Study of MeRes100 in the Treatment of Patient With Coronary Artery Disease. (MeReS100-China)
A Multi Center Randomized Control Study of MeRes100 Sirolimus Eluting BioResorbable Vascular Scaffold System in Treatment of Coronary Artery Disease Patients: MeRes - China.
Study Overview
Status
Status
Conditions
Conditions
Intervention / Treatment
Intervention / Treatment
Detailed Description
This is a Multi Center Randomized (MeRes:XIENCE=1:1) Control Study of MeRes100 Sirolimus Eluting BioResorbable Vascular Scaffold System in treatment of approximately 484 subjects with Coronary Artery Disease in China. The post marketed XIENCE Everolimus Eluting Coronary Stent System will serve as the control device, to evaluate the safety and efficacy of MeRes100 Sirolimus Eluting BioResorbable Vascular Scaffold System in coronary artery disease, and to support the product registration approval of the China Food and Drug Administration (CFDA).
Eligible subjects shall have up to two de novo lesions in different epicardial vessels ( One lesion in each coronary artery or it's major branch),Target lesion shall have visually estimated diameter stenosis ≥ 70% (or ≥ 50% and has clinical evidence of myocardial ischemia), lesion length ≤ 24mm, reference vascular diameter visually ≥ 2.75mm and ≤ 4.0mm. And subjects must meet all the study inclusion / exclusion criteria before enrolment in the study.
All the subjects shall accept angiography follow up at 12 months post procedure.
OCT subgroup ((N=80, 40 from test group and 40 from control group) shall accept follow up of OCT at immediately and 12 months post procedure. In segment late lumen loss at 12 months post procedure is the primary endpoint, and the percentage of stent strut neointimal coverage assessed with OCT at 12 months post procedure is the key secondary endpoint.
Study Type
Study Type
Enrollment (Anticipated)
Enrollment
Phase
Phase
- Not Applicable
Contacts and Locations
Study Contact
Study Contact
- Name: Imrankhan M Lohani, M.Sc.
- Phone Number: 9619608544
- Email: Imrankhan.lohani@merillife.com
Study Contact Backup
- Name: Ashok Thakkar, Ph.D
- Phone Number: 9879443584
- Email: ashok.thakkar@merillife.com
Participation Criteria
Eligibility Criteria
Eligibility Criteria
Ages Eligible for Study
Accepts Healthy Volunteers
Genders Eligible for Study
Description
Inclusion Criteria:
General inclusion criteria:
- Age ≥ 18 and ≤75years old, man or non-pregnant or non-lactating woman.
- Subjects with asymptomatic ischemic evidence, stable or unstable angina, or old myocardial infarction >7 days, suitable for selective PCI. Subjects without contraindications of coronary artery bypass grafting.
- Subjects be able to understand the purpose of this study, volunteer to participate and sign informed consent, willing to accept invasive imaging follow-up.
Angiographic inclusion criteria:
- One de novo target lesion or up-to two de novo target lesions in different epicardial vessels: Different epicardial vessels are defined as left anterior descending artery (LAD) and its branches, left circumflex artery (LCX) arteries and its branches, and right coronary arteries (RCA) and its branches. Thus, for example, the subject must not have two target lesions required to be treated at the LAD and its branches at the same time.
- Target lesion diameter stenosis ≥ 70% (or ≥ 50% simultaneously shall have clinical evidence of myocardial ischemia), and TIMI flow grade ≥1.
- Target lesion length ≤ 24mm (estimated visually); reference vessel diameter between ≥ 2.75 mm to ≤ 4.00 mm.
- Each target lesion can be fully covered by one scaffold.
Exclusion Criteria:
General exclusion criteria:
- Any non-target lesion in coronary artery needs to be treated simultaneously or selectively (anticipation within 1 year), non-target lesion defined as any lesion that does not fulfill the inclusion/exclusion criteria.
- Any newly onset acute myocardial infarction within 1 week (< 7days) or, myocardial enzyme has not return to normal level (clinically non-significant) after myocardial infarction.
- Subject has undergone prior PCI within the target vessel during the last 12months or plans to receive another PCI within 6months or subject with history of CABG.
- Subjects with severe heart failure, such as ≥ grade III NYHA or left ventricular ejection fraction <30% (accessed by ECHO/ultrasound or left ventricular angiography).
- Blood tests shows the platelet count is less than 100 × 109 / L, or more than 700 ×109 / L, the white blood cells count is less than 3 × 109 / L.
Pre-procedure severe liver or kidney functional damaged: serum creatinine> 1.5mg
/ dl (132.6μmol / L) or subject is receiving hemodialysis, ALT or AST were three times greater than the upper limit of normal.
