- ICH GCP
- US Clinical Trials Registry
- Clinical Trial NCT06860230
Philips FAST Evaluation (NELFST)
July 23, 2025 updated by: Medtronic - MITG
Evaluation of NellcorTM Pulse Oximetry Accuracy With the Philips picoSAT X SpO2 Board With Fourier Artifact Suppression Technology in Adult Volunteers
Evaluation of Philips FAST technology with Nellcor sensors
Study Overview
Status
Completed
Conditions
Intervention / Treatment
Detailed Description
Evaluate saturation accuracy claims for the Philips picoSAT X SpO2 board with FAST algorithm, paired with Nellcor™ sensors in a diverse subject population, over a specified saturation range
Study Type
Observational
Enrollment (Actual)
27
Contacts and Locations
This section provides the contact details for those conducting the study, and information on where this study is being conducted.
Study Locations
-
-
Colorado
-
Denver, Colorado, United States, 80218
- Medtronic Clinical Physiology Lab
-
-
Participation Criteria
Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.
Eligibility Criteria
Ages Eligible for Study
- Adult
Accepts Healthy Volunteers
Yes
Sampling Method
Non-Probability Sample
Study Population
Healthy subjects ≥ 18 and < 51 years of age, with varying skin tones and diversity across ethnicity, race, gender, and sex assigned at birth.
Description
Inclusion Criteria:
- Subjects ≥ 18 and < 51 years of age
- Subject is willing and able to comply with study procedures and duration
- Subject is willing to sign an informed consent
- Subject weighs >40kg
- Subject is a non-smoker or has not smoked within 36 hours prior to the study
- Cleared same day health assessment form and health screening
Exclusion Criteria:
- Subject is considered as being morbidly obese (defined as BMI >39.5)
- Compromised circulation, injury, or physical malformation of fingers, hands, ears or forehead/skull or other sensor sites which would limit the ability to test sites needed for the study (Note: Certain malformations may still allow subjects to participate if the condition is noted and would not affect the sites utilized)
- Participants of childbearing potential who are pregnant, who are trying to get pregnant, or who have a urine test positive for pregnancy on the day of the study
- Subjects with COHb levels >3% as assessed by CO-Oximetry during the procedure
- tHb < 10 g/dl as assessed by CO-Oximetry during the procedure
- MetHb ≥ 2% as assessed by CO-Oximetry during the procedure
Subjects with known respiratory conditions such as:
- uncontrolled / severe asthma
- flu or influenza type infection
- pneumonia / bronchitis
- shortness of breath / respiratory distress
- unresolved respiratory or lung surgery
- emphysema, COPD, lung disease
- recent COVID (last 2 months)
Subjects with known heart or cardiovascular conditions such as:
- hypertension: systolic pressure >140mmHg, or diastolic pressure >90mmHg on 3 consecutive readings
- previous cardiovascular surgery
- chest pain (angina)
- heart rhythms other than a normal sinus rhythm or with respiratory sinus arrhythmia (reviewed during health screen)
- previous heart attack
- blocked artery
- unexplained shortness of breath
- congestive heart failure (CHF)
- history of stroke
- transient ischemic attack
- carotid artery disease
- myocardial ischemia
- myocardial infarction
- cardiomyopathy
- implantable active medical device such as pacemaker or automatic defibrillator
Self-reported health conditions as identified in the Health Assessment Form
- diabetes
- uncontrolled thyroid disease
- kidney disease / chronic renal impairment
- history of seizures (except childhood febrile seizures)
- epilepsy
- history of unexplained syncope
- recent history of frequent migraine headaches
- recent symptomatic head injury, within the last 2 months
- cancer requiring chemotherapy, radiation, or current treatment
- subjects with known clotting disorders
- history of bleeding disorders or personal history of prolonged bleeding from injury
- history of blood clots
- hemophilia
- sickle cell trait or disease
- current use of blood thinner: prescription or daily use of aspirin
- subjects with Severe contact allergies to standard adhesives, latex or other materials found in pulse oximetry sensors, ECG electrodes, respiration monitor electrodes or other medical sensors
- participants with severe allergy to iodine (only applicable if iodine is used)
- subjects with prior or known severe allergies to lidocaine (or similar pharmacological agents, e.g. Novocain) or heparin
- arterial cannulation within the last 30 days prior to study date, (this may exclude only one radial artery site, left or right)
- history of clinically significant complications from previous arterial cannulation
- Failure of Allen's test
- unwillingness or inability to remove colored nail polish or colored artificial nails other than clear from test digits
- other known health condition, upon disclosure in Health Assessment form at PI's discretion
Study Plan
This section provides details of the study plan, including how the study is designed and what the study is measuring.
