- ICH GCP
- US Clinical Trials Registry
- Clinical Trial NCT05796219
Automated Low Dose Risk Assessment Mammography (ALDRAM) (ALDRAM)
Technical Development of Automated Low Dose Risk Assessment Mammography (ALDRAM) in Women Attending for Annual Mammography Through a Family History Clinic
Study Overview
Detailed Description
The more prolonged compression of the breast required to acquire the LD images may cause some additional breast discomfort. However, in a similar study in The Netherlands only 1% of women could not tolerate the procedure.
Participants will receive an extra dose of ionising radiation amounting to 20% of the radiation dose of a single FFDM view (of which 4 are usually taken, 2 on each breast). The risks associated with this are described in section B and entered into the PIS.
The research question for this study is whether an automated, low dose mammogram can be developed to provide an accurate assessment of mammographic density, and thus breast cancer risk, in women aged 30-45.
The two key objectives are:
- To develop machine learning methods for automated quantification of mammographic density using low dose mammograms in a study of 600 women attending for annual Full Field Digital Mammography (FFDM).
- To validate Automated Low Dose Risk Assessment Mammography (ALDRAM) against automated FFDM assessment that has been shown to predict BC risk.
Breast cancer is the commonest of all causes of death in young women. Currently the only factor triggering referral to risk assessment clinics is the presence of a family history. Although this is known to confer increased risk, only 20% of young women who develop BC actually have such a family history of the disease. Thus 80% of cases come 'out of the blue' and methods to assess risk in the general population are required if we are to improve survival (screening in those at increased risk aged 35-45 has been shown to reduce mortality). High mammographic density has been shown in multiple studies to confer significantly increased risk of BC, however population screening with full field digital mammograms is not recommended in those <40 primarily due to the reduced sensitivity and use of a moderate doses of ionising radiation. There is thus an urgent need for safe, high throughput, techniques to accurately define mammographic density in young women - the purpose of this study. Following consent and initial Case Report Form (CRF) completion, mammography will commence as standard. The right breast craniocaudal (CC) view will be performed first and the steps below followed:
- The breast will be positioned as standard for the CC view.
- The Automatic Optimisation of Parameters (AOP; also known as Automatic Exposure Control (AEC)) will be performed for target, filter, Peak KiloVoltage (kVp) and MilliAmpere second (mAs) selection for Full Field Digital Mammography (FFDM).
- The full field CC view will be performed as standard.
- The breast will remain compressed whilst the radiographer manually reduces the mAs by 90% or to the minimum value of 4 mAs (the machine does not deliver less than 4 mAs). No other parameters will be changed. A repeat CC exposure at this reduced dose will be performed.
- The machine will be reset to automatic and the subsequent mediolateral oblique (MLO) view will be obtained and the low dose view performed again as above.
- The left breast CC and MLO views will then be obtained as standard.
- This concludes the woman's participation in the study. The images from the low dose and FFDMs will then be processed to determine whether mammographic density from low dose mammograms correlates with that from FFDM and what modifications are required to automated algorithms for future studies.
Study Type
Enrollment (Actual)
Phase
- Not Applicable
Contacts and Locations
Study Locations
-
-
-
Manchester, United Kingdom, M23 9LT
- The Nightingale Centre
-
-
Participation Criteria
Eligibility Criteria
Ages Eligible for Study
Accepts Healthy Volunteers
Genders Eligible for Study
Description
Inclusion Criteria:
- Females with moderate to high risk of breast cancer, identified through and attending for annual review and mammography at the family history clinic, Nightingale Centre, Manchester
- Current age 30-45 years
- Capable of providing informed consent to a patient information sheet written in English
Exclusion Criteria:
- Prior breast cancer
- Prior breast augmentation or reduction
- Participation in the TARA-Prev study which included an additional mammogram and thus radiation dose
Study Plan
How is the study designed?
Design Details
- Primary Purpose: Screening
- Allocation: N/A
- Interventional Model: Single Group Assignment
- Masking: None (Open Label)
Arms and Interventions
Participant Group / Arm |
Intervention / Treatment |
|---|---|
|
Other: Low dose mammogram
Low dose mammo to compare with standard mammo
|
Low dose mammogram
|
What is the study measuring?
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Correlation between full dose and low dose percent mammographic density estimates
Time Frame: The low dose mammogram will be taken immediately after the full dose exposure, under the same breast compression.
|
Full dose and Low dose mammograms will be analysed using an established AI technique, the predicted Visual Analogue Score (pVAS), to determine the percent mammographic density (%MD).
Correlation between the full and low dose image derived %MD across the whole study population will be assessed using Pearson correlation coefficient.
|
The low dose mammogram will be taken immediately after the full dose exposure, under the same breast compression.
|
Collaborators and Investigators
Study record dates
Study Major Dates
Study Start (Actual)
Primary Completion (Actual)
Study Completion (Actual)
Study Registration Dates
First Submitted
First Submitted That Met QC Criteria
First Posted (Actual)
Study Record Updates
Last Update Posted (Actual)
Last Update Submitted That Met QC Criteria
Last Verified
More Information
Terms related to this study
Keywords
Other Study ID Numbers
- B00394
Plan for Individual participant data (IPD)
Plan to Share Individual Participant Data (IPD)?
IPD Plan Description
Drug and device information, study documents
Studies a U.S. FDA-regulated drug product
Studies a U.S. FDA-regulated device product
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.