Safety and Effectiveness of the External Fastening System for Temporary Active Fixation Pacemakers (TEMPACE) (TEMPACE)

April 16, 2025 updated by: ICU Medical Technologies S.L.
Temporary cardiac pacing with active-fixation leads (TPAFL) using a reusable permanent pacemaker generator has been shown to be safer than lead systems without fixation. However, TPAFL requires the off-label (non-certificated) use of Pacemaker leads and generators. It is designed a fastening system to ensure the safety and efficacy of the procedure: the KronoSafe System®. The aim of this study is to demonstrate the safety and effectiveness of the KronoSafe System® for temporary pacing in a series of patients receiving TPAFL.

Study Overview

Status

Completed

Conditions

Intervention / Treatment

Study Type

Interventional

Enrollment (Actual)

30

Phase

  • Not Applicable

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Locations

      • Alicante, Spain, 03010
        • Hospital General Universitario de Alicante Dr.Balmis
      • Barcelona, Spain, 08035
        • Hospital Universitari Vall d'Hebron

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

18 years and older (Adult, Older Adult)

Accepts Healthy Volunteers

No

Description

Inclusion Criteria:

  • Patient older than 18 years.
  • Need for temporary cardiac stimulation
  • Temporary active fixation pacemaker implantation through left or right jugular access.
  • Signature of informed consent
  • Correct functioning of the electrode after the implantation: sensing of the R wave above 5 millivolts. Ventricular pacing threshold below 1 volt. Electrode impedance between 400-1200 Ohm.
  • Pacemaker programming with stimulation output at 5 volts intensity and 1 millisecond pulse duration.
  • Placement of the KronoSafe® fastening system after electrode implantation.

Exclusion Criteria:

  • Complication during the procedure: pneumothorax, hematoma at the puncture site, cardiac perforation, accidental arterial puncture or other that, in the investigator's opinion, poses an added risk of pacemaker dysfunction.
  • Need to transfer the patient to another center with loss of follow-up.
  • Absence of surveillance through continuous electrocardiographic monitoring with detection of arrhythmic events.

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Primary Purpose: Supportive Care
  • Allocation: N/A
  • Interventional Model: Single Group Assignment
  • Masking: None (Open Label)

Arms and Interventions

Participant Group / Arm
Intervention / Treatment
Experimental: Temporary Cardiac Pacing with Active-Fixation Leads
All patients included in the study will have a temporary pacemaker with active fixation leads, held by the KronoSafe® fixation system.

