Standardisering af Trykmålinger ved Synkning

29. oktober 2026 opdateret af: University of Wisconsin, Madison

Standardisering af metrikker for pharyngeal højopløsningsmanometri

This study aims to directly compare and contrast swallowing pressure and impedance measurements obtained from two high-resolution manometry (HRM) systems: the Laborie LMT Pharyngeal System and the Medtronic ManoScan ESO HRM System in healthy adult volunteers. The primary objective is to determine how differences in catheter diameter, sensor configuration, and acquisition platforms influence measurement outcomes. Twenty-five healthy adults will participate in a single, approximately 2-hour visit; the study will be open for one year for data collection and two years for data analysis.

Studieoversigt

Detaljeret beskrivelse

The primary objective is to obtain, compare, and contrast directly the measured swallow between the Laborie Pharyngeal HRM impedance system with the 3.7 mm catheter and the Medtronic HRM impedance system with the 4.2 mm catheter in the same setting with normal healthy volunteers.

Undersøgelsestype

Interventionel

Tilmelding (Anslået)

25

Fase

  • Ikke anvendelig

Kontakter og lokationer

Dette afsnit indeholder kontaktoplysninger for dem, der udfører undersøgelsen, og oplysninger om, hvor denne undersøgelse udføres.

Studiekontakt

Studiesteder

    • Wisconsin
      • Madison, Wisconsin, Forenede Stater, 53792
        • University of Wisconsin

Deltagelseskriterier

Forskere leder efter personer, der passer til en bestemt beskrivelse, kaldet berettigelseskriterier. Nogle eksempler på disse kriterier er en persons generelle helbredstilstand eller tidligere behandlinger.

Berettigelseskriterier

Aldre berettiget til at studere

18 år til 99 år (Voksen, Ældre voksen)

Tager imod sunde frivillige

Ja

Køn, der er berettiget til at studere

Alle

Beskrivelse

Inclusion Criteria:

  • without known swallowing disorders

Exclusion Criteria:

  • Known swallowing disorder
  • Allergy to topical anesthetic
  • Allergy to food relevant to study participation (e.g. lactose intolerance)
  • Inability to follow direction or remain calm during the experiment.

Studieplan

Dette afsnit indeholder detaljer om studieplanen, herunder hvordan undersøgelsen er designet, og hvad undersøgelsen måler.

Hvordan er undersøgelsen tilrettelagt?

Design detaljer

  • Primært formål: Diagnostisk
  • Tildeling: Randomiseret
  • Interventionel model: Crossover opgave
  • Maskning: Ingen (Åben etiket)

Våben og indgreb

Deltagergruppe / Arm
Intervention / Behandling
Eksperimentel: Laborie then Medtronic

This study will include a data collection session that involves swallowing tasks during HRM.

The Solar GI HRM system is intended to record, store, view, and analyze pressure, impedance, EMG, swallow and respiration, and various auxiliary input device data online anywhere in the gastrointestinal tract (pharynx; esophagus; stomach; duodenum; sphincter of Oddi; small bowel; colon and anorectal area, including rectum and pelvic floor) to assist in the diagnosis and evaluation of gastrointestinal and swallowing disorders. Used with 3.7 mm catheter.

Andre navne:
  • Laborie HRM-impedanssystem
  • Laborie Solar™ GI HRM Faststof, Vognsystem (G3-9)
  • Laborie Solid State HRIM Catheter (K83659-E-1820-D)

The ManoScan™ ESO high resolution manometry system enables full evaluation of the motor functions of the esophagus. The system provides useful information to support diagnosis of conditions like dysphagia, achalasia, and hiatal hernia. A 36-channel circumferential solid-state catheter with 4.2 mm diameter, 1.0 cm spacing between adjacent sensors.

Andre navne:
  • Medtronic HRM impedanssystem
  • Medtronic ManoScan™ ESO højopløsningsmanometrisystem
  • Medtronic ManoScan360 HRM-kateter
Eksperimentel: Medtronic then Laborie

This study will include a data collection session that involves swallowing tasks during HRM.

The Solar GI HRM system is intended to record, store, view, and analyze pressure, impedance, EMG, swallow and respiration, and various auxiliary input device data online anywhere in the gastrointestinal tract (pharynx; esophagus; stomach; duodenum; sphincter of Oddi; small bowel; colon and anorectal area, including rectum and pelvic floor) to assist in the diagnosis and evaluation of gastrointestinal and swallowing disorders. Used with 3.7 mm catheter.

Andre navne:
  • Laborie HRM-impedanssystem
  • Laborie Solar™ GI HRM Faststof, Vognsystem (G3-9)
  • Laborie Solid State HRIM Catheter (K83659-E-1820-D)

The ManoScan™ ESO high resolution manometry system enables full evaluation of the motor functions of the esophagus. The system provides useful information to support diagnosis of conditions like dysphagia, achalasia, and hiatal hernia. A 36-channel circumferential solid-state catheter with 4.2 mm diameter, 1.0 cm spacing between adjacent sensors.

Andre navne:
  • Medtronic HRM impedanssystem
  • Medtronic ManoScan™ ESO højopløsningsmanometrisystem
  • Medtronic ManoScan360 HRM-kateter

Hvad måler undersøgelsen?

