3D Versus Low-Temperature Thermoplastic Relative Motion Orthosis: MP Joint Stabilization

3 de noviembre de 2026 actualizado por: Enver Guven

Relative motion flexion (RMF) orthoses are used in hand rehabilitation to position finger joints in a controlled degree of flexion. These orthoses are traditionally made from low-temperature thermoplastic materials. Three-dimensional (3D) printing offers an alternative method of producing individually designed orthoses using different filament materials.

The purpose of this study was to compare finger joint stabilization and user satisfaction among RMF orthoses made from two 3D-printing materials, thermoplastic polyurethane (TPU) and polylactic acid (PLA), and a conventional low-temperature thermoplastic (LTT) material.

The study included 30 healthy adult volunteers. Each participant used all three types of RMF orthoses for two consecutive days per orthosis, with the order of application randomly determined. The orthoses were designed to position the metacarpophalangeal (MCP) joint at 30 degrees of flexion.

Joint positioning was evaluated using a goniometer, and user satisfaction was assessed using the Quebec User Evaluation of Satisfaction with Assistive Technology (QUEST 2.0).

The study aimed to provide comparative information about orthosis materials and fabrication methods that may help guide future orthotic applications in hand rehabilitation.

Descripción general del estudio

Estado

Terminado

Condiciones

Intervención / Tratamiento

Tipo de estudio

Intervencionista

Inscripción (Actual)

30

Fase

  • No aplica

Contactos y Ubicaciones

Esta sección proporciona los datos de contacto de quienes realizan el estudio e información sobre dónde se lleva a cabo este estudio.

Ubicaciones de estudio

      • Ankara, Turquía (Türkiye)
        • Ankara University, Faculty of Health Sciences

Criterios de participación

Los investigadores buscan personas que se ajusten a una determinada descripción, denominada criterio de elegibilidad. Algunos ejemplos de estos criterios son el estado de salud general de una persona o tratamientos previos.

Criterio de elegibilidad

Edades elegibles para estudiar

18 años a 65 años (Adulto, Adulto Mayor)

Acepta Voluntarios Saludables

Sí

Géneros elegibles para el estudio

Todos

Descripción

Inclusion Criteria:

  • Healthy adult volunteers.
  • No orthopedic or neurological impairments involving the upper extremities.
  • No reported history of systemic illness or previous surgical interventions.
  • Provision of written informed consent before enrollment.

Exclusion Criteria:

  • Current upper extremity pain.
  • Joint deformities.
  • Skin integrity problems.
  • Any condition that could interfere with RMF orthosis application or wearing.

Plan de estudios

Esta sección proporciona detalles del plan de estudio, incluido cómo está diseñado el estudio y qué mide el estudio.

¿Cómo está diseñado el estudio?

Detalles de diseño

  • Propósito principal: Otro
  • Asignación: Aleatorizado
  • Modelo Intervencionista: Asignación cruzada
  • Enmascaramiento: Ninguno (etiqueta abierta)

Armas e Intervenciones

Grupo de participantes/brazo
Intervención / Tratamiento
Experimental: TPU-PLA-LTT

Participants received three RMF orthoses in the following order: 3D-printed TPU, 3D-printed PLA, and conventional LTT. Each orthosis was worn for two consecutive days.

A custom relative motion flexion orthosis fabricated using thermoplastic polyurethane (TPU) filament. Individual hand geometry was obtained by 3D scanning, and the orthosis was digitally designed using Meshmixer software and manufactured using a Creality CR-10 Max 3D printer. The orthosis positioned the third metacarpophalangeal joint at 30 degrees of flexion and was worn for two consecutive days.

Açıklamayı kopyala

A custom relative motion flexion orthosis fabricated using polylactic acid (PLA) filament. Individual hand geometry was obtained by 3D scanning, and the orthosis was digitally designed using Meshmixer software and manufactured using a Creality CR-10 Max 3D printer. The orthosis positioned the third metacarpophalangeal joint at 30 degrees of flexion and was worn for two consecutive days.

A conventional relative motion flexion orthosis fabricated using Orficast More low-temperature thermoplastic material (Orfit Industries, Belgium). The orthosis was manually molded directly on the participant's hand, positioning the third metacarpophalangeal joint at 30 degrees of flexion, verified using a goniometer. The orthosis was worn for two consecutive days.

Experimental: TPU-LTT-PLA

Participants received three RMF orthoses in the following order: 3D-printed TPU, conventional LTT, and 3D-printed PLA. Each orthosis was worn for two consecutive days.

A custom relative motion flexion orthosis fabricated using thermoplastic polyurethane (TPU) filament. Individual hand geometry was obtained by 3D scanning, and the orthosis was digitally designed using Meshmixer software and manufactured using a Creality CR-10 Max 3D printer. The orthosis positioned the third metacarpophalangeal joint at 30 degrees of flexion and was worn for two consecutive days.

Açıklamayı kopyala

A custom relative motion flexion orthosis fabricated using polylactic acid (PLA) filament. Individual hand geometry was obtained by 3D scanning, and the orthosis was digitally designed using Meshmixer software and manufactured using a Creality CR-10 Max 3D printer. The orthosis positioned the third metacarpophalangeal joint at 30 degrees of flexion and was worn for two consecutive days.

