Clinical Performance of Habitual Hydrogel vs. Silicone Hydrogel Toric Contact Lenses
July 19, 2020 updated by: CooperVision, Inc.
Clinical Performance of Habitual Wearers of Hydrogel Toric Lenses When Refitted With Avaira Toric Silicone Hydrogel Lenses
Evaluate the clinical performance of existing wearers of hydrogel toric lenses when refitted with Avaira toric silicone hydrogel contact lenses over 1 week of wear.
Study Overview
Detailed Description
This is a 60-subject, single masked, bilateral, parallel study design comparing the fitting characteristics of enfilcon A toric lenses against the subjects habitual hydrogel toric lenses (omafilcon A, ocufilcon D or methafilcon B wearers).
Subject's habitual toric lenses will be evaluated at the first visit and then re-fitted with a pair of Avaira toric lenses.
After 1 week of daily wear, subjects will return for a second and final evaluation.
Study Type
Interventional
Enrollment (Actual)
60
Phase
- Not Applicable
Contacts and Locations
This section provides the contact details for those conducting the study, and information on where this study is being conducted.
Study Locations
-
-
-
Valencia, Spain, 46100
- Optometry Research Group (GIO) Optics Department, University of Valencia
-
-
Participation Criteria
Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.
Eligibility Criteria
Ages Eligible for Study
18 years to 40 years (Adult)
Accepts Healthy Volunteers
No
Genders Eligible for Study
All
Description
Inclusion Criteria:
- Is between 18 and 40 years of age (inclusive)
- Has had a self-reported visual exam in the last two years
- Is an adapted soft toric contact lens wearer
- Has a contact lens spherical prescription between +6.00 to - 8.00 (inclusive)
- Have no less than 0.75D of astigmatism and no more than 1.75 D in both eyes.
- Can achieve best corrected spectacle distance visual acuity of 20/25 (0.10 logMAR) or better in each eye.
- Can achieve a distance visual acuity of 20/30 (0.18 logMAR) or better in each eye with the study contact lenses.
- Has clear corneas and no active ocular disease
- Has read, understood and signed the information consent letter.
- Patient contact lens refraction should fit within the available parameters of the study lenses.
- Is willing to comply with the wear schedule (at least 5 days per week, > 8 hours/day assuming there are no contraindications for doing so).
- Is willing to comply with the visit schedule
Exclusion Criteria:
- Has a contact lens prescription outside the range of the available parameters of the study lenses.
- Has a spectacle cylinder less than -0.75D or more than -1.75D of cylinder in either eye.
- Has a history of not achieving comfortable contact lens wear (5 days per week; > 8 hours/day)
- Has contact lens best corrected distance vision worse than 20/25 (0.10 logMAR) in either eye.
- Presence of clinically significant (grade 2-4) anterior segment abnormalities
- Presence of ocular or systemic disease or need of medications which might interfere with contact lens wear.
Slit lamp findings that would contraindicate contact lens wear such as:
- Pathological dry eye or associated findings
- Pterygium, pinguecula, or corneal scars within the visual axis
- Neovascularization > 0.75 mm in from of the limbus
- Giant papillary conjunctivitis (GCP) worse than grade 1
- Anterior uveitis or iritis (past or present)
- Seborrheic eczema, Seborrheic conjunctivitis
- History of corneal ulcers or fungal infections
- Poor personal hygiene
- Has a known history of corneal hypoesthesia (reduced corneal sensitivity)
- Has aphakia, keratoconus or a highly irregular cornea.
- Has Presbyopia or has dependence on spectacles for near work over the contact lenses.
- Has undergone corneal refractive surgery.
- Is participating in any other type of eye related clinical or research study.
Study Plan
This section provides details of the study plan, including how the study is designed and what the study is measuring.
How is the study designed?
Design Details
- Primary Purpose: Other
- Allocation: Non-Randomized
- Interventional Model: Parallel Assignment
- Masking: Single
Arms and Interventions
Participant Group / Arm |
Intervention / Treatment |
|---|---|
|
Active Comparator: Omafilcon A/Enfilcon A
Subject's habitual hydrogel toric lenses Omafilcon A will be evaluated at the first visit and then re-fitted with a pair of Enfilcon A lenses.
|
Silicone hydrogel toric lenses.
Other Names:
|
|
Active Comparator: Ocufilcon D/Enfilcon A
Subject's habitual hydrogel toric lenses Ocufilcon D will be evaluated at the first visit and then re-fitted with a pair of Enfilcon A lenses.
|
Silicone hydrogel toric lenses.
