Frequency- and Intensity-Dependent After-Effects of High-Definition Transcranial Alternating Current Stimulation on Resting-State Electroencephalography
Frequency- and Intensity-Dependent After-Effects of High-Definition Transcranial Alternating Current Stimulation on Resting-State Electroencephalography in Patients With Auditory Hallucinations and Healthy Controls
Studieoversikt
Status
Status
Forhold
Forhold
Intervensjon / Behandling
Intervensjon / Behandling
Studietype
Studietype
Registrering (Faktiske)
Registrering
Fase
Fase
- Ikke aktuelt
Kontakter og plasseringer
Studiesteder
-
-
-
Fuyang, Kina
- Fuyang Mental Heath Center
-
-
Deltakelseskriterier
Kvalifikasjonskriterier
Kvalifikasjonskriterier
Alder som er kvalifisert for studier
- Voksen
- Eldre voksen
Tar imot friske frivillige
Beskrivelse
Inclusion Criteria:
- Age 18 to 65 years;
- Meet the diagnostic criteria for schizophrenia according to the Diagnostic and Statistical Manual of Mental Disorders, Fifth Edition (DSM-5);
- Adequate verbal communication skills;
- At least one auditory hallucination episode within the past week;
- Willing to participate in the study and able to complete all trial procedures.
Exclusion Criteria:
- History of neurological disorders, severe physical illness, traumatic brain injury, or craniotomy;
- History of epileptic seizures or other neuropsychiatric disorders;
- At risk for suicide, self-injury, or impulsive behavior.
Studieplan
Hvordan er studiet utformet?
Designdetaljer
- Primært formål: Behandling
- Tildeling: Randomisert
- Intervensjonsmodell: Crossover-oppdrag
- Masking: Dobbelt
Antall våpen
Våpen og intervensjoner
Deltakergruppe / ArmDeltakergruppe / Arm |
Intervensjon / BehandlingIntervensjon / Behandling |
|---|---|
|
Eksperimentell: 0.4mA-10Hz HD-tACS
For 0.4mA-10Hz HD-tACS prevention, we will stimulate the personalized left or right temporal lobe channel at 10 Hz and 0.4mA.
|
The active intervention consists of high-definition transcranial alternating current stimulation (HD-tACS) delivered using a 4×1 ring configuration (one center electrode surrounded by four return electrodes) over the personalized left or right temporal lobe.
The return electrodes deliver one-quarter of the central electrode current with opposite polarity.
Stimulation is administered for 90 seconds per condition, with the frequency and current intensity set according to the experimental protocol.
A 10-second ramp-up and a 5-second ramp-down period are included in each stimulation condition.
The stimulus waveform is sinusoidal.
Stimulation is delivered using saline-soaked sponge electrodes.
Electrode impedance is maintained below 10 kΩ throughout the stimulation session.
|
|
Eksperimentell: 0.4mA-20Hz HD-tACS
For 0.4mA-20Hz HD-tACS prevention, we will stimulate the personalized left or right temporal lobe channel at 20 Hz and 0.4mA.
|
The active intervention consists of high-definition transcranial alternating current stimulation (HD-tACS) delivered using a 4×1 ring configuration (one center electrode surrounded by four return electrodes) over the personalized left or right temporal lobe.
The return electrodes deliver one-quarter of the central electrode current with opposite polarity.
Stimulation is administered for 90 seconds per condition, with the frequency and current intensity set according to the experimental protocol.
A 10-second ramp-up and a 5-second ramp-down period are included in each stimulation condition.
The stimulus waveform is sinusoidal.
Stimulation is delivered using saline-soaked sponge electrodes.
Electrode impedance is maintained below 10 kΩ throughout the stimulation session.
|
|
Eksperimentell: 0.4mA-40Hz HD-tACS
For 0.4mA-40Hz HD-tACS prevention, we will stimulate the personalized left or right temporal lobe channel at 20 Hz and 0.4mA.
|
The active intervention consists of high-definition transcranial alternating current stimulation (HD-tACS) delivered using a 4×1 ring configuration (one center electrode surrounded by four return electrodes) over the personalized left or right temporal lobe.
The return electrodes deliver one-quarter of the central electrode current with opposite polarity.
Stimulation is administered for 90 seconds per condition, with the frequency and current intensity set according to the experimental protocol.
A 10-second ramp-up and a 5-second ramp-down period are included in each stimulation condition.
The stimulus waveform is sinusoidal.
Stimulation is delivered using saline-soaked sponge electrodes.