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Ventricular R wave detection
Time Frame: Immediately after the procedure of lead implantation
Ventricular R wave, measured in millivolts with pacemaker generator programmer
Immediately after the procedure of lead implantation
Lead impedance detection
Time Frame: Immediately after the procedure of lead implantation
Lead impedance, measured in ohms with pacemaker generator programmer
Immediately after the procedure of lead implantation
Right ventricle threshold detection
Time Frame: Immediately after the procedure of lead implantation
Right ventricle threshold, measured in volts with pacemaker generator programmer
Immediately after the procedure of lead implantation
Ventricular R wave detection
Time Frame: 24 hours after the procedure of lead implantation
Ventricular R wave, measured in millivolts with pacemaker generator programmer
24 hours after the procedure of lead implantation
Ventricular R wave detection
Time Frame: day 3 after the procedure of lead implantation
Ventricular R wave, measured in millivolts with pacemaker generator programmer
day 3 after the procedure of lead implantation
Ventricular R wave detection
Time Frame: day 5 after the procedure of lead implantation
Ventricular R wave, measured in millivolts with pacemaker generator programmer
day 5 after the procedure of lead implantation
Ventricular R wave detection
Time Frame: day 7 after the procedure of lead implantation
Ventricular R wave, measured in millivolts with pacemaker generator programmer
day 7 after the procedure of lead implantation
Ventricular R wave detection
Time Frame: day 9 after the procedure of lead implantation
Ventricular R wave, measured in millivolts with pacemaker generator programmer
day 9 after the procedure of lead implantation
Ventricular R wave detection
Time Frame: day 11 after the procedure of lead implantation
Ventricular R wave, measured in millivolts with pacemaker generator programmer
day 11 after the procedure of lead implantation
Ventricular R wave detection
Time Frame: day 13 after the procedure of lead implantation
Ventricular R wave, measured in millivolts with pacemaker generator programmer
day 13 after the procedure of lead implantation
Ventricular R wave detection
Time Frame: day 15 after the procedure of lead implantation
Ventricular R wave, measured in millivolts with pacemaker generator programmer
day 15 after the procedure of lead implantation
Ventricular R wave detection
Time Frame: day 17 after the procedure of lead implantation
Ventricular R wave, measured in millivolts with pacemaker generator programmer
day 17 after the procedure of lead implantation
Ventricular R wave detection
Time Frame: day 19 after the procedure of lead implantation
Ventricular R wave, measured in millivolts with pacemaker generator programmer
day 19 after the procedure of lead implantation
Ventricular R wave detection
Time Frame: day 21 after the procedure of lead implantation
Ventricular R wave, measured in millivolts with pacemaker generator programmer
day 21 after the procedure of lead implantation
Ventricular R wave detection
Time Frame: day 23 after the procedure of lead implantation
Ventricular R wave, measured in millivolts with pacemaker generator programmer
day 23 after the procedure of lead implantation
Ventricular R wave detection
Time Frame: day 25 after the procedure of lead implantation
Ventricular R wave, measured in millivolts with pacemaker generator programmer
day 25 after the procedure of lead implantation
Ventricular R wave detection
Time Frame: day 27 after the procedure of lead implantation
Ventricular R wave, measured in millivolts with pacemaker generator programmer
day 27 after the procedure of lead implantation
Ventricular R wave detection
Time Frame: day 29 after the procedure of lead implantation
Ventricular R wave, measured in millivolts with pacemaker generator programmer
day 29 after the procedure of lead implantation
Ventricular R wave detection
Time Frame: Immediately before the procedure of lead removal
Ventricular R wave, measured in millivolts with pacemaker generator programmer
Immediately before the procedure of lead removal
Lead impedance detection
Time Frame: 24 hours after the procedure of lead implantation
Lead impedance, measured in ohms with pacemaker generator programmer
24 hours after the procedure of lead implantation
Lead impedance detection
Time Frame: day 3 after the procedure of lead implantation
Lead impedance, measured in ohms with pacemaker generator programmer
day 3 after the procedure of lead implantation
Lead impedance detection
Time Frame: day 5 after the procedure of lead implantation
Lead impedance, measured in ohms with pacemaker generator programmer
day 5 after the procedure of lead implantation
Lead impedance detection
Time Frame: day 7 after the procedure of lead implantation
Lead impedance, measured in ohms with pacemaker generator programmer
day 7 after the procedure of lead implantation
Lead impedance detection
Time Frame: day 9 after the procedure of lead implantation
Lead impedance, measured in ohms with pacemaker generator programmer
day 9 after the procedure of lead implantation
Lead impedance detection
Time Frame: day 11 after the procedure of lead implantation
Lead impedance, measured in ohms with pacemaker generator programmer
day 11 after the procedure of lead implantation
Lead impedance detection
Time Frame: day 13 after the procedure of lead implantation
Lead impedance, measured in ohms with pacemaker generator programmer
day 13 after the procedure of lead implantation
Lead impedance detection
Time Frame: day 15 after the procedure of lead implantation
Lead impedance, measured in ohms with pacemaker generator programmer
day 15 after the procedure of lead implantation
Lead impedance detection
Time Frame: day 17 after the procedure of lead implantation
Lead impedance, measured in ohms with pacemaker generator programmer
day 17 after the procedure of lead implantation
Lead impedance detection
Time Frame: day 19 after the procedure of lead implantation
Lead impedance, measured in ohms with pacemaker generator programmer
day 19 after the procedure of lead implantation
Lead impedance detection
Time Frame: day 21 after the procedure of lead implantation
Lead impedance, measured in ohms with pacemaker generator programmer
day 21 after the procedure of lead implantation
Lead impedance detection
Time Frame: day 23 after the procedure of lead implantation
Lead impedance, measured in ohms with pacemaker generator programmer
day 23 after the procedure of lead implantation
Lead impedance detection
Time Frame: day 25 after the procedure of lead implantation
Lead impedance, measured in ohms with pacemaker generator programmer
day 25 after the procedure of lead implantation
Lead impedance detection
Time Frame: day 27 after the procedure of lead implantation
Lead impedance, measured in ohms with pacemaker generator programmer
day 27 after the procedure of lead implantation
Lead impedance detection
Time Frame: day 29 after the procedure of lead implantation
Lead impedance, measured in ohms with pacemaker generator programmer
day 29 after the procedure of lead implantation
Lead impedance detection
Time Frame: Immediately before the procedure of lead removal
Lead impedance, measured in ohms with pacemaker generator programmer
Immediately before the procedure of lead removal
Right ventricle threshold detection
Time Frame: 24 hours after the procedure of lead implantation
Right ventricle threshold, measured in volts with pacemaker generator programmer
24 hours after the procedure of lead implantation
Right ventricle threshold detection
Time Frame: day 3 after the procedure of lead implantation
Right ventricle threshold, measured in volts with pacemaker generator programmer
day 3 after the procedure of lead implantation
Right ventricle threshold detection
Time Frame: day 5 after the procedure of lead implantation
Right ventricle threshold, measured in volts with pacemaker generator programmer
day 5 after the procedure of lead implantation
Right ventricle threshold detection
Time Frame: day 7 after the procedure of lead implantation
Right ventricle threshold, measured in volts with pacemaker generator programmer
day 7 after the procedure of lead implantation
Right ventricle threshold detection
Time Frame: day 9 after the procedure of lead implantation
Right ventricle threshold, measured in volts with pacemaker generator programmer
day 9 after the procedure of lead implantation
Right ventricle threshold detection
Time Frame: day 11 after the procedure of lead implantation
Right ventricle threshold, measured in volts with pacemaker generator programmer
day 11 after the procedure of lead implantation
Right ventricle threshold detection
Time Frame: day 13 after the procedure of lead implantation
Right ventricle threshold, measured in volts with pacemaker generator programmer
day 13 after the procedure of lead implantation
Right ventricle threshold detection
Time Frame: day 15 after the procedure of lead implantation
Right ventricle threshold, measured in volts with pacemaker generator programmer
day 15 after the procedure of lead implantation
Right ventricle threshold detection
Time Frame: day 17 after the procedure of lead implantation
Right ventricle threshold, measured in volts with pacemaker generator programmer
day 17 after the procedure of lead implantation
Right ventricle threshold detection
Time Frame: day 19 after the procedure of lead implantation
Right ventricle threshold, measured in volts with pacemaker generator programmer
day 19 after the procedure of lead implantation
Right ventricle threshold detection
Time Frame: day 21 after the procedure of lead implantation
Right ventricle threshold, measured in volts with pacemaker generator programmer
day 21 after the procedure of lead implantation
Right ventricle threshold detection
Time Frame: day 23 after the procedure of lead implantation
Right ventricle threshold, measured in volts with pacemaker generator programmer
day 23 after the procedure of lead implantation
Right ventricle threshold detection
Time Frame: day 25 after the procedure of lead implantation
Right ventricle threshold, measured in volts with pacemaker generator programmer
day 25 after the procedure of lead implantation
Right ventricle threshold detection
Time Frame: day 27 after the procedure of lead implantation
Right ventricle threshold, measured in volts with pacemaker generator programmer
day 27 after the procedure of lead implantation
Right ventricle threshold detection
Time Frame: day 29 after the procedure of lead implantation
Right ventricle threshold, measured in volts with pacemaker generator programmer
day 29 after the procedure of lead implantation
Right ventricle threshold detection
Time Frame: Immediately before the procedure of lead removal
Right ventricle threshold, measured in volts with pacemaker generator programmer
Immediately before the procedure of lead removal
Lead displacement detection
Time Frame: Immediately before the procedure of lead removal
Visual observation of 2 cm electrode displacement through vein acces by clinical examination
Immediately before the procedure of lead removal
Capture failure detection
Time Frame: Immediately before the procedure of lead removal
Right ventricle threshold above 5 volts during 1 millisecond, measured with pacemaker generator programmer.
Immediately before the procedure of lead removal
Sensing failure detection
Time Frame: Immediately before the procedure of lead removal
Ventricular R wave below 1 volt, measured with pacemaker generator programmer
Immediately before the procedure of lead removal
Lead insulation breakdown detection
Time Frame: Immediately before the procedure of lead removal
Visual observation of lead insulation breakdown by clinical examination
Immediately before the procedure of lead removal
Generator detachment detection
Time Frame: Immediately before the procedure of lead removal
Visual observation of generator detachment by clinical examination. removal of the generator from the casing.
Immediately before the procedure of lead removal
Cardiac peroration detection
Time Frame: Immediately before the procedure of lead removal
New pericardial effusion detected by Echocardiography with signs of cardiac tamponade: respiratory variation of transmitral filling flow greater than 25%
Immediately before the procedure of lead removal

