Nielaisun painemittausten standardointi
Kurkun korkean resoluution manometrian mittareiden standardisointi
This study aims to directly compare and contrast swallowing pressure and impedance measurements obtained from two high-resolution manometry (HRM) systems: the Laborie LMT Pharyngeal System and the Medtronic ManoScan ESO HRM System in healthy adult volunteers. The primary objective is to determine how differences in catheter diameter, sensor configuration, and acquisition platforms influence measurement outcomes. Twenty-five healthy adults will participate in a single, approximately 2-hour visit; the study will be open for one year for data collection and two years for data analysis.
Tutkimuksen yleiskatsaus
Tila
Ehdot
Yksityiskohtainen kuvaus
The primary objective is to obtain, compare, and contrast directly the measured swallow between the Laborie Pharyngeal HRM impedance system with the 3.7 mm catheter and the Medtronic HRM impedance system with the 4.2 mm catheter in the same setting with normal healthy volunteers.
Opintotyyppi
Ilmoittautuminen (Arvioitu)
Vaihe
- Ei sovellettavissa
Yhteystiedot ja paikat
Opiskeluyhteys
- Nimi: Suzan Abdelhalim, MD
- Puhelinnumero: 608-265-2470
- Sähköposti: abdelhalim@surgery.wisc.edu
Opiskelupaikat
-
-
Wisconsin
-
Madison, Wisconsin, Yhdysvallat, 53792
- University of Wisconsin
-
-
Osallistumiskriteerit
Kelpoisuusvaatimukset
Opintokelpoiset iät
Hyväksyy terveitä vapaaehtoisia
Sukupuolet, jotka voivat opiskella
Kuvaus
Inclusion Criteria:
- without known swallowing disorders
Exclusion Criteria:
- Known swallowing disorder
- Allergy to topical anesthetic
- Allergy to food relevant to study participation (e.g. lactose intolerance)
- Inability to follow direction or remain calm during the experiment.
Opintosuunnitelma
Miten tutkimus on suunniteltu?
Suunnittelun yksityiskohdat
- Ensisijainen käyttötarkoitus: Diagnostiikka
- Jako: Satunnaistettu
- Inventiomalli: Crossover-tehtävä
- Naamiointi: Ei mitään (avoin tarra)
Aseet ja interventiot
Osallistujaryhmä / Arm |
Interventio / Hoito |
|---|---|
|
Kokeellinen: Laborie then Medtronic
This study will include a data collection session that involves swallowing tasks during HRM. |
The Solar GI HRM system is intended to record, store, view, and analyze pressure, impedance, EMG, swallow and respiration, and various auxiliary input device data online anywhere in the gastrointestinal tract (pharynx; esophagus; stomach; duodenum; sphincter of Oddi; small bowel; colon and anorectal area, including rectum and pelvic floor) to assist in the diagnosis and evaluation of gastrointestinal and swallowing disorders. Used with 3.7 mm catheter.
Muut nimet:
The ManoScan™ ESO high resolution manometry system enables full evaluation of the motor functions of the esophagus. The system provides useful information to support diagnosis of conditions like dysphagia, achalasia, and hiatal hernia. A 36-channel circumferential solid-state catheter with 4.2 mm diameter, 1.0 cm spacing between adjacent sensors.
Muut nimet:
|
|
Kokeellinen: Medtronic then Laborie
This study will include a data collection session that involves swallowing tasks during HRM. |
The Solar GI HRM system is intended to record, store, view, and analyze pressure, impedance, EMG, swallow and respiration, and various auxiliary input device data online anywhere in the gastrointestinal tract (pharynx; esophagus; stomach; duodenum; sphincter of Oddi; small bowel; colon and anorectal area, including rectum and pelvic floor) to assist in the diagnosis and evaluation of gastrointestinal and swallowing disorders. Used with 3.7 mm catheter.
Muut nimet:
The ManoScan™ ESO high resolution manometry system enables full evaluation of the motor functions of the esophagus. The system provides useful information to support diagnosis of conditions like dysphagia, achalasia, and hiatal hernia. A 36-channel circumferential solid-state catheter with 4.2 mm diameter, 1.0 cm spacing between adjacent sensors.
Muut nimet:
|
Mitä tutkimuksessa mitataan?
