Estandarización de las Mediciones de Presión de Deglución
Estandarización de Métricas de Manometría de Alta Resolución Faríngea
This study aims to directly compare and contrast swallowing pressure and impedance measurements obtained from two high-resolution manometry (HRM) systems: the Laborie LMT Pharyngeal System and the Medtronic ManoScan ESO HRM System in healthy adult volunteers. The primary objective is to determine how differences in catheter diameter, sensor configuration, and acquisition platforms influence measurement outcomes. Twenty-five healthy adults will participate in a single, approximately 2-hour visit; the study will be open for one year for data collection and two years for data analysis.
Descripción general del estudio
Estado
Condiciones
Descripción detallada
The primary objective is to obtain, compare, and contrast directly the measured swallow between the Laborie Pharyngeal HRM impedance system with the 3.7 mm catheter and the Medtronic HRM impedance system with the 4.2 mm catheter in the same setting with normal healthy volunteers.
Tipo de estudio
Inscripción (Estimado)
Fase
- No aplica
Contactos y Ubicaciones
Estudio Contacto
- Nombre: Suzan Abdelhalim, MD
- Número de teléfono: 608-265-2470
- Correo electrónico: abdelhalim@surgery.wisc.edu
Ubicaciones de estudio
-
-
Wisconsin
-
Madison, Wisconsin, Estados Unidos, 53792
- University of Wisconsin
-
-
Criterios de participación
Criterio de elegibilidad
Edades elegibles para estudiar
Acepta Voluntarios Saludables
Géneros elegibles para el estudio
Descripción
Inclusion Criteria:
- without known swallowing disorders
Exclusion Criteria:
- Known swallowing disorder
- Allergy to topical anesthetic
- Allergy to food relevant to study participation (e.g. lactose intolerance)
- Inability to follow direction or remain calm during the experiment.
Plan de estudios
¿Cómo está diseñado el estudio?
Detalles de diseño
- Propósito principal: Diagnóstico
- Asignación: Aleatorizado
- Modelo Intervencionista: Asignación cruzada
- Enmascaramiento: Ninguno (etiqueta abierta)
Armas e Intervenciones
Grupo de participantes/brazo |
Intervención / Tratamiento |
|---|---|
|
Experimental: Laborie then Medtronic
This study will include a data collection session that involves swallowing tasks during HRM. |
The Solar GI HRM system is intended to record, store, view, and analyze pressure, impedance, EMG, swallow and respiration, and various auxiliary input device data online anywhere in the gastrointestinal tract (pharynx; esophagus; stomach; duodenum; sphincter of Oddi; small bowel; colon and anorectal area, including rectum and pelvic floor) to assist in the diagnosis and evaluation of gastrointestinal and swallowing disorders. Used with 3.7 mm catheter.
Otros nombres:
The ManoScan™ ESO high resolution manometry system enables full evaluation of the motor functions of the esophagus. The system provides useful information to support diagnosis of conditions like dysphagia, achalasia, and hiatal hernia. A 36-channel circumferential solid-state catheter with 4.2 mm diameter, 1.0 cm spacing between adjacent sensors.
Otros nombres:
|
|
Experimental: Medtronic then Laborie
This study will include a data collection session that involves swallowing tasks during HRM. |
The Solar GI HRM system is intended to record, store, view, and analyze pressure, impedance, EMG, swallow and respiration, and various auxiliary input device data online anywhere in the gastrointestinal tract (pharynx; esophagus; stomach; duodenum; sphincter of Oddi; small bowel; colon and anorectal area, including rectum and pelvic floor) to assist in the diagnosis and evaluation of gastrointestinal and swallowing disorders. Used with 3.7 mm catheter.
Otros nombres:
The ManoScan™ ESO high resolution manometry system enables full evaluation of the motor functions of the esophagus. The system provides useful information to support diagnosis of conditions like dysphagia, achalasia, and hiatal hernia. A 36-channel circumferential solid-state catheter with 4.2 mm diameter, 1.0 cm spacing between adjacent sensors.
Otros nombres:
|
¿Qué mide el estudio?