Secondary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Detection of Bacteremia
Time Frame: Immediately after the procedure of lead removal
Isolation of the same bacteria in the pacemaker lead and 2 blood cultures from the patient
Immediately after the procedure of lead removal
Detection of Agitation
Time Frame: Immediately after the procedure of lead removal
Record of antipsychotic and anxiolytic treatment of the patient
Immediately after the procedure of lead removal

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Investigators

  • Principal Investigator: Raimundo Vicente, MD, Intensive Care Physician at Dr. Balmis University Hospital (Alicante)

Publications and helpful links

The person responsible for entering information about the study voluntarily provides these publications. These may be about anything related to the study.

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start (Actual)

January 5, 2023

Primary Completion (Actual)

November 18, 2024

Study Completion (Actual)

December 31, 2024

Study Registration Dates

First Submitted

March 28, 2022

First Submitted That Met QC Criteria

April 25, 2022

First Posted (Actual)

April 28, 2022

Study Record Updates

Last Update Posted (Actual)

April 20, 2025

Last Update Submitted That Met QC Criteria

April 16, 2025

Last Verified

April 1, 2025

More Information

Terms related to this study

Plan for Individual participant data (IPD)

Plan to Share Individual Participant Data (IPD)?

NO

Drug and device information, study documents

Studies a U.S. FDA-regulated drug product

No

Studies a U.S. FDA-regulated device product

No

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.