Transducer Levelling Errors in Everyday ICU Practice - A Multicentre Blinded Observational Study
Invasive pressure monitoring (arterial blood pressure via arterial catheter and central venous pressure via central venous catheter) is fundamental to clinical decision-making in intensive care. Treatment with vasopressors, fluids, and nursing interventions depends on accurate measurements. Because pressure transducers are hydrostatic, even small levelling errors can produce clinically relevant deviations (approximately 1 mmHg per 1.3 cm of vertical misplacement). Such errors may contribute to inappropriate therapy, for example under-recognition of hypotension, unnecessary vasopressor escalation, or missed venous congestion.
This prospective, multicenter, blinded observational study will quantify transducer levelling deviations during routine ICU care and evaluate whether clinically relevant deviations are associated with treatment decisions. Adult mechanically ventilated ICU patients with an arterial catheter, a central venous catheter, and ongoing vasopressor therapy will be included at hospitals in Västra Götalandsregionen, Sweden. Two blinded reference lines/sensors will be placed at predefined physiological zero levels for MAP and CVP and connected in parallel to the patient monitoring system. Continuous deviation (mmHg) between the clinical transducer position and the blinded reference level will be recorded for 8 hours while clinical staff remain unaware of the reference setup. MAP/CVP-related treatment decisions (e.g., vasopressor adjustments, fluid therapy, and nursing interventions) will be recorded with timestamps.
The study will provide real-world data on the magnitude and frequency of invasive pressure transducer misalignment in daily ICU practice and its potential relationship to patient management.
Study Overview
Status
Conditions
Detailed Description
This is a prospective, multicenter, blinded observational study evaluating the accuracy of invasive pressure transducer levelling for arterial pressure (mean arterial pressure, MAP) and central venous pressure (CVP) during routine intensive care. Adult mechanically ventilated ICU patients with an indwelling arterial catheter and central venous catheter receiving vasopressor therapy are observed without any protocol-mandated changes to standard clinical care.
To quantify levelling deviations without influencing clinical behavior, a parallel reference measurement setup is connected to the existing monitoring system. Two fluid-filled reference lines are connected in parallel to the pressure dome used for invasive monitoring and secured to the chest wall at the fourth intercostal space at half of the anteroposterior thoracic diameter, one on each side of the chest. This configuration allows continuous estimation of vertical deviation (mmHg) between the clinical transducer position and a predefined physiological reference level. During lateral positioning, deviation is calculated using the midpoint between the two reference lines as the reference. Reference lines are covered with dressings to maintain blinding of clinical staff to the reference levels.
Physiological waveforms and calculated deviations are recorded continuously for an 8-hour monitoring period for each participant. In parallel, treatment actions plausibly influenced by interpretation of MAP or CVP (including vasopressor titration, fluid administration or removal, and nursing interventions related to positioning or transducer handling) are documented with timestamps to allow temporal association analyses. Relevant contextual variables are collected to support interpretation of deviations and clinical responses.
Data quality is ensured through predefined measurement procedures, continuous digital waveform acquisition, synchronization of reference and clinical signals, and post hoc review of signal integrity. Periods affected by technical artifacts or signal loss are flagged according to predefined criteria. Data segments that are missing, uninterpretable, or invalid due to technical failure are treated as missing and excluded from analyses, without imputation.
The study is designed to quantify the magnitude and frequency of invasive pressure transducer misalignment during routine ICU care and to explore whether clinically relevant deviations are associated with contemporaneous treatment decisions, with the aim of informing patient safety and quality improvement initiatives.
Study Type
Enrollment (Estimated)
Contacts and Locations
Study Contact
- Name: Carl Sjödin, PhD student
- Phone Number: +46313421096
- Email: carl.sjodin@vgregion.se
Study Contact Backup
- Name: Lotta Johansson, PhD
- Phone Number: 0708199664
- Email: lotta.k.johansson@vgregion.se
Participation Criteria
Eligibility Criteria
Ages Eligible for Study
- Adult
- Older Adult
Accepts Healthy Volunteers
Sampling Method
Study Population
Description
Inclusion Criteria:
- Age ≥18 years
- ICU admission
- Mechanically ventilated
- Arterial catheter for MAP
- Central venous catheter for CVP
- Ongoing vasopressor therapy
Exclusion Criteria:
- Severe hemodynamic or respiratory instability
- Contraindications for lateral positioning
- Expected survival <24 hours
Study Plan
How is the study designed?
Design Details
Cohorts and Interventions
Group / Cohort |
|---|
|
Invasive Pressure Monitoring Cohort
Mechanically ventilated adult ICU patients undergoing routine invasive arterial blood pressure and central venous pressure monitoring.
Pressure transducer positioning and associated pressure deviations are observed using blinded reference sensors without altering clinical care.
|
What is the study measuring?
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Mean deviation in transducer position for MAP and CVP
Time Frame: During the 8-hour continuous monitoring period following initiation of blinded reference monitoring
|
Continuous deviation (mmHg) between the clinical transducer position and the blinded reference zero point.
