Análise biomecânica da córnea usando microscopia de Brillouin
The objective of this study is to measure the Brillouin biomechanical properties in keratoconic corneas and characterize biomechanical alterations that occur after corneal procedures that inherently strengthen or weaken the cornea by evaluating the change in Brillouin metrics before and after treatments.
Visão geral do estudo
Status
Status
Condições
Condições
Intervenção / Tratamento
Intervenção / Tratamento
Descrição detalhada
Surgical correction of myopia and keratoconus identification/management are separate but tightly intertwined issues of major significance. For both, there is an unmet need for direct measurements to evaluate corneal stiffness (i.e. its resistance to deformation). The prevalence of myopia is expected to double, affecting more than 50% of the US population, by 2050. Laser in situ keratomileusis (LASIK) is one of the most popular and successful surgeries in the world and compares favorably to long-term contact lens wear use for myopia correction. However, only ~10% of eligible patients undergo LASIK currently; the others cite safety concerns as a major factor in their decision. The primary risk for poor refractive surgery outcomes is biomechanical failure due to unidentified (subclinical) ectasia (i.e. keratoconus). Patients presenting for LASIK evaluation with atypical, suspicious corneal curvature but with undetermined true risk represent the leading reason for surgery screening failures. This results in good candidates being denied surgery, while up to 10% of truly poor candidates are still missed using current screening algorithms.
Keratoconus is up to 10 times more prevalent than the previously reported 1/2000 figure. Corneal cross-linking (CXL) is now FDA approved in the US for keratoconus treatment and is effective at stiffening the cornea and halting ectasia progression. Early identification of keratoconus is critical, but current tests in the clinic are morphological, not biomechanical, and therefore do not allow a definitive diagnosis at the earliest stages resulting in vision loss before CXL treatment is initiated. Thus, the need for accurate identification of subclinical ectasia has never been greater.
In the past years, newly developed technology, Brillouin microscopy, has emerged as the most promising tool to address this clinical need. This study will systemically address the critical gap in current knowledge by linking Brillouin mapping of corneal biomechanical alterations to abnormal morphological behavior and testing the findings in conditions where corneal biomechanics are abruptly altered, by: 1) weakening with refractive surgery procedures, and 2) strengthening through corneal cross-linking.
It is anticipated that a clinical tool assessing the mechanical state of the cornea will improve early diagnosis and management of keratoconus as well as refractive surgery planning. Ultimately, this will lead to predictive models where Brillouin measurements could be an accurate predictor of postoperative outcomes and thus aid in developing individualized surgical parameters.
Tipo de estudo
Tipo de estudo
Inscrição (Estimado)
Inscrição
Contactos e Locais
Locais de estudo
-
-
Ohio
-
Cleveland, Ohio, Estados Unidos, 44195
- Cleveland Clinic Cole Eye Institute
-
-
Critérios de participação
Critérios de elegibilidade
Critérios de elegibilidade
Idades elegíveis para estudo
Aceita Voluntários Saudáveis
Gêneros Elegíveis para o Estudo
Método de amostragem
População do estudo
Group 1: Patients with normal corneas Group 2: Patients with keratoconus Group 3: Patients with normal corneas undergoing LASIK Group 4: Patients with normal corneas undergoing PRK Group 5: Patients with normal corneas undergoing SMILE Group 6: Patients with keratoconus undergoing CXL
Descrição
Inclusion Criteria:
- patients aged 18-60 with keratoconus
- patients aged 18-60 with normal corneas,
- patients aged 18-60 undergoing refractive surgery (LASIK, PRK, SMILE)
- patients aged 18-60 with keratoconus undergoing CXL
Exclusion Criteria:
- outside age range
- history of previous ocular surgeries
- unable to cooperate for the Brillouin microscopic examination
- unable to provide informed consent
Plano de estudo
Como o estudo é projetado?