A conventional relative motion flexion orthosis fabricated using Orficast More low-temperature thermoplastic material (Orfit Industries, Belgium). The orthosis was manually molded directly on the participant's hand, positioning the third metacarpophalangeal joint at 30 degrees of flexion, verified using a goniometer. The orthosis was worn for two consecutive days.

Experimental: PLA-TPU-LTT

Participants received three RMF orthoses in the following order: 3D-printed PLA, 3D-printed TPU, and conventional LTT. Each orthosis was worn for two consecutive days.

Açıklamayı kopyala

A custom relative motion flexion orthosis fabricated using thermoplastic polyurethane (TPU) filament. Individual hand geometry was obtained by 3D scanning, and the orthosis was digitally designed using Meshmixer software and manufactured using a Creality CR-10 Max 3D printer. The orthosis positioned the third metacarpophalangeal joint at 30 degrees of flexion and was worn for two consecutive days.

Açıklamayı kopyala

A custom relative motion flexion orthosis fabricated using polylactic acid (PLA) filament. Individual hand geometry was obtained by 3D scanning, and the orthosis was digitally designed using Meshmixer software and manufactured using a Creality CR-10 Max 3D printer. The orthosis positioned the third metacarpophalangeal joint at 30 degrees of flexion and was worn for two consecutive days.

A conventional relative motion flexion orthosis fabricated using Orficast More low-temperature thermoplastic material (Orfit Industries, Belgium). The orthosis was manually molded directly on the participant's hand, positioning the third metacarpophalangeal joint at 30 degrees of flexion, verified using a goniometer. The orthosis was worn for two consecutive days.

Experimental: PLA-LTT-TPU

Participants received three RMF orthoses in the following order: 3D-printed PLA, conventional LTT, and 3D-printed TPU. Each orthosis was worn for two consecutive days.

A custom relative motion flexion orthosis fabricated using thermoplastic polyurethane (TPU) filament. Individual hand geometry was obtained by 3D scanning, and the orthosis was digitally designed using Meshmixer software and manufactured using a Creality CR-10 Max 3D printer. The orthosis positioned the third metacarpophalangeal joint at 30 degrees of flexion and was worn for two consecutive days.

Açıklamayı kopyala

A custom relative motion flexion orthosis fabricated using polylactic acid (PLA) filament. Individual hand geometry was obtained by 3D scanning, and the orthosis was digitally designed using Meshmixer software and manufactured using a Creality CR-10 Max 3D printer. The orthosis positioned the third metacarpophalangeal joint at 30 degrees of flexion and was worn for two consecutive days.

A conventional relative motion flexion orthosis fabricated using Orficast More low-temperature thermoplastic material (Orfit Industries, Belgium). The orthosis was manually molded directly on the participant's hand, positioning the third metacarpophalangeal joint at 30 degrees of flexion, verified using a goniometer. The orthosis was worn for two consecutive days.

Experimental: LTT-TPU-PLA

Participants received three RMF orthoses in the following order: conventional LTT, 3D-printed TPU, and 3D-printed PLA. Each orthosis was worn for two consecutive days.

A custom relative motion flexion orthosis fabricated using thermoplastic polyurethane (TPU) filament. Individual hand geometry was obtained by 3D scanning, and the orthosis was digitally designed using Meshmixer software and manufactured using a Creality CR-10 Max 3D printer. The orthosis positioned the third metacarpophalangeal joint at 30 degrees of flexion and was worn for two consecutive days.

Açıklamayı kopyala

A custom relative motion flexion orthosis fabricated using polylactic acid (PLA) filament. Individual hand geometry was obtained by 3D scanning, and the orthosis was digitally designed using Meshmixer software and manufactured using a Creality CR-10 Max 3D printer. The orthosis positioned the third metacarpophalangeal joint at 30 degrees of flexion and was worn for two consecutive days.

A conventional relative motion flexion orthosis fabricated using Orficast More low-temperature thermoplastic material (Orfit Industries, Belgium). The orthosis was manually molded directly on the participant's hand, positioning the third metacarpophalangeal joint at 30 degrees of flexion, verified using a goniometer. The orthosis was worn for two consecutive days.

Experimental: LTT-PLA-TPU

Participants received three RMF orthoses in the following order: conventional LTT, 3D-printed PLA, and 3D-printed TPU. Each orthosis was worn for two consecutive days.

Açıklamayı kopyala

A custom relative motion flexion orthosis fabricated using thermoplastic polyurethane (TPU) filament. Individual hand geometry was obtained by 3D scanning, and the orthosis was digitally designed using Meshmixer software and manufactured using a Creality CR-10 Max 3D printer. The orthosis positioned the third metacarpophalangeal joint at 30 degrees of flexion and was worn for two consecutive days.

Açıklamayı kopyala

A custom relative motion flexion orthosis fabricated using polylactic acid (PLA) filament. Individual hand geometry was obtained by 3D scanning, and the orthosis was digitally designed using Meshmixer software and manufactured using a Creality CR-10 Max 3D printer. The orthosis positioned the third metacarpophalangeal joint at 30 degrees of flexion and was worn for two consecutive days.