Continuous (mmHg).
|
During the 8-hour continuous monitoring period following initiation of blinded reference monitoring
|
Secondary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Frequency of clinically relevant deviations
Time Frame: During the same 8-hour continuous monitoring period following initiation of blinded reference monitoring
|
Proportion of monitoring time (%) with deviation >2 mmHg (CVP) or >5 mmHg (MAP).
Percentage of total monitoring time.
|
During the same 8-hour continuous monitoring period following initiation of blinded reference monitoring
|
|
Association between deviations and treatment decisions
Time Frame: During the 8-hour continuous monitoring period, with treatment decisions assessed within 5 minutes following detection of a deviation
|
Relationship between deviation size and treatment decisions (vasopressor adjustments, fluid therapy, nursing interventions).
Binary outcome (decision yes/no), regression analysis.
|
During the 8-hour continuous monitoring period, with treatment decisions assessed within 5 minutes following detection of a deviation
|
|
Differences in deviation magnitude and frequency across hospitals and ICU specialties.
Time Frame: During the 8-hour continuous monitoring period following initiation of blinded reference monitoring
|
Differences in deviation magnitude and frequency across hospitals and ICU specialties.
Continuous (mmHg), categorical (% above threshold).
|
During the 8-hour continuous monitoring period following initiation of blinded reference monitoring
|
|
Hemodynamic context of deviations
Time Frame: At baseline and during the subsequent 8-hour continuous monitoring period
|
Association between deviation magnitude and concurrent patient status (SOFA, MAP, CVP, CO, SVR, temperature).
Continuous and categorical variables.
|
At baseline and during the subsequent 8-hour continuous monitoring period
|
Collaborators and Investigators
Sponsor
Investigators
- Principal Investigator: Lotta Johansson, PhD, Sahlgrenska University Hospital
Publications and helpful links
General Publications
- 1. Cecconi, M., et al., Consensus on circulatory shock and hemodynamic monitoring. Task force of the European Society of Intensive Care Medicine. Intensive Care Med, 2014. 40(12): p. 1795-815. 2. Vincent, J.L., Critical care - where have we been and where are we going? New England Journal of Medicine, 2013. 369(19): p. 1839-1841. 3. Vincent, J.L. and D. De Backer, Circulatory shock. New England Journal of Medicine, 2013. 369(18): p. 1726-1734. 4. Giusti, G.D., Critical care nurses' perspectives on cardiac monitoring: barriers and training needs. BMC Nursing, 2024. 23: p. 742. 5. Alastalo, T.T., Critical care nurses' observational skills in detecting early patient deterioration. Intensive and Critical Care Nursing, 2017. 43: p. 97-103. 6. Magder, S., Right Atrial Pressure in the Critically Ill: How to Measure, What Is the Value, What Are the Limitations? Chest, 2017. 151(4): p. 908-916. 7. Sjödin, C., S. Sondergaard, and L. Johansson, Variability in alignment of central venous pressure transducer to physiologic reference point in the intensive care unit-A descriptive and correlational study. Aust Crit Care, 2019. 32(3): p. 213-217. 8. Figg, K.K. and E.C. Nemergut, Error in central venous pressure measurement. Anesth Analg, 2009. 108(4): p. 1209-11. 9. Oh, C., et al., Errors in pressure measurements due to changes in pressure transducer levels during adult cardiac surgery: a prospective observational study. BMC Anesthesiol, 2023. 23(1): p. 8. 10. Jacobs, K., et al., Accuracy of intra-arterial line transducer levelling practice in a general intensive care unit. Aust Crit Care, 2024. 37(1): p. 51-57.
Study record dates
Study Major Dates
Study Start (Estimated)
Primary Completion (Estimated)
Study Completion (Estimated)
Study Registration Dates
First Submitted
First Submitted That Met QC Criteria
First Posted (Actual)
Study Record Updates
Last Update Posted (Actual)
Last Update Submitted That Met QC Criteria
Last Verified
More Information
Terms related to this study
Keywords
Additional Relevant MeSH Terms
Other Study ID Numbers
- 2024-06578-01
Plan for Individual participant data (IPD)
Plan to Share Individual Participant Data (IPD)?
IPD Plan Description
IPD Sharing Time Frame
IPD Sharing Access Criteria
IPD Sharing Supporting Information Type
- STUDY_PROTOCOL
- SAP
- ICF
Drug and device information, study documents
Studies a U.S. FDA-regulated drug product
Studies a U.S. FDA-regulated device product
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.
Clinical Trials on Critically Ill
-
Third Military Medical UniversityNot yet recruiting
-
Niveus Medical, Inc.Terminated
-
Mahidol UniversityNot yet recruiting
-
China Medical University HospitalRecruitingCritically Ill PatientsTaiwan
-
Hospital Sao DomingosCompletedCritically Ill PatientsBrazil
-
Ain Shams UniversityUnknownCritically-ill PatientsEgypt
-
Hôpital Edouard HerriotCompletedCritically Ill PatientsFrance
-
Johannes Gutenberg University MainzUnknown
-
Kaohsiung Veterans General Hospital.CompletedCritically Ill PatientsTaiwan
-
St. Antonius HospitalCompletedCritically Ill Patients