Medidas de resultado primarias
Medida de resultado |
Medida Descripción |
Periodo de tiempo |
|---|---|---|
|
Base of tongue pressure
Periodo de tiempo: data collected during 1 study visit (up to 2 hours)
|
Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison. |
data collected during 1 study visit (up to 2 hours)
|
|
Hypopharyngeal peak pressure
Periodo de tiempo: data collected during 1 study visit (up to 2 hours)
|
Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison. |
data collected during 1 study visit (up to 2 hours)
|
|
Pharyngeal Contractile Integral (PhCI)
Periodo de tiempo: data collected during 1 study visit (up to 2 hours)
|
Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison. |
data collected during 1 study visit (up to 2 hours)
|
|
Pressure rise time
Periodo de tiempo: data collected during 1 study visit (up to 2 hours)
|
Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison. |
data collected during 1 study visit (up to 2 hours)
|
|
Contractile front velocity
Periodo de tiempo: data collected during 1 study visit (up to 2 hours)
|
Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison. |
data collected during 1 study visit (up to 2 hours)
|
|
Upper Esophageal Sphincter (UES) resting pressure
Periodo de tiempo: data collected during 1 study visit (up to 2 hours)
|
Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison. |
data collected during 1 study visit (up to 2 hours)
|
|
UES peak contraction pressure
Periodo de tiempo: data collected during 1 study visit (up to 2 hours)
|
Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison. |
data collected during 1 study visit (up to 2 hours)
|
|
UES relaxation nadir pressure
Periodo de tiempo: data collected during 1 study visit (up to 2 hours)
|
Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison. |
data collected during 1 study visit (up to 2 hours)
|
|
UES relaxation duration
Periodo de tiempo: data collected during 1 study visit (up to 2 hours)
|
Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison. |
data collected during 1 study visit (up to 2 hours)
|
|
Nadir impedance
Periodo de tiempo: data collected during 1 study visit (up to 2 hours)
|
Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison. |
data collected during 1 study visit (up to 2 hours)
|
|
Bolus presence time
Periodo de tiempo: data collected during 1 study visit (up to 2 hours)
|
Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison. |
data collected during 1 study visit (up to 2 hours)
|
|
Flow interval
Periodo de tiempo: data collected during 1 study visit (up to 2 hours)
|
Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison. |
data collected during 1 study visit (up to 2 hours)
|
|
Pressure-flow index component
Periodo de tiempo: data collected during 1 study visit (up to 2 hours)
|
Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison. |
data collected during 1 study visit (up to 2 hours)
|
|
Timing relationships (onset-to-peak intervals)
Periodo de tiempo: data collected during 1 study visit (up to 2 hours)
|
Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison. |
data collected during 1 study visit (up to 2 hours)
|
|
Construction of Spatial pressure distribution maps
Periodo de tiempo: data collected during 1 study visit (up to 2 hours)
|
Multi-level modeling for repeated measures across bolus types, volumes, and repetitions. Sensor-level and region-level analyses (BOT, hypopharynx, UES). Agreement thresholds and bias definitions added to facilitate interpretability and ensure clarity for cross-platform comparison. Map layers will be summarized as a proxy for this measure. |
data collected during 1 study visit (up to 2 hours)
|
Colaboradores e Investigadores
Patrocinador
Investigadores
- Investigador principal: Timothy McCulloch, MD, University of Wisconsin, Madison
Fechas de registro del estudio
Fechas importantes del estudio
Inicio del estudio (Estimado)
Finalización primaria (Estimado)
Finalización del estudio (Estimado)
Fechas de registro del estudio
Enviado por primera vez
Primero enviado que cumplió con los criterios de control de calidad
Publicado por primera vez (Actual)
Actualizaciones de registros de estudio
Última actualización publicada (Actual)
Última actualización enviada que cumplió con los criterios de control de calidad
Última verificación
Más información
Términos relacionados con este estudio
Palabras clave
Términos MeSH relevantes adicionales
Otros números de identificación del estudio
- 2025-1642
- SMPH | Otolaryngology (Otro identificador: UW Madison)
- Protocol Version 4/28/26 (Otro identificador: UW Madison)
Plan de datos de participantes individuales (IPD)
¿Planea compartir datos de participantes individuales (IPD)?
Tipo de información de apoyo para compartir IPD
- PROTOCOLO DE ESTUDIO
- SAVIA
- CIF
Información sobre medicamentos y dispositivos, documentos del estudio
Estudia un producto farmacéutico regulado por la FDA de EE. UU.
Estudia un producto de dispositivo regulado por la FDA de EE. UU.
producto fabricado y exportado desde los EE. UU.
Esta información se obtuvo directamente del sitio web clinicaltrials.gov sin cambios. Si tiene alguna solicitud para cambiar, eliminar o actualizar los detalles de su estudio, comuníquese con register@clinicaltrials.gov. Tan pronto como se implemente un cambio en clinicaltrials.gov, también se actualizará automáticamente en nuestro sitio web. .