A Comparative, Controlled, Clinical Investigation and Quality Control of a New Hearing Aid
A Comparative, Controlled, Clinical Investigation and Quality Control of a New Hearing Aid in Comparison to the Currently Marketed Device
The aim of this study is to investigate the performance of the hearing aid under laboratory as well as everyday conditions. In addition, the performance data of the hearing aid are compared with a Bernafon hearing aid already available on the market.
The hearing aids are adapted according to the user requirements. The data collected in the study is intended to show that the benefit of the hearing aid being examined is identical or better than the current CE-marked hearing aid. A further aim is to improve the adaptation of the hearing device in such a way that the benefit for people with hearing impairments can be increased.
Study Overview
Status
Conditions
Intervention / Treatment
Detailed Description
The reason for this study is to evaluate a new hearing aid. The goal is to evaluate the audiological performance and usability as well as features and functions in comparison to the hearing aid already CE marked. Furthermore, it is important to identify unexpected or unwanted behavior from the devices. This clinical investigation is a validation testing and is designed to evaluate the new hearing instrument system. As human subjects are involved, this validation test falls under the definition of a clinical investigation. The validation addresses the performance of the device with new functionality, and additionally addresses whether the end user can understand speech as well as with the current marketed device.
End users will be tested in the lab with speech tests and usability tests. They will also participate in field tests after which they will answer questionnaires concerning their experience with the instruments. Evaluating the overall performance of the hearing instrument is important to verify that the end user is satisfied with the device and that all user requirements are fulfilled. Many of the features available in the new device have already been validated and are used in devices currently on the market.
The purpose of this research is to show that the performance of the new device is as good as or better than the current device. Furthermore, speech should not be negatively affected with the addition of new features, the usability of the devices (that end users are able to handle the devices) should be as good as the current device, and there should be no artifacts or unwanted noises. Some measurements serve as quality control prior to product launch.
Study Type
Enrollment (Actual)
Phase
- Not Applicable
Contacts and Locations
Study Locations
-
-
-
Bern, Switzerland, 3018
- Bernafon AG
-
-
Participation Criteria
Eligibility Criteria
Ages Eligible for Study
Accepts Healthy Volunteers
Genders Eligible for Study
Description
Inclusion Criteria:
- All classifications of hearing loss (sensorineural, conductive, mixed)
- If the hearing loss is conductive or mixed it must be approved for amplification by a physician
- All shapes of hearing loss (flat, sloping, reverse slope, notch)
- Severity ranging from mild to severe
- German speaking
- Ability and willingness to sign the consent form
Exclusion Criteria:
- Contraindications for amplification
- Active ear disease
- Inability to follow the procedures of the study due to language problems, psychological disorders, dementia, or other cognitive problems of the participant,
- A reduced mobility making them unable to attend weekly study appointments
- A reduced ability to describe auditory impressions and the usage of the hearing aids
- Uncooperative so that it is not possible to record a valid pure tone audiogram
- A strongly reduced dexterity
- Central hearing disorders
- Bernafon employees
- Family members of Bernafon employees
Study Plan
How is the study designed?
Design Details
- Primary Purpose: TREATMENT
- Allocation: NA
- Interventional Model: SINGLE_GROUP
- Masking: NONE
Arms and Interventions
Participant Group / Arm |
Intervention / Treatment |
|---|---|
|
EXPERIMENTAL: Successor hearing aid to Juna
The intervention is the new device which is the successor to the Juna device.
The participants will use their current device as a control.
The will wear the intervention device for approximately two weeks.
|
The participants will all be experienced hearing instrument device users and will be switched from their current (control) devices to the new test device which succeeds the Juna device.
|
What is the study measuring?
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Speech Intelligibility Performance
Time Frame: 2 hour period of speech testing during a lab appointment after wearing the devices for approximately 2 weeks.
|
Speech test scores will be measured with speech reception thresholds (SRTs) of speech in noise in three conditions: unaided, aided with the current device, and aided with the new device.
The score is measured by the the signal-to-noise ratio (SNR) at which 50% of words in a sentence list are correctly repeated.
The speech is always presented at 65 dB and the background noise varies to maintain the 50% correct.
A lower SNR score indicates a better score.
The maximum score will be 15 dB and the minimum will be -5 dB.
|
2 hour period of speech testing during a lab appointment after wearing the devices for approximately 2 weeks.
|
Secondary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Subjective Performance of Aided Benefit
Time Frame: A 2 week time period during which they will wear the devices and answer questionnaires about the experience.
|
The subjective performance for the new device will be measured with the APHAB questionnaire (Abbreviated Hearing Aid Benefit Profile). The questionnaire is a list of 24 statement that the subjects must agree or disagree with using a scale of units from A to G (A being Always and G being Never). Each letter has a corresponding score used for the calculation: A receives 99, B receives 87, C receives 75, D receives 50, E receives 25, F recevies 12, and G receives 1. An average for each subscale is calculated, and a global or overall score can be calculated by taking the mean of the three positive subscales (Ease of communication, Background noise, and Reverberation). For these three subscales a higher score indicates a better performance. The highest possible score is 99 and the lowest is 1. The fourth scale (Aversiveness) is a negative scale meaning that a higher score means a worse performance. The highest possible score is 99 and the lowest is 1. |
A 2 week time period during which they will wear the devices and answer questionnaires about the experience.
|
|
Speech Intelligibility Performance With New Feature
Time Frame: 2 hour period of speech testing during a lab appointment after wearing the devices for approximately 2 weeks.
