LMA Protector vs I-Gel in Low-Flow Anesthesia for Laparoscopic Cholecystectomy
Comparison of LMA Protector and I-Gel Use in Low-Flow Anesthesia for Laparoscopic Cholecystectomy: A Prospective Randomized Study
Visão geral do estudo
Status
Status
Condições
Condições
Intervenção / Tratamento
Intervenção / Tratamento
Descrição detalhada
Tipo de estudo
Tipo de estudo
Inscrição (Real)
Inscrição
Estágio
Estágio
- Não aplicável
Contactos e Locais
Locais de estudo
-
-
-
Istanbul, Turquia (Türkiye), 34752
- Fatih Sultan Mehmet Training and Research Hospital
-
-
Critérios de participação
Critérios de elegibilidade
Critérios de elegibilidade
Idades elegíveis para estudo
- Adulto
- Adulto mais velho
Aceita Voluntários Saudáveis
Descrição
Inclusion Criteria:
- Patients aged 18-65 years
- American Society of Anesthesiologists (ASA) physical status I or II
- Scheduled for elective laparoscopic cholecystectomy
Exclusion Criteria:
- Known or predicted difficult airway or history of difficult intubation
- Body Mass Index (BMI) greater than 35 kg/m²
- High risk of gastroesophageal reflux or aspiration
- Severe cardiac, respiratory, renal, or hepatic disease
- Pregnancy
- Patient refusal to participate
Plano de estudo
Como o estudo é projetado?
Detalhes do projeto
- Finalidade Principal: Tratamento
- Alocação: Randomizado
- Modelo Intervencional: Atribuição Paralela
- Mascaramento: Solteiro
Número de braços
Armas e Intervenções
Grupo de Participantes / BraçoGrupo de Participantes / Braço |
Intervenção / TratamentoIntervenção / Tratamento |
|---|---|
|
Experimental: Group I: I-Gel
Patients managed with the I-Gel supraglottic airway device during low-flow desflurane anesthesia for laparoscopic cholecystectomy.
|
Placement of an appropriate size of i-gel supraglottic airway device for airway management during general anesthesia.
Anesthesia maintenance is provided with desflurane in a fresh gas flow of 1 L/min (low-flow anesthesia).
Outros nomes:
|
|
Experimental: Group P: LMA Protector
Patients managed with the LMA Protector supraglottic airway device during low-flow desflurane anesthesia for laparoscopic cholecystectomy.
|
Placement of an appropriate size of LMA Protector supraglottic airway device for airway management during general anesthesia.
Anesthesia maintenance is provided with desflurane in a fresh gas flow of 1 L/min (low-flow anesthesia).
Outros nomes:
|
O que o estudo está medindo?
Medidas de resultados primários
Medidas de resultados primários
Medida de resultado |
Descrição da medida |
Prazo |
|---|---|---|
|
Dynamic Lung Compliance (Cdyn)
Prazo: Measured intraoperatively at 6 specific time points from 5 minutes after device insertion until immediately before device removal.
|
Dynamic compliance of the respiratory system, calculated automatically by the ventilator as tidal volume divided by (peak inspiratory pressure minus positive end-expiratory pressure).
|
Measured intraoperatively at 6 specific time points from 5 minutes after device insertion until immediately before device removal.
|
Medidas de resultados secundários
Medidas de resultados secundários
Medida de resultado |
Descrição da medida |
Prazo |
|---|---|---|
|
Peak Inspiratory Pressure (PIP)
Prazo: Measured intraoperatively at 6 specific time points from 5 minutes after device insertion until immediately before device removal.
|
Maximum airway pressure generated during the inspiratory phase, monitored to assess airway dynamics under general anesthesia.
|
Measured intraoperatively at 6 specific time points from 5 minutes after device insertion until immediately before device removal.
|
|
Plateau Pressure (Pplateau)
Prazo: Measured intraoperatively at 6 specific time points from 5 minutes after device insertion until immediately before device removal.
|
Pressure applied to the small airways and alveoli during a period of zero gas flow, monitored to assess risk of barotrauma.
|
Measured intraoperatively at 6 specific time points from 5 minutes after device insertion until immediately before device removal.
|
|
Airway Resistance
Prazo: Measured intraoperatively at 6 specific time points from 5 minutes after device insertion until immediately before device removal.
|
The resistance of the respiratory tract to airflow during mechanical ventilation, calculated using peak and plateau pressures.
|
Measured intraoperatively at 6 specific time points from 5 minutes after device insertion until immediately before device removal.
|
Colaboradores e Investigadores
Patrocinador
Patrocinador
Colaboradores
Colaboradores
Datas de registro do estudo
Datas Principais do Estudo
Início do estudo (Real)
Início do estudo
Conclusão Primária (Real)
Conclusão Primária
Conclusão do estudo (Real)
Conclusão do estudo
Datas de inscrição no estudo
Enviado pela primeira vez
Enviado pela primeira vez
Enviado pela primeira vez que atendeu aos critérios de CQ
Enviado pela primeira vez que atendeu aos critérios de CQ
Primeira postagem (Real)
Primeira postagem
Atualizações de registro de estudo
Última Atualização Postada (Real)
Última Atualização Postada
Última atualização enviada que atendeu aos critérios de controle de qualidade
Última atualização enviada que atendeu aos critérios de controle de qualidade
Última verificação
Última verificação
Mais Informações
Termos relacionados a este estudo
Palavras-chave
Outros números de identificação do estudo
Outros números de identificação do estudo
- FSM KAEK-2020/79
Plano para dados de participantes individuais (IPD)
Planeja compartilhar dados de participantes individuais (IPD)?
Informações sobre medicamentos e dispositivos, documentos de estudo
Estuda um medicamento regulamentado pela FDA dos EUA
Estuda um produto de dispositivo regulamentado pela FDA dos EUA
Essas informações foram obtidas diretamente do site clinicaltrials.gov sem nenhuma alteração. Se você tiver alguma solicitação para alterar, remover ou atualizar os detalhes do seu estudo, entre em contato com register@clinicaltrials.gov. Assim que uma alteração for implementada em clinicaltrials.gov, ela também será atualizada automaticamente em nosso site .