Electrode impedance is maintained below 10 kΩ throughout the stimulation session.
|
|
Eksperimentell: 0.4mA-140Hz HD-tACS
For 0.4mA-140Hz HD-tACS prevention, we will stimulate the personalized left or right temporal lobe channel at 10 Hz and 0.4mA.
|
The active intervention consists of high-definition transcranial alternating current stimulation (HD-tACS) delivered using a 4×1 ring configuration (one center electrode surrounded by four return electrodes) over the personalized left or right temporal lobe.
The return electrodes deliver one-quarter of the central electrode current with opposite polarity.
Stimulation is administered for 90 seconds per condition, with the frequency and current intensity set according to the experimental protocol.
A 10-second ramp-up and a 5-second ramp-down period are included in each stimulation condition.
The stimulus waveform is sinusoidal.
Stimulation is delivered using saline-soaked sponge electrodes.
Electrode impedance is maintained below 10 kΩ throughout the stimulation session.
|
|
Eksperimentell: 1mA-10Hz HD-tACS
For 1mA-10Hz HD-tACS prevention, we will stimulate the personalized left or right temporal lobe channel at 10 Hz and 0.4mA.
|
The active intervention consists of high-definition transcranial alternating current stimulation (HD-tACS) delivered using a 4×1 ring configuration (one center electrode surrounded by four return electrodes) over the personalized left or right temporal lobe.
The return electrodes deliver one-quarter of the central electrode current with opposite polarity.
Stimulation is administered for 90 seconds per condition, with the frequency and current intensity set according to the experimental protocol.
A 10-second ramp-up and a 5-second ramp-down period are included in each stimulation condition.
The stimulus waveform is sinusoidal.
Stimulation is delivered using saline-soaked sponge electrodes.
Electrode impedance is maintained below 10 kΩ throughout the stimulation session.
|
|
Eksperimentell: 1mA-20Hz HD-tACS
For 1mA-20Hz HD-tACS prevention, we will stimulate the personalized left or right temporal lobe channel at 10 Hz and 0.4mA.
|
The active intervention consists of high-definition transcranial alternating current stimulation (HD-tACS) delivered using a 4×1 ring configuration (one center electrode surrounded by four return electrodes) over the personalized left or right temporal lobe.
The return electrodes deliver one-quarter of the central electrode current with opposite polarity.
Stimulation is administered for 90 seconds per condition, with the frequency and current intensity set according to the experimental protocol.
A 10-second ramp-up and a 5-second ramp-down period are included in each stimulation condition.
The stimulus waveform is sinusoidal.
Stimulation is delivered using saline-soaked sponge electrodes.
Electrode impedance is maintained below 10 kΩ throughout the stimulation session.
|
|
Eksperimentell: 1mA-40Hz HD-tACS
For 1mA-40Hz HD-tACS prevention, we will stimulate the personalized left or right temporal lobe channel at 10 Hz and 0.4mA.
|
The active intervention consists of high-definition transcranial alternating current stimulation (HD-tACS) delivered using a 4×1 ring configuration (one center electrode surrounded by four return electrodes) over the personalized left or right temporal lobe.
The return electrodes deliver one-quarter of the central electrode current with opposite polarity.
Stimulation is administered for 90 seconds per condition, with the frequency and current intensity set according to the experimental protocol.
A 10-second ramp-up and a 5-second ramp-down period are included in each stimulation condition.
The stimulus waveform is sinusoidal.
Stimulation is delivered using saline-soaked sponge electrodes.
Electrode impedance is maintained below 10 kΩ throughout the stimulation session.
|
|
Eksperimentell: 1mA-140Hz HD-tACS
For 1mA-140Hz HD-tACS prevention, we will stimulate the personalized left or right temporal lobe channel at 10 Hz and 0.4mA.
|
The active intervention consists of high-definition transcranial alternating current stimulation (HD-tACS) delivered using a 4×1 ring configuration (one center electrode surrounded by four return electrodes) over the personalized left or right temporal lobe.
The return electrodes deliver one-quarter of the central electrode current with opposite polarity.
Stimulation is administered for 90 seconds per condition, with the frequency and current intensity set according to the experimental protocol.
A 10-second ramp-up and a 5-second ramp-down period are included in each stimulation condition.
The stimulus waveform is sinusoidal.
Stimulation is delivered using saline-soaked sponge electrodes.
Electrode impedance is maintained below 10 kΩ throughout the stimulation session.
|
|
Eksperimentell: 2mA-10Hz HD-tACS
For 2mA-10Hz HD-tACS prevention, we will stimulate the personalized left or right temporal lobe channel at 10 Hz and 0.4mA.
|
The active intervention consists of high-definition transcranial alternating current stimulation (HD-tACS) delivered using a 4×1 ring configuration (one center electrode surrounded by four return electrodes) over the personalized left or right temporal lobe.