Primære resultatmål

Resultatmål
Foranstaltningsbeskrivelse
Tidsramme
Base of tongue pressure
Tidsramme: data collected during 1 study visit (up to 2 hours)

Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison.

data collected during 1 study visit (up to 2 hours)
Hypopharyngeal peak pressure
Tidsramme: data collected during 1 study visit (up to 2 hours)

Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison.

data collected during 1 study visit (up to 2 hours)
Pharyngeal Contractile Integral (PhCI)
Tidsramme: data collected during 1 study visit (up to 2 hours)

Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison.

data collected during 1 study visit (up to 2 hours)
Pressure rise time
Tidsramme: data collected during 1 study visit (up to 2 hours)

Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison.

data collected during 1 study visit (up to 2 hours)
Contractile front velocity
Tidsramme: data collected during 1 study visit (up to 2 hours)

Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison.

data collected during 1 study visit (up to 2 hours)
Upper Esophageal Sphincter (UES) resting pressure
Tidsramme: data collected during 1 study visit (up to 2 hours)

Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison.

data collected during 1 study visit (up to 2 hours)
UES peak contraction pressure
Tidsramme: data collected during 1 study visit (up to 2 hours)

Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison.

data collected during 1 study visit (up to 2 hours)
UES relaxation nadir pressure
Tidsramme: data collected during 1 study visit (up to 2 hours)

Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison.

data collected during 1 study visit (up to 2 hours)
UES relaxation duration
Tidsramme: data collected during 1 study visit (up to 2 hours)

Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison.

data collected during 1 study visit (up to 2 hours)
Nadir impedance
Tidsramme: data collected during 1 study visit (up to 2 hours)

Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison.

data collected during 1 study visit (up to 2 hours)
Bolus presence time
Tidsramme: data collected during 1 study visit (up to 2 hours)

Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison.

data collected during 1 study visit (up to 2 hours)
Flow interval
Tidsramme: data collected during 1 study visit (up to 2 hours)

Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison.

data collected during 1 study visit (up to 2 hours)
Pressure-flow index component
Tidsramme: data collected during 1 study visit (up to 2 hours)

Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison.

data collected during 1 study visit (up to 2 hours)
Timing relationships (onset-to-peak intervals)
Tidsramme: data collected during 1 study visit (up to 2 hours)

Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison.

data collected during 1 study visit (up to 2 hours)
Construction of Spatial pressure distribution maps
Tidsramme: data collected during 1 study visit (up to 2 hours)

Multi-level modeling for repeated measures across bolus types, volumes, and repetitions. Sensor-level and region-level analyses (BOT, hypopharynx, UES). Agreement thresholds and bias definitions added to facilitate interpretability and ensure clarity for cross-platform comparison. Map layers will be summarized as a proxy for this measure.

data collected during 1 study visit (up to 2 hours)

Samarbejdspartnere og efterforskere

Det er her, du vil finde personer og organisationer, der er involveret i denne undersøgelse.

Efterforskere

  • Ledende efterforsker: Timothy McCulloch, MD, University of Wisconsin, Madison

Datoer for undersøgelser

Disse datoer sporer fremskridtene for indsendelser af undersøgelsesrekord og resumeresultater til ClinicalTrials.gov. Studieregistreringer og rapporterede resultater gennemgås af National Library of Medicine (NLM) for at sikre, at de opfylder specifikke kvalitetskontrolstandarder, før de offentliggøres på den offentlige hjemmeside.

Studer store datoer

Studiestart (Anslået)

1. november 2026

Primær færdiggørelse (Anslået)

1. september 2027

Studieafslutning (Anslået)

1. september 2027

Datoer for studieregistrering

Først indsendt

11. marts 2026

Først indsendt, der opfyldte QC-kriterier

11. marts 2026

Først opslået (Faktiske)

16. marts 2026

Opdateringer af undersøgelsesjournaler

Sidste opdatering sendt (Faktiske)

2. november 2026

Sidste opdatering indsendt, der opfyldte kvalitetskontrolkriterier

29. oktober 2026

Sidst verificeret

1. oktober 2026

Mere information

Begreber relateret til denne undersøgelse

Andre undersøgelses-id-numre

  • 2025-1642
  • SMPH | Otolaryngology (Anden identifikator: UW Madison)
  • Protocol Version 4/28/26 (Anden identifikator: UW Madison)

Plan for individuelle deltagerdata (IPD)

Planlægger du at dele individuelle deltagerdata (IPD)?

JA

IPD-deling Understøttende informationstype

  • STUDY_PROTOCOL
  • SAP
  • ICF

Lægemiddel- og udstyrsoplysninger, undersøgelsesdokumenter

Studerer et amerikansk FDA-reguleret lægemiddelprodukt

Ingen

Studerer et amerikansk FDA-reguleret enhedsprodukt

Ja

produkt fremstillet i og eksporteret fra U.S.A.

Ja

Disse oplysninger blev hentet direkte fra webstedet clinicaltrials.gov uden ændringer. Hvis du har nogen anmodninger om at ændre, fjerne eller opdatere dine undersøgelsesoplysninger, bedes du kontakte register@clinicaltrials.gov. Så snart en ændring er implementeret på clinicaltrials.gov, vil denne også blive opdateret automatisk på vores hjemmeside .