Safety and Effectiveness of the External Fastening System for Temporary Active Fixation Pacemakers (TEMPACE) (TEMPACE)
April 16, 2025 updated by: ICU Medical Technologies S.L.
Temporary cardiac pacing with active-fixation leads (TPAFL) using a reusable permanent pacemaker generator has been shown to be safer than lead systems without fixation.
However, TPAFL requires the off-label (non-certificated) use of Pacemaker leads and generators.
It is designed a fastening system to ensure the safety and efficacy of the procedure: the KronoSafe System®.
The aim of this study is to demonstrate the safety and effectiveness of the KronoSafe System® for temporary pacing in a series of patients receiving TPAFL.
Study Overview
Status
Completed
Conditions
Intervention / Treatment
Study Type
Interventional
Enrollment (Actual)
30
Phase
- Not Applicable
Contacts and Locations
This section provides the contact details for those conducting the study, and information on where this study is being conducted.
Study Locations
-
-
-
Alicante, Spain, 03010
- Hospital General Universitario de Alicante Dr.Balmis
-
Barcelona, Spain, 08035
- Hospital Universitari Vall d'Hebron
-
-
Participation Criteria
Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.
Eligibility Criteria
Ages Eligible for Study
18 years and older (Adult, Older Adult)
Accepts Healthy Volunteers
No
Description
Inclusion Criteria:
- Patient older than 18 years.
- Need for temporary cardiac stimulation
- Temporary active fixation pacemaker implantation through left or right jugular access.
- Signature of informed consent
- Correct functioning of the electrode after the implantation: sensing of the R wave above 5 millivolts. Ventricular pacing threshold below 1 volt. Electrode impedance between 400-1200 Ohm.
- Pacemaker programming with stimulation output at 5 volts intensity and 1 millisecond pulse duration.
- Placement of the KronoSafe® fastening system after electrode implantation.
Exclusion Criteria:
- Complication during the procedure: pneumothorax, hematoma at the puncture site, cardiac perforation, accidental arterial puncture or other that, in the investigator's opinion, poses an added risk of pacemaker dysfunction.
- Need to transfer the patient to another center with loss of follow-up.
- Absence of surveillance through continuous electrocardiographic monitoring with detection of arrhythmic events.
Study Plan
This section provides details of the study plan, including how the study is designed and what the study is measuring.
How is the study designed?
Design Details
- Primary Purpose: Supportive Care
- Allocation: N/A
- Interventional Model: Single Group Assignment
- Masking: None (Open Label)
Arms and Interventions
Participant Group / Arm |
Intervention / Treatment |
|---|---|
|
Experimental: Temporary Cardiac Pacing with Active-Fixation Leads
|
All patients included in the study will have a temporary pacemaker with active fixation leads, held by the KronoSafe® fixation system.
|
What is the study measuring?
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Ventricular R wave detection
Time Frame: Immediately after the procedure of lead implantation
|
Ventricular R wave, measured in millivolts with pacemaker generator programmer
|
Immediately after the procedure of lead implantation
|
|
Lead impedance detection
Time Frame: Immediately after the procedure of lead implantation
|
Lead impedance, measured in ohms with pacemaker generator programmer
|
Immediately after the procedure of lead implantation
|
|
Right ventricle threshold detection
Time Frame: Immediately after the procedure of lead implantation
|
Right ventricle threshold, measured in volts with pacemaker generator programmer
|
Immediately after the procedure of lead implantation
|
|
Ventricular R wave detection
Time Frame: 24 hours after the procedure of lead implantation
|
Ventricular R wave, measured in millivolts with pacemaker generator programmer
|
24 hours after the procedure of lead implantation
|
|
Ventricular R wave detection
Time Frame: day 3 after the procedure of lead implantation
|
Ventricular R wave, measured in millivolts with pacemaker generator programmer
|
day 3 after the procedure of lead implantation
|
|
Ventricular R wave detection
Time Frame: day 5 after the procedure of lead implantation
|
Ventricular R wave, measured in millivolts with pacemaker generator programmer
|
day 5 after the procedure of lead implantation
|
|
Ventricular R wave detection
Time Frame: day 7 after the procedure of lead implantation
|
Ventricular R wave, measured in millivolts with pacemaker generator programmer
|
day 7 after the procedure of lead implantation
|
|
Ventricular R wave detection
Time Frame: day 9 after the procedure of lead implantation
|
Ventricular R wave, measured in millivolts with pacemaker generator programmer
|
day 9 after the procedure of lead implantation
|
|
Ventricular R wave detection
Time Frame: day 11 after the procedure of lead implantation