Other Names:
|
|
Active Comparator: Methafilcon B/ Enfilcon A
Subject's habitual hydrogel toric lenses Methafilcon B will be evaluated at the first visit and then re-fitted with a pair of Enfilcon A lenses.
|
Silicone hydrogel toric lenses.
Other Names:
|
What is the study measuring?
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Participant's Subjective Rating for Lens Initial Comfort
Time Frame: Baseline and 1 week
|
Surveyed for habitual pair at baseline.
Dispensed pair after insertion/settling at 1 week.
Rated on a visual analog scale (VAS).
(0-10, 0=poor, 10= can't feel)
|
Baseline and 1 week
|
|
Participant's Subjective Rating for Lens Comfort Prior to Removal
Time Frame: Baseline and 1 week
|
Surveyed prior to removal of each lens pair.
Habitual pair at baseline.
Dispensed pair at 1 week.
Rated on a visual analog scale (VAS).
(0-10, 0=poor, 10= can't feel)
|
Baseline and 1 week
|
|
Participant's Subjective Rating for Overall Lens Comfort
Time Frame: Baseline and 1 week
|
Surveyed for each lens pair.
Habitual pair at baseline.
Dispensed pair at 1 week.
Rated on a visual analog scale (VAS).
(0-10, 0=poor, 10= can't feel)
|
Baseline and 1 week
|
|
Participant's Subjective Rating for Dryness During the Day
Time Frame: Baseline and 1 week
|
Surveyed for each lens pair.
Habitual pair at baseline.
Dispensed pair at 1 week.
Rated on a visual analog scale (VAS).
(0-10, 0=very dry, 10= no dryness)
|
Baseline and 1 week
|
|
Participant's Subjective Rating for Dryness Prior to Removal
Time Frame: Baseline and 1 week
|
Surveyed prior to removal of each lens pair.
Habitual pair at baseline .
Dispensed pair at 1 week.
Rated on a visual analog scale (VAS).
(0-10, 0=very dry, 10= no dryness)
|
Baseline and 1 week
|
|
Participant's Subjective Rating for Overall Dryness
Time Frame: Baseline and 1 week
|
Surveyed for each lens pair.
Habitual pair at baseline.
Dispensed pair at 1 week .
Rated on a visual analog scale (VAS).
(0-10, 0=very dry, 10= no dryness)
|
Baseline and 1 week
|
|
Participant's Subjective Rating for Lens Handling - Insertion
Time Frame: Baseline and 1 week
|
Surveyed for habitual pair at baseline.
Dispensed pair after insertion/settling at 1 week.
Rated on a visual analog scale (VAS).
(0-10, 0=poor, 10= very easy)
|
Baseline and 1 week
|
|
Participant's Subjective Rating for Overall Lens Fit Stability
Time Frame: Baseline and 1 week
|
Surveyed for each lens pair.
Habitual pair at baseline.
Dispensed pair at 1 week.
Rated on a visual analog scale (VAS).
(0-10, 0=very unstable / excessive movement, 10= very stable / good movement)
|
Baseline and 1 week
|
|
Participant's Subjective Rating for Overall Vision Satisfaction
Time Frame: Baseline and 1 week
|
Surveyed for each lens pair.
Habitual pair at baseline.
Dispensed pair at 1 week.
Rated on a visual analog scale (VAS).
(0-10, 0=completely dissatisfied, 10= very satisfied)
|
Baseline and 1 week
|
|
Participant's Subjective Rating for Vision Quality at Insertion
Time Frame: Baseline and 1 week
|
Surveyed for habitual pair at baseline.
Dispensed pair after insertion/settling at 1 week .Rated on a visual analog scale (VAS).
(0-100, 0=extremely poor vision totally blurred, 100= excellent vision totally sharp)
|
Baseline and 1 week
|
|
Participant's Subjective Rating for Vision Quality During the Day
Time Frame: Baseline and 1 week
|
Surveyed for each lens pair.
Habitual pair at baseline.
Dispensed pair at 1 week .
Rated on a visual analog scale (VAS).
(0-100, 0=extremely poor vision totally blurred, 100= excellent vision totally sharp)
|
Baseline and 1 week
|
|
Participant's Subjective Rating for Vision Quality End of the Day
Time Frame: Baseline and 1 week
|
Surveyed for each lens pair.
Habitual pair at baseline.
