嚥下圧測定の標準化

2026年10月29日 更新者:University of Wisconsin, Madison

咽頭高解像度マノメトリーの指標標準化

This study aims to directly compare and contrast swallowing pressure and impedance measurements obtained from two high-resolution manometry (HRM) systems: the Laborie LMT Pharyngeal System and the Medtronic ManoScan ESO HRM System in healthy adult volunteers. The primary objective is to determine how differences in catheter diameter, sensor configuration, and acquisition platforms influence measurement outcomes. Twenty-five healthy adults will participate in a single, approximately 2-hour visit; the study will be open for one year for data collection and two years for data analysis.

調査の概要

詳細な説明

The primary objective is to obtain, compare, and contrast directly the measured swallow between the Laborie Pharyngeal HRM impedance system with the 3.7 mm catheter and the Medtronic HRM impedance system with the 4.2 mm catheter in the same setting with normal healthy volunteers.

研究の種類

介入

入学 (推定)

25

段階

  • 適用できない

連絡先と場所

このセクションには、調査を実施する担当者の連絡先の詳細と、この調査が実施されている場所に関する情報が記載されています。

研究連絡先

研究場所

    • Wisconsin
      • Madison、Wisconsin、アメリカ、53792
        • University of Wisconsin

参加基準

研究者は、適格基準と呼ばれる特定の説明に適合する人を探します。これらの基準のいくつかの例は、人の一般的な健康状態または以前の治療です。

適格基準

就学可能な年齢

18年~99年 (大人、高齢者)

健康ボランティアの受け入れ

はい

受講資格のある性別

全て

説明

Inclusion Criteria:

  • without known swallowing disorders

Exclusion Criteria:

  • Known swallowing disorder
  • Allergy to topical anesthetic
  • Allergy to food relevant to study participation (e.g. lactose intolerance)
  • Inability to follow direction or remain calm during the experiment.

研究計画

このセクションでは、研究がどのように設計され、研究が何を測定しているかなど、研究計画の詳細を提供します。

研究はどのように設計されていますか?

デザインの詳細

  • 主な目的:診断
  • 割り当て:ランダム化
  • 介入モデル:クロスオーバー割り当て
  • マスキング:なし(オープンラベル)

武器と介入

参加者グループ / アーム
介入・治療
実験的:Laborie then Medtronic

This study will include a data collection session that involves swallowing tasks during HRM.

The Solar GI HRM system is intended to record, store, view, and analyze pressure, impedance, EMG, swallow and respiration, and various auxiliary input device data online anywhere in the gastrointestinal tract (pharynx; esophagus; stomach; duodenum; sphincter of Oddi; small bowel; colon and anorectal area, including rectum and pelvic floor) to assist in the diagnosis and evaluation of gastrointestinal and swallowing disorders. Used with 3.7 mm catheter.

他の名前:
  • Laborie HRM インピーダンスシステム
  • Laborie Solar™ GI HRM 固体状態、トロリーシステム (G3-9)
  • Laborie Solid State HRIM Catheter (K83659-E-1820-D)

The ManoScan™ ESO high resolution manometry system enables full evaluation of the motor functions of the esophagus. The system provides useful information to support diagnosis of conditions like dysphagia, achalasia, and hiatal hernia. A 36-channel circumferential solid-state catheter with 4.2 mm diameter, 1.0 cm spacing between adjacent sensors.

他の名前:
  • メドトロニックHRMインピーダンスシステム
  • メドトロニック ManoScan™ ESO 高解像度マノメトリーシステム
  • Medtronic ManoScan360 HRMカテーテル
実験的:Medtronic then Laborie

This study will include a data collection session that involves swallowing tasks during HRM.

The Solar GI HRM system is intended to record, store, view, and analyze pressure, impedance, EMG, swallow and respiration, and various auxiliary input device data online anywhere in the gastrointestinal tract (pharynx; esophagus; stomach; duodenum; sphincter of Oddi; small bowel; colon and anorectal area, including rectum and pelvic floor) to assist in the diagnosis and evaluation of gastrointestinal and swallowing disorders. Used with 3.7 mm catheter.

他の名前:
  • Laborie HRM インピーダンスシステム
  • Laborie Solar™ GI HRM 固体状態、トロリーシステム (G3-9)
  • Laborie Solid State HRIM Catheter (K83659-E-1820-D)

The ManoScan™ ESO high resolution manometry system enables full evaluation of the motor functions of the esophagus. The system provides useful information to support diagnosis of conditions like dysphagia, achalasia, and hiatal hernia. A 36-channel circumferential solid-state catheter with 4.2 mm diameter, 1.0 cm spacing between adjacent sensors.

他の名前:
  • メドトロニックHRMインピーダンスシステム
  • メドトロニック ManoScan™ ESO 高解像度マノメトリーシステム
  • Medtronic ManoScan360 HRMカテーテル

この研究は何を測定していますか?

