Evaluation of the safety and efficacy of the new Enova® preloaded monofocal intraocular lens and cartridge injector system in cataract surgery

Studie Overzicht

Toestand

Voltooid

Conditie

Interventie / Behandeling

Gedetailleerde beschrijving

In this prospective and single-center clinical study, the safety and efficacy of the Enova PGF3 intraocular lens (IOL) and cartridge injector system will be evaluated.

The Enova PGF3 cartridge injector system is designed to be used in cataract surgeries. Cataract patients aged 40 years and older, both male and female, without any serious ocular disease other than cataracts, will undergo IOL implantation following cataract surgery. It is planned to perform a total of 200 implantations with either unilateral or bilateral implantation of the participants' eyes.

Postoperative visual acuity, refraction, patient satisfaction, and adverse events will be monitored during follow-up visits at day 1, week 1, and month 1.

The primary objectives of the study are to achieve an uncorrected distance visual acuity of 20/20 or better, and to maintain an error and complication rate related to the preloaded system below 7.5%. Secondary objectives include evaluating patient satisfaction and assessing whether surgeons find the preloaded system easy to use and if it meets their expectations.

The primary endpoints are refractive stability within ±0.50 D, uncorrected distance visual acuity of 20/20 or better, and a general clinical performance rating of the preloaded system with a score of 95% or higher. Secondary endpoints include the duration of surgery, accuracy of lens positioning, surgical complications, surgeon-reported ease of use, and patient satisfaction survey results meeting expectations.

As a safety parameter, adverse events reported in postoperative surgical reports will be evaluated.

Studietype

Ingrijpend

Inschrijving (Werkelijk)

200

Fase

  • Niet toepasbaar

Contacten en locaties

In dit gedeelte vindt u de contactgegevens van degenen die het onderzoek uitvoeren en informatie over waar dit onderzoek wordt uitgevoerd.

Studie Locaties

    • Fatih
      • Istanbul, Fatih, Turkije (Türkiye), 34250
        • Istanbul University Istanbul Medical Faculty Ophthalmology Department

Deelname Criteria

Onderzoekers zoeken naar mensen die aan een bepaalde beschrijving voldoen, de zogenaamde geschiktheidscriteria. Enkele voorbeelden van deze criteria zijn iemands algemene gezondheidstoestand of eerdere behandelingen.

Geschiktheidscriteria

Leeftijden die in aanmerking komen voor studie

40 jaar en ouder (Volwassen, Oudere volwassene)

Accepteert gezonde vrijwilligers

Nee

Geslachten die in aanmerking komen voor studie

Allemaal

Beschrijving

Inclusion Criteria:

  • Adults of either sex.

Age 40 years or older.

Willing and able to provide written informed consent and comply with planned study visits and procedures.

Healthy ocular status, with no ocular pathology other than cataract and mild refractive error.

Potential visual acuity of 20/32 (0.2 logMAR) or better in both eyes.

Presence of regular corneal astigmatism suitable for treatment with the study device.

Candidates for uncomplicated cataract surgery using the study device.

Exclusion Criteria:

  • Irregular corneal astigmatism (e.g., keratoconus).

Corneal pathology (e.g., scar, dystrophy, pterygium, moderate to severe dry eye).

Monocular conditions (e.g., amblyopia).

Previous corneal surgery, including radial keratotomy, corneal refractive surgery, corneal transplantation, DSAEK, or lamellar keratoplasty.

Previous anterior or posterior segment surgery (e.g., vitrectomy, laser iridotomy).

Diabetic retinopathy.

Macular pathology (e.g., age-related macular degeneration [ARMD], epiretinal membrane [ERM]).

History of retinal detachment.

Surgical complications that may affect study outcomes, such as:

Decentration due to capsulorhexis issues

Improper lens placement

Inadequate or missing postoperative capsular support

Posterior capsular rupture

Vitreous loss

Iris damage (tear, rupture, etc.)

Acute or chronic ocular diseases known to affect visual acuity or study outcomes (e.g., immunosuppressed state, connective tissue disease, clinically significant atopic disease, diabetes, or other systemic or ocular conditions).

Participation in any investigational drug or device study within 30 days prior to study start, or ongoing participation in another study.

Studie plan

Dit gedeelte bevat details van het studieplan, inclusief hoe de studie is opgezet en wat de studie meet.

Hoe is de studie opgezet?

Ontwerpdetails

  • Primair doel: Behandeling
  • Toewijzing: NVT
  • Interventioneel model: Opdracht voor een enkele groep
  • Masker: Geen (open label)

Wapens en interventies

Deelnemersgroep / Arm
Interventie / Behandeling
Experimenteel: Patients who need cataract surgery

Participants will undergo standard cataract surgery using the new preloaded intraocular lens (IOL) system. The intervention involves implantation of the preloaded IOL following routine phacoemulsification techniques.

This intervention uses a preloaded IOL delivery system, which allows for a single-use, sterile implantation during standard phacoemulsification cataract surgery. Unlike standard manually loaded IOLs, this system reduces preparation time and minimizes handling of the lens, potentially improving surgical efficiency and maintaining lens sterility.

Wat meet het onderzoek?

Primaire uitkomstmaten

Uitkomstmaat
Maatregel Beschrijving
Tijdsspanne
Uncorrected Distance Visual Acuity (UDVA) - Photopic Conditions
Tijdsspanne: 1 month postoperatively

Proportion of eyes achieving 20/20 or better uncorrected distance visual acuity under photopic conditions.