Detalhes do projeto
- Modelos de observação: CASE_CONTROL
- Perspectivas de Tempo: CROSS_SECTIONAL
Número de grupos/coortes
Coortes e Intervenções
Grupo / CoorteGrupo / Coorte |
Intervenção / TratamentoIntervenção / Tratamento |
|---|---|
|
1: Normal Controls
Patients with normal corneas without any prior surgery to serve as the control group |
The Brillouin clinical instrument is comprised of three parts: a human interface, a laser-scanning confocal microscope, and an etalon-based spectrometer. The human interface is a modified ophthalmic slit-lamp instrument with chin support and headrest. The light source is a single longitudinal mode CW laser at 780 nm. A polarizing beam splitter and quarter-wave plate assembly sends the laser beam to the human interface. To focus light into the eye, a long-working distance microscope objective is used. Brillouin scattered light from the eye is collected with a single-mode optical fiber. For spectral analysis, a two-stage VIPA-etalon spectrometer configured with the cross-axis cascade principle and the spectrum is measured on a EM-CCD camera. |
|
2 Keratoconus
Patients with various stages of keratoconus |
The Brillouin clinical instrument is comprised of three parts: a human interface, a laser-scanning confocal microscope, and an etalon-based spectrometer. The human interface is a modified ophthalmic slit-lamp instrument with chin support and headrest. The light source is a single longitudinal mode CW laser at 780 nm. A polarizing beam splitter and quarter-wave plate assembly sends the laser beam to the human interface. To focus light into the eye, a long-working distance microscope objective is used. Brillouin scattered light from the eye is collected with a single-mode optical fiber. For spectral analysis, a two-stage VIPA-etalon spectrometer configured with the cross-axis cascade principle and the spectrum is measured on a EM-CCD camera. |
|
3: LASIK
Patients with normal corneas who are undergoing laser in situ keratomileusis (LASIK) |
The Brillouin clinical instrument is comprised of three parts: a human interface, a laser-scanning confocal microscope, and an etalon-based spectrometer. The human interface is a modified ophthalmic slit-lamp instrument with chin support and headrest. The light source is a single longitudinal mode CW laser at 780 nm. A polarizing beam splitter and quarter-wave plate assembly sends the laser beam to the human interface. To focus light into the eye, a long-working distance microscope objective is used. Brillouin scattered light from the eye is collected with a single-mode optical fiber. For spectral analysis, a two-stage VIPA-etalon spectrometer configured with the cross-axis cascade principle and the spectrum is measured on a EM-CCD camera. |
|
Group 4: PRK
Patients with normal corneas who are undergoing photorefractive keratectomy (PRK) |
The Brillouin clinical instrument is comprised of three parts: a human interface, a laser-scanning confocal microscope, and an etalon-based spectrometer. The human interface is a modified ophthalmic slit-lamp instrument with chin support and headrest. The light source is a single longitudinal mode CW laser at 780 nm. A polarizing beam splitter and quarter-wave plate assembly sends the laser beam to the human interface. To focus light into the eye, a long-working distance microscope objective is used. Brillouin scattered light from the eye is collected with a single-mode optical fiber. For spectral analysis, a two-stage VIPA-etalon spectrometer configured with the cross-axis cascade principle and the spectrum is measured on a EM-CCD camera. |
|
5: SMILE
Patients with normal corneas who are undergoing small incision lenticular extraction (SMILE) |
The Brillouin clinical instrument is comprised of three parts: a human interface, a laser-scanning confocal microscope, and an etalon-based spectrometer. The human interface is a modified ophthalmic slit-lamp instrument with chin support and headrest. The light source is a single longitudinal mode CW laser at 780 nm. A polarizing beam splitter and quarter-wave plate assembly sends the laser beam to the human interface. To focus light into the eye, a long-working distance microscope objective is used. Brillouin scattered light from the eye is collected with a single-mode optical fiber. For spectral analysis, a two-stage VIPA-etalon spectrometer configured with the cross-axis cascade principle and the spectrum is measured on a EM-CCD camera. |
|
6: CXL
Patients with keratoconus who are undergoing corneal cross-linking (CXL) |
The Brillouin clinical instrument is comprised of three parts: a human interface, a laser-scanning confocal microscope, and an etalon-based spectrometer. The human interface is a modified ophthalmic slit-lamp instrument with chin support and headrest. The light source is a single longitudinal mode CW laser at 780 nm. A polarizing beam splitter and quarter-wave plate assembly sends the laser beam to the human interface. To focus light into the eye, a long-working distance microscope objective is used. Brillouin scattered light from the eye is collected with a single-mode optical fiber. For spectral analysis, a two-stage VIPA-etalon spectrometer configured with the cross-axis cascade principle and the spectrum is measured on a EM-CCD camera. |
O que o estudo está medindo?