- Subjects with bleeding tendency, active gastrointestinal ulcers, history of cerebral hemorrhage or history of subarachnoid hemorrhage, history within six months of ischemic stroke, contraindications of anti-platelet agents and anticoagulants treatment, and subjects cannot receive anti-thrombolytic therapy.
- Hypersensitive or allergic to antiplatelet drugs(such as aspirin and clopidogrel), anticoagulant drugs(such as heparin), contrast media, and scaffold components(such as PLLA,PDLLA, sirolimus, everolimus, iridium and platinum).
- The subject's life expectancy is less than 12 months.
- Subjects who anticipated in other drug or medical device clinical trial but haven't finished the primary endpoint evaluation visit.
- The investigators judged that subjects were poorly compliance and unable to complete the study as required.
- Subjects have accepted substantial organ transplant or ready to undergo organ transplant.
- Subjects have unstable arrhythmia, such as high-risk ventricular premature beats, and ventricular tachycardia.
- Subjects are receiving or planning to receive chemotherapy.
- Subjects have received or planning to receive radiotherapy.
- Subjects with immunosuppressive or autoimmune diseases, are receiving or planning to receive immunosuppressive therapy.
- Patients suffering from HIV, HBV or HCV infections
- Subjects are planning to receive or are receiving long-term anticoagulation therapy, such as heparin, warfarin and similar agents.
- Subjects are planning to accept selective surgery within 12 months, need to discontinue aspirin or clopidogrel (DAPT therapy).
- Subjects with diffuse peripheral vascular disease, cannot use 6F catheter.
Angiographic exclusion criteria:
- Subjects with coronary artery triple vessel lesion in LAD, LCX or RCA, all need to be treated.
- Target lesion locates in left main.
- Target lesion locates in the aorto-ostial of RCA (within 3 mm of the origin of the RCA).
- Target lesion locates within 3 mm of the origin of the LM, LAD and LCX.
Target lesion involving a bifurcation with a:
- Side branch ≥2.0 mm in diameter, or
- Side branch with diameter stenosis ≥ 50%, or
- Side branch requiring protection guide wire, or
- Side branch requiring pre-dilatation.
Anatomy proximal to or within the lesion that may impair delivery of the
MeRes100TM or Xience EES, including:
- Extreme angulation (≥ 90°) proximal to or within the target lesion
- Excessive tortuosity (≥ two 45° angles) proximal to or within the target lesion
- Moderate or heavy calcification proximal to or within the target lesion
- Target lesion involves a myocardial bridge.
- Target vessel contains thrombus as indicated in the angiographic images or IVUS.
- Target vessel has been previously treated with a stent at any time prior to the index procedure such that the MeRes100TM or XIENCE stent would need to cross the stent to reach the target lesion..
- Target vessel has been previously treated with a stent and the target lesion is within 5 mm proximal to a previously treated lesion.
- Chronic complete occlusive lesion (preoperative TIMI grade 0 flow).
Target lesion which prevents complete balloon pre-dilatation, defined as full balloon expansion with the following outcomes:
- Residual %DS is < 40% (per visual estimation), ( ≤ 20% is strongly recommended).
- TIMI Grade-3 flow (per visual estimation).
- No angiographic complications (e.g., no-reflow, distal embolization, side branch closure).
- No dissections NHLBI grade D-F.
- No chest pain lasting > 5 minutes.
- No ST depression or elevation lasting > 5 minutes.
Study Plan
How is the study designed?
Design Details
- Primary Purpose: Treatment
- Allocation: Randomized
- Interventional Model: Parallel Assignment
- Masking: None (Open Label)
Number of Arms
Arms and Interventions
Participant Group / ArmParticipant Group / Arm |
Intervention / TreatmentIntervention / Treatment |
|---|---|
|
Experimental: MeRes100 BRS
MeRes100 Sirolimus Eluting BioResorbable Vascular Scaffold System indicated for the treatment of coronary artery disease along with standard percutaneous angioplasty procedure.
|
MeRes100 Sirolimus-Eluting BioResorbable Vascular Scaffold System used in the treatment of patients with coronary artery disease along with standard percutaneous angioplasty procedure.
Xience EES Everolimus Eluting Coronary Stent System indicated for the treatment of coronary artery disease along with standard percutaneous angioplasty procedure.
|
|
Active Comparator: Xience EES
Xience EES is a Everolimus Eluting Coronary Stent System indicated for the treatment of coronary artery disease along with standard percutaneous angioplasty procedure.
|
MeRes100 Sirolimus-Eluting BioResorbable Vascular Scaffold System used in the treatment of patients with coronary artery disease along with standard percutaneous angioplasty procedure.
Xience EES Everolimus Eluting Coronary Stent System indicated for the treatment of coronary artery disease along with standard percutaneous angioplasty procedure.
|
What is the study measuring?