How is the study designed?
Design Details
Cohorts and Interventions
Group / Cohort |
Intervention / Treatment |
|---|---|
|
Healthy Volunteers
All subjects who meet the inclusion criteria are enrolled into the test group and participate in data collection have the Philips FAST SpO2 with Nellcor Pulse Oximetry Sensors
|
Noninvasive pulse oximeter
|
What is the study measuring?
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Evaluation of Saturation Accuracy
Time Frame: 2 - 3 hours
|
Performance will be determined by calculating the accuracy in terms of root mean square difference (ARMS) comparing the noninvasive oxygen saturation measurement (SpO2) to the arterial oxygen saturation (SaO2) value obtained from a reference blood sample.
|
2 - 3 hours
|
Collaborators and Investigators
This is where you will find people and organizations involved with this study.
Sponsor
Collaborators
Investigators
- Principal Investigator: Randall M Clark, M.D., Medtronic Clinical Physiology Lab
Study record dates
These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.
Study Major Dates
Study Start (Actual)
March 11, 2025
Primary Completion (Actual)
April 11, 2025
Study Completion (Actual)
July 11, 2025
Study Registration Dates
First Submitted
February 28, 2025
First Submitted That Met QC Criteria
February 28, 2025
First Posted (Actual)
March 5, 2025
Study Record Updates
Last Update Posted (Actual)
July 25, 2025
Last Update Submitted That Met QC Criteria
July 23, 2025
Last Verified
January 1, 2025
More Information
Terms related to this study
Keywords
Other Study ID Numbers
- MDT24049NELFST
Drug and device information, study documents
Studies a U.S. FDA-regulated drug product
No
Studies a U.S. FDA-regulated device product
No
product manufactured in and exported from the U.S.
Yes
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.
Clinical Trials on Oxygen Saturation Measurement
-
Brno University HospitalMasaryk UniversityNot yet recruitingOxygen Saturation MeasurementCzechia
-
Perin Health DevicesRecruitingOxygen Saturation MeasurementUnited States
-
Philips Clinical & Medical Affairs GlobalCompletedSpO2 (Peripheral Oxygen Saturation) Measurement ValidationUnited States
-
GE HealthcareNot yet recruitingSpO2 | SpO2 (Peripheral Oxygen Saturation) Measurement ValidationUnited States
-
Centre Hospitalier Régional d'OrléansRecruitingExtubation | Rox Index | Oxygen Saturation Measurement | Mechanical Ventilation With Oral Intubation | Weaning Invasive Mechanical VentilationFrance
-
Medtronic - MITGCompleted
-
GE HealthcareCompletedSpO2 | Peripheral Oxygen Saturation MeasurementUnited States
-
Sakarya UniversityNot yet recruiting
-
Medtronic - MITGNot yet recruitingOxygen Saturation
-
Medtronic - MITGCompletedOxygen SaturationUnited States
Clinical Trials on Philips FAST SpO2 with Nellcor Pulse Oximetry Sensors
-
Masimo CorporationCompleted
-
University of Alabama at BirminghamEunice Kennedy Shriver National Institute of Child Health and Human Development...RecruitingPulmonary Hypertension | Bronchopulmonary DysplasiaUnited States
-
Medtronic - MITGCompleted