Evaluating the Accuracy of an Integrated Vital Sign Measurement Platform

August 29, 2020 updated by: Joseph C. Kvedar, Massachusetts General Hospital
Partners Connected Health will conduct a clinical validation study to evaluate the accuracy of measurement for each vital sign with the Vital Moto Mod platform, compared to appropriate reference devices used in the hospital setting. The study will recruit volunteers from the Partners Healthcare network of clinics and hospitals and the general public.

Study Overview

Detailed Description

To assess the comparative clinical performance of the Vital Moto Mod device in the measurement of five vital signs (body temperature, respiratory rate, heart rate, oxygen saturation and blood pressure) in correlation with a reference device used in the hospital environment for vital sign measurement.

Vital sign measurement is a core process in delivering healthcare. Body temperature, respiratory rate, heart rate, oxygen saturation and blood pressure are some of the most frequently measured vital signs in inpatient as well as outpatient encounters. Currently, separate devices exist for measurement of each of these vital signs, each of which may have to be measured several times in a day. Consequently, measurement of vital signs is a time and human resource intensive process. Moreover, most of the devices currently used to measure vital signs cannot transmit data to the electronic medical record of patient, thereby creating opportunities for human error. VVVital USA has used proprietary sensor technology to develop Vital Moto Mod - a vital sign measurement platform that can non-invasively measure body temperature, respiratory rate, heart rate, oxygen saturation and blood pressure from a person's index finger, and allows this information to be pushed into their electronic medical/health record from the point of care

The primary market focus use of the Vital Moto Mod is to replace the vital sign cart in hospitals. Vital sign carts are expensive and cumbersome to move around the hospital. At a fraction of the cost, the Vital Moto Mod provides the same functionality including bar code scanning at a fraction of the cost. Additionally, the Vital Moto Mod can also be used at home without the need for a trained nurse. The Vital Moto Mod App will guide a user through the workflow of performing the measurements. The Vital Moto Mod is connected to a smartphone for storage, and has a USB interface for charging, thereby requiring no electrical connections during use. All measurements are taken by the device independent of the phone.

Partners Connected Health will conduct a clinical validation study to evaluate the accuracy of measurement for each vital sign with the Vital Moto Mod platform, compared to appropriate reference devices used in the hospital setting. The study will recruit volunteers from the Partners Healthcare network of clinics and hospitals and the general public.

Study Type

Interventional

Enrollment (Actual)

265

Phase

  • Not Applicable

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Locations

    • Massachusetts
      • Boston, Massachusetts, United States, 02114
        • Partners Connected Health

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

16 years and older (Adult, Older Adult)

Accepts Healthy Volunteers

No

Genders Eligible for Study

All

Description

Inclusion Criteria:

For volunteers with one or more vital signs presumed to be within normal physiological range, at rest:

  • Age ≥ 18 years
  • Willing to allow arm and finger size to be measured
  • Willing to allow measurement of skin color (optional)a
  • Willing to undergo exercise testing (optional);
  • Willing to allow an arterial blood gas draw (optional);

For volunteers with one or more vital signs expected to be outside of normal physiological range, at rest:

  • Age≥18 years
  • Willing to allow arm and finger size to be measured
  • Willing to allow measurement of skin color (optional)

Exclusion Criteria:

For Volunteers Undergoing Heart Rate and Oxygen Saturation Measurement:

  • Nail paint or artificial nail on the index finger of the left hand
  • SpO2 less than 94%

For Volunteers Undergoing Exercise Testing with Vital Signs within Normal Physiological Range, at rest:

  • Volunteers with vital signs outside of normal physiological range (ascertained during measurements conducted at study site before exercise testing)
  • Age 65 years or older
  • Consumption of food, alcohol or caffeine (coffee, tea or caffeinated soft-drinks) in a 3-hour period prior to testing
  • Uncomfortable clothing, or inappropriate footwear for exercise testing
  • Self-report of any of the following:
  • History of fall or fracture in the past year
  • History of arthritis
  • History of dizziness
  • History of neurological illness
  • History of chest pain or myocardial infarction
  • History of hypertension or diabetes mellitus
  • History of heart disease
  • History of stroke or transient ischemic attack
  • Active treatment for anemia, electrolyte imbalance, or hyperthyroidism
  • Physical disability that precludes safe and adequate testing
  • Mental impairment with limited ability to cooperate4
  • Pregnancy (Volunteer will be excluded if either of the following is true: a) self-report of being pregnant; b) positive urine pregnancy test)

For Volunteers undergoing an Arterial Blood Gas Draw with Vital Signs within Normal Physiological Range, at rest:

  • Self-reported history of a clotting disorder of the blood, or active treatment with medicines affecting blood clotting (e.g. Warfarin).

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Primary Purpose: Other
  • Allocation: Non-Randomized
  • Interventional Model: Parallel Assignment
  • Masking: None (Open Label)

Arms and Interventions

Participant Group / Arm
Intervention / Treatment
Other: Participants with Vital Signs Within Normal Range
All participants will be asked to have an arm and finger measurement conducted and undergo vital sign measurements using the Vital Moto Mod and reference devices for blood pressure, heart rate, respiration rate, blood oxygen saturation, and temperature. Participants with vital signs within the normal physiological range will also be asked to perform an optional exercise testing and/or undergo arterial blood gas measurement.
All participants receive vital signs measurement using the Vital Moto Mod device and reference devices. The vital signs measured are blood pressure, heart rate, respiratory rate, oxygen saturation and temperature.
This is the FDA approved reference device to measure heart rate, SpO2 and temperature.
This is the reference device to be used to obtain etCO2 waveforms for respiratory rate assessment.
These reference devices will be used together for blood pressure measurements.
This is the device that will be used for exercise testing
Other: Participants with Vital Signs Outside of Normal Range
All participants will be asked to have an arm and finger measurement conducted and undergo vital sign measurements using the Vital Moto Mod and reference devices for blood pressure, heart rate, respiration rate, blood oxygen saturation, and temperature. Participants with vital signs outside of the normal physiological range would be documented.
All participants receive vital signs measurement using the Vital Moto Mod device and reference devices. The vital signs measured are blood pressure, heart rate, respiratory rate, oxygen saturation and temperature.
This is the FDA approved reference device to measure heart rate, SpO2 and temperature.
This is the reference device to be used to obtain etCO2 waveforms for respiratory rate assessment.
These reference devices will be used together for blood pressure measurements.

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Number of Vital Moto Mod vital signs measurements in agreement with reference device vital signs measurements
Time Frame: 1 day
Vital signs measurements for each participant would be compared using the Vital Moto Mod and reference devices for blood pressure, heart rate, respiratory rate, oxygen saturation and temperature
1 day

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Collaborators

Investigators

  • Principal Investigator: Kamal Jethwani, Massachusetts General Hospital

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start (Actual)

July 23, 2018

Primary Completion (Actual)

March 8, 2019

Study Completion (Actual)

December 30, 2019

Study Registration Dates

First Submitted

June 4, 2018

First Submitted That Met QC Criteria

July 5, 2018

First Posted (Actual)

July 18, 2018

Study Record Updates

Last Update Posted (Actual)

September 1, 2020

Last Update Submitted That Met QC Criteria

August 29, 2020

Last Verified

August 1, 2020

More Information

Terms related to this study

Keywords

Other Study ID Numbers

  • 2018P000152

Drug and device information, study documents

Studies a U.S. FDA-regulated drug product

No

Studies a U.S. FDA-regulated device product

No

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.