Philips FAST Evaluation (NELFST)

July 23, 2025 updated by: Medtronic - MITG

Evaluation of NellcorTM Pulse Oximetry Accuracy With the Philips picoSAT X SpO2 Board With Fourier Artifact Suppression Technology in Adult Volunteers

Evaluation of Philips FAST technology with Nellcor sensors

Study Overview

Detailed Description

Evaluate saturation accuracy claims for the Philips picoSAT X SpO2 board with FAST algorithm, paired with Nellcor™ sensors in a diverse subject population, over a specified saturation range

Study Type

Observational

Enrollment (Actual)

27

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Locations

    • Colorado
      • Denver, Colorado, United States, 80218
        • Medtronic Clinical Physiology Lab

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

  • Adult

Accepts Healthy Volunteers

Yes

Sampling Method

Non-Probability Sample

Study Population

Healthy subjects ≥ 18 and < 51 years of age, with varying skin tones and diversity across ethnicity, race, gender, and sex assigned at birth.

Description

Inclusion Criteria:

  1. Subjects ≥ 18 and < 51 years of age
  2. Subject is willing and able to comply with study procedures and duration
  3. Subject is willing to sign an informed consent
  4. Subject weighs >40kg
  5. Subject is a non-smoker or has not smoked within 36 hours prior to the study
  6. Cleared same day health assessment form and health screening

Exclusion Criteria:

  1. Subject is considered as being morbidly obese (defined as BMI >39.5)
  2. Compromised circulation, injury, or physical malformation of fingers, hands, ears or forehead/skull or other sensor sites which would limit the ability to test sites needed for the study (Note: Certain malformations may still allow subjects to participate if the condition is noted and would not affect the sites utilized)
  3. Participants of childbearing potential who are pregnant, who are trying to get pregnant, or who have a urine test positive for pregnancy on the day of the study
  4. Subjects with COHb levels >3% as assessed by CO-Oximetry during the procedure
  5. tHb < 10 g/dl as assessed by CO-Oximetry during the procedure
  6. MetHb ≥ 2% as assessed by CO-Oximetry during the procedure
  7. Subjects with known respiratory conditions such as:

    1. uncontrolled / severe asthma
    2. flu or influenza type infection
    3. pneumonia / bronchitis
    4. shortness of breath / respiratory distress
    5. unresolved respiratory or lung surgery
    6. emphysema, COPD, lung disease
    7. recent COVID (last 2 months)
  8. Subjects with known heart or cardiovascular conditions such as:

    1. hypertension: systolic pressure >140mmHg, or diastolic pressure >90mmHg on 3 consecutive readings
    2. previous cardiovascular surgery
    3. chest pain (angina)
    4. heart rhythms other than a normal sinus rhythm or with respiratory sinus arrhythmia (reviewed during health screen)
    5. previous heart attack
    6. blocked artery
    7. unexplained shortness of breath
    8. congestive heart failure (CHF)
    9. history of stroke
    10. transient ischemic attack
    11. carotid artery disease
    12. myocardial ischemia
    13. myocardial infarction
    14. cardiomyopathy
    15. implantable active medical device such as pacemaker or automatic defibrillator
  9. Self-reported health conditions as identified in the Health Assessment Form

    1. diabetes
    2. uncontrolled thyroid disease
    3. kidney disease / chronic renal impairment
    4. history of seizures (except childhood febrile seizures)
    5. epilepsy
    6. history of unexplained syncope
    7. recent history of frequent migraine headaches
    8. recent symptomatic head injury, within the last 2 months
    9. cancer requiring chemotherapy, radiation, or current treatment
    10. subjects with known clotting disorders
    11. history of bleeding disorders or personal history of prolonged bleeding from injury
    12. history of blood clots
    13. hemophilia
    14. sickle cell trait or disease
    15. current use of blood thinner: prescription or daily use of aspirin
    16. subjects with Severe contact allergies to standard adhesives, latex or other materials found in pulse oximetry sensors, ECG electrodes, respiration monitor electrodes or other medical sensors
    17. participants with severe allergy to iodine (only applicable if iodine is used)
    18. subjects with prior or known severe allergies to lidocaine (or similar pharmacological agents, e.g. Novocain) or heparin
    19. arterial cannulation within the last 30 days prior to study date, (this may exclude only one radial artery site, left or right)
    20. history of clinically significant complications from previous arterial cannulation
    21. Failure of Allen's test
    22. unwillingness or inability to remove colored nail polish or colored artificial nails other than clear from test digits
    23. other known health condition, upon disclosure in Health Assessment form at PI's discretion

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

Cohorts and Interventions

Group / Cohort
Intervention / Treatment
Healthy Volunteers
All subjects who meet the inclusion criteria are enrolled into the test group and participate in data collection have the Philips FAST SpO2 with Nellcor Pulse Oximetry Sensors
Noninvasive pulse oximeter

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Evaluation of Saturation Accuracy
Time Frame: 2 - 3 hours
Performance will be determined by calculating the accuracy in terms of root mean square difference (ARMS) comparing the noninvasive oxygen saturation measurement (SpO2) to the arterial oxygen saturation (SaO2) value obtained from a reference blood sample.
2 - 3 hours

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Collaborators

Investigators

  • Principal Investigator: Randall M Clark, M.D., Medtronic Clinical Physiology Lab

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start (Actual)

March 11, 2025

Primary Completion (Actual)

April 11, 2025

Study Completion (Actual)

July 11, 2025

Study Registration Dates

First Submitted

February 28, 2025

First Submitted That Met QC Criteria

February 28, 2025

First Posted (Actual)

March 5, 2025

Study Record Updates

Last Update Posted (Actual)

July 25, 2025

Last Update Submitted That Met QC Criteria

July 23, 2025

Last Verified

January 1, 2025

More Information

Terms related to this study

Other Study ID Numbers

  • MDT24049NELFST

Drug and device information, study documents

Studies a U.S. FDA-regulated drug product

No

Studies a U.S. FDA-regulated device product

No

product manufactured in and exported from the U.S.

Yes

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.

Clinical Trials on Oxygen Saturation Measurement

Clinical Trials on Philips FAST SpO2 with Nellcor Pulse Oximetry Sensors

Subscribe