A conventional relative motion flexion orthosis fabricated using Orficast More low-temperature thermoplastic material (Orfit Industries, Belgium). The orthosis was manually molded directly on the participant's hand, positioning the third metacarpophalangeal joint at 30 degrees of flexion, verified using a goniometer. The orthosis was worn for two consecutive days.

¿Qué mide el estudio?

Medidas de resultado primarias

Medida de resultado
Medida Descripción
Periodo de tiempo
MCP Joint Stabilization
Periodo de tiempo: At the end of each 2-day orthosis wearing period (3 assessments per participant).

The metacarpophalangeal (MCP) joint angle of the third digit of the dominant hand was measured in degrees using a goniometer after each orthosis wearing period. All orthoses were designed to maintain the MCP joint at 30 degrees of flexion. The measured joint angles were compared among three relative motion flexion (RMF) orthoses fabricated from 3D-printed thermoplastic polyurethane (TPU), 3D-printed polylactic acid (PLA), and conventional low-temperature thermoplastic (LTT). Values closer to 30 degrees indicated more accurate maintenance of the target joint position.

At the end of each 2-day orthosis wearing period (3 assessments per participant).

Medidas de resultado secundarias

Medida de resultado
Medida Descripción
Periodo de tiempo
QUEST 2.0 Total Satisfaction Score
Periodo de tiempo: At the end of each 2-day orthosis wearing period (3 assessments per participant).

Overall user satisfaction was assessed using the Quebec User Evaluation of Satisfaction with Assistive Technology (QUEST 2.0). The questionnaire includes 12 items covering device-related satisfaction (8 items) and service-related satisfaction (4 items). Each item is rated on a 5-point Likert scale, with higher scores indicating greater satisfaction. Total satisfaction scores were compared across TPU, PLA, and LTT orthoses.

At the end of each 2-day orthosis wearing period (3 assessments per participant).
QUEST 2.0 Device Satisfaction Score
Periodo de tiempo: At the end of each 2-day orthosis wearing period (3 assessments per participant).

Device-related satisfaction was assessed using the eight device items of the QUEST 2.0 questionnaire: dimensions, weight, adjustability, safety, durability, ease of use, comfort, and effectiveness. Each item is rated on a 5-point Likert scale, with higher scores indicating greater satisfaction. Device satisfaction scores were compared among TPU, PLA, and LTT orthoses.

At the end of each 2-day orthosis wearing period (3 assessments per participant).
QUEST 2.0 Service Satisfaction Score
Periodo de tiempo: At the end of each 2-day orthosis wearing period (3 assessments per participant).

Service-related satisfaction was assessed using the four service items of the QUEST 2.0 questionnaire, covering maintenance, repair, professional services, and follow-up. Each item is rated on a 5-point Likert scale, with higher scores indicating greater satisfaction. Service satisfaction scores were compared among TPU, PLA, and LTT orthoses.

At the end of each 2-day orthosis wearing period (3 assessments per participant).

Colaboradores e Investigadores

Aquí es donde encontrará personas y organizaciones involucradas en este estudio.

Patrocinador

Fechas de registro del estudio

Estas fechas rastrean el progreso del registro del estudio y los envíos de resultados resumidos a ClinicalTrials.gov. Los registros del estudio y los resultados informados son revisados ​​por la Biblioteca Nacional de Medicina (NLM) para asegurarse de que cumplan con los estándares de control de calidad específicos antes de publicarlos en el sitio web público.

Fechas importantes del estudio

Inicio del estudio (Actual)

12 de junio de 2024

Finalización primaria (Actual)

21 de febrero de 2025

Finalización del estudio (Actual)

27 de febrero de 2025

Fechas de registro del estudio

Enviado por primera vez

3 de noviembre de 2026

Primero enviado que cumplió con los criterios de control de calidad

3 de noviembre de 2026

Publicado por primera vez (Actual)

9 de noviembre de 2026

Actualizaciones de registros de estudio

Última actualización publicada (Actual)

9 de noviembre de 2026

Última actualización enviada que cumplió con los criterios de control de calidad

3 de noviembre de 2026

Última verificación

1 de octubre de 2026

Más información

Términos relacionados con este estudio

Otros números de identificación del estudio

  • AU-RMF-2023-620

Plan de datos de participantes individuales (IPD)

¿Planea compartir datos de participantes individuales (IPD)?

NO

Información sobre medicamentos y dispositivos, documentos del estudio

Estudia un producto farmacéutico regulado por la FDA de EE. UU.

No

Estudia un producto de dispositivo regulado por la FDA de EE. UU.

No

Esta información se obtuvo directamente del sitio web clinicaltrials.gov sin cambios. Si tiene alguna solicitud para cambiar, eliminar o actualizar los detalles de su estudio, comuníquese con register@clinicaltrials.gov. Tan pronto como se implemente un cambio en clinicaltrials.gov, también se actualizará automáticamente en nuestro sitio web. .