主要な結果の測定

結果測定
メジャーの説明
時間枠
Base of tongue pressure
時間枠:data collected during 1 study visit (up to 2 hours)

Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison.

data collected during 1 study visit (up to 2 hours)
Hypopharyngeal peak pressure
時間枠:data collected during 1 study visit (up to 2 hours)

Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison.

data collected during 1 study visit (up to 2 hours)
Pharyngeal Contractile Integral (PhCI)
時間枠:data collected during 1 study visit (up to 2 hours)

Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison.

data collected during 1 study visit (up to 2 hours)
Pressure rise time
時間枠:data collected during 1 study visit (up to 2 hours)

Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison.

data collected during 1 study visit (up to 2 hours)
Contractile front velocity
時間枠:data collected during 1 study visit (up to 2 hours)

Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison.

data collected during 1 study visit (up to 2 hours)
Upper Esophageal Sphincter (UES) resting pressure
時間枠:data collected during 1 study visit (up to 2 hours)

Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison.

data collected during 1 study visit (up to 2 hours)
UES peak contraction pressure
時間枠:data collected during 1 study visit (up to 2 hours)

Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison.

data collected during 1 study visit (up to 2 hours)
UES relaxation nadir pressure
時間枠:data collected during 1 study visit (up to 2 hours)

Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison.

data collected during 1 study visit (up to 2 hours)
UES relaxation duration
時間枠:data collected during 1 study visit (up to 2 hours)

Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison.

data collected during 1 study visit (up to 2 hours)
Nadir impedance
時間枠:data collected during 1 study visit (up to 2 hours)

Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison.

data collected during 1 study visit (up to 2 hours)
Bolus presence time
時間枠:data collected during 1 study visit (up to 2 hours)

Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison.

data collected during 1 study visit (up to 2 hours)
Flow interval
時間枠:data collected during 1 study visit (up to 2 hours)

Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison.

data collected during 1 study visit (up to 2 hours)
Pressure-flow index component
時間枠:data collected during 1 study visit (up to 2 hours)

Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison.

data collected during 1 study visit (up to 2 hours)
Timing relationships (onset-to-peak intervals)
時間枠:data collected during 1 study visit (up to 2 hours)

Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison.

data collected during 1 study visit (up to 2 hours)
Construction of Spatial pressure distribution maps
時間枠:data collected during 1 study visit (up to 2 hours)

Multi-level modeling for repeated measures across bolus types, volumes, and repetitions. Sensor-level and region-level analyses (BOT, hypopharynx, UES). Agreement thresholds and bias definitions added to facilitate interpretability and ensure clarity for cross-platform comparison. Map layers will be summarized as a proxy for this measure.

data collected during 1 study visit (up to 2 hours)

協力者と研究者

ここでは、この調査に関係する人々や組織を見つけることができます。

捜査官

  • 主任研究者:Timothy McCulloch, MD、University of Wisconsin, Madison

研究記録日

これらの日付は、ClinicalTrials.gov への研究記録と要約結果の提出の進捗状況を追跡します。研究記録と報告された結果は、国立医学図書館 (NLM) によって審査され、公開 Web サイトに掲載される前に、特定の品質管理基準を満たしていることが確認されます。

主要日程の研究

研究開始 (推定)

2026年11月1日

一次修了 (推定)

2027年9月1日

研究の完了 (推定)

2027年9月1日

試験登録日

最初に提出

2026年3月11日

QC基準を満たした最初の提出物

2026年3月11日

最初の投稿 (実際)

2026年3月16日

学習記録の更新

投稿された最後の更新 (実際)

2026年11月2日

QC基準を満たした最後の更新が送信されました

2026年10月29日

最終確認日

2026年10月1日

詳しくは

本研究に関する用語

その他の研究ID番号

  • 2025-1642
  • SMPH | Otolaryngology (その他の識別子:UW Madison)
  • Protocol Version 4/28/26 (その他の識別子:UW Madison)

個々の参加者データ (IPD) の計画

個々の参加者データ (IPD) を共有する予定はありますか?

はい

IPD 共有サポート情報タイプ

  • STUDY_PROTOCOL
  • SAP
  • ICF

医薬品およびデバイス情報、研究文書

米国FDA規制医薬品の研究

いいえ

米国FDA規制機器製品の研究

はい

米国で製造され、米国から輸出された製品。

はい

この情報は、Web サイト clinicaltrials.gov から変更なしで直接取得したものです。研究の詳細を変更、削除、または更新するリクエストがある場合は、register@clinicaltrials.gov。 までご連絡ください。 clinicaltrials.gov に変更が加えられるとすぐに、ウェブサイトでも自動的に更新されます。

購読する