A Comparison of the Detection of Hypoventilation During Deep Sedation Utilizing Nasal End Title CO@ Versus Transcutaneous CO2 Measurement Techniques
The estimation of the partial pressure of carbon dioxide (PCO2) in the arterial blood is used to judge the adequacy of ventilation during spontaneous and controlled ventilation. Although the gold standard for monitoring PCO2 remains arterial blood gas sampling, this requires an invasive procedure and provides only an intermittent estimate of what is frequently a continuously changing value. The solution to this problem has been the development and validation of accurate noninvasive monitoring techniques which provide a continuous intraoperative estimate of PCO2.The most commonly used noninvasive technique to monitor PCO2 is measurement of the end tidal CO2 (PECO2) . However, sampling errors and patient -related issues such as ventilation-perfusion mismatch, patient positioning or decreases in pulmonary blood flow may influence the accuracy of PECO2 monitoring (1-3). Nasal capnography has been proved to be an accurate monitor during the post-operative period (4) but its ability to accurately detect hypoventilation associated with deep sedation has not been studied.
According to the American Society of Anesthesiologist standards for basic monitoring, continuous capnography is required for all patients undergoing general anesthesia but it is optional for MAC/sedation cases. The need for CO2 monitoring has been studied by other medical specialties that use procedural sedation, including gastroenterology (12) and emergency medicine (13, 14) and many specialties now recommend capnography as a standard monitor.
Patients receiving supplemental oxygen may experience significant persistent hypoventilation leading to progressive hypercarbia and acidosis which may go undetected for a significant time interval since the routinely monitored SpO2 may be maintained within normal range. A recent study has shown that despite the fact that end tidal CO2 is reliable in detecting apnea , increasing oxygen flow rates decrease the amplitude of measured CO2, probably via dilution, making the quantitative value less reliable as an assessment of adequacy of ventilation (15). Furthermore, during hypoventilation there is reduced alveolar ventilation and the end tidal CO2 is not a true reflection of arterial CO2.
Transcutaneous measurement of PCO2(PtcCO2) is a non-invasive method of measuring PCO2 that has been used much less frequently due to technical difficulties with earlier transcutaneous electrodes. Preliminary studies of the reliability of the current PtcCO2 electrodes (TOSCA, Linde Medical Sensors, and Basel, Switzerland) have shown good correlation of arterial and transcutaneous measurements in both adult volunteers and anesthetized subjects (5). PtcCO2 is measured with a sensor attached by a low pressure clip to an earlobe. The sensor probe heats the earlobe to 42 degrees Celsius to enhance blood flow. The current sensors have also been evaluated in anesthetized children (7, 8), anesthetized adults (9, 10) and critically ill neonates (11) and all these studies revealed a good correlation between PtCO2 and PaCO2.
調査の概要
詳細な説明
A group of _40 patients scheduled to undergo Hysteroscopy surgery with deep sedation will be studied. Standard monitors will be utilized. Additionally a transcutaneous CO2 electrode will be attached to the ear lobe and a BIS monitor will be applied to the forehead.
The anesthetic management will be left to the discretion of the anesthesia provider (Anesthesia resident or CRNA, under the supervision of a faculty attending anesthesiologist).All patients will receive supplemental oxygen at 3L/ minute via nasal cannula, to maintain oxygen saturation greater than 90%.
Both the end tidal CO2 (ETCO2) and transcutaneous CO2 (TcCO2) (Linde Medical Sensors, Basel , Switzerland ) will be calibrated before use, according to the manufacturer's instruction.. The anesthesia provider will be blinded to the TcCO2 monitor but will have access to the ETCO2 values. The ETCO2 and TcCO2 values will be recorded simultaneously by an independent observer.
One arterial blood samples will be collected from each subject for measurement of PaCO2 at a deep level of sedation defined by a modified Ramsey score greater or equal to 5 (1= anxious/agitated/restless; 2= cooperative/oriented/tranquil; 3= drowsy/responds to commands only; 4= brisk response to shaking/loud sound; 5= sluggish response to shaking/loud sound; 6= no response) The ETCO2 and .TcCO2 will be recorded throughout the time of obtaining the blood sample every 15 seconds and the mean value will be used for comparison. Hypoventilation will be defined as a PaCO2 level greater than 45mmHg.
