Pilot Clinical Study on a Low-power Electromagnetic Wave Breast Imaging Device for Cancer Screening Purposes.
Pilot Clinical Evaluation of a Microwave Imaging System for Early Breast Cancer Detection
The trialed investigational medical imaging device is a low-power microwave breast imaging system for cancer screening purposes. It is an active device which uses non-ionizing radiation.
Microwave imaging is an emerging imaging modality for the early detection of the breast cancer. The physical basis of microwave imaging is the dielectric contrast between healthy and cancerous breast tissues at microwave frequencies. Microwave imaging can potentially be used for monitoring neoadjuvant chemotherapy treatment, breast health monitoring, and for routine screening and diagnosis of the breast cancer at the early-stage. The non-invasive and the non-ionizing characteristics of microwaves should allow for frequent scans of the breast using microwave imaging, unlike X-ray mammography. In addition to safety, microwave imaging does not require uncomfortable breast compression and it is potentially a lower-cost modality.
This is a first-in-human clinical test of the investigational device, which has been so far tested only with experimental phantoms modelling the human female breast.
The clinical data that will be collected in the context of this study is intended to provide early safety information for the investigational medical imaging device. In addition, this exploratory data will guide the refinement of the device hardware and the imaging algorithm design, before decision to proceed (or not) with further clinical tests.
Furthermore, this study will be used to guide sample size calculation for a subsequent study designed to evaluate efficacy should that appear warranted once this study is completed.
Study Overview
Status
Status
Conditions
Conditions
Intervention / Treatment
Intervention / Treatment
Detailed Description
This is a single-site early-phase pilot clinical study, taking place at Galway University Hospital/HRB Clinical Research Facility, Galway.
Patients will have a conventional history and breast examinations (Mammogram and / or Ultrasound and Clinical assessment) performed by the physician, as per normal practice in the Symptomatic Breast Unit of University Hospital Galway.
If patients are interested in the study, provide written informed consent and are deemed eligible, they will attend the Clinical Research Facility Galway for a Microwave Breast Imaging (MBI) scan. The data from the scan will be collected and stored securely. Patients will be followed up 7-21 days after the microwave breast investigation or before surgery, whichever comes first. Patients will be assessed for their experience of the microwave breast investigation, and for any adverse events.
The investigational medical imaging device consists of two subsystems, both performing a non-invasive examination:
- the Microwave Breast Imaging subsystem (MBI subsystem);
- the Optical Breast Contour Detection subsystem (OBCD subsystem);
The MBI subsystem is an active device which uses non-ionizing radiation. It illuminates the breast with low-power electromagnetic waves in the microwave frequency spectrum, which penetrate the breast under examination. The subsystem collects the scattered electromagnetic waves and recovers useful information on the breast tissue consistency, given the dielectric contrast of these tissues.
Multi-static radar detection technology is employed in the MBI subsystem for breast image formation.The well-established Microwave Vision Group (MVG) technology for fast antenna measurement, using multiple sensors in a vertical arch configuration, has been transposed to a horizontal arch of sensors. In addition, vertical translation of the horizontal arch has been enabled, such that 3D multi-static short-range radar imaging is possible.
The sensors are in contact with a cylindrical container filled with a liquid; the so-called coupling liquid has been designed to have electromagnetic (EM) properties appropriately selected such that the EM wave penetration in the breast is maximized.
During the MBI scan, the patient is lying in a face down position on a special bed, integrated with the MBI subsystem. The breast under examination is immersed in the coupling liquid, through a dedicated circular opening of the bed. The breast is then scanned. This vertical scan takes approximately 10-15 minutes to complete, depending on the size of each breast.
In order to compute the required volume of coupling liquid, such that the container of the MBI subsystem is optimally filled after immersion of the breast, a simple process for optical assessment of the total volume of the breast takes place just before starting the MBI scan.
The Optical Breast Contour Detection (OBCD) subsystem serves to provide the total volume of the breast, and also its external contour, as a priori information to the MBI subsystem. The OBCD subsystem consists of a 3D optical camera placed below the examination table, at a distance of several tens of centimetres below the breast. An azimuthal scan of the 3D camera permits to reconstruct the external surface of the breast.
In order for the breast contour to be a useful a priori information for the MBI subsystem, it is important that the patient is lying in the same face down position during both the MBI and OBCD scans. Thus, an identical examination table, as the one integrated with the MBI subsystem, is also integrated with the OBCD subsystem. During the OBCD scan, the patient is lying on the examination table, with her breast under examination inserted in the circular opening of the examination table. For this scan, there is no coupling liquid; the breast is in the air, hanging below the examination table.
Both the breast bearing the palpable lump and the contralateral breast will be scanned. The contralateral breast scan will serve as control in the data analysis.
The MBI scan duration of each breast should not exceed 15 minutes. The recorded data will be checked and in the case of non-optimal positioning of the patient, the scan may be repeated.