The return electrodes deliver one-quarter of the central electrode current with opposite polarity.
Stimulation is administered for 90 seconds per condition, with the frequency and current intensity set according to the experimental protocol.
A 10-second ramp-up and a 5-second ramp-down period are included in each stimulation condition.
The stimulus waveform is sinusoidal.
Stimulation is delivered using saline-soaked sponge electrodes.
Electrode impedance is maintained below 10 kΩ throughout the stimulation session.
|
|
Eksperimentell: 2mA-20Hz HD-tACS
For 2mA-20Hz HD-tACS prevention, we will stimulate the personalized left or right temporal lobe channel at 10 Hz and 0.4mA.
|
The active intervention consists of high-definition transcranial alternating current stimulation (HD-tACS) delivered using a 4×1 ring configuration (one center electrode surrounded by four return electrodes) over the personalized left or right temporal lobe.
The return electrodes deliver one-quarter of the central electrode current with opposite polarity.
Stimulation is administered for 90 seconds per condition, with the frequency and current intensity set according to the experimental protocol.
A 10-second ramp-up and a 5-second ramp-down period are included in each stimulation condition.
The stimulus waveform is sinusoidal.
Stimulation is delivered using saline-soaked sponge electrodes.
Electrode impedance is maintained below 10 kΩ throughout the stimulation session.
|
|
Eksperimentell: 2mA-40Hz HD-tACS
For 2mA-40Hz HD-tACS prevention, we will stimulate the personalized left or right temporal lobe channel at 10 Hz and 0.4mA.
|
The active intervention consists of high-definition transcranial alternating current stimulation (HD-tACS) delivered using a 4×1 ring configuration (one center electrode surrounded by four return electrodes) over the personalized left or right temporal lobe.
The return electrodes deliver one-quarter of the central electrode current with opposite polarity.
Stimulation is administered for 90 seconds per condition, with the frequency and current intensity set according to the experimental protocol.
A 10-second ramp-up and a 5-second ramp-down period are included in each stimulation condition.
The stimulus waveform is sinusoidal.
Stimulation is delivered using saline-soaked sponge electrodes.
Electrode impedance is maintained below 10 kΩ throughout the stimulation session.
|
|
Eksperimentell: 2mA-140Hz HD-tACS
For 2mA-140Hz HD-tACS prevention, we will stimulate the personalized left or right temporal lobe channel at 10 Hz and 0.4mA.
|
The active intervention consists of high-definition transcranial alternating current stimulation (HD-tACS) delivered using a 4×1 ring configuration (one center electrode surrounded by four return electrodes) over the personalized left or right temporal lobe.
The return electrodes deliver one-quarter of the central electrode current with opposite polarity.
Stimulation is administered for 90 seconds per condition, with the frequency and current intensity set according to the experimental protocol.
A 10-second ramp-up and a 5-second ramp-down period are included in each stimulation condition.
The stimulus waveform is sinusoidal.
Stimulation is delivered using saline-soaked sponge electrodes.
Electrode impedance is maintained below 10 kΩ throughout the stimulation session.
|
|
Eksperimentell: sham HD-tACS
For sham HD-tACS, we will stimulate the personalized left or right temporal lobe channel using sham HD-tACS pattern.
|
The sham intervention consists of high-definition transcranial alternating current stimulation (HD-tACS) delivered using a 4×1 ring configuration (one center electrode surrounded by four return electrodes) over the personalized left or right temporal lobe.
The return electrodes deliver one-quarter of the central electrode current with opposite polarity.
To mimic the sensory experience of active stimulation, the sham condition includes a 10-second ramp-up and a 5-second ramp-down period at 0.4 mA and 140 Hz, with no current delivered during the intervening 90-second period.
|
Hva måler studien?
Primære resultatmål
Primære resultatmål
Resultatmål |
Tiltaksbeskrivelse |
Tidsramme |
|---|---|---|
|
Beta2-band EEG power at the stimulation site
Tidsramme: 180-second recording after each 90-second stimulation
|
Changes in EEG power(dB) in beta2 band within different frequency band measured at the temporal lobe during the 180-second resting-state EEG recording immediately following each stimulation condition.
|
180-second recording after each 90-second stimulation
|
Sekundære resultatmål
Sekundære resultatmål
Resultatmål |
Tiltaksbeskrivelse |
Tidsramme |
|---|---|---|
|
Alpha-band EEG power at the stimulation site
Tidsramme: 180-second recording after each 90-second stimulation
|
Changes in EEG power(dB) in alpha band within different frequency band measured at the temporal lobe during the 180-second resting-state EEG recording immediately following each stimulation condition.
