Clinical Equivalence of Masimo INVSENSOR00061 to RD SET Neo

June 11, 2024 updated by: Masimo Corporation
This study is designed to compare the equivalence of the Masimo INVSENSOR00061 to RD SET Neo for SpO2 and performance of PR and RRp as the subjects undergo a controlled desaturation protocol. This type of study design is required to evaluate the performance of pulse oximeters based upon the ISO 80601-2-61 standard which is also adopted and referenced by the FDA.

Study Overview

Status

Completed

Conditions

Intervention / Treatment

Study Type

Interventional

Enrollment (Actual)

62

Phase

  • Not Applicable

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Locations

    • California
      • Irvine, California, United States, 92618
        • Masimo Corporation

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

  • Adult

Accepts Healthy Volunteers

Yes

Description

Inclusion Criteria:

  • Subject is 18 to 50 years of age.
  • Subject weighs a minimum of 110 lbs.
  • Hemoglobin value ≥ 11 g/dL.
  • Baseline heart rate ≥ 45 bpm and ≤ 85 bpm.
  • Blood Pressure: Systolic BP ≤ 140 mmHg and ≥ 90 mmHg, Diastolic BP ≤ 90 mmHg and ≥ 50 mmHg, and if systolic BP is lower than 100 mmHg and/or diastolic BP is lower than 60 mmHg, subject passes an orthostatic blood pressure test.
  • CO value ≤ 3.0% FCOHb.
  • Subject is able to read and communicate in English and understands the study and the risks involved

Exclusion Criteria:

  • Subject is pregnant.
  • Subject has a BMI > 35.
  • Subject is wearing nail polish that cannot be removed, gel nails, and/or acrylic nails that can interfere with study device's placement is deemed ineligible at the discretion of investigator/study staff. *
  • Subject has a history of fainting (vasovagal syncope), blacking out or losing consciousness during or after a blood draw, or has a fear of blood draws.
  • Subject has open wounds, inflamed tattoos or piercings, and/or has any visible healing wounds that a medical professional determines may place them at an increased risk for participation. *
  • Subject has known drug or alcohol abuse.
  • Subject uses recreational drugs. *
  • Subject experiences frequent or severe headaches and/or migraine headaches, migraine auras, altitude sickness, and/or headaches accompanied by visual changes or sensitivity to light or sound.
  • Subject has experienced a concussion or head injury with loss of consciousness within the past 12 months.
  • Subject has any history of a stroke, myocardial infarction (heart attack), and/or seizures.
  • Subject has any chronic bleeding disorder (e.g., hemophilia).
  • Subject has taken anticoagulant medication within the past 30 days (excluding nonsteroidal anti-inflammatory drugs (NSAIDS)).
  • Subject has donated blood within the past 4 weeks.
  • Subject has Wolff-Parkinson-White Syndrome or Stokes-Adams Syndrome.
  • Subject has any cardiac dysrhythmia (e.g., atrial fibrillation) and has not received clearance from their physician to participate.
  • Subject has a known neurological and/or psychiatric disorder (e.g., schizophrenia, bipolar disorder, Multiple Sclerosis, Huntington's disease) that interferes with the subject's level of consciousness. *
  • Subject has taken opioid pain medication 24 hours before the study.
  • Subject has any active signs and/or symptoms of infectious disease (e.g., Hepatitis, HIV, Tuberculosis, Flu, Malaria, Measles, etc.). *
  • Subject is taking medications known to treat any type of infectious disease.
  • Subject has either signs or history of peripheral ischemia or carpal tunnel syndrome.
  • Subject has had invasive surgery within the past year, including but not limited to major dental surgery, appendectomy, plastic surgery, jaw surgery, major ENT surgery, major abdominal and/or pelvic surgery, heart surgery, or thoracic surgery. *
  • Subject has symptoms of congestion, head cold, or other illnesses.
  • Subject has been in a severe car accident(s) or a similar type of accident(s) requiring hospitalization within the past 12 months.
  • Subject has any cancer or history of cancer (not including skin cancer). *
  • Subject has chronic unresolved asthma, lung disease (including COPD) and/or respiratory disease.
  • Subject is allergic to lidocaine, chlorhexidine, latex, adhesives, or plastic.
  • Subject has a heart condition, insulin-dependent diabetes, or uncontrolled hypertension.
  • Subject has delivered vaginally, has had a pregnancy terminated, a miscarriage with hospitalization or had a C-section within the past 6 months.
  • Subject intends on participating in any heavy lifting, repetitive movement of their wrist (including riding a motorcycle, tennis), exercise (working out, riding a bike, riding a skateboard, etc.), or any activity that will put additional stress on the wrist within 24 hours following a study that involves an arterial line.
  • Subject has any medical condition which in the judgment of the investigator and/or medical staff, renders them ineligible for participation in this study or subject is deemed ineligible by the discretion of the investigator/study staff.

(*)Physician's discretion.

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Primary Purpose: Other
  • Allocation: N/A
  • Interventional Model: Single Group Assignment
  • Masking: None (Open Label)

Arms and Interventions

Participant Group / Arm
Intervention / Treatment
Experimental: Test Subject
All subjects who are enrolled into the test group and participate in data collection receive the noninvasive INVSENSOR00061 sensor
Noninvasive pulse oximeter sensor

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
SpO2 Accuracy
Time Frame: 1-3 hours
SpO2 accuracy was determined by calculating the root mean-squared (rms) difference between the measured values (SpO2i) to the reference values (SpO2ref_i) in accordance with ISO-80601-2-61. Arms= √(∑(i=1 to n) ((SpO2_i-SpO2ref_i )^2 ))/n
1-3 hours

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start (Actual)

November 18, 2022

Primary Completion (Actual)

December 7, 2022

Study Completion (Actual)

December 7, 2022

Study Registration Dates

First Submitted

December 8, 2022

First Submitted That Met QC Criteria

December 21, 2022

First Posted (Actual)

January 6, 2023

Study Record Updates

Last Update Posted (Actual)

June 13, 2024

Last Update Submitted That Met QC Criteria

June 11, 2024

Last Verified

June 1, 2024

More Information

Terms related to this study

Other Study ID Numbers

  • CIP-1056

Drug and device information, study documents

Studies a U.S. FDA-regulated drug product

No

Studies a U.S. FDA-regulated device product

Yes

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.

Clinical Trials on Healthy

Clinical Trials on INVSENSOR00061

Subscribe