It is expected that the duration of the microwave breast investigation (including OBCD scan of both breasts for breast volume assessment, preparation of the MBI subsystem, MBI scan of both breasts, and data verification) should not exceed 1 hour. If the Data Quality Check suggests repetition of the MBI scan, and upon agreement by the patient, the total duration of the microwave breast investigation will be extended.
The written radiology reports from conventional imaging will be acquired and used to evaluate the performance of the MBI subsystem in terms of detecting and estimating the size and consistency of the breast lump.
MBI data analysis will be performed off-line, at MVG premises (France), and is expected to be completed few months after the end-date of data collection. Following the assessment of the MBI results by the medical experts involved in the study, a decision will be made regarding the potential of this emerging imaging modality and the interest in proceeding with clinical studies involving larger sample sizes. Based on these results, MVG will decide and plan actions for refinement of the system, towards an upgraded version to be used for potential future tests with a more significant impact.
Study Type
Study Type
Enrollment (Actual)
Enrollment
Phase
Phase
- Not Applicable
Contacts and Locations
Study Locations
-
-
-
Galway, Ireland
- HRB Clinical Research Facility Galway
-
-
Participation Criteria
Eligibility Criteria
Eligibility Criteria
Ages Eligible for Study
Accepts Healthy Volunteers
Genders Eligible for Study
Description
Inclusion Criteria:
- Subjects must be able and willing to give written informed consent and to comply with the requirements of this study protocol
- Subject must have attended the Symptomatic Breast Unit with a palpable breast lump
- Subjects must have had a mammogram in the clinical assessment period (≤ 6 weeks before the microwave breast investigation)
- Subjects must be able to comfortably lie reasonably still in a prone position for up to 15 minutes
- Subjects with bra size larger than 32B and cup size larger or equal to B.
- Subjects whose breast size is adapted to the cylindrical container of the MBI system with sufficient margin to allow the presence of transition liquid around the breast. Final decision to be taken by the physicians based on their judgement.
Exclusion Criteria:
- Subjects unable to provide written informed consent
- Subjects who are pregnant or breast-feeding
- Subjects who have had previous surgery to the breast
- Subjects who have previously received chemotherapy or radiotherapy to the breast
- Subjects who have had a breast biopsy less than two weeks prior to imaging
- Subjects with any active or metallic implant (e.g. cardiac pacemaker, stents, internal cardiac defibrillator, cardiac resynchronisation device, nerve stimulator…), or subjects bearing any non-removable metallic object (e.g. piercing) on their torso
- Post-biopsy patients whose breast tissue is not healed sufficiently for the imaging procedure, in the opinion of the investigator
- Patients who have had or plan to have a breast cyst aspiration before MBI.
- Subjects with significant co-morbidities which, in the opinion of the investigator, may influence the result of the study
- Subjects with prior or concurrent malignancy
- Subjects under the age of 18 years old
- Subjects with evidence of inflammation and/or erythema of the breast as well as any subjects who have a break in the skin which would be in contact with the coupling fluid
- Subjects who would be unsuitable for an MBI scan, unlikely to attend a follow up visit, or would otherwise be unsuitable for such an investigation, in the opinion of the Investigator
Study Plan
How is the study designed?
Design Details
- Primary Purpose: Diagnostic
- Allocation: Non-Randomized
- Interventional Model: Parallel Assignment
- Masking: None (Open Label)
Number of Arms
Arms and Interventions
Participant Group / ArmParticipant Group / Arm |
Intervention / TreatmentIntervention / Treatment |
|---|---|
|
Experimental: Pre-diagnosed breast cancer
Low-power microwave breast imaging system. Core needle biopsy performed ≥14 days before the microwave breast investigation |
Investigate the capacity of microwave imaging to detect and characterise diagnosed palpable breast lump
|
|
Experimental: Pre-diagnosed breast cyst
Low-power microwave breast imaging system. No prior biopsy |
Investigate the capacity of microwave imaging to detect and characterise diagnosed palpable breast lump
|
|
Experimental: Pre-diagnosed benign lesion
Low-power microwave breast imaging system. Core needle biopsy performed ≥14 days before the microwave breast investigation |
Investigate the capacity of microwave imaging to detect and characterise diagnosed palpable breast lump
|
What is the study measuring?
Primary Outcome Measures
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Number of Serious Adverse Events, Directly Related to Normal Functioning of the Device
Time Frame: Through study completion, up to 21 days
|
To provide early safety information for the proposed investigational medical imaging device.
|
Through study completion, up to 21 days
|
Other Outcome Measures
Other Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Capacity of the Low-power Microwave Breast Imaging (MBI) System to Detect Pre-diagnosed Palpable Breast Lesions. Detection Rate for Malignant and Benign Breast Lesions.
