Non-invasive Detection of Coronary Artery Stenosis (NIDCAS)

December 24, 2024 updated by: Stephen Greenwald, Nilocas

The aim of this viability study is to investigate using a novel medical device, the potential for detecting partial blockages in the arteries within the wall of the heart (coronary artery stenosis). A prototype device will be used in the proposed study. It is a stick-on patch containing sensors which measure the vibration produced at the chest surface by flowing blood as it accelerates through the narrowed artery in the heart emerging downstream, hitting the inner wall of the artery and causing it to vibrate. Some of the energy in the vibrations reaches the skin of the chest, allowing the blockage to be detected non-invasively. The principle has been proved in the laboratory using a simple chest model but to date no measurements have been made in people. Therefore, the proposed trial will serve as the first in-human measurements and will involve a small number of patients and healthy volunteers, with the objective of establishing if it is possible to determine the presence or absence of the stenosis.

The patch will be tested by recording acoustic signatures in healthy volunteers, then in patients undergoing coronary CT angiography. Results will be validated against the CT angiography gold standard. At this early stage no clinical decisions will be made based on the patch recordings and the data will be used only to assess the feasibility of continuing the development of the device.

Study Overview

Status

Not yet recruiting

Detailed Description

This feasibility study is a non-randomised, proof of concept study being undertaken at a single clinical centre. It involves a novel, non-invasive battery powered medical device that is not CE marked. As a non-CE marked device, review and approval of the proposed clinical investigation a UK Competent Authority, the HRA, is required in addition to ethical review and approval.

The device, a patch containing microphones and accelerometers applied to the chest, is battery powered and will not introduce any energy or substances into the subjects involved in the study. The device will listen for sounds and movement within the body of the individual being examined. The signals detected by the microphones and accelerometers are transmitted wirelessly to a data acquisition unit which is connected to a laptop computer more than 2 metres away from the patient, which displays the signals in real time and stores them for later analysis off-line.

Data (acoustic signatures) will be collected from patients with known or suspected coronary artery disease. Results will be compared with the current gold standard of coronary angiography or coronary CT angiography.

The risk to participants is minimal. No specific risks associated with the use of the device have been identified and no requirement for additional testing has been identified. There are no anticipated interactions with concomitant medical treatments.

The results of this study will not be used to manage the treatment pathway of the patient involved.

If proof of concept is successfully demonstrated with the new device, a larger randomised study will be undertaken to obtain the clinical data necessary to support commercial and regulatory requirements.

Having conducted in-vitro proof of principle experiments (as listed in the references section) and before conducting a formal validation trial, we wish to determine the viability of the new patch device in an initial study on a small number of patients undergoing coronary angiography. We aim firstly, to prove the principle that it is possible to distinguish between those with and those without coronary artery stenosis and, secondly to assess how well the device is tolerated by the patient and how easy it is to deploy the patch and take readings 'in the field'. We believe that this small trial is essential to provide first in human data to support applications for funding a larger validation trial.

Potential study participants will be recruited from patients attending, the Cleveland Clinic who have agreed to elective angiography or CTa investigation. All potential participants will be screened against the inclusion and exclusion criteria detailed elsewhere. This is a non-randomised study. Accordingly, participants will be recruited i on a sequential basis until a total sample size of at least 20 is reached.

Written Informed Consent will be obtained from all participants before recruitment into the study and this will be recorded on the consent form. The nature of the measurements and the nature of the clinical investigation, together with all procedures and risks will be fully explained to each prospective participant by one of the Investigators or members of the research team. If the patient meets all inclusion/exclusion criteria and agrees to participate in the study, he/she will be invited to sign the approved CF.

When the potential subject has agreed to take part, they will undergo the following procedures (approximate times in brackets).

  • Height, weight and blood pressure recording. (15 minutes)
  • CTa or angiography scan as normal. (30 minutes)
  • Patch attached and wireless connection to remote receiver established.
  • The patch will be adjusted to ensure that the patient is comfortable
  • Data recorded. (15 minutes)
  • Patch removed. (5 minutes)
  • Subject asked to suggest any aspects of the procedure that might be improved. All comments noted. (10 minutes)

The patch data will be analysed blind by a member of our team who would not take part in the recording sessions. For the analysis the time series data will be transformed to the frequency domain and the relative acoustic power detected by each sensor over various frequency range will be tabulated. For this small preliminary study, the agreement between the reference and test method to determine the categorical variable namely, the presence or absence of a coronary artery stenosis, will be assessed by Cohen's Kappa test. This measures how much the agreement between the two methods exceeds what would be expected by chance. At this early stage no further statistical tests are planned.

