light or sound.
  • Subject has experienced a concussion or head injury with loss of consciousness within the past 12 months.
  • Subject has any history of a stroke, myocardial infarction (heart attack), and/or seizures.
  • Subject has any chronic bleeding disorder (e.g. hemophilia).
  • Subject has taken anticoagulant medication within the past 30 days (excluding nonsteroidal anti-inflammatory drugs (NSAIDS)).
  • Subject has donated blood within the past 4 weeks.
  • Subject has Wolff-Parkinson-White Syndrome or Stokes-Adams Syndrome.
  • Subject has any symptomatic cardiac dysrhythmia (e.g. atrial fibrillation) and has not received clearance from their physician to participate.
  • Subject has a known neurological and/or psychiatric disorder (e.g. schizophrenia, bipolar disorder, multiple sclerosis, Huntington's disease) that interferes with the subject's level of consciousness.*
  • Subject has taken opioid pain medication 24 hours before the study.
  • Subject has any active signs and/or symptoms of infectious disease (e.g. hepatitis, HIV, tuberculosis, flu, malaria, measles, etc.).*
  • Subject is taking medications known to treat any type of infectious disease.
  • Subject has either signs or history of peripheral ischemia or carpal tunnel syndrome.
  • Subject has had invasive surgery within the past year, including but not limited to major dental surgery, appendectomy, plastic surgery, jaw surgery, major ENT surgery, major abdominal and/or pelvic surgery, heart surgery, or thoracic surgery.*
  • Subject has symptoms of congestion, head cold, or other illnesses.
  • Subject has been in a severe car accident(s) or a similar type of accident(s) requiring hospitalization within the past 12 months.
  • Subject has any cancer or history of cancer (not including skin cancer).*
  • Subject has chronic unresolved asthma, lung disease (including COPD) and/or respiratory disease.
  • Subject is allergic to lidocaine, chlorhexidine, latex, adhesives, or plastic.
  • Subject has a heart condition, insulin-dependent diabetes, or uncontrolled hypertension.
  • Subject has delivered vaginally, has had a pregnancy terminated, a miscarriage with hospitalization, or had a C-section within the past 6 months.
  • Subject intends on participating in any heavy lifting, repetitive movement of their wrist (including riding a motorcycle, tennis), exercise (working out, riding a bike, riding a skateboard, etc.), or any activity that will put additional stress on the wrist within 24 hours following a study that involves an arterial line.
  • Subject has any medical condition which in the judgment of the investigator and/or medical staff, renders them ineligible for participation in this study or subject is deemed ineligible by the discretion of the investigator/study staff.

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Primary Purpose: Other
  • Allocation: N/A
  • Interventional Model: Single Group Assignment
  • Masking: None (Open Label)

Arms and Interventions

Participant Group / Arm
Intervention / Treatment
Experimental: INVSENSOR00040
All subjects who are enrolled into the test group and participate in data collection receive the noninvasive INVSENSOR00040 device.
Noninvasive wearable health monitoring device

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Oxygen Saturation (SpO2) Accuracy of INVSENSOR00040
Time Frame: 1-5 hours

SpO2 performance will be determined by calculating the arithmetic root square mean (ARMS) value through the comparison of the noninvasive SpO2 measurement obtained from INVSENSOR00040 to the arterial oxygen saturation (SaO2) value obtained from a reference blood sample.

The square root of the sum of the squares of bias and precision is computed as the ARMS Error Value.

The desaturation range over which the accuracy is performed is 70-100%.

1-5 hours
Pulse Rate (PR) Accuracy of INVSENSOR00040
Time Frame: 1-5 hours

Pulse Rate (PR) performance will be determined by calculating the ARMS value through the comparison of the PR values obtained from INVSENSOR00040 to the heart rate (HR) values obtained from a reference standard of care EKG monitor.

The square root of the sum of the squares of bias and precision is computed as the ARMS Error Value.

The PR/HR range over which the accuracy is calculated is determined by convenience sampling in a generally healthy population undergoing a desaturation procedure.

1-5 hours

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start (Actual)

April 14, 2022

Primary Completion (Actual)

April 13, 2023

Study Completion (Actual)

April 13, 2023

Study Registration Dates

First Submitted

April 11, 2022

First Submitted That Met QC Criteria

April 11, 2022

First Posted (Actual)

April 18, 2022

Study Record Updates

Last Update Posted (Actual)

June 11, 2024

Last Update Submitted That Met QC Criteria

May 15, 2024

Last Verified

January 1, 2024

More Information

Terms related to this study

Other Study ID Numbers

  • CIP-1019

Plan for Individual participant data (IPD)

Plan to Share Individual Participant Data (IPD)?

NO

Drug and device information, study documents

Studies a U.S. FDA-regulated drug product

No

Studies a U.S. FDA-regulated device product

Yes

product manufactured in and exported from the U.S.

Yes

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.

Clinical Trials on Healthy

Subscribe