Diagnostic Accuracy of A Diquat Quantitative Detection Kit and A Portable Mass Spectrometry System for Quantifying Diquat Concentrations in Human Blood Samples

Accuracy, Safety, and Clinical Performance of a Diquat Quantitative Detection Kit (In-Situ Ionization Mass Spectrometry) and a Portable Mass Spectrometry System for Quantifying Diquat Concentrations in Human Blood Samples (Whole Blood/Plasma)

This is an observational, non-interventional diagnostic accuracy study designed to evaluate a diquat quantitative detection kit (ambient ionization mass spectrometry method) and a portable mass spectrometry analysis system for measuring diquat concentrations in human blood samples (whole blood/plasma), using LC-MS/MS as the clinical gold standard for comparison.

Study Overview

Study Type

Observational

Enrollment (Estimated)

60

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Contact

Study Locations

    • Jiangsu
      • Nanjing, Jiangsu, China, 210008
        • Nanjing Drum Tower Hospital
        • Contact:

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

  • Child
  • Adult
  • Older Adult

Accepts Healthy Volunteers

No

Sampling Method

Non-Probability Sample

Study Population

Participants will be recruited from patients presenting to the emergency department during the study period. The study population includes individuals with suspected or clinically diagnosed acute diquat poisoning, stratified into low, medium, and high concentration groups (15 participants per group), as well as a negative control group consisting of non-diquat-poisoning patients (15 participants), for a total enrollment of 60 participants.

Description

Inclusion Criteria:

  1. Patients with suspected or clinically diagnosed acute diquat poisoning, providing whole blood and/or plasma samples, including qualified residual specimens retained after prior clinical testing when available.
  2. The participant or their legally authorized representative can fully understand the study purpose and procedures, voluntarily agrees to participate, and is willing and able to comply with the study requirements.
  3. Sample collection is performed according to routine clinical standards, with no apparent ethical concerns related to sample acquisition.

Exclusion Criteria:

  1. Abnormal sample appearance, such as visible flocculent material or other gross abnormalities.
  2. The participant is unable to provide a specimen, or the specimen does not meet testing requirements.
  3. Any participant considered inappropriate for inclusion by the investigator.

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

Cohorts and Interventions

Group / Cohort
Intervention / Treatment
Diquat Poisoning Group (Low Concentration)
Participants presenting to the emergency department with clinically diagnosed acute diquat poisoning whose diquat concentration is classified into the low concentration stratum.
The portable mass spectrometry analysis system is an in vitro diagnostic device used with a diquat quantitative detection kit based on an in-situ/ambient ionization mass spectrometry method to quantify diquat concentrations in human blood samples. Whole blood specimens collected in routine clinical care will be analyzed using this device, and the quantitative results will be compared against those obtained using the reference standard method, liquid chromatography-tandem mass spectrometry (LC-MS/MS), to evaluate analytical accuracy and agreement. Each specimen will be tested repeatedly (three measurements per sample), and the mean value will be used for statistical analysis. The study is observational and non-interventional, and test results generated by the portable mass spectrometry system are used for research evaluation purposes and do not alter routine clinical diagnosis or treatment decisions
Diquat Poisoning Group (Medium Concentration)
Participants presenting to the emergency department with clinically diagnosed acute diquat poisoning whose diquat concentration is classified into the medium concentration stratum.
The portable mass spectrometry analysis system is an in vitro diagnostic device used with a diquat quantitative detection kit based on an in-situ/ambient ionization mass spectrometry method to quantify diquat concentrations in human blood samples. Whole blood specimens collected in routine clinical care will be analyzed using this device, and the quantitative results will be compared against those obtained using the reference standard method, liquid chromatography-tandem mass spectrometry (LC-MS/MS), to evaluate analytical accuracy and agreement. Each specimen will be tested repeatedly (three measurements per sample), and the mean value will be used for statistical analysis. The study is observational and non-interventional, and test results generated by the portable mass spectrometry system are used for research evaluation purposes and do not alter routine clinical diagnosis or treatment decisions
Diquat Poisoning Group (High Concentration)
Participants presenting to the emergency department with clinically diagnosed acute diquat poisoning whose diquat concentration is classified into the high concentration stratum.
The portable mass spectrometry analysis system is an in vitro diagnostic device used with a diquat quantitative detection kit based on an in-situ/ambient ionization mass spectrometry method to quantify diquat concentrations in human blood samples. Whole blood specimens collected in routine clinical care will be analyzed using this device, and the quantitative results will be compared against those obtained using the reference standard method, liquid chromatography-tandem mass spectrometry (LC-MS/MS), to evaluate analytical accuracy and agreement. Each specimen will be tested repeatedly (three measurements per sample), and the mean value will be used for statistical analysis. The study is observational and non-interventional, and test results generated by the portable mass spectrometry system are used for research evaluation purposes and do not alter routine clinical diagnosis or treatment decisions
Non-Diquat Poisoning Group (Negative Control)
Participants presenting to the emergency department who are not diagnosed with acute diquat poisoning will be enrolled as negative controls. Their blood samples will be tested in parallel to support evaluation of method performance and to confirm negative results in non-diquat cases.
The portable mass spectrometry analysis system is an in vitro diagnostic device used with a diquat quantitative detection kit based on an in-situ/ambient ionization mass spectrometry method to quantify diquat concentrations in human blood samples. Whole blood specimens collected in routine clinical care will be analyzed using this device, and the quantitative results will be compared against those obtained using the reference standard method, liquid chromatography-tandem mass spectrometry (LC-MS/MS), to evaluate analytical accuracy and agreement. Each specimen will be tested repeatedly (three measurements per sample), and the mean value will be used for statistical analysis. The study is observational and non-interventional, and test results generated by the portable mass spectrometry system are used for research evaluation purposes and do not alter routine clinical diagnosis or treatment decisions

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Bland-Altman agreement
Time Frame: Baseline
Assesses agreement and bias between the two methods; requires most data points within preset LOA
Baseline
Correlation coefficient between portable MS and LC-MS/MS
Time Frame: Baseline
Measures linear association of quantitative results; target orrelation coefficient ≥ 0.95
Baseline
Relative deviation at medical decision levels
Time Frame: Baseline
Evaluates clinically acceptable error; target relative deviatio within ±15%
Baseline

Secondary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Recovery rate (spiked mixed samples vs fresh samples)
Time Frame: Baseline
Target recovery 85%-115%
Baseline
Precision (Coefficient of Variation, CV)
Time Frame: Baseline
Within-run CV ≤ 10%; between-run CV ≤ 15%
Baseline

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start (Estimated)

September 1, 2026

Primary Completion (Estimated)

December 31, 2027

Study Completion (Estimated)

December 31, 2028

Study Registration Dates

First Submitted

January 26, 2026

First Submitted That Met QC Criteria

February 2, 2026

First Posted (Actual)

February 10, 2026

Study Record Updates

Last Update Posted (Actual)

May 27, 2026

Last Update Submitted That Met QC Criteria

May 24, 2026

Last Verified

February 1, 2026

More Information

Terms related to this study

Other Study ID Numbers

  • 2025-1207-02

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.