Clinical Investigation on Assessing the Repeatability and Reproducibility of Central and Peripheral Refractive Error
Clinical Investigation on Assessing the Repeatability and Reproducibility of Central and Peripheral Refractive Error Measurement With the Grand Seiko WAM-5500
Study Overview
Status
Status
Conditions
Conditions
Intervention / Treatment
Intervention / Treatment
Study Type
Study Type
Enrollment (Actual)
Enrollment
Contacts and Locations
Study Contact
Study Contact
- Name: Study Contact
- Phone Number: 1-800-843-2020
- Email: anixon1@its.jnj.com
Study Locations
-
-
Florida
-
Jacksonville, Florida, United States, 32256
- VRC
-
-
Participation Criteria
Eligibility Criteria
Eligibility Criteria
Ages Eligible for Study
- Adult
Accepts Healthy Volunteers
Sampling Method
Study Population
Description
Inclusion Criteria:
Potential subjects must satisfy all of the following criteria to be enrolled in the study:
- Read, understand, and sign the STATEMENT OF INFORMED CONSENT and receive a fully executed copy of the form.
- Appear able and willing to adhere to the instructions set forth in this clinical protocol.
- Be between 18 and 50 (inclusive) years of age at the time of screening.
- The spherical component of the subject's non-vertex-corrected distance subjective refraction must be between +1.50 and -6.00DS (inclusive), with the magnitude of the cylinder 2.00 DC or less in the right eye.
- The subject must have a Van Herrick angle grade of at least 3 or greater in the right eye.
- The subject must have a BSCVA of at least 20/30 (defined as reading at least 3 out of the 5 letters on the 20/30 line) in the right eye.
Exclusion Criteria:
Potential subjects who meet any of the following criteria will be excluded from participating in the study:
- Be currently pregnant or lactating.
- Be diabetic or pre-diabetic by self-report.
- Have a history of ocular trauma, systemic disease or medication use known to cause variability in refractive error.
- Have had or have planned (within the study period) any ocular or intraocular surgery (e.g., radial keratotomy, PRK, LASIK, iridotomy, retinal laser photocoagulation, etc.).
- Have habitually worn rigid gas permeable (RGP) lenses, orthokeratology lenses, scleral lenses, or hybrid lenses (e.g. SynergEyes, SoftPerm) within the past 3 months.
- Be an employee (e.g., Investigator, Coordinator, Technician) or immediate family member of an employee (including partner, child, parent, grandparent, grandchild or sibling of the employee or their spouse) of the clinical site.
- Take ocular and systemic medications known to interact with Tropicamide ophthalmic solution or systemic medications that can cause myopia.
- Have a history of an allergic reaction to Tropicamide or Proparacaine.
- Have clinically significant (grade 3 or higher on the FDA grading scale) slit lamp findings (e.g., corneal edema, neovascularization or staining, tarsal abnormalities, or bulbar injection) or other corneal or ocular disease or abnormalities that may compromise study endpoints (including entropion, ectropion, chalazia, recurrent styes, glaucoma, corneal scar or opacity ≥ 0.3 mm, history of recurrent corneal erosions, aphakia, moderate or above corneal distortion, herpetic keratitis).
- Have fluctuations in vision due to clinically significant dry eye or other ocular conditions.
- Have a history of strabismus, amblyopia, nystagmus, or any condition that affects fixation.
Study Plan
How is the study designed?
Design Details
What is the study measuring?
Primary Outcome Measures
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Central (on-axis) cycloplegic auto-refraction
Time Frame: Up to 10-day follow-up visit
|
Central (on-axis) cycloplegic auto-refraction will be measured to evaluate the intra-visit repeatability, inter-visit repeatability, and reproducibility of central refractive error measurements with the Grand Seiko WAM-5500.
|
Up to 10-day follow-up visit
|
|
Peripheral (10° off-axis on the temporal retina) cycloplegic auto-refraction
Time Frame: Up to 10-day follow-up visit
|
Peripheral cycloplegic auto-refraction will be measured to evaluate the intra-visit repeatability, inter-visit repeatability, and reproducibility of peripheral refractive error measurements with the Grand Seiko WAM-5500 at 10° temporal retina.
|
Up to 10-day follow-up visit
|
|
Peripheral (20° off-axis on the temporal retina) cycloplegic auto-refraction
Time Frame: Up to 10-day follow-up visit
|
Peripheral cycloplegic auto-refraction will be measured to evaluate the intra-visit repeatability, inter-visit repeatability, and reproducibility of peripheral refractive error measurements with the Grand Seiko WAM-5500 at 20° temporal retina.
|
Up to 10-day follow-up visit
|
|
Peripheral (30° off-axis on the temporal retina) cycloplegic auto-refraction
Time Frame: Up to 10-day follow-up visit
|
Peripheral cycloplegic auto-refraction will be measured to evaluate the intra-visit repeatability, inter-visit repeatability, and reproducibility of peripheral refractive error measurements with the Grand Seiko WAM-5500 at 30° temporal retina.
|
Up to 10-day follow-up visit
|
|
Peripheral (10° off-axis on the nasal retina) cycloplegic auto-refraction
Time Frame: Up to 10-day follow-up visit
|
Peripheral cycloplegic auto-refraction will be measured to evaluate the intra-visit repeatability, inter-visit repeatability, and reproducibility of peripheral refractive error measurements with the Grand Seiko WAM-5500 at 10° nasal retina.
|
Up to 10-day follow-up visit
|
|
Peripheral (20° off-axis on the nasal retina) cycloplegic auto-refraction:
Time Frame: Up to 10-day follow-up visit
|
Peripheral cycloplegic auto-refraction will be measured to evaluate the intra-visit repeatability, inter-visit repeatability, and reproducibility of peripheral refractive error measurements with the Grand Seiko WAM-5500 at 20° nasal retina.
|
Up to 10-day follow-up visit
|
|
Peripheral (30° off-axis on the nasal retina) cycloplegic auto-refraction
Time Frame: Up to 10-day follow-up visit
|
Peripheral cycloplegic auto-refraction will be measured to evaluate the intra-visit repeatability, inter-visit repeatability, and reproducibility of peripheral refractive error measurements with the Grand Seiko WAM-5500 at 30° nasal retina.
|
Up to 10-day follow-up visit
|
Collaborators and Investigators
Sponsor
Sponsor
Investigators
Investigators
- Study Director: Johnson & Johnson Vision Care, Inc. Clinical Trial, Johnson & Johnson Vision Care, Inc.
Study record dates
Study Major Dates
Study Start (Actual)
Study Start
Primary Completion (Actual)
Primary Completion
Study Completion (Actual)
Study Completion
Study Registration Dates
First Submitted
First Submitted
First Submitted That Met QC Criteria
First Submitted That Met QC Criteria
First Posted (Actual)
First Posted
Study Record Updates
Last Update Posted (Actual)
Last Update Posted
Last Update Submitted That Met QC Criteria
Last Update Submitted That Met QC Criteria
Last Verified
Last Verified
More Information
Terms related to this study
Additional Relevant MeSH Terms
Other Study ID Numbers
Other Study ID Numbers
- CR-6559
Plan for Individual participant data (IPD)
Plan to Share Individual Participant Data (IPD)?
IPD Plan Description
Drug and device information, study documents
Studies a U.S. FDA-regulated drug product
Studies a U.S. FDA-regulated device product
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.