Time Frame: Duration of study - 19 months
|
Percentage of total number of biopsy-confirmed malignant breast lesions being detected with MBI; Percentage of total number of pre-diagnosed breast cysts being detected with MBI; Percentage of total number of biopsy-confirmed benign breast lesions being detected with MBI.
|
Duration of study - 19 months
|
Collaborators and Investigators
Sponsor
Sponsor
Collaborators
Collaborators
Investigators
Investigators
- Principal Investigator: Michael Kerin, Prof., Galway University Hospital
Publications and helpful links
Study record dates
Study Major Dates
Study Start (Actual)
Study Start
Primary Completion (Actual)
Primary Completion
Study Completion (Actual)
Study Completion
Study Registration Dates
First Submitted
First Submitted
First Submitted That Met QC Criteria
First Submitted That Met QC Criteria
First Posted (Actual)
First Posted
Study Record Updates
Last Update Posted (Actual)
Last Update Posted
Last Update Submitted That Met QC Criteria
Last Update Submitted That Met QC Criteria
Last Verified
Last Verified
More Information
Terms related to this study
Additional Relevant MeSH Terms
Other Study ID Numbers
Other Study ID Numbers
- TN.32.1.17.SATF
Plan for Individual participant data (IPD)
Plan to Share Individual Participant Data (IPD)?
Drug and device information, study documents
Studies a U.S. FDA-regulated drug product
Studies a U.S. FDA-regulated device product
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.
Clinical Trials on Breast Cancer
-
NCT05020860RecruitingBreast Cancer | Breast Neoplasm | Triple Negative Breast Cancer | Triple Negative Breast Neoplasms | HER2-positive Breast Cancer | Breast Cancer Stage II | Breast Cancer Female | Breast Cancer Stage III | Estrogen Receptor-positive Breast Cancer | Hormone Receptor-positive Breast Cancer
-
NCT02580448CompletedBreast Cancer | Advanced Breast Cancer | Metastatic Breast Cancer | Triple Negative Breast Cancer | Male Breast Cancer | ER+ Breast Cancer | Cancer of the Breast
-
NCT00003199CompletedInflammatory Breast Cancer | Male Breast Cancer | Stage IV Breast Cancer | Stage IIIB Breast Cancer | Estrogen Receptor-negative Breast Cancer | Estrogen Receptor-positive Breast Cancer | Progesterone Receptor-negative Breast Cancer | Progesterone Receptor-positive Breast Cancer
-
NCT01372579UnknownMale Breast Cancer | Stage II Breast Cancer | Stage IIIA Breast Cancer | Stage IIIB Breast Cancer | Triple-negative Breast Cancer | Stage IA Breast Cancer | Stage IB Breast Cancer | Stage IIIC Breast Cancer | Estrogen Receptor-negative Breast Cancer | Progesterone Receptor-negative Breast Cancer
-
NCT01881048CompletedStage IV Breast Cancer | Stage II Breast Cancer | Stage IIIA Breast Cancer | Stage IIIB Breast Cancer | Stage IA Breast Cancer | Stage IB Breast Cancer | Stage IIIC Breast Cancer
-
NCT01151449TerminatedMale Breast Cancer | Stage IV Breast Cancer | Stage IIIA Breast Cancer | Stage IIIB Breast Cancer | Triple-negative Breast Cancer | Stage IIIC Breast Cancer | Recurrent Breast Cancer | Estrogen Receptor-negative Breast Cancer | Progesterone Receptor-negative Breast Cancer | HER2-negative Breast Cancer
-
NCT01632332CompletedHER2-positive Breast Cancer | Male Breast Cancer | Stage II Breast Cancer | Stage IIIA Breast Cancer | Stage IIIB Breast Cancer | Stage IIIC Breast Cancer
-
NCT02315196Active, not recruitingStage IIIA Breast Cancer | Stage IIIB Breast Cancer | Triple-negative Breast Cancer | Stage IIA Breast Cancer | Stage IIB Breast Cancer | Stage IIIC Breast Cancer | Estrogen Receptor-negative Breast Cancer | Progesterone Receptor-negative Breast Cancer | HER2-negative Breast Cancer
-
NCT01222377TerminatedMale Breast Cancer | Stage IV Breast Cancer | Stage II Breast Cancer | Stage IIIA Breast Cancer | Stage IIIB Breast Cancer | Stage IA Breast Cancer | Stage IB Breast Cancer | Stage IIIC Breast Cancer | Recurrent Breast Cancer
-
NCT07214883RecruitingBreast Cancer | Breast Cancer Female | Breast Cancer Diagnosis | Breast Cancer Survivors | Breast Cancer Detection | Breast Cancer Awareness
Clinical Trials on Low-power microwave breast imaging system
-
NCT02566499Withdrawn
-
NCT00267449Unknown
-
NCT05946200CompletedIntracranial Hypertension
-
NCT06969820Not yet recruiting
-
NCT00494676CompletedHomonymous Hemianopia
-
NCT05263752Completed
-
NCT04757714Withdrawn
-
NCT01000064TerminatedTraumatic Brain Injury | Attention Deficit Disorder