Evaluation of the XACT Robotic System for Image-guided Percutaneous Lung Procedures

October 30, 2020 updated by: Xact Robotics Ltd.

Evaluation of the Safety, Effectiveness and Usability of the XACT Robotic System for Image Guided Percutaneous Lung Procedures

This is a prospective, single arm study, to evaluate the safety, effectiveness and usability of the XACT device. Subjects undergoing CT-guided, minimally invasive percutaneous lung procedures in the interventional radiology suite. eg. core, biopsy, will participate in the study. System and clinical accuracy will be the primary efficacy endpoint. Usability and safety will also be evaluated.

Study Overview

Detailed Description

This is a prospective, single-arm study is to evaluate the safety, effectiveness and usability of the XACT device. The study will be approved by the Institutional Review Board (IRB) at each of the participating centers prior to patient enrollment.

Subjects undergoing CT-guided, minimally invasive percutaneous lung procedures in the interventional radiology suite, e.g., core biopsy, fine needle aspiration (FNA), tumor ablation, etc., and willing to sign an informed consent document will be screened for compliance with the study inclusion and exclusion criteria.

A total of thirty (41) subjects will be enrolled in the study at 5 medical centers. The intention is to recruit subjects, which will cover a variety of CT-guided interventional lung procedures and a variety of different procedural tools that may be used with the device. Investigators will screen patients based on the inclusion/exclusion criteria described below and the subjects' demographic, general medical history, medical condition/indication, coagulation factors, concomitant medications and vital signs will be obtained.

The system accuracy will be the primary efficacy endpoint and is defined as the measured distance from the tip of the needle/tool to the target, once the XACT robot reaches the pre-defined target .

Clinical accuracy will be the secondary efficacy endpoint and is defined as the ability to place the instrument or procedural tool at a location suitable for the planned intervention. The investigator will review the final instrument position on the post-placement CT images to determine if the pre-operative planned target was reached. This information will be used to calculate success rate.

Additionally, the usability of the XACT device will be evaluated using a rating scale by assessing the ease of device setup, device operation, pre-operative planning, robot positioning, guiding and needle advancement. The total time of the procedure will be recorded, as will the number of CT scans performed and the radiation dose (DLP and CTDI).

Study Type

Interventional

Phase

  • Not Applicable

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Locations

      • Jerusalem, Israel, 91120
        • Hadassah Medical Center
    • Massachusetts
      • Burlington, Massachusetts, United States, 01805
        • Lahey Hospital and Medical Center

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

18 years and older (Adult, Older Adult)

Accepts Healthy Volunteers

No

Genders Eligible for Study

All

Description

Inclusion Criteria:

  • Men and women over 18 years of age
  • Subjects undergoing CT-guided, minimally invasive percutaneous procedures in the interventional radiology suite, e.g., core biopsy, fine needle aspiration (FNA), tumor ablation, etc.
  • Subject is capable and willing to provide informed consent.
  • Subject is capable and willing to adhere to the study procedures.

Exclusion Criteria:

  • Subjects in whom the target is written 1 cm of a major blood vessel or major nerve.
  • Subject with lesion in the central and peripheral nervous system and the spine.
  • Subjects in whom the target is within 1 cm of a major blood vessel or major nerve.
  • Subject with significant coagulopathy
  • Subjects with a preexisting conditions, which, in the opinion of the investigator, may interfere with the conduct of the study.
  • Subjects with an unstable medical condition, e.g. unstable hypertension, unstable cardiac disease, etc.
  • Subjects who are uncooperative or cannot follow instructions.
  • Subjects with a mental state that may preclude completion of the study procedure.
  • Female subjects who are pregnant or nursing.

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Primary Purpose: Treatment
  • Allocation: N/A
  • Interventional Model: Single Group Assignment
  • Masking: None (Open Label)

Arms and Interventions

Participant Group / Arm
Intervention / Treatment
Experimental: Procedures with XACT Robotic System
Device: XACT Robotic System The XACT device is a real-time, CT image guided, 3-dimensional robotic system. The XACT device is intended for use as an image guided positioning and steering system for insertion of clinical tools, such as biopsy needles, ablation needles, etc., during minimally invasive percutaneous lung procedures. The system is defined to guide (i.e., position and steer) the tool according to a predefined trajectory following a registration process between the device's coordinate system and real-time CT images.
CT-guided Minimally Invasive Procedures e.g., Biopsies

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
System accuracy will be determined by the measured distance from the tip of the needle/tool to the target.
Time Frame: At the end of each interventional procedure, approximately 1 hour
The primary endpoint of the study is to evaluate the system accuracy of the XACT System. System accuracy of the study will be determined by the measured distance from the tip of the needle/tool to the target, once the XACT Robot reaches the pre-defined target.
At the end of each interventional procedure, approximately 1 hour

Secondary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Clinical accuracy of the study will be determined by the ability of the XACT system to reach the pre-defined target in each procedure.
Time Frame: At the end of each interventional procedure, approximately 1 hour
The secondary endpoint of the study is to evaluate the clinical accuracy of the XACT system. Clinical accuracy of the study will be determined by the ability of the XACT system to reach the pre-defined target in each procedure.
At the end of each interventional procedure, approximately 1 hour

Other Outcome Measures

Outcome Measure
Measure Description
Time Frame
Device Usability
Time Frame: At the end of each interventional procedure, approximately 1 hour
Usability of the XACT device performance in the hands of users will be evaluated using a rating scale for assessing the ease of device performance.
At the end of each interventional procedure, approximately 1 hour
Evaluation of Safety
Time Frame: Through study completion (2 weeks)
Evaluation of safety by assessing frequency, severity and causality of device related adverse events.
Through study completion (2 weeks)

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Investigators

  • Study Director: Ahava Stein, A. Stein - Regulatory Affairs Consulting Ltd.

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start (Anticipated)

September 1, 2018

Primary Completion (Anticipated)

September 1, 2019

Study Completion (Anticipated)

December 1, 2019

Study Registration Dates

First Submitted

February 14, 2018

First Submitted That Met QC Criteria

October 30, 2020

First Posted (Actual)

November 3, 2020

Study Record Updates

Last Update Posted (Actual)

November 3, 2020

Last Update Submitted That Met QC Criteria

October 30, 2020

Last Verified

October 1, 2020

More Information

Terms related to this study

Other Study ID Numbers

  • CLN-002-00

Plan for Individual participant data (IPD)

Plan to Share Individual Participant Data (IPD)?

NO

Drug and device information, study documents

Studies a U.S. FDA-regulated drug product

No

Studies a U.S. FDA-regulated device product

Yes

product manufactured in and exported from the U.S.

No

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.