Primary Outcome Measures
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
In-segment Late Lumen Loss
Time Frame: 12 Months
|
Angiographic analysis for measuring Late Lumen Loss in both the treatment group
|
12 Months
|
Secondary Outcome Measures
Secondary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
The percentage of stent strut neointimal coverage assessed with OCT
Time Frame: 12 Months
|
OCT analysis will be done to identify percentage of Neointimal coverage
|
12 Months
|
Collaborators and Investigators
Sponsor
Sponsor
Investigators
Investigators
- Principal Investigator: Gao Runlin, MD, Fuwai Hospital CAMS
Study record dates
Study Major Dates
Study Start (Anticipated)
Study Start
Primary Completion (Anticipated)
Primary Completion
Study Completion (Anticipated)
Study Completion
Study Registration Dates
First Submitted
First Submitted
First Submitted That Met QC Criteria
First Submitted That Met QC Criteria
First Posted (Actual)
First Posted
Study Record Updates
Last Update Posted (Actual)
Last Update Posted
Last Update Submitted That Met QC Criteria
Last Update Submitted That Met QC Criteria
Last Verified
Last Verified
More Information
Terms related to this study
Additional Relevant MeSH Terms
- Heart Diseases
- Cardiovascular Diseases
- Vascular Diseases
- Arteriosclerosis
- Arterial Occlusive Diseases
- Coronary Artery Disease
- Myocardial Ischemia
- Coronary Disease
- Physiological Effects of Drugs
- Anti-Infective Agents
- Antineoplastic Agents
- Immunosuppressive Agents
- Immunologic Factors
- Anti-Bacterial Agents
- Antibiotics, Antineoplastic
- Antifungal Agents
- Everolimus
- Sirolimus
Other Study ID Numbers
Other Study ID Numbers
- MeRes-201706
Plan for Individual participant data (IPD)
Plan to Share Individual Participant Data (IPD)?
Drug and device information, study documents
Studies a U.S. FDA-regulated drug product
Studies a U.S. FDA-regulated device product
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.
Clinical Trials on Coronary Artery Disease
-
NCT07163858RecruitingCoronary Artery Bypass | Coronary Artery Disease(CAD) | Off Pump Coronary Artery Bypass Surgery | Hemodynamic Optimization | Hemodynamic Management | Off Pump Coronary Artery Bypass Graft | Coronary Artery Disease With Need for Bypass Surgery | Noradrenaline
-
NCT07388030Not yet recruitingCoronary Artery Disease | Coronary Artery Calcification | Severe Coronary Artery Disease
-
NCT07491107Not yet recruitingCoronary Artery Disease (CAD) | Multivessel Coronary Artery Disease | Complex Coronary Lesions | Calcific Coronary Arteriosclerosis | Small Vessel Ischemic Disease | Stenosis Coronary
-
NCT07596706Active, not recruitingCoronary Artery Disease (CAD) | Coronary Bifurcation Lesion | Left Main Coronary Artery Stenosis
-
NCT07172308CompletedCoronary Artery Disease (CAD) | Atherosclerosis of Coronary Artery
-
NCT07357675Not yet recruitingCoronary Artery Disease (CAD) | Coronary Artery Bypass Graft Surgery(CABG)
-
NCT07354399RecruitingCoronary Artery Disease With Myocardial Infarction
-
NCT05464147Active, not recruitingCoronary Artery Disease | Chronic Total Occlusion of Coronary Artery | Multi Vessel Coronary Artery Disease | Bifurcation of Coronary Artery | Long Lesions Coronary Artery Disease
-
NCT07407738Not yet recruitingCalcified Coronary Artery Disease | Coronary Arterial Disease
-
NCT07512206Not yet recruitingCoronary Artery Disease (CAD) | Coronary Calcification | Coronary Calcified Disease | Coronary Calcified Nodules
Clinical Trials on MeRes100 Sirolimus-Eluting BioResorbable Vascular Scaffold System
-
NCT03033862WithdrawnCoronary Artery Disease
-
NCT02663323Unknown
-
NCT03236415WithdrawnCoronary Artery Disease | Diabetes Mellitus
-
NCT02622100TerminatedCardiovascular Diseases | Coronary Artery Disease | Coronary Disease | Arterial Occlusive Diseases
-
NCT02771288Unknown
-
NCT02699671TerminatedCardiovascular Diseases | Coronary Artery Disease | Arterial Occlusive Diseases
-
NCT03180931CompletedDrug-Eluting Stents | Coronary Thrombosis | Tomography, Optical Coherence
-
NCT02672878UnknownCoronary In-stent Restenosis
-
NCT02319733TerminatedVulnerable Plaque
-
NCT02086045Completed