Study Type

Observational

Enrollment (Estimated)

20

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Contact

Study Contact Backup

  • Name: Mark Aichroth
  • Phone Number: ‭+44 7767 640986‬

Study Locations

      • London, United Kingdom, W1B 1LU
        • Cleveland Clinic Portland Place Outpatient Centre 24 Portland Place

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

  • Adult
  • Older Adult

Accepts Healthy Volunteers

Yes

Sampling Method

Non-Probability Sample

Study Population

Approximately equal numbers of symptomatic and asymptomatic subjects undergoing angiography or CT angiography at the Cleveland Clinic, London, UK

Description

Inclusion Criteria:

  1. Patient is able to provide written informed consent
  2. Age ≥18 years;
  3. Investigation for suspected stable coronary artery disease/stenosis isolated from other cardiac comorbidities.
  4. Subjects of either sex. Female subjects of child-bearing potential will be included if deemed suitable for clinically mandated coronary CTa.
  5. No infectious disease or neoplasia.

Exclusion Criteria:

  1. Indications to coronary CTa different from isolated suspected stable coronary artery disease, such as acute coronary syndrome, unstable angina, congenital heart disease, cardiomyopathy, heart valve disease, aortic coarctation, electrophysiological disorder/arrhythmia, screening study, unstable clinical condition, left ventricular ejection fraction <35%, previous coronary revascularization (PCI or CABG surgery), previous heart valve surgery or transcatheter valve replacement, pace maker or implantable defibrillator.
  2. If female and of child-bearing potential: evidence of a positive pregnancy test or a stated intention to become pregnant in the next 6-12 months;
  3. Currently participating in another device or drug study or has done so in previous 3 months.

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

Cohorts and Interventions

Group / Cohort
Coronary Artery Disease (CAD)
Patients in whom occlusive coronary artery disease has been diagnosed by angiography or CT angiography
Control
Healthy volunteers and/or patients in whom no CAD has been identified on angiography or CT angiography

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Accuracy
Time Frame: Immediately following patch recording and imaging

Diagnostic accuracy of the device (stick-on patch) for indicating the presence or absence of one or more coronary arterial stenoses as indicated by the reference method (angiography/CT angiography results). This will be expressed as: Np/(Na+Np) x100.

Where:

Np is the number of positive diagnoses of coronary artery disease derived from the patch measurements Na is the number of positive diagnoses of coronary artery disease (with at least one lesion showing a reduction in maximum stenosis lumen diameter 0.8 of more) derived from the angiography images.

This is a dimensionless quantity

Immediately following patch recording and imaging

Secondary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Stenosis severity
Time Frame: Immediately following the patch recordings and imaging

In subjects in whom one or more stenoses are identified by the reference method:

Comparison of the magnitude of the disturbed flow to the severity of the stenosis as measured from the CTa images.

The comparison will be made by assessing the significance of the linear correlation coefficient between the average power of the patch signals in the frequency domain (in the frequency range 20 - 500 Hz) and the stenosis severity defined as the ratio of the minimum stenosis diameter to the mean diameter of the coronary artery immediately proximal to the stenosis. (Note that this is at best, a semi-quantitative measure and with the small number of subjects proposed for this viability trial the results are for guidance only.)

Immediately following the patch recordings and imaging
Patch wearability
Time Frame: Immediately following the patch recordings and imaging
Assessment of the subject's comfort when wearing the device. This will be quantified using a Visual Analog Score between 1 and 10, 1 being the least comfortable and 10, the most.
Immediately following the patch recordings and imaging
Ease of patch removal
Time Frame: Immediately following the patch recordings and imaging
Ease of removal at the end of a measurements session. This will be quantified quantified using a Visual Analog Score between 1 and 10, 1 being the least comfortable and 10, the most.
Immediately following the patch recordings and imaging

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Sponsor

Investigators

  • Principal Investigator: Francesca Pugliese, MD, Queen Mary University of London

Publications and helpful links

The person responsible for entering information about the study voluntarily provides these publications. These may be about anything related to the study.

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start (Estimated)

May 1, 2025

Primary Completion (Estimated)

October 1, 2025

Study Completion (Estimated)

December 1, 2025

Study Registration Dates

First Submitted

December 17, 2024

First Submitted That Met QC Criteria

December 24, 2024

First Posted (Actual)

March 25, 2025

Study Record Updates

Last Update Posted (Actual)

March 25, 2025

Last Update Submitted That Met QC Criteria

December 24, 2024

Last Verified

December 1, 2024

More Information

Terms related to this study

Plan for Individual participant data (IPD)

Plan to Share Individual Participant Data (IPD)?

NO

IPD Plan Description

The results of this preliminary viability study will be described in an internal report and this will be used to provided evidence when seeking funding for a larger validation trial. They will not therefore be published.

Drug and device information, study documents

Studies a U.S. FDA-regulated drug product

No

Studies a U.S. FDA-regulated device product

No

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.

Clinical Trials on Coronary Arterial Disease (CAD)

Subscribe