|
Ventricular R wave, measured in millivolts with pacemaker generator programmer
|
day 11 after the procedure of lead implantation
|
|
Ventricular R wave detection
Time Frame: day 13 after the procedure of lead implantation
|
Ventricular R wave, measured in millivolts with pacemaker generator programmer
|
day 13 after the procedure of lead implantation
|
|
Ventricular R wave detection
Time Frame: day 15 after the procedure of lead implantation
|
Ventricular R wave, measured in millivolts with pacemaker generator programmer
|
day 15 after the procedure of lead implantation
|
|
Ventricular R wave detection
Time Frame: day 17 after the procedure of lead implantation
|
Ventricular R wave, measured in millivolts with pacemaker generator programmer
|
day 17 after the procedure of lead implantation
|
|
Ventricular R wave detection
Time Frame: day 19 after the procedure of lead implantation
|
Ventricular R wave, measured in millivolts with pacemaker generator programmer
|
day 19 after the procedure of lead implantation
|
|
Ventricular R wave detection
Time Frame: day 21 after the procedure of lead implantation
|
Ventricular R wave, measured in millivolts with pacemaker generator programmer
|
day 21 after the procedure of lead implantation
|
|
Ventricular R wave detection
Time Frame: day 23 after the procedure of lead implantation
|
Ventricular R wave, measured in millivolts with pacemaker generator programmer
|
day 23 after the procedure of lead implantation
|
|
Ventricular R wave detection
Time Frame: day 25 after the procedure of lead implantation
|
Ventricular R wave, measured in millivolts with pacemaker generator programmer
|
day 25 after the procedure of lead implantation
|
|
Ventricular R wave detection
Time Frame: day 27 after the procedure of lead implantation
|
Ventricular R wave, measured in millivolts with pacemaker generator programmer
|
day 27 after the procedure of lead implantation
|
|
Ventricular R wave detection
Time Frame: day 29 after the procedure of lead implantation
|
Ventricular R wave, measured in millivolts with pacemaker generator programmer
|
day 29 after the procedure of lead implantation
|
|
Ventricular R wave detection
Time Frame: Immediately before the procedure of lead removal
|
Ventricular R wave, measured in millivolts with pacemaker generator programmer
|
Immediately before the procedure of lead removal
|
|
Lead impedance detection
Time Frame: 24 hours after the procedure of lead implantation
|
Lead impedance, measured in ohms with pacemaker generator programmer
|
24 hours after the procedure of lead implantation
|
|
Lead impedance detection
Time Frame: day 3 after the procedure of lead implantation
|
Lead impedance, measured in ohms with pacemaker generator programmer
|
day 3 after the procedure of lead implantation
|
|
Lead impedance detection
Time Frame: day 5 after the procedure of lead implantation
|
Lead impedance, measured in ohms with pacemaker generator programmer
|
day 5 after the procedure of lead implantation
|
|
Lead impedance detection
Time Frame: day 7 after the procedure of lead implantation
|
Lead impedance, measured in ohms with pacemaker generator programmer
|
day 7 after the procedure of lead implantation
|
|
Lead impedance detection
Time Frame: day 9 after the procedure of lead implantation
|
Lead impedance, measured in ohms with pacemaker generator programmer
|
day 9 after the procedure of lead implantation
|
|
Lead impedance detection
Time Frame: day 11 after the procedure of lead implantation
|
Lead impedance, measured in ohms with pacemaker generator programmer
|
day 11 after the procedure of lead implantation
|
|
Lead impedance detection
Time Frame: day 13 after the procedure of lead implantation
|
Lead impedance, measured in ohms with pacemaker generator programmer
|
day 13 after the procedure of lead implantation
|
|
Lead impedance detection
Time Frame: day 15 after the procedure of lead implantation
|
Lead impedance, measured in ohms with pacemaker generator programmer
|
day 15 after the procedure of lead implantation
|
|
Lead impedance detection
Time Frame: day 17 after the procedure of lead implantation
|
Lead impedance, measured in ohms with pacemaker generator programmer
|
day 17 after the procedure of lead implantation
|
|
Lead impedance detection
Time Frame: day 19 after the procedure of lead implantation
|
Lead impedance, measured in ohms with pacemaker generator programmer
|
day 19 after the procedure of lead implantation
|
|
Lead impedance detection
Time Frame: day 21 after the procedure of lead implantation
|
Lead impedance, measured in ohms with pacemaker generator programmer
|
day 21 after the procedure of lead implantation
|
|
Lead impedance detection
Time Frame: day 23 after the procedure of lead implantation
|
Lead impedance, measured in ohms with pacemaker generator programmer
|
day 23 after the procedure of lead implantation
|
|
Lead impedance detection