Dispensed pair at 1 week end of the day.
Rated on a visual analog scale (VAS).
(0-100, 0=extremely poor vision totally blurred, 100= excellent vision totally sharp)
|
Baseline and 1 week
|
|
Participant's Subjective Rating for Night Vision Quality
Time Frame: Baseline and 1 week
|
Surveyed for each lens pair.
Habitual pair at baseline.
Study pair dispensed pair at 1 week.
Rated on a visual analog scale (VAS).
(0-100, 0=extremely poor vision totally blurred, 100= excellent vision totally sharp)
|
Baseline and 1 week
|
|
Participant's Subjective Rating for Vision Stability at Insertion
Time Frame: Baseline and 1 week
|
Surveyed for habitual pair at baseline.
Dispensed pair after insertion/settling at 1 week.
Rated on a visual analog scale (VAS).
(0-100, 0=Totally unstable Fluctuating/changing, 100= perfectly stable Not fluctuating/changing)
|
Baseline and 1 week
|
|
Participant's Subjective Rating for Vision Stability During the Day
Time Frame: Baseline and 1 week
|
Surveyed for each lens pair.
Habitual pair at baseline.
Dispensed pair at 1 week.
Rated on a visual analog scale (VAS).
(0-100, 0=Totally unstable Fluctuating/changing, 100= perfectly stable Not fluctuating/changing)
|
Baseline and 1 week
|
|
Participant's Subjective Rating for Vision Stability at End of Day
Time Frame: Baseline and 1 week
|
Surveyed for each lens pair.
Habitual pair at baseline.
Dispensed pair at 1 week.
Rated on a visual analog scale (VAS).
(0-100, 0=Totally unstable Fluctuating/changing, 100= perfectly stable Not fluctuating/changing)
|
Baseline and 1 week
|
|
Participant's Subjective Rating for Lens Pair Preference
Time Frame: 1 week
|
Participants subjective preference in relation to comfort, dryness, handling, vision, lens fit and overall of the patient after wearing the study and their habitual lenses.
|
1 week
|
|
Participant's Subjective Rating for Overall Satisfaction - Dryness
Time Frame: Baseline and 1 week
|
Surveyed for each lens pair.
Habitual pair at baseline.
Dispensed pair at 1 week .
Likert scale.
(4 choices - completely satisfied, somewhat satisfied, somewhat dissatisfied, completely dissatisfied)
|
Baseline and 1 week
|
|
Participant's Subjective Rating for Overall Satisfaction - Handling
Time Frame: Baseline and 1 week
|
Surveyed for each lens pair.
Habitual pair at baseline.
Dispensed pair at 1 week .
Likert scale.
(4 choices - completely satisfied, somewhat satisfied, somewhat dissatisfied, completely dissatisfied)
|
Baseline and 1 week
|
|
Participant's Subjective Rating for Overall Satisfaction - Vision
Time Frame: Baseline and 1 week
|
Surveyed for each lens pair.
Habitual pair at baseline.
Dispensed pair at 1 week .
Likert scale.
(4 choices - completely satisfied, somewhat satisfied, somewhat dissatisfied, completely dissatisfied)
|
Baseline and 1 week
|
|
Participant's Subjective Rating for Overall Satisfaction - Lens Fit
Time Frame: Baseline and 1 week
|
Surveyed for each lens pair.
Habitual pair at baseline.
Dispensed pair at 1 week.
Likert scale.
(4 choices - completely satisfied, somewhat satisfied, somewhat dissatisfied, completely dissatisfied)
|
Baseline and 1 week
|
|
Participant's Subjective Rating for Overall Satisfaction - Overall
Time Frame: Baseline and 1 week
|
Surveyed for each lens pair.
Habitual pair at baseline.
Dispensed pair at 1 week.
Likert scale.
(4 choices - completely satisfied, somewhat satisfied, somewhat dissatisfied, completely dissatisfied)
|
Baseline and 1 week
|
|
Clinician's Objective Assessment Monocular High Contrast Distance Visual
Time Frame: Baseline and 1 week
|
Assessed for each lens pair.
Habitual pair at baseline.
Dispensed pair at 1 week.
LogMAR - Positive values denote poorer vision, negative values denote better vision than baseline 20/20 value.
|
Baseline and 1 week
|
|
Clinician's Objective Assessment Binocular High Contrast Distance Visual Acuity
Time Frame: Baseline and 1 week
|
Assessed for each lens pair.
