- ICH GCP
- US Clinical Trials Registry
- Clinical Trial NCT06148623
Clinical Performance Testing of Philips FAST SpO2 With Masimo Pulse Oximetry Sensors Across Skin Pigmentation
Study Overview
Status
Conditions
Intervention / Treatment
Study Type
Enrollment (Actual)
Phase
- Not Applicable
Contacts and Locations
Study Locations
-
-
California
-
Irvine, California, United States, 92618
- Masimo Corporation
-
-
Participation Criteria
Eligibility Criteria
Ages Eligible for Study
- Adult
Accepts Healthy Volunteers
Description
Inclusion Criteria (Eligible Subjects):
- Subject is 18 to 50 years of age.
- Subject weighs a minimum of 110 lbs.
- Baseline heart rate ≥ 45 bpm and ≤ 90 bpm.
- Blood Pressure: Systolic BP ≤ 140 mmHg and ≥ 90 mmHg, Diastolic BP ≤ 90 mmHg and ≥ 50 mmHg, and if systolic BP is lower than 100 mmHg and/or diastolic BP is lower than 60 mmHg, subject passes an orthostatic blood pressure test.
- Hemoglobin value ≥ 11 g/dL.
- CO value ≤ 3.0% FCOHb.
- Subject can read and communicate in English and understands the study and the risks involved.
Exclusion Criteria (Ineligible Subjects) (* = per clinician discretion):
- Subject is pregnant or breastfeeding.
- Subject has a BMI > 35.
- Subject has a history of fainting (vasovagal syncope), blacking out or losing consciousness during or after a blood draw, or has a fear of blood draws.
- Subject has open wounds, inflamed tattoos, or piercings, and/or has any visible healing wounds that a medical professional determines may place them at an increased risk for participation. *
- Subject has finger deformities, nail deformities, nail polish, and/or gel/acrylic that can interfere with study device placement. *
- Subject has known drug or alcohol abuse.
- Subject uses recreational drugs. *
- Subject experiences frequent or severe headaches and/or migraine headaches, migraine auras, altitude sickness, and/or headaches accompanied by visual changes or sensitivity to light or sound.
- Subject has experienced a concussion or head injury with loss of consciousness within the past 12 months.
- Subject has any history of a stroke, myocardial infarction (heart attack), and/or seizures.
- Subject has any chronic bleeding disorder (e.g., hemophilia).
- Subject has taken anticoagulant medication within the past 30 days (excluding nonsteroidal anti-inflammatory drugs (NSAIDs)).
- Subject has donated blood within the past 4 weeks.
- Subject has any cardiac dysrhythmia (e.g., atrial fibrillation) and has not received clearance from their physician to participate.
- Subject has a known neurological and/or psychiatric disorder (e.g., schizophrenia, bipolar disorder, Multiple Sclerosis, Huntington's disease) that interferes with the subject's level of consciousness. *
- Subject has taken opioid pain medication 24 hours before the study.
- Subject has any infectious disease (e.g., Hepatitis, HIV, Tuberculosis, Flu, Malaria, Measles). *
- Subject is taking medications known to treat any type of infectious disease.
- Subject has either signs or history of peripheral ischemia or carpal tunnel syndrome.
- Subject has had invasive surgery within the past year, including but not limited to major dental surgery, appendectomy, plastic surgery, jaw surgery, major ENT surgery, major abdominal and/or pelvic surgery, heart surgery, or thoracic surgery. *
- Subject has symptoms of congestion, head cold, or other illnesses.
- Subject has been in a severe car accident(s) or a similar type of accident(s) requiring hospitalization within the past 12 months.
- Subject has any cancer or history of cancer (not including skin cancer). *
- Subject has chronic unresolved asthma, lung disease (including COPD) and/or respiratory disease.
- Subject is allergic to lidocaine, chlorhexidine, latex, adhesives, or plastic.
- Subject has a heart condition, insulin-dependent diabetes, uncontrolled hypertension, or hypercholesterolemia.
- Subject delivered vaginally, had a pregnancy terminated, had a miscarriage with hospitalization, or had a C-section within the past 6 months.
- Subject intends on participating in any heavy lifting, repetitive movement of their wrist (including riding a motorcycle, tennis), exercise (e.g., working out, riding a bike, riding a skateboard), or any activity that will put additional stress on the wrist within 24 hours following a study that involves an arterial line.
- Difficulty inserting an intravenous line in the subject's hand or arm and/or an arterial line in the radial artery of the subject's wrist.