|
180-second recording after each 90-second stimulation
|
|
Delta-band EEG power at the stimulation site
Tidsramme: 180-second recording after each 90-second stimulation
|
Changes in EEG power(dB) in delta band within different frequency band measured at the temporal lobe during the 180-second resting-state EEG recording immediately following each stimulation condition.
|
180-second recording after each 90-second stimulation
|
|
Theta-band EEG power at the stimulation site
Tidsramme: 180-second recording after each 90-second stimulation
|
Changes in EEG power(dB) in theta band within different frequency band measured at the temporal lobe during the 180-second resting-state EEG recording immediately following each stimulation condition.
|
180-second recording after each 90-second stimulation
|
|
Gamma-band EEG power at the stimulation site
Tidsramme: 180-second recording after each 90-second stimulation
|
Changes in EEG power(dB) in gamma band within different frequency band measured at the temporal lobe during the 180-second resting-state EEG recording immediately following each stimulation condition.
|
180-second recording after each 90-second stimulation
|
|
Beta2-band functional connectivity measured by EEG wPLI
Tidsramme: 180-second recording after each 90-second stimulation
|
Beta2-band functional connectivity between predefined EEG electrodes, quantified using the weighted Phase Lag Index (wPLI), during the 180-second resting-state EEG recording immediately following each stimulation condition.
|
180-second recording after each 90-second stimulation
|
|
Alpha-band functional connectivity measured by EEG wPLI
Tidsramme: 180-second recording after each 90-second stimulation
|
Alpha-band functional connectivity between predefined EEG electrodes, quantified using the weighted Phase Lag Index (wPLI), during the 180-second resting-state EEG recording immediately following each stimulation condition.
|
180-second recording after each 90-second stimulation
|
|
Delta-band functional connectivity measured by EEG wPLI
Tidsramme: 180-second recording after each 90-second stimulation
|
Delta-band functional connectivity between predefined EEG electrodes, quantified using the weighted Phase Lag Index (wPLI), during the 180-second resting-state EEG recording immediately following each stimulation condition.
|
180-second recording after each 90-second stimulation
|
|
Theta-band functional connectivity measured by EEG wPLI
Tidsramme: 180-second recording after each 90-second stimulation
|
Theta-band functional connectivity between predefined EEG electrodes, quantified using the weighted Phase Lag Index (wPLI), during the 180-second resting-state EEG recording immediately following each stimulation condition.
|
180-second recording after each 90-second stimulation
|
|
Gamma-band functional connectivity measured by EEG wPLI
Tidsramme: 180-second recording after each 90-second stimulation
|
Gamma-band functional connectivity between predefined EEG electrodes, quantified using the weighted Phase Lag Index (wPLI), during the 180-second resting-state EEG recording immediately following each stimulation condition.
|
180-second recording after each 90-second stimulation
|
Samarbeidspartnere og etterforskere
Sponsor
Sponsor
Studierekorddatoer
Studer hoveddatoer
Studiestart (Faktiske)
Studiestart
Primær fullføring (Faktiske)
Primær fullføring
Studiet fullført (Faktiske)
Studiet fullført
Datoer for studieregistrering
Først innsendt
Først innsendt
Først innsendt som oppfylte QC-kriteriene
Først innsendt som oppfylte QC-kriteriene
Først lagt ut (Faktiske)
Først lagt ut
Oppdateringer av studieposter
Sist oppdatering lagt ut (Faktiske)
Sist oppdatering lagt ut
Siste oppdatering sendt inn som oppfylte QC-kriteriene
Siste oppdatering sendt inn som oppfylte QC-kriteriene
Sist bekreftet
Sist bekreftet
Mer informasjon
Begreper knyttet til denne studien
Andre studie-ID-numre
Andre studie-ID-numre
- HFJiang-006
Plan for individuelle deltakerdata (IPD)
Planlegger du å dele individuelle deltakerdata (IPD)?
IPD-planbeskrivelse
Legemiddel- og utstyrsinformasjon, studiedokumenter
Studerer et amerikansk FDA-regulert medikamentprodukt
Studerer et amerikansk FDA-regulert enhetsprodukt
Denne informasjonen ble hentet direkte fra nettstedet clinicaltrials.gov uten noen endringer. Hvis du har noen forespørsler om å endre, fjerne eller oppdatere studiedetaljene dine, vennligst kontakt register@clinicaltrials.gov. Så snart en endring er implementert på clinicaltrials.gov, vil denne også bli oppdatert automatisk på nettstedet vårt. .