Time Frame: day 25 after the procedure of lead implantation
|
Lead impedance, measured in ohms with pacemaker generator programmer
|
day 25 after the procedure of lead implantation
|
|
Lead impedance detection
Time Frame: day 27 after the procedure of lead implantation
|
Lead impedance, measured in ohms with pacemaker generator programmer
|
day 27 after the procedure of lead implantation
|
|
Lead impedance detection
Time Frame: day 29 after the procedure of lead implantation
|
Lead impedance, measured in ohms with pacemaker generator programmer
|
day 29 after the procedure of lead implantation
|
|
Lead impedance detection
Time Frame: Immediately before the procedure of lead removal
|
Lead impedance, measured in ohms with pacemaker generator programmer
|
Immediately before the procedure of lead removal
|
|
Right ventricle threshold detection
Time Frame: 24 hours after the procedure of lead implantation
|
Right ventricle threshold, measured in volts with pacemaker generator programmer
|
24 hours after the procedure of lead implantation
|
|
Right ventricle threshold detection
Time Frame: day 3 after the procedure of lead implantation
|
Right ventricle threshold, measured in volts with pacemaker generator programmer
|
day 3 after the procedure of lead implantation
|
|
Right ventricle threshold detection
Time Frame: day 5 after the procedure of lead implantation
|
Right ventricle threshold, measured in volts with pacemaker generator programmer
|
day 5 after the procedure of lead implantation
|
|
Right ventricle threshold detection
Time Frame: day 7 after the procedure of lead implantation
|
Right ventricle threshold, measured in volts with pacemaker generator programmer
|
day 7 after the procedure of lead implantation
|
|
Right ventricle threshold detection
Time Frame: day 9 after the procedure of lead implantation
|
Right ventricle threshold, measured in volts with pacemaker generator programmer
|
day 9 after the procedure of lead implantation
|
|
Right ventricle threshold detection
Time Frame: day 11 after the procedure of lead implantation
|
Right ventricle threshold, measured in volts with pacemaker generator programmer
|
day 11 after the procedure of lead implantation
|
|
Right ventricle threshold detection
Time Frame: day 13 after the procedure of lead implantation
|
Right ventricle threshold, measured in volts with pacemaker generator programmer
|
day 13 after the procedure of lead implantation
|
|
Right ventricle threshold detection
Time Frame: day 15 after the procedure of lead implantation
|
Right ventricle threshold, measured in volts with pacemaker generator programmer
|
day 15 after the procedure of lead implantation
|
|
Right ventricle threshold detection
Time Frame: day 17 after the procedure of lead implantation
|
Right ventricle threshold, measured in volts with pacemaker generator programmer
|
day 17 after the procedure of lead implantation
|
|
Right ventricle threshold detection
Time Frame: day 19 after the procedure of lead implantation
|
Right ventricle threshold, measured in volts with pacemaker generator programmer
|
day 19 after the procedure of lead implantation
|
|
Right ventricle threshold detection
Time Frame: day 21 after the procedure of lead implantation
|
Right ventricle threshold, measured in volts with pacemaker generator programmer
|
day 21 after the procedure of lead implantation
|
|
Right ventricle threshold detection
Time Frame: day 23 after the procedure of lead implantation
|
Right ventricle threshold, measured in volts with pacemaker generator programmer
|
day 23 after the procedure of lead implantation
|
|
Right ventricle threshold detection
Time Frame: day 25 after the procedure of lead implantation
|
Right ventricle threshold, measured in volts with pacemaker generator programmer
|
day 25 after the procedure of lead implantation
|
|
Right ventricle threshold detection
Time Frame: day 27 after the procedure of lead implantation
|
Right ventricle threshold, measured in volts with pacemaker generator programmer
|
day 27 after the procedure of lead implantation
|
|
Right ventricle threshold detection
Time Frame: day 29 after the procedure of lead implantation
|
Right ventricle threshold, measured in volts with pacemaker generator programmer
|
day 29 after the procedure of lead implantation
|
|
Right ventricle threshold detection
Time Frame: Immediately before the procedure of lead removal
|
Right ventricle threshold, measured in volts with pacemaker generator programmer
|
Immediately before the procedure of lead removal
|
|
Lead displacement detection
Time Frame: Immediately before the procedure of lead removal
|
Visual observation of 2 cm electrode displacement through vein acces by clinical examination
|
Immediately before the procedure of lead removal
|
|
Capture failure detection
Time Frame: Immediately before the procedure of lead removal
|
Right ventricle threshold above 5 volts during 1 millisecond, measured with pacemaker generator programmer.