- Subject has any medical condition which in the judgment of the investigator and/or medical staff, renders them ineligible for participation in this study or subject is deemed ineligible by the discretion of the investigator/study staff.
Study Plan
How is the study designed?
Design Details
- Primary Purpose: Other
- Allocation: N/A
- Interventional Model: Single Group Assignment
- Masking: None (Open Label)
Arms and Interventions
Participant Group / Arm |
Intervention / Treatment |
|---|---|
|
Experimental: Test Subject
All subjects who are enrolled into the test group and participate in data collection have the Philips FAST SpO2 with Masimo Pulse Oximetry Sensors
|
Noninvasive pulse oximeter
|
What is the study measuring?
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Oxygen Saturation Performance of Philips FAST SpO₂ With Masimo Pulse Oximeter Sensor (RD SET® Adt)
Time Frame: 1 hour
|
The SpO₂ accuracy of Philips FAST SpO₂ with Masimo Pulse Oximeter Sensor (RD SET® Adt) was determined by calculating the accuracy root mean-squared (ARMS) difference between the measured values (SpO₂) to the reference values (SaO₂) in accordance with ISO-80601-2-61. ARMS is a standard statistical method used in pulse oximetry performance evaluations to quantify the overall accuracy of a device. Arms= √(∑(i=1 to n) ((SpO₂_i-SaO₂_i )^2 ))/n. Participants in this arm could receive one or more sensor types used for data collection. |
1 hour
|
|
Oxygen Saturation Performance of Philips FAST SpO₂ With Masimo Pulse Oximeter Sensor (RD SET® Adt CS)
Time Frame: 1 hour
|
The SpO₂ accuracy of Philips FAST SpO₂ with Masimo Pulse Oximeter Sensor (RD SET® Adt CS) was determined by calculating the accuracy root mean-squared (ARMS) difference between the measured values (SpO₂) to the reference values (SaO₂) in accordance with ISO-80601-2-61. ARMS is a standard statistical method used in pulse oximetry performance evaluations to quantify the overall accuracy of a device. Arms= √(∑(i=1 to n) ((SpO₂_i-SaO₂_i )^2 ))/n. Participants in this arm could receive one or more sensor types used for data collection. |
1 hour
|
|
Oxygen Saturation Performance of Philips FAST SpO₂ With Masimo Pulse Oximeter Sensor (RD SET® YI)
Time Frame: 1 hour
|
The SpO₂ accuracy of Philips FAST SpO₂ with Masimo Pulse Oximeter Sensor (RD SET® YI) was determined by calculating the accuracy root mean-squared (ARMS) difference between the measured values (SpO₂) to the reference values (SaO₂) in accordance with ISO-80601-2-61. ARMS is a standard statistical method used in pulse oximetry performance evaluations to quantify the overall accuracy of a device. Arms= √(∑(i=1 to n) ((SpO₂_i-SaO₂_i )^2 ))/n. Participants in this arm could receive one or more sensor types used for data collection. |
1 hour
|
|
Oxygen Saturation Performance of Philips FAST SpO₂ With Masimo Pulse Oximeter Sensor (RD SET® DBI)
Time Frame: 1 hour
|
The SpO₂ accuracy of Philips FAST SpO₂ with Masimo Pulse Oximeter Sensor (RD SET® DBI) was determined by calculating the accuracy root mean-squared (ARMS) difference between the measured values (SpO₂) to the reference values (SaO₂) in accordance with ISO-80601-2-61. ARMS is a standard statistical method used in pulse oximetry performance evaluations to quantify the overall accuracy of a device. Arms= √(∑(i=1 to n) ((SpO₂_i-SaO₂_i )^2 ))/n. Participants in this arm could receive one or more sensor types used for data collection. |
1 hour
|
|
Oxygen Saturation Performance of Philips FAST SpO₂ With Masimo Pulse Oximeter Sensor (LNCS® Adtx)
Time Frame: 1 hour
|
The SpO₂ accuracy of Philips FAST SpO₂ with Masimo Pulse Oximeter Sensor (LNCS® Adtx) was determined by calculating the accuracy root mean-squared (ARMS) difference between the measured values (SpO₂) to the reference values (SaO₂) in accordance with ISO-80601-2-61. ARMS is a standard statistical method used in pulse oximetry performance evaluations to quantify the overall accuracy of a device. Arms= √(∑(i=1 to n) ((SpO₂_i-SaO₂_i )^2 ))/n. Participants in this arm could receive one or more sensor types used for data collection. |