|
Word recognition test scores will be measured with two signal-to-noise ratio (SNR) levels in percent correct in two conditions, with the new feature on and the new feature off. The test is a speech test with the words stated in noise at either a 5 dB SNR or a 15 dB SNR. The subject is presented with 5 rhyming words and must choose which word they heard. It is a forced choice meaning that the test will not continue with the next word until they've made a selection. They cannot leave any blank and are not penalized for wrong answers. The score is a percentage of words correct with the maximum score being 100% and the minimum being 0%. A higher score means a better outcome. |
2 hour period of speech testing during a lab appointment after wearing the devices for approximately 2 weeks.
|
|
Handling/Usability Performance
Time Frame: 2 hour period of usability testing with tasks and questionnaires.
|
The usability of the new device will be measured with a 3-point rating scale used by the tester to evaluate tasks performed by the end user as well as questionnaires and interviews. The questionnaire asks subjects to perform tasks, and the clinician rates the ease with which they complete the tasks. 2 is the highest score meaning that they could complete without help, 1 means that they required some assistance, and 0 means that they could not complete the task. A higher score means a better outcome. |
2 hour period of usability testing with tasks and questionnaires.
|
|
Events
Time Frame: A 2 hour period of interviews and questionnaires after having worn the devices for approximately 2 weeks and keeping a diary to make notes of device behavior.
|
Events will be measured by the reporting of unexpected sounds or behavior of the device through questionnaires, diaries, and interviews.
The number of events reported including AEs related to the device will be used to measure the outcome.
Less events is a better outcome.
|
A 2 hour period of interviews and questionnaires after having worn the devices for approximately 2 weeks and keeping a diary to make notes of device behavior.
|
Collaborators and Investigators
Sponsor
Study record dates
Study Major Dates
Study Start (ACTUAL)
Primary Completion (ACTUAL)
Study Completion (ACTUAL)
Study Registration Dates
First Submitted
First Submitted That Met QC Criteria
First Posted (ACTUAL)
Study Record Updates
Last Update Posted (ACTUAL)
Last Update Submitted That Met QC Criteria
Last Verified
More Information
Terms related to this study
Additional Relevant MeSH Terms
Other Study ID Numbers
- BF001-1611
Plan for Individual participant data (IPD)
Plan to Share Individual Participant Data (IPD)?
IPD Plan Description
Drug and device information, study documents
Studies a U.S. FDA-regulated drug product
Studies a U.S. FDA-regulated device product
product manufactured in and exported from the U.S.
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.
Clinical Trials on Hearing Loss
-
Oticon MedicalCompletedEar Diseases | Hearing Loss, Conductive | Hearing Loss Mixed | Hearing Disability | Conductive Hearing Loss | Conductive Hearing Loss, Bilateral | Conductive Hearing Loss, UnilateralUnited Kingdom
-
MED-EL Elektromedizinische Geräte GesmbHCompletedHearing Loss | Hearing Loss, Sensorineural | Hearing Loss, Bilateral | Hearing Loss, Conductive | Hearing Loss, Unilateral | Hearing Loss, MixedAustria, Germany, United Kingdom
-
Oticon MedicalNot yet recruitingSensorineural Hearing Loss, Bilateral | Sensorineural Hearing Loss, Severe | Sensorineural Hearing Loss, Profound
-
University of California, San FranciscoPatient-Centered Outcomes Research InstituteRecruitingHearing Loss | Hearing Loss, Sensorineural | Hearing Loss, Bilateral | Hearing Loss, Conductive | Hearing Loss, Noise-Induced | Hearing Loss, Unilateral | Hearing Loss, Mixed | Hearing Disorders in ChildrenUnited States
-
Manchester University NHS Foundation TrustUniversity of ManchesterCompletedCochlear Hearing Loss | Sensorineural Hearing Loss, BilateralUnited Kingdom
-
Truway Health, Inc.Enrolling by invitationSensorineural Hearing Loss | Tinnitus | Sudden Hearing Loss | Acoustic Trauma | Inner Ear Injury | Noise-Induced Hearing Loss | Vestibular DysfunctionUnited States
-
Wonju Severance Christian HospitalRecruitingHearing Loss, Sudden | Sudden Hearing Loss | Hearing Loss, Idiopathic Sudden Sensorineural | SSNHL(Sudden Sensory Neural Hearing Loss)Korea, Republic of
-
Envoy Medical CorporationActive, not recruitingSensorineural Hearing Loss | Sensorineural Hearing Loss (Disorder) | Sensorineural Hearing Loss, Bilateral | Sensorineural Hearing Loss, Severe | Sensorineural Hearing Loss, Profound | Sensori-Neural DeafnessUnited States
-
Oticon MedicalCompletedConductive Hearing Loss | Conductive and Sensori-neural Hearing Loss in the Same Ear | Unilateral, Profound Sensori-neural Hearing LossUnited States
-
Frequency TherapeuticsCompletedHearing Loss, Sensorineural | Presbycusis | Noise Induced Hearing Loss | Sudden Hearing LossUnited States
Clinical Trials on Successor hearing aid to Juna
-
Sonova AGCompletedHearing Loss, Bilateral Sensorineural, ProgressiveSwitzerland
-
Sonova AGCompleted
-
Sonova AGCompleted
-
Northwestern UniversityNational Institute on Deafness and Other Communication Disorders (NIDCD)Completed
-
Purdue UniversityOticonRecruitingHearing Loss, CochlearUnited States
-
Sonova AGCompletedHearing Loss, Sensorineural | Hearing Loss, BilateralCanada
-
Bernafon AGCompletedHearing LossSwitzerland
-
HearX SAWithdrawnHearing Loss | Hearing Impairment
-
Northwestern UniversityEnrolling by invitation