Medidas de resultados primários
Medidas de resultados primários
Medida de resultado |
Descrição da medida |
Prazo |
|---|---|---|
|
Change in Brillouin Metrics
Prazo: Difference between baseline and 3 months after intervention
|
Brillouin metrics to be evaluated include localized Mean Brillouin modulus measure across the cornea and at each depth of the corneal stroma |
Difference between baseline and 3 months after intervention
|
Colaboradores e Investigadores
Patrocinador
Patrocinador
Colaboradores
Colaboradores
Publicações e links úteis
Publicações Gerais
- Zhang H, Roozbahani M, Piccinini AL, Golan O, Hafezi F, Scarcelli G, Randleman JB. Depth-Dependent Reduction of Biomechanical Efficacy of Contact Lens-Assisted Corneal Cross-linking Analyzed by Brillouin Microscopy. J Refract Surg. 2019 Nov 1;35(11):721-728. doi: 10.3928/1081597X-20191004-01.
- Webb JN, Langille E, Hafezi F, Randleman JB, Scarcelli G. Biomechanical Impact of Localized Corneal Cross-linking Beyond the Irradiated Treatment Area. J Refract Surg. 2019 Apr 1;35(4):253-260. doi: 10.3928/1081597X-20190304-01.
- Randleman JB, Su JP, Scarcelli G. Biomechanical Changes After LASIK Flap Creation Combined With Rapid Cross-Linking Measured With Brillouin Microscopy. J Refract Surg. 2017 Jun 1;33(6):408-414. doi: 10.3928/1081597X-20170421-01.
- Scarcelli G, Pineda R, Yun SH. Brillouin optical microscopy for corneal biomechanics. Invest Ophthalmol Vis Sci. 2012 Jan 20;53(1):185-90. doi: 10.1167/iovs.11-8281.
- Scarcelli G, Kling S, Quijano E, Pineda R, Marcos S, Yun SH. Brillouin microscopy of collagen crosslinking: noncontact depth-dependent analysis of corneal elastic modulus. Invest Ophthalmol Vis Sci. 2013 Feb 19;54(2):1418-25. doi: 10.1167/iovs.12-11387.
- Scarcelli G, Besner S, Pineda R, Yun SH. Biomechanical characterization of keratoconus corneas ex vivo with Brillouin microscopy. Invest Ophthalmol Vis Sci. 2014 Jun 17;55(7):4490-5. doi: 10.1167/iovs.14-14450.
- Scarcelli G, Besner S, Pineda R, Kalout P, Yun SH. In vivo biomechanical mapping of normal and keratoconus corneas. JAMA Ophthalmol. 2015 Apr;133(4):480-2. doi: 10.1001/jamaophthalmol.2014.5641. No abstract available.
Datas de registro do estudo
Datas Principais do Estudo
Início do estudo (Real)
Início do estudo
Conclusão Primária (Estimado)
Conclusão Primária
Conclusão do estudo (Estimado)
Conclusão do estudo
Datas de inscrição no estudo
Enviado pela primeira vez
Enviado pela primeira vez
Enviado pela primeira vez que atendeu aos critérios de CQ
Enviado pela primeira vez que atendeu aos critérios de CQ
Primeira postagem (Real)
Primeira postagem
Atualizações de registro de estudo
Última Atualização Postada (Real)
Última Atualização Postada
Última atualização enviada que atendeu aos critérios de controle de qualidade
Última atualização enviada que atendeu aos critérios de controle de qualidade
Última verificação
Última verificação
Mais Informações
Termos relacionados a este estudo
Palavras-chave
Termos MeSH relevantes adicionais
Outros números de identificação do estudo
Outros números de identificação do estudo
- 20-355
Plano para dados de participantes individuais (IPD)
Planeja compartilhar dados de participantes individuais (IPD)?
Descrição do plano IPD
there is no plan to share data
Informações sobre medicamentos e dispositivos, documentos de estudo
Estuda um medicamento regulamentado pela FDA dos EUA
Estuda um produto de dispositivo regulamentado pela FDA dos EUA
produto fabricado e exportado dos EUA
Essas informações foram obtidas diretamente do site clinicaltrials.gov sem nenhuma alteração. Se você tiver alguma solicitação para alterar, remover ou atualizar os detalhes do seu estudo, entre em contato com register@clinicaltrials.gov. Assim que uma alteração for implementada em clinicaltrials.gov, ela também será atualizada automaticamente em nosso site .