1 hour
|
|
Oxygen Saturation Performance of Philips FAST SpO₂ With Masimo Pulse Oximeter Sensor (RD SET® TC-I)
Time Frame: 1 hour
|
The SpO₂ accuracy of Philips FAST SpO₂ with Masimo Pulse Oximeter Sensor (RD SET® TC-I) was determined by calculating the accuracy root mean-squared (ARMS) difference between the measured values (SpO₂) to the reference values (SaO₂) in accordance with ISO-80601-2-61. ARMS is a standard statistical method used in pulse oximetry performance evaluations to quantify the overall accuracy of a device. Arms= √(∑(i=1 to n) ((SpO₂_i-SaO₂_i )^2 ))/n. Participants in this arm could receive one or more sensor types used for data collection. |
1 hour
|
|
Oxygen Saturation Performance of Philips FAST SpO₂ With Masimo Pulse Oximeter Sensor (RD SET® DCI)
Time Frame: 1 hour
|
The SpO₂ accuracy of Philips FAST SpO₂ with Masimo Pulse Oximeter Sensor (RD SET® DCI) was determined by calculating the accuracy root mean-squared (ARMS) difference between the measured values (SpO₂) to the reference values (SaO₂) in accordance with ISO-80601-2-61. ARMS is a standard statistical method used in pulse oximetry performance evaluations to quantify the overall accuracy of a device. Arms= √(∑(i=1 to n) ((SpO₂_i-SaO₂_i )^2 ))/n. Participants in this arm could receive one or more sensor types used for data collection. |
1 hour
|
Secondary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Difference in oxygen saturation performance between three population subgroups
Time Frame: 1-3 hours
|
Average root mean square (ARMS) will be computed for each pigmentation subgroup for light, medium, and dark skin pigmentation and the difference between the subgroup ARMS will be calculated.
|
1-3 hours
|
Collaborators and Investigators
Sponsor
Study record dates
Study Major Dates
Study Start (Actual)
Primary Completion (Actual)
Study Completion (Actual)
Study Registration Dates
First Submitted
First Submitted That Met QC Criteria
First Posted (Actual)
Study Record Updates
Last Update Posted (Actual)
Last Update Submitted That Met QC Criteria
Last Verified
More Information
Terms related to this study
Other Study ID Numbers
- CIP-1088
Drug and device information, study documents
Studies a U.S. FDA-regulated drug product
Studies a U.S. FDA-regulated device product
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.
Clinical Trials on Healthy
-
University of Vermont Medical CenterAvocado Nutrition CenterRecruitingHealthy | Healthy Volunteers | Healthy Subjects | Healthy Volunteer | Healthy Adult | Healthy Volunteers Only | Healthy Male and Female Subjects | Healthy Non-smokersUnited States
-
Dragonfly TherapeuticsRecruitingHealthy | Healthy Participants | Healthy Adult Females | Volunteer | Healthy Adult MaleAustralia
-
Umm Al-Qura UniversityActive, not recruitingHealthy | Healthy Participants | Healthy Adult | Healthy Women | Healthy Adult Females | Healthy Adult Participants | Healthy Young Adults | Healthy Adult Female Participants | Healthy Adult Male | Poor Sleep Quality | Healthy (Controls) | Poor Sleeping Quality | Healthy Adult Male Subjects | Health Adult SubjectsSaudi Arabia
-
University of PalermoCompletedHealthy | Healthy Volunteers | Healthy Subjects | Healthy Participants | Static Stretching | Stretch | StretchingItaly
-
Prevent Age Resort "Pervaya Liniya"RecruitingHealthy Aging | Healthy Diet | Healthy LifestyleRussian Federation
-
University of PalermoCompletedHealthy Participants | Healthy Adult Participants | Healthy Young AdultsItaly
-
Maastricht University Medical CenterCompletedHealthy Volunteers | Healthy Subjects | Healthy AdultsNetherlands
-
Yale UniversityNot yet recruitingHealth-related Benefits of Introducing Table Olives Into the Diet of Young Adults: Olives For HealthHealthy Diet | Healthy Lifestyle | Healthy Nutrition | CholesterolUnited States
-
PfizerNot yet recruitingHealthy | Healthy AdultsUnited States
-
Atisama TherapeuticsRecruitingHealthy | Healthy SmokerAustralia
Clinical Trials on Philips FAST SpO2 with Masimo Pulse Oximetry Sensors
-
Medtronic - MITGPhilips Medical SystemsCompleted
-
Masimo CorporationRecruiting