|
Immediately before the procedure of lead removal
|
|
Sensing failure detection
Time Frame: Immediately before the procedure of lead removal
|
Ventricular R wave below 1 volt, measured with pacemaker generator programmer
|
Immediately before the procedure of lead removal
|
|
Lead insulation breakdown detection
Time Frame: Immediately before the procedure of lead removal
|
Visual observation of lead insulation breakdown by clinical examination
|
Immediately before the procedure of lead removal
|
|
Generator detachment detection
Time Frame: Immediately before the procedure of lead removal
|
Visual observation of generator detachment by clinical examination.
removal of the generator from the casing.
|
Immediately before the procedure of lead removal
|
|
Cardiac peroration detection
Time Frame: Immediately before the procedure of lead removal
|
New pericardial effusion detected by Echocardiography with signs of cardiac tamponade: respiratory variation of transmitral filling flow greater than 25%
|
Immediately before the procedure of lead removal
|
Secondary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Detection of Bacteremia
Time Frame: Immediately after the procedure of lead removal
|
Isolation of the same bacteria in the pacemaker lead and 2 blood cultures from the patient
|
Immediately after the procedure of lead removal
|
|
Detection of Agitation
Time Frame: Immediately after the procedure of lead removal
|
Record of antipsychotic and anxiolytic treatment of the patient
|
Immediately after the procedure of lead removal
|
Collaborators and Investigators
This is where you will find people and organizations involved with this study.
Sponsor
Collaborators
Investigators
- Principal Investigator: Raimundo Vicente, MD, Intensive Care Physician at Dr. Balmis University Hospital (Alicante)
Publications and helpful links
The person responsible for entering information about the study voluntarily provides these publications. These may be about anything related to the study.
General Publications
- Vicente-Miralles R, Martin-Langerwerf DA, Nunez-Martinez JM, Marco-Juan A, Sangio-Ferreyros AD, Davila-Arias A, Montes-Jacobo L, Barrios V, Palazon-Bru A, Gil-Guillen VF, Bertomeu-Gonzalez V. A new fastening system for temporary pacing with active-fixation leads: effectiveness and safety. Eur Heart J Acute Cardiovasc Care. 2022 Mar 16;11(3):224-229. doi: 10.1093/ehjacc/zuab119.
- Zei PC, Eckart RE, Epstein LM. Modified temporary cardiac pacing using transvenous active fixation leads and external re-sterilized pulse generators. J Am Coll Cardiol. 2006 Apr 4;47(7):1487-9. doi: 10.1016/j.jacc.2006.01.006. Epub 2006 Mar 15. No abstract available.
- Kawata H, Pretorius V, Phan H, Mulpuru S, Gadiyaram V, Patel J, Steltzner D, Krummen D, Feld G, Birgersdotter-Green U. Utility and safety of temporary pacing using active fixation leads and externalized re-usable permanent pacemakers after lead extraction. Europace. 2013 Sep;15(9):1287-91. doi: 10.1093/europace/eut045. Epub 2013 Mar 12.
- Betts TR. Regional survey of temporary transvenous pacing procedures and complications. Postgrad Med J. 2003 Aug;79(934):463-5. doi: 10.1136/pmj.79.934.463.
Study record dates
These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.
Study Major Dates
Study Start (Actual)
January 5, 2023
Primary Completion (Actual)
November 18, 2024
Study Completion (Actual)
December 31, 2024
Study Registration Dates
First Submitted
March 28, 2022
First Submitted That Met QC Criteria
April 25, 2022
First Posted (Actual)
April 28, 2022
Study Record Updates
Last Update Posted (Actual)
April 20, 2025
Last Update Submitted That Met QC Criteria
April 16, 2025
Last Verified
April 1, 2025
More Information
Terms related to this study
Additional Relevant MeSH Terms
Other Study ID Numbers
- TEMPACE
Plan for Individual participant data (IPD)
Plan to Share Individual Participant Data (IPD)?
NO
Drug and device information, study documents
Studies a U.S. FDA-regulated drug product
No
Studies a U.S. FDA-regulated device product
No
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.