Sedation score (OASS) will be recorded on admission to recovery area. Time to meet discharge criteria will be recorded.
The experimental procedures: ABG draw, ETCO2 monitor placement on the ear, 2 Ramey scale assessments.
研究の種類
入学 (実際)
段階
- 適用できない
連絡先と場所
研究場所
-
-
Illinois
-
Chicago、Illinois、アメリカ、60611
- Northwestern University
-
-
参加基準
適格基準
就学可能な年齢
健康ボランティアの受け入れ
受講資格のある性別
説明
Inclusion Criteria:
- Age>18 years of age
- Sex: Non-pregnant, Female
- ASA PS: I, II
- Surgery: Elective Hysteroscopy surgery
- Consent: Obtained
Exclusion Criteria:
- Age < 18 years old
- Patients who refuses participation
- History of lung disease
- History of Obstructive sleep apnea
研究計画
研究はどのように設計されていますか?
デザインの詳細
- 主な目的:診断
- 割り当て:非ランダム化
- 介入モデル:単一グループの割り当て
- マスキング:なし(オープンラベル)
武器と介入
参加者グループ / アーム |
介入・治療 |
---|---|
実験的:All participants
All participants, one arterial blood draw
|
One arterial blood draw
|
この研究は何を測定していますか?
主要な結果の測定
結果測定 |
メジャーの説明 |
時間枠 |
---|---|---|
TcCo2 vs PACo2 Difference
時間枠:1 hour
|
Evaluate the correlation between PaCO2- TcCO2 in detecting hypoventilation for patients undergoing deep sedation Absolute mean difference between TcCo2 and the PA Co2
|
1 hour
|
協力者と研究者
スポンサー
捜査官
- スタディディレクター:Robert McCarthy, PharmD、Northwestern University
出版物と役立つリンク
一般刊行物
- Whitesell R, Asiddao C, Gollman D, Jablonski J. Relationship between arterial and peak expired carbon dioxide pressure during anesthesia and factors influencing the difference. Anesth Analg. 1981 Jul;60(7):508-12.
- Yamanaka MK, Sue DY. Comparison of arterial-end-tidal PCO2 difference and dead space/tidal volume ratio in respiratory failure. Chest. 1987 Nov;92(5):832-5. doi: 10.1378/chest.92.5.832.
- Grenier B, Verchere E, Mesli A, Dubreuil M, Siao D, Vandendriessche M, Cales J, Maurette P. Capnography monitoring during neurosurgery: reliability in relation to various intraoperative positions. Anesth Analg. 1999 Jan;88(1):43-8. doi: 10.1097/00000539-199901000-00009.
研究記録日
主要日程の研究
研究開始
一次修了 (実際)
研究の完了 (実際)
試験登録日
最初に提出
QC基準を満たした最初の提出物
最初の投稿 (見積もり)
学習記録の更新
投稿された最後の更新 (見積もり)
QC基準を満たした最後の更新が送信されました
最終確認日
詳しくは
この情報は、Web サイト clinicaltrials.gov から変更なしで直接取得したものです。研究の詳細を変更、削除、または更新するリクエストがある場合は、register@clinicaltrials.gov。 までご連絡ください。 clinicaltrials.gov に変更が加えられるとすぐに、ウェブサイトでも自動的に更新されます。
手術の臨床試験
-
Dong Yangわからない腹腔鏡補助下手術 | 切開を行わない全腹腔鏡手術(Natural Orifice Transluminal Endoscopic Surgery、NOSES)中国
Arterial blood drawの臨床試験
-
University of UtahAlbert Einstein College of Medicine; University of California, San Francisco; National Human Genome... と他の協力者完了新生児スクリーニング
-
Ischemia Care LLC完了虚血性脳卒中 | 心房細動 | 血栓性脳卒中 | 一過性脳虚血発作 | 心塞栓性脳卒中 | 脳底動脈の脳卒中 | 一過性脳血管イベントアメリカ
-
Applied Science & Performance Institute完了
-
Bedford Hospital NHS TrustAnglia Ruskin Universityわからない