Effectiveness and Safety Study of the DexCom™ G4 Continuous Glucose Monitoring System

February 7, 2017 updated by: DexCom, Inc.

The purpose of this study is to evaluate the safety and effectiveness of the G4 System when worn for up to 7-days by subjects >18 years-old with diabetes mellitus requiring insulin.

Device performance will be primarily evaluated in terms of the proportion of G4 System values that are within ±20% of YSI reference value for glucose levels >80 mg/dL and ±20 mg/dL at meter glucose levels <80 mg/dL. These G4-YSI matched pairs will be collected during in-clinic sessions on Days 1, 4, and 7 of Sensor wear. The proportion of matched pairs from this study (PTL-900360) meeting the above criteria will be compared to the proportion of such points observed in the DexCom clinical study that supported approval of the SEVEN PLUS System (PTL 300012).

Safety data of the G4 System will also be collected and safety will be characterized by the incidence of Adverse Device Effects, Serious Adverse Device Events, and Unanticipated Adverse Device Effects experienced by study participants.

Study Overview

Status

Completed

Conditions

Detailed Description

This will be a non-randomized, single-armed study conducted in the United States. Following screening, a minimum of 60 to a maximum of 80 subjects at up to 4 clinical sites will be enrolled. All participants will wear the G4 Sensor for one 7-day wear period (~168 hours). A subset of 20-30 subjects will wear two G4 Systems (one of which will be blinded) during the 7-day wear period for the purposes of assessing precision. During home use, subjects will be asked to use the meter provided to them to take fingersticks for calibration and diabetes management. Subjects will also be asked to confirm high and low G4 System glucose alerts/alarms by taking a comparative fingerstick immediately after receiving an alert/alarm.

All subjects will participate in one, 8-hour in-clinic session on Day 1, 4, or 7 of the study to collect accuracy information against a laboratory standard (YSI) and against SMBG. During the in-clinic session, subjects will have venous blood draws for evaluation of the YSI blood glucose measurements. Carbohydrate consumption and insulin dosing may be manipulated during the in-clinic session to ensure a wide range of glucose values. For the duration of this study, subjects will be instructed to use G4 System information as an adjunct to (and not as a replacement for) standard SMBG guidance of diabetes self-management.

Subjects will perform G4 Sensor insertions at the clinic under supervision of study staff. Sensor insertion sites will be examined after removal of the Sensor(s). Study staff will document any irritation and/or Adverse Device Effects.

Study Type

Interventional

Enrollment (Actual)

60

Phase

  • Not Applicable

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

18 years and older (ADULT, OLDER_ADULT)

Accepts Healthy Volunteers

No

Genders Eligible for Study

All

Description

Inclusion Criteria:

  1. Age 18 years or older;
  2. Have been diagnosed with insulin-requiring diabetes and are on multiple daily injections (MDI) or Continuous Subcutaneous Insulin Infusion (CSII) insulin therapy;
  3. Willing not to inject insulin or wear an insulin pump insertion set within 3 inches from the Sensor site during Sensor wear;
  4. Willing to use only the blood glucose meter provided to them for self-monitoring of blood glucose (SMBG) during Sensor wear;
  5. Willing to participate in one, 8-hour in-clinic session and be willing to take 4 fingersticks per hour and have 4 blood draws per hour for the entire 8-hour duration;
  6. Willing to take a minimum of 6 fingersticks per day during home use (2 for calibration purposes, 4 for comparative purposes);
  7. Willing to refrain from the use of acetaminophen during the Sensor insertion period and for at least 24-hours prior to Sensor insertion;
  8. Willing not to schedule a magnetic resonance (MRI) scan, computed tomography (CT) scan, or x-ray, for the duration of the study;
  9. Able to speak, read, and write English.

Exclusion Criteria:

  1. Have extensive skin changes/diseases that preclude wearing the Sensor on normal skin (e.g. extensive psoriasis, recent burns or severe sunburn, extensive eczema, extensive scarring, extensive tattoos, dermatitis herpetiformis) at the proposed wear sites;
  2. Subjects who have a known allergy to medical-grade adhesives;
  3. Are pregnant as demonstrated by a positive pregnancy test within 72 hours of insertion;
  4. Have a hematocrit that is less than 30%, or greater than 55%;
  5. Current participation in another investigational study protocol (if a subject has recently completed participation in another drug study, the subject must have completed that study at least 30 days prior to being enrolled in this study);
  6. Have any condition that, in the opinion of the Investigator, would interfere with their participation in the trial or pose an excessive risk to study staff handling venous blood samples (e.g., known history of hepatitis B or C).

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Primary Purpose: OTHER
  • Allocation: NA
  • Interventional Model: SINGLE_GROUP
  • Masking: NONE

Arms and Interventions

Participant Group / Arm
Intervention / Treatment
EXPERIMENTAL: CGM
continuous glucose monitoring system
Continuous Glucose Monitoring System
Other Names:
  • Continuous Glucose Monitoring System

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
the Proportion of G4 CGM System in Agreement With the Reference Standard
Time Frame: Assessment done on either Day 1, 4 or 7 of the sensor wear period
The proportion of G4 System values within (±) 20% of YSI reference value for glucose levels >80 mg/dL and (±) 20 mg/dL at YSI glucose levels <80 mg/dL. This is primarayly laboratory measurement outcome,and it is not measured by clinical outcome, for example, diagnosis, treatement and complication, and clinical effectiveness, etc.
Assessment done on either Day 1, 4 or 7 of the sensor wear period

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Sponsor

Investigators

  • Study Chair: Andrew Balo, DexCom, Inc.

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start

February 1, 2010

Primary Completion (ACTUAL)

April 1, 2010

Study Completion (ACTUAL)

April 1, 2010

Study Registration Dates

First Submitted

April 22, 2010

First Submitted That Met QC Criteria

April 26, 2010

First Posted (ESTIMATE)

April 27, 2010

Study Record Updates

Last Update Posted (ACTUAL)

March 27, 2017

Last Update Submitted That Met QC Criteria

February 7, 2017

Last Verified

August 1, 2015

More Information

Terms related to this study

Plan for Individual participant data (IPD)

Plan to Share Individual Participant Data (IPD)?

NO

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.