Remote Blood Pressure Monitoring With the Wearable SENBIOSYS Photoplethysmographic Device

October 14, 2021 updated by: Stéphane Cook, Prof
During the last couple of years, a growing number of wearable devices evolved to provide accurate, cheap and non-invasive monitoring of vitals parameters.This connected care technology could be helpful for treatment and care during a pandemic such as COVID-19. The use of these non-invasive remote monitoring devices can help health care providers to assess patient's vital signs and symptom progression, reducing reducing patient and healthcare provider contact and exposure to COVID-19 during this pandemic.

Study Overview

Status

Completed

Conditions

Intervention / Treatment

Detailed Description

Several cuff-less blood pressure monitoring devices have emerged in patient care using photoplethysmography. PPG utilizes an infrared or visible light to measure peripheral volumetric variations of blood circulation and its waveform has been proven to have a good correlation with BP waveform.Nonetheless, the accuracy of these instruments is of fundamental importance.

Several validation procedures for assessing the precision of BP monitoring devices have been developed, including protocols by the Association for the Advancement of Medical Instrumentation, International Organization for Standardization and European Society of Hypertension. Adherence to these guidelines are essential to ensure the effectiveness of BP measurement of the wearable devices.

Therefore, the investigators propose to test wearable low-noise, low-power SENBIOSYS photoplethysmography signals for estimation of BP.

Study Type

Interventional

Enrollment (Actual)

35

Phase

  • Not Applicable

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Locations

      • Fribourg, Switzerland, 1700
        • University and Hospital Fribourg

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

18 years and older (ADULT, OLDER_ADULT)

Accepts Healthy Volunteers

Yes

Genders Eligible for Study

All

Description

Participants fulfilling the following inclusion criteria are eligible for the study:

  • Age ≥ 18 years
  • Patients referred for coronarography
  • Patient in the intensive care unit requiring invasive blood pressure monitoring and with an arterial catheter in place at the time of inclusion to the study

The presence of any of the following exclusion criteria will lead to exclusion of the participant:

  • Unable or unwilling to provide written informed consent
  • Coronarography in patients with myocardial infarction
  • Patient with suspected or certified COVID-infection
  • Patients with atrial fibrillation

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Primary Purpose: OTHER
  • Allocation: NON_RANDOMIZED
  • Interventional Model: SINGLE_GROUP
  • Masking: NONE

Arms and Interventions

Participant Group / Arm
Intervention / Treatment
EXPERIMENTAL: Standardized measures in the cardiac catheterization laborator
Immediately after coronary angiography, intra-arterial BP waveforms will be recorded using a fluid-filled catheter via right femoral or radial access. The catheter will be flushed before any waveform recordings is made. At first, the catheter will be positioned in the aorta for 3 minutes of stable BP waveforms recording. Intracoronary nitroglycerin will be administered at a dose of 300 µg, newly preceding 3 minute of recording. At the end of the coronary angiography, additional 3 minutes of recording will be performed in the aorta.
The patients will undergo blood pressure estimation as per clinical routine and will simultaneously wear the device.
EXPERIMENTAL: Standardized measures in the intensive care unit
Invasive BP monitoring is a commonly used technique in the ICU and is used to guide many intensive care unit therapies. Enrolled patients must have had an arterial catheter in place at the time of inclusion to the study. Arterial catheterization will be performed by the intensive care team according to current medical guidelines. No arterial catheters were placed for the sole purpose of this study. The Senbiosys device will be placed on the opposite arm of the arterial catheter for simultaneous measurements.
The patients will undergo blood pressure estimation as per clinical routine and will simultaneously wear the device.

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Feasibility of non-invasive BP monitors: blood pressure
Time Frame: 10 to 15 minutes per patient (once)
The primary outcome is the assessment of mean bias (± 95% CI or precision of bias) for systolic blood pressure (SBP), diastolic blood pressure (DBP) and mean blood pressure (MBP) between invasive and non-invasive BP measurements. The standard deviation of the bias (±95% limits of agreement), will be assessed for SBP, DBP, and MBP measurements.
10 to 15 minutes per patient (once)

Secondary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Percentage of signal with artefact,
Time Frame: 10 to 15 minutes per patient (once)
Secondary outcomes include reliability index (Qualification Index QI) for blood pressure epochs and count of qualified epochs.
10 to 15 minutes per patient (once)

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Investigators

  • Study Chair: Stéphane Cook, MD, University of Freiburg

Publications and helpful links

The person responsible for entering information about the study voluntarily provides these publications. These may be about anything related to the study.

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start (ACTUAL)

May 1, 2020

Primary Completion (ACTUAL)

May 31, 2020

Study Completion (ACTUAL)

May 31, 2020

Study Registration Dates

First Submitted

April 22, 2020

First Submitted That Met QC Criteria

May 4, 2020

First Posted (ACTUAL)

May 8, 2020

Study Record Updates

Last Update Posted (ACTUAL)

October 20, 2021

Last Update Submitted That Met QC Criteria

October 14, 2021

Last Verified

May 1, 2020

More Information

Terms related to this study

Other Study ID Numbers

  • SENBIOSYS trial

Plan for Individual participant data (IPD)

Plan to Share Individual Participant Data (IPD)?

NO

Drug and device information, study documents

Studies a U.S. FDA-regulated drug product

No

Studies a U.S. FDA-regulated device product

No

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.

Clinical Trials on Blood Pressure

Subscribe