Habitual pair at baseline.
Dispensed pair at 1 week.
LogMAR - Positive values denote poorer vision, negative values denote better vision than baseline 20/20 value
|
Baseline and 1 week
|
|
Clinician's Assessment Overall Lens Stability-Right Eye
Time Frame: Baseline and 1 week
|
Surveyed for habitual pair at baseline.
Dispensed pair after insertion/settling at 1 week.
Overall performance of the lens in terms of axis stability on primary gaze, during lateral gaze, rotational recovery, and mislocation of the lens on blinking.
Likert scale.
(0-4, 0=very poor,1=poor, 2=moderate, 3=good, 4= excellent)
|
Baseline and 1 week
|
|
Clinician's Assessment Overall Lens Stability-Left Eye
Time Frame: Baseline and 1 week
|
Surveyed for habitual pair at baseline.
Dispensed pair after insertion/settling at 1 week.
Overall performance of the lens in terms of axis stability on primary gaze, during lateral gaze, rotational recovery, and mislocation of the lens on blinking.
Likert scale.
(0-4, 0=very poor,1=poor, 2=moderate, 3=good, 4= excellent)
|
Baseline and 1 week
|
|
Clinician's Assessment Overall Fit Acceptance- Right Eye
Time Frame: Baseline and 1 week
|
Surveyed for habitual pair at baseline.
Dispensed pair after insertion/settling at 1 week.
Overall fit acceptance based on lens fit alone (not comfort or vision).
Likert scale.
(0-4, 0=Very poor 1=Poor 2=Moderate, 3=Good, 4= Excellent)
|
Baseline and 1 week
|
|
Clinician's Assessment Overall Fit Acceptance- Left Eye
Time Frame: Baseline and 1 week
|
Surveyed for habitual pair at baseline.
Dispensed pair after insertion/settling at 1 week.
Overall fit acceptance based on lens fit alone (not comfort or vision).
Likert scale.
(0-4, 0=Very poor 1=Poor 2=Moderate, 3=Good, 4= Excellent)
|
Baseline and 1 week
|
|
Clinician's Assessment Lens Centration-Right Eye
Time Frame: Baseline and 1 week
|
Surveyed for habitual pair at baseline.
Dispensed pair after insertion/settling at 1 week .
Lens centration recorded by degree and direction in the primary position.
(0-2, 0=centered/optimal, 1=decentered slightly, 2=substantially decentered (>0.5mm))
|
Baseline and 1 week
|
|
Clinician's Assessment Lens Centration- Left Eye
Time Frame: Baseline and 1 week
|
Surveyed for habitual pair at baseline.
Dispensed pair after insertion/settling at 1 week.
Lens centration recorded by degree and direction in the primary positions.
(0-2, 0=centered/optimal, 1=decentered slightlty, 2=substantially decentered (>0.5mm)
|
Baseline and 1 week
|
|
Clinician's Assessment Corneal Coverage-Right Eye
Time Frame: Baseline and 1 week
|
Surveyed for habitual pair at baseline.
Dispensed pair after insertion/settling at 1 week.
Assessed in primary gaze.
(Y=yes, full corneal coverage at all times, N=no, incomplete corneal coverage)
|
Baseline and 1 week
|
|
Clinicians Assessment Corneal Coverage-Left Eye
Time Frame: Baseline and 1 week
|
Surveyed for habitual pair at baseline.
Dispensed pair after insertion/settling at 1 week.
Assessed in primary gaze.
(Y=yes, full corneal coverage at all times, N=no incomplete corneal coverage)
|
Baseline and 1 week
|
|
Clinician's Assessment Post-Blink Movement- Right Eye
Time Frame: Baseline and 1 week
|
Surveyed for habitual pair at baseline.
Dispensed pair after insertion/settling at 1 week.
Assessed immediately after the blink.
(0-4, 0=insufficient, unacceptable movement, 1=minimal, but acceptable movement, 2=optimal movement, 3=moderate, but acceptable movement, 4=excessive, unacceptable movement
|
Baseline and 1 week
|
|
Clinician's Assessment Post-Blink Movement-Left Eye
Time Frame: Baseline and 1 week
|
Surveyed for habitual pair at baseline.
Dispensed pair after insertion/settling at 1 week.
Assessed immediately after the blink.
