This study aims to directly compare and contrast swallowing pressure and impedance measurements obtained from two high-resolution manometry (HRM) systems: the Laborie LMT Pharyngeal System and the Medtronic ManoScan ESO HRM System in healthy adult volunteers. The primary objective is to determine how differences in catheter diameter, sensor configuration, and acquisition platforms influence measurement outcomes. Twenty-five healthy adults will participate in a single, approximately 2-hour visit; the study will be open for one year for data collection and two years for data analysis.
A tanulmány áttekintése
Állapot
Körülmények
Részletes leírás
The primary objective is to obtain, compare, and contrast directly the measured swallow between the Laborie Pharyngeal HRM impedance system with the 3.7 mm catheter and the Medtronic HRM impedance system with the 4.2 mm catheter in the same setting with normal healthy volunteers.
Tanulmány típusa
Beiratkozás (Becsült)
Fázis
- Nem alkalmazható
Kapcsolatok és helyek
Tanulmányi kapcsolat
- Név: Suzan Abdelhalim, MD
- Telefonszám: 608-265-2470
- E-mail: abdelhalim@surgery.wisc.edu
Tanulmányi helyek
-
-
Wisconsin
-
Madison, Wisconsin, Egyesült Államok, 53792
- University of Wisconsin
-
-
Részvételi kritériumok
Jogosultsági kritériumok
Tanulmányozható életkorok
Egészséges önkénteseket fogad
Tanulmányozható nemek
Leírás
Inclusion Criteria:
- without known swallowing disorders
Exclusion Criteria:
- Known swallowing disorder
- Allergy to topical anesthetic
- Allergy to food relevant to study participation (e.g. lactose intolerance)
- Inability to follow direction or remain calm during the experiment.
Tanulási terv
Hogyan készül a tanulmány?
Tervezési részletek
- Elsődleges cél: Diagnosztikai
- Kiosztás: Véletlenszerűsített
- Beavatkozó modell: Crossover kiosztás
- Maszkolás: Nincs (Open Label)
Fegyverek és beavatkozások
Résztvevő csoport / kar |
Beavatkozás / kezelés |
|---|---|
|
Kísérleti: Laborie then Medtronic
This study will include a data collection session that involves swallowing tasks during HRM. |
The Solar GI HRM system is intended to record, store, view, and analyze pressure, impedance, EMG, swallow and respiration, and various auxiliary input device data online anywhere in the gastrointestinal tract (pharynx; esophagus; stomach; duodenum; sphincter of Oddi; small bowel; colon and anorectal area, including rectum and pelvic floor) to assist in the diagnosis and evaluation of gastrointestinal and swallowing disorders. Used with 3.7 mm catheter.
Más nevek:
The ManoScan™ ESO high resolution manometry system enables full evaluation of the motor functions of the esophagus. The system provides useful information to support diagnosis of conditions like dysphagia, achalasia, and hiatal hernia. A 36-channel circumferential solid-state catheter with 4.2 mm diameter, 1.0 cm spacing between adjacent sensors.
Más nevek:
|
|
Kísérleti: Medtronic then Laborie
This study will include a data collection session that involves swallowing tasks during HRM. |
The Solar GI HRM system is intended to record, store, view, and analyze pressure, impedance, EMG, swallow and respiration, and various auxiliary input device data online anywhere in the gastrointestinal tract (pharynx; esophagus; stomach; duodenum; sphincter of Oddi; small bowel; colon and anorectal area, including rectum and pelvic floor) to assist in the diagnosis and evaluation of gastrointestinal and swallowing disorders. Used with 3.7 mm catheter.
Más nevek:
The ManoScan™ ESO high resolution manometry system enables full evaluation of the motor functions of the esophagus. The system provides useful information to support diagnosis of conditions like dysphagia, achalasia, and hiatal hernia. A 36-channel circumferential solid-state catheter with 4.2 mm diameter, 1.0 cm spacing between adjacent sensors.
Más nevek:
|
Mit mér a tanulmány?