Ensisijaiset tulostoimenpiteet
Tulosmittaus |
Toimenpiteen kuvaus |
Aikaikkuna |
|---|---|---|
|
Base of tongue pressure
Aikaikkuna: data collected during 1 study visit (up to 2 hours)
|
Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison. |
data collected during 1 study visit (up to 2 hours)
|
|
Hypopharyngeal peak pressure
Aikaikkuna: data collected during 1 study visit (up to 2 hours)
|
Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison. |
data collected during 1 study visit (up to 2 hours)
|
|
Pharyngeal Contractile Integral (PhCI)
Aikaikkuna: data collected during 1 study visit (up to 2 hours)
|
Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison. |
data collected during 1 study visit (up to 2 hours)
|
|
Pressure rise time
Aikaikkuna: data collected during 1 study visit (up to 2 hours)
|
Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison. |
data collected during 1 study visit (up to 2 hours)
|
|
Contractile front velocity
Aikaikkuna: data collected during 1 study visit (up to 2 hours)
|
Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison. |
data collected during 1 study visit (up to 2 hours)
|
|
Upper Esophageal Sphincter (UES) resting pressure
Aikaikkuna: data collected during 1 study visit (up to 2 hours)
|
Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison. |
data collected during 1 study visit (up to 2 hours)
|
|
UES peak contraction pressure
Aikaikkuna: data collected during 1 study visit (up to 2 hours)
|
Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison. |
data collected during 1 study visit (up to 2 hours)
|
|
UES relaxation nadir pressure
Aikaikkuna: data collected during 1 study visit (up to 2 hours)
|
Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison. |
data collected during 1 study visit (up to 2 hours)
|
|
UES relaxation duration
Aikaikkuna: data collected during 1 study visit (up to 2 hours)
|
Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison. |
data collected during 1 study visit (up to 2 hours)
|
|
Nadir impedance
Aikaikkuna: data collected during 1 study visit (up to 2 hours)
|
Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison. |
data collected during 1 study visit (up to 2 hours)
|
|
Bolus presence time
Aikaikkuna: data collected during 1 study visit (up to 2 hours)
|
Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison. |
data collected during 1 study visit (up to 2 hours)
|
|
Flow interval
Aikaikkuna: data collected during 1 study visit (up to 2 hours)
|
Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison. |
data collected during 1 study visit (up to 2 hours)
|
|
Pressure-flow index component
Aikaikkuna: data collected during 1 study visit (up to 2 hours)
|
Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison. |
data collected during 1 study visit (up to 2 hours)
|
|
Timing relationships (onset-to-peak intervals)
Aikaikkuna: data collected during 1 study visit (up to 2 hours)
|
Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison. |
data collected during 1 study visit (up to 2 hours)
|
|
Construction of Spatial pressure distribution maps
Aikaikkuna: data collected during 1 study visit (up to 2 hours)
|
Multi-level modeling for repeated measures across bolus types, volumes, and repetitions. Sensor-level and region-level analyses (BOT, hypopharynx, UES). Agreement thresholds and bias definitions added to facilitate interpretability and ensure clarity for cross-platform comparison. Map layers will be summarized as a proxy for this measure. |
data collected during 1 study visit (up to 2 hours)
|
Yhteistyökumppanit ja tutkijat
Sponsori
Tutkijat
- Päätutkija: Timothy McCulloch, MD, University of Wisconsin, Madison
Opintojen ennätyspäivät
Opi tärkeimmät päivämäärät
Opiskelun aloitus (Arvioitu)
Ensisijainen valmistuminen (Arvioitu)
Opintojen valmistuminen (Arvioitu)
Opintoihin ilmoittautumispäivät
Ensimmäinen lähetetty
Ensimmäinen toimitettu, joka täytti QC-kriteerit
Ensimmäinen Lähetetty (Todellinen)
Tutkimustietojen päivitykset
Viimeisin päivitys julkaistu (Todellinen)
Viimeisin lähetetty päivitys, joka täytti QC-kriteerit
Viimeksi vahvistettu
Lisää tietoa
Tähän tutkimukseen liittyvät termit
Avainsanat
Muita asiaankuuluvia MeSH-ehtoja
Muut tutkimustunnusnumerot
- 2025-1642
- SMPH | Otolaryngology (Muu tunniste: UW Madison)
- Protocol Version 4/28/26 (Muu tunniste: UW Madison)
Yksittäisten osallistujien tietojen suunnitelma (IPD)
Aiotko jakaa yksittäisten osallistujien tietoja (IPD)?
IPD-jakamista tukeva tietotyyppi
- STUDY_PROTOCOL
- MAHLA
- ICF
Lääke- ja laitetiedot, tutkimusasiakirjat
Tutkii yhdysvaltalaista FDA sääntelemää lääkevalmistetta
Tutkii yhdysvaltalaista FDA sääntelemää laitetuotetta
Yhdysvalloissa valmistettu ja sieltä viety tuote
Nämä tiedot haettiin suoraan verkkosivustolta clinicaltrials.gov ilman muutoksia. Jos sinulla on pyyntöjä muuttaa, poistaa tai päivittää tutkimustietojasi, ota yhteyttä register@clinicaltrials.gov. Heti kun muutos on otettu käyttöön osoitteessa clinicaltrials.gov, se päivitetään automaattisesti myös verkkosivustollemme .