(0-4, 0=insufficient, unacceptable movement, 1=minimal, but acceptable movement, 2=optimal movement, 3=moderate, but acceptable movement, 4=excessive, unacceptable movement
|
Baseline and 1 week
|
|
Clinician's Assessment Lens Orientation in Primary Position of Gaze- Right Eye
Time Frame: Baseline and 1 week
|
Assessed for habitual pair at baseline.
Dispensed pair after insertion/settling at 1 week .
Slit lamp, with 10x magnification.
Nasal mislocation is recorded as (+) and temporal as (-).
Mislocation of the axis mark on the lens relative to the 6 o'clock position, zero rotation, measured in degrees.
|
Baseline and 1 week
|
|
Clinician's Assessment Lens Orientation in Primary Position of Gaze-Left Eye
Time Frame: Baseline and 1 week
|
Assessed for habitual pair at baseline.
Dispensed pair after insertion/settling at 1 week .
Slit lamp, with 10x magnification.
Nasal mislocation is recorded as (+) and temporal as (-).
Mislocation of the axis mark on the lens relative to the 6 o'clock position, zero rotation, measured in degrees.
|
Baseline and 1 week
|
|
Clinician's Assessment Rotational Recovery in Degrees After 60 Seconds-Right Eye
Time Frame: Baseline and 1 week
|
Assessed for habitual pair at baseline.
Dispensed pair after insertion/settling at 1 week .
Slit lamp, with 10x magnification.
Lens ability to return to its original position measured 60 seconds after manually rotating the lens 45 degrees temporally.
|
Baseline and 1 week
|
|
Clinician's Assessment Rotational Recovery in Degrees After 60 Seconds-Left Eye
Time Frame: Baseline and 1 week
|
Assessed for habitual pair at baseline.
Dispensed pair after insertion/settling at 1 week .
Slit lamp, with 10x magnification.
Lens ability to return to its original position measured 60 seconds after manually rotating the lens 45 degrees temporally.
|
Baseline and 1 week
|
|
Participant's Subjective Rating for Overall Satisfaction - Comfort
Time Frame: Baseline and 1 week
|
Surveyed for each lens pair.
Habitual pair at baseline.
Dispensed pair at 1 week.
Likert scale.
(4 choices - completely satisfied, somewhat satisfied, somewhat dissatisfied, completely dissatisfied)
|
Baseline and 1 week
|
Collaborators and Investigators
This is where you will find people and organizations involved with this study.
Sponsor
Investigators
- Principal Investigator: Robert Montés-Mico, OD MPhil PhD, Optometry Research Group (GIO) Optics Department, University of Valencia
Study record dates
These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.
Study Major Dates
Study Start
May 1, 2014
Primary Completion (Actual)
June 1, 2014
Study Completion (Actual)
July 1, 2014
Study Registration Dates
First Submitted
May 16, 2014
First Submitted That Met QC Criteria
May 19, 2014
First Posted (Estimate)
May 20, 2014
Study Record Updates
Last Update Posted (Actual)
July 28, 2020
Last Update Submitted That Met QC Criteria
July 19, 2020
Last Verified
July 1, 2020
More Information
Terms related to this study
Additional Relevant MeSH Terms
Other Study ID Numbers
- EX-MKTG-48
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.
Clinical Trials on Astigmatism
-
Bausch & Lomb IncorporatedNot yet recruiting
-
London Vision ClinicOptana GmbHActive, not recruitingCorneal AstigmatismUnited Kingdom
-
Eye Hospital Pristina KosovoActive, not recruiting
-
Yonsei UniversityUnknownMyopic AstigmatismKorea, Republic of
-
Abbott Medical OpticsCompletedRefractive AstigmatismAustria
-
Ziemer Ophthalmic Systems AGTigermed Consulting Co., LtdRecruitingMyopic AstigmatismChina
-
Bausch & Lomb IncorporatedCompleted
-
CooperVision, Inc.CompletedAstigmatism BilateralCanada
-
University Clinic FrankfurtCompleted
Clinical Trials on enfilcon A
-
CooperVision, Inc.Completed
-
CooperVision, Inc.CompletedAstigmatism | Myopia | HyperopiaCanada, United States
-
CooperVision, Inc.Completed
-
CooperVision, Inc.Completed
-
CooperVision, Inc.Completed
-
CooperVision, Inc.Completed
-
CooperVision, Inc.TerminatedMyopiaUnited Kingdom, United States
-
CooperVision, Inc.Completed
-
CooperVision, Inc.CompletedAmetropiaUnited States
-
Johnson & Johnson Vision Care, Inc.Completed