Elsődleges eredményintézkedések
Eredménymérő |
Intézkedés leírása |
Időkeret |
|---|---|---|
|
Base of tongue pressure
Időkeret: data collected during 1 study visit (up to 2 hours)
|
Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison. |
data collected during 1 study visit (up to 2 hours)
|
|
Hypopharyngeal peak pressure
Időkeret: data collected during 1 study visit (up to 2 hours)
|
Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison. |
data collected during 1 study visit (up to 2 hours)
|
|
Pharyngeal Contractile Integral (PhCI)
Időkeret: data collected during 1 study visit (up to 2 hours)
|
Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison. |
data collected during 1 study visit (up to 2 hours)
|
|
Pressure rise time
Időkeret: data collected during 1 study visit (up to 2 hours)
|
Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison. |
data collected during 1 study visit (up to 2 hours)
|
|
Contractile front velocity
Időkeret: data collected during 1 study visit (up to 2 hours)
|
Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison. |
data collected during 1 study visit (up to 2 hours)
|
|
Upper Esophageal Sphincter (UES) resting pressure
Időkeret: data collected during 1 study visit (up to 2 hours)
|
Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison. |
data collected during 1 study visit (up to 2 hours)
|
|
UES peak contraction pressure
Időkeret: data collected during 1 study visit (up to 2 hours)
|
Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison. |
data collected during 1 study visit (up to 2 hours)
|
|
UES relaxation nadir pressure
Időkeret: data collected during 1 study visit (up to 2 hours)
|
Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison. |
data collected during 1 study visit (up to 2 hours)
|
|
UES relaxation duration
Időkeret: data collected during 1 study visit (up to 2 hours)
|
Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison. |
data collected during 1 study visit (up to 2 hours)
|
|
Nadir impedance
Időkeret: data collected during 1 study visit (up to 2 hours)
|
Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison. |
data collected during 1 study visit (up to 2 hours)
|
|
Bolus presence time
Időkeret: data collected during 1 study visit (up to 2 hours)
|
Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison. |
data collected during 1 study visit (up to 2 hours)
|
|
Flow interval
Időkeret: data collected during 1 study visit (up to 2 hours)
|
Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison. |
data collected during 1 study visit (up to 2 hours)
|
|
Pressure-flow index component
Időkeret: data collected during 1 study visit (up to 2 hours)
|
Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison. |
data collected during 1 study visit (up to 2 hours)
|
|
Timing relationships (onset-to-peak intervals)
Időkeret: data collected during 1 study visit (up to 2 hours)
|
Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison. |
data collected during 1 study visit (up to 2 hours)
|
|
Construction of Spatial pressure distribution maps
Időkeret: data collected during 1 study visit (up to 2 hours)
|
Multi-level modeling for repeated measures across bolus types, volumes, and repetitions. Sensor-level and region-level analyses (BOT, hypopharynx, UES). Agreement thresholds and bias definitions added to facilitate interpretability and ensure clarity for cross-platform comparison. Map layers will be summarized as a proxy for this measure. |
data collected during 1 study visit (up to 2 hours)
|
Együttműködők és nyomozók
Szponzor
Nyomozók
- Kutatásvezető: Timothy McCulloch, MD, University of Wisconsin, Madison
Tanulmányi rekorddátumok
Tanulmány főbb dátumok
Tanulmány kezdete (Becsült)
Elsődleges befejezés (Becsült)
A tanulmány befejezése (Becsült)
Tanulmányi regisztráció dátumai
Először benyújtva
Először nyújtották be, amely megfelel a minőségbiztosítási kritériumoknak
Első közzététel (Tényleges)
Tanulmányi rekordok frissítései
Utolsó frissítés közzétéve (Tényleges)
Az utolsó frissítés elküldve, amely megfelel a minőségbiztosítási kritériumoknak
Utolsó ellenőrzés
Több információ
A tanulmányhoz kapcsolódó kifejezések
Kulcsszavak
További vonatkozó MeSH feltételek
Egyéb vizsgálati azonosító számok
- 2025-1642
- SMPH | Otolaryngology (Egyéb azonosító: UW Madison)
- Protocol Version 4/28/26 (Egyéb azonosító: UW Madison)
Terv az egyéni résztvevői adatokhoz (IPD)
Tervezi megosztani az egyéni résztvevői adatokat (IPD)?
Az IPD megosztását támogató információ típusa
- STUDY_PROTOCOL
- NEDV
- ICF
Gyógyszer- és eszközinformációk, tanulmányi dokumentumok
Egy amerikai FDA által szabályozott gyógyszerkészítményt tanulmányoz
Egy amerikai FDA által szabályozott eszközterméket tanulmányoz
az Egyesült Államokban gyártott és onnan exportált termék
Ezt az információt közvetlenül a clinicaltrials.gov webhelyről szereztük be, változtatás nélkül. Ha bármilyen kérése van vizsgálati adatainak módosítására, eltávolítására vagy frissítésére, kérjük, írjon a következő címre: register@clinicaltrials.gov. Amint a változás bevezetésre kerül a clinicaltrials.gov oldalon, ez a webhelyünkön is automatikusan frissül. .