Impact of Acetaminophen on Performance of Guardian™ Sensor (3) in Adults

August 12, 2023 updated by: Medtronic Diabetes
The purpose of this study is to characterize the impact of acetaminophen ingestion on the performance of the Guardian™ Sensor (3) (i.e., C algorithm and Zeus algorithm) in subjects age 18 - 80 years.

Study Overview

Detailed Description

The study is a multi-center, prospective, study without controls. The purpose of this study is to characterize the impact of acetaminophen ingestion on the performance of the Guardian™ Sensor (3) (i.e., C algorithm and Zeus algorithm) in subjects age 18 - 80 years.

Study Type

Interventional

Enrollment (Actual)

104

Phase

  • Not Applicable

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Locations

    • California
      • Escondido, California, United States, 92026
        • AMCR Institue
      • Santa Barbara, California, United States, 93105
        • Sansum Diabetes Research Institute
      • Walnut Creek, California, United States, 92025
        • Diablo Clinical Research
    • Colorado
      • Aurora, Colorado, United States, 80045
        • Barbara Davis Center - Adults
    • Georgia
      • Atlanta, Georgia, United States, 30318
        • "Atlanta Diabetes Associates
    • Washington
      • Renton, Washington, United States, 98057
        • Rainier Clinical Research Center, Inc.

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

14 years to 76 years (Adult, Older Adult)

Accepts Healthy Volunteers

No

Description

Inclusion Criteria:

  1. Individual is 18 - 80 years of age at time of enrollment.
  2. Subject has a clinical diagnosis of type 1 or type 2 diabetes for a minimum of 6 months duration as determined via medical record/ source documentation by an individual qualified to make a medical diagnosis.
  3. Subject has adequate venous access as assessed by investigator or appropriate staff.

Exclusion Criteria:

  1. Subject has history of allergy to acetaminophen or has been told by a health care provider they may not ingest acetaminophen
  2. Subject reports history of liver cirrhosis or liver problems that a health care provider told them they should not use acetaminophen because of liver disorder.
  3. Subject will not tolerate tape adhesive in the area of Guardian™ Sensor (3) placement as assessed by a qualified individual.
  4. Subject has any unresolved adverse skin condition in the area of sensor or device placement (e.g., psoriasis, rash, Staphylococcus infection).
  5. Subject is actively participating in an investigational study (e.g., drug or device) wherein he/she has received treatment from an investigational study (drug or device) in the last 2 weeks prior to Visit 1. (Please note participation in an observational study is acceptable.)
  6. Subject is female of child-bearing potential and has a pregnancy screening test that is positive.
  7. Subject is female of child-bearing potential and who is sexually active should be excluded if she is not using a form of contraception deemed reliable by investigator.
  8. Subject is female and plans to become pregnant during the course of the study.
  9. Subject is breast feeding.
  10. Subject has a chronic heavy alcohol use as determined by investigator.
  11. Subject has a history of a seizure disorder.
  12. Subject has a hematocrit (Hct) more than 10% below the lower limit of normal reference range (please note that patients may use prior blood draw from routine care as long as done within 6 months of screening and report of lab placed with subject source documents).
  13. Subject has a history of adrenal insufficiency.
  14. Subject is a member of the research staff involved with the study.

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Primary Purpose: Other
  • Allocation: Non-Randomized
  • Interventional Model: Sequential Assignment
  • Masking: None (Open Label)

Arms and Interventions

Participant Group / Arm
Intervention / Treatment
Experimental: Subjects with diabetes wearing Guardian™ Sensor (3). C algorithm applied retrospectively
Subjects wear Guardian™ Sensor (3) over 7 days and undergo one frequent sample test (FST). C sensor algorithm applied retrospectively to raw sensor data.
Characterize the impact of acetaminophen ingestion on the accuracy of Guardian™ Sensor (3) when C sensor algorithm is used.
One time administration of acetaminophen.
Experimental: Subjects with diabetes wearing Guardian™ Sensor (3). Zeus algorithm applied retrospectively
Subjects wear Guardian™ Sensor (3) over 7 days and undergo one FST. Zeus sensor algorithm applied retrospectively to raw sensor data.
One time administration of acetaminophen.
Characterize the impact of acetaminophen ingestion on the accuracy of Guardian™ Sensor (3) when Zeus sensor algorithm is used.

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Bias (mg/dL) Between the Guardian™ Sensor (3) Values and Yellow Springs Instrument™ (YSI™) Plasma Glucose
Time Frame: Frequent sample testing recordings from the 6-hour period leading to and after Acetaminophen ingestion on day 4 or day 5 reported

Bias (mg/dL) between the Guardian™ Sensor (3) values and YSI™ plasma glucose values 1 hour before through 5 hours after ingestion of acetaminophen; results are reported hourly.

Bias = Mean of [Guardian™ Sensor (3) values - YSI™ plasma glucose values].

Frequent sample testing recordings from the 6-hour period leading to and after Acetaminophen ingestion on day 4 or day 5 reported

Secondary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Mean Absolute Relative Difference (MARD, %) Between the Guardian™ Sensor (3) Values and YSI™ Plasma Glucose
Time Frame: Frequent sample testing recordings from the 6-hour period leading to and after Acetaminophen ingestion on day 4 or day 5 reported

Mean absolute relative difference (MARD, %) between the Guardian™ Sensor (3) values and YSI™ plasma glucose 1 hour before through 5 hours after ingestion of acetaminophen; results are reported hourly.

Mean Absolute Relative Difference = Mean of ([absolute difference of Guardian™ Sensor (3) values and YSI™ plasma glucose / YSI™ plasma glucose] * 100%).

Frequent sample testing recordings from the 6-hour period leading to and after Acetaminophen ingestion on day 4 or day 5 reported
Percentage of Guardian™ Sensor (3) Values That Were Within 20% of YSI™ Plasma Glucose Values
Time Frame: Frequent sample testing recordings from the 6-hour period leading to and after Acetaminophen ingestion on day 4 or day 5 reported

Percentage of Guardian™ Sensor (3) values within 20% of YSI™ plasma glucose value (±20 mg/dL [1.1 mmol/L] when YSI™ value less than or equal to (≤) 80 mg/dL [4.4 mmol/L]) during the 1 hour before through 5 hours after ingestion of acetaminophen; results are reported hourly.

20% agreement = Number of Guardian™ Sensor (3) values within 20% (±20 mg/dL) of YSI™ plasma glucose value / Total number of paired Guardian™ Sensor (3) and YSI™ plasma glucose points * 100%.

Frequent sample testing recordings from the 6-hour period leading to and after Acetaminophen ingestion on day 4 or day 5 reported

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start (Actual)

May 18, 2020

Primary Completion (Actual)

July 17, 2020

Study Completion (Actual)

July 17, 2020

Study Registration Dates

First Submitted

May 4, 2020

First Submitted That Met QC Criteria

May 4, 2020

First Posted (Actual)

May 7, 2020

Study Record Updates

Last Update Posted (Actual)

September 8, 2023

Last Update Submitted That Met QC Criteria

August 12, 2023

Last Verified

August 1, 2023

More Information

Terms related to this study

Drug and device information, study documents

Studies a U.S. FDA-regulated drug product

No

Studies a U.S. FDA-regulated device product

Yes

product manufactured in and exported from the U.S.

No

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.

Clinical Trials on Type 2 Diabetes Mellitus

Clinical Trials on Guardian™ Sensor (3) connected to a Guardian™ Connect Transmitter. C sensor algorithm applied retrospectively to raw sensor data.

Subscribe