1 month postoperatively
Subjective Refractive Error
Tijdsspanne: 1 month postoperatively

Eyes with a postoperative refractive error within ±0.50 diopters of the intended target.

1 month postoperatively
Overall Clinical Performance of the Preloaded IOL System
Tijdsspanne: 1 month after surgery

Surgeon-rated assessment of the general clinical performance of the preloaded system by The ENOVA PGF3 Questionnaire. This questionnaire is a 14-item, study-specific, non-validated survey developed to evaluate surgeons' experience with an intraocular lens (IOL) preloaded delivery system. Eleven items are rated on a 5-point Likert scale with the following scoring structure:

  1. = Low
  2. = Below Expectation
  3. = Meets Expectation
  4. = Above Expectation
  5. = Outstanding Performance

Some additional items are Yes/No questions and do not contribute to the numerical total score.

The total numerical score is calculated by summing the 11 Likert-scaled items. Total score range: 11 to 55, with higher scores indicating better performance and more favorable user experience.

1 month after surgery

Secundaire uitkomstmaten

Uitkomstmaat
Maatregel Beschrijving
Tijdsspanne
Preloaded System Evaluation Patient Questionnaire
Tijdsspanne: Baseline (preoperative) and 1 month postoperatively

Patient-reported outcomes assessing whether expectations for visual improvement and overall satisfaction were met.The questionnaire consists of 14 items assessing a patient's self-reported difficulty in performing vision-related daily activities. Each activity is rated using a 0-4 scale, where:

4 = No difficulty

3 = Mild difficulty

2 = Moderate difficulty

1 = Severe difficulty

0 = Unable to perform the activity

The total score is calculated by summing responses across all 14 items. Total possible score range: 0 to 56

Higher scores indicate better visual function and less activity limitation.

Lower scores indicate worse visual function.

Baseline (preoperative) and 1 month postoperatively
Surgical Procedure Assessment
Tijdsspanne: Intraoperative

Surgeon-reported ease of lens implantation and accuracy of lens placement, evaluating whether expectations for procedural handling were met.The surgeon notes, during the operative visit, the location of the incision, the size of the incision, whether the implantation system is compatible with the wound site, how many seconds the implantation time is shortened, and whether there were any issues with the unfolding of the intraocular lens or its placement into the capsular bag.

Intraoperative

Andere uitkomstmaten

Uitkomstmaat
Maatregel Beschrijving
Tijdsspanne
Serious Adverse Events
Tijdsspanne: 1 month postoperatively

Any serious adverse events occurring during surgery or within 1 month after surgery.

1 month postoperatively
Postoperative Inflammation
Tijdsspanne: 1 week and 1 month postoperatively

Assessment of inflammation levels by biomicroscopic anterior chamber and vitreus examination according to Standardization of Uveitis Nomenclature (SUN) International Workshop Grading System

1 week and 1 month postoperatively
Intraocular Pressure,
Tijdsspanne: 1 week and 1 month postoperatively

Intraocular pressure is assessed via Goldmann Applanation Tonometry. Normal levels are accepted as 10-21 mmHG.

1 week and 1 month postoperatively

Medewerkers en onderzoekers

Hier vindt u mensen en organisaties die betrokken zijn bij dit onderzoek.

Sponsor

Publicaties en nuttige links

De persoon die verantwoordelijk is voor het invoeren van informatie over het onderzoek stelt deze publicaties vrijwillig ter beschikking. Dit kan gaan over alles wat met het onderzoek te maken heeft.

Studie record data

Deze datums volgen de voortgang van het onderzoeksdossier en de samenvatting van de ingediende resultaten bij ClinicalTrials.gov. Studieverslagen en gerapporteerde resultaten worden beoordeeld door de National Library of Medicine (NLM) om er zeker van te zijn dat ze voldoen aan specifieke kwaliteitscontrolenormen voordat ze op de openbare website worden geplaatst.

Bestudeer belangrijke data

Studie start (Werkelijk)

10 mei 2025

Primaire voltooiing (Werkelijk)

1 juni 2026

Studie voltooiing (Werkelijk)

1 juli 2026

Studieregistratiedata

Eerst ingediend

26 augustus 2025

Eerst ingediend dat voldeed aan de QC-criteria

3 december 2025

Eerst geplaatst (Werkelijk)

17 december 2025

Updates van studierecords

Laatste update geplaatst (Werkelijk)

4 november 2026

Laatste update ingediend die voldeed aan QC-criteria

3 november 2026

Laatst geverifieerd

1 november 2026

Meer informatie

Termen gerelateerd aan deze studie

Andere studie-ID-nummers

  • IstanbulUni

Plan Individuele Deelnemersgegevens (IPD)

Bent u van plan om gegevens van individuele deelnemers (IPD) te delen?

ONBESLIST

Beschrijving IPD-plan

Individual participant data will not be shared outside the study.

Informatie over medicijnen en apparaten, studiedocumenten

Bestudeert een door de Amerikaanse FDA gereguleerd geneesmiddel

Nee

Bestudeert een door de Amerikaanse FDA gereguleerd apparaatproduct

Nee

Deze informatie is zonder wijzigingen rechtstreeks van de website clinicaltrials.gov gehaald. Als u verzoeken heeft om uw onderzoeksgegevens te wijzigen, te verwijderen of bij te werken, neem dan contact op met register@clinicaltrials.gov. Zodra er een wijziging wordt doorgevoerd op clinicaltrials.gov, wordt deze ook automatisch bijgewerkt op onze website .