Continuous glucose monitors (CGMs) are currently used in the inpatient setting at many hospitals for the care of patients with diabetes and for clinical research studies. This includes Boston Children's Hospital, however, the CGM data is not accessible to clinical or research teams in real-time. CGMs measure and display a patient's glucose level every 5 minutes and provide alerts for predicted or actual low or high glucose levels. The CGM data is currently only accessible on bedside handheld receivers or a mobile device app, neither of which sync with the hospital EMR or other hospital databases. This observational study tests a companion mobile Android device app that wirelessly receives real-time outputs from the Dexcom G7 CGM app and transmits the data to a centralized hospital database. That database in turn feeds into an internal BCH web dashboard which allows live data viewing and integration into future clinical decision support tools.
The investigators are testing the hypothesis that the companion mobile app can accurately and efficiently transmit CGM data to a hospital database. Data will be analyzed for accuracy and efficiency by comparing CGM readings captured by the companion app to CGM readings documented in the formal Dexcom database. Accuracy (difference between CGM readings from both sources) and efficiency (number of CGM readings captured) will be assessed. The findings will be generalizable to using the companion app in future research studies.
Přehled studie
Postavení
Detailní popis
Continuous glucose monitors are currently used in the inpatient setting at Boston Children's Hospital (BCH) for the care of patients with diabetes and for research studies. CGMs measure and display patient's glucose level every 5 minutes and provide alerts for predicted or actual low or high glucose levels. BCH policy allows for use of Dexcom G6 and G7 CGM for glucose monitoring and insulin dosing, so long as the CGM accuracy is verified daily.
Currently, CGM data is not accessible to clinical or research teams in real-time. The CGM data is currently only accessible on bedside handheld receivers or a cell phone app, neither of which syncs with the hospital EMR or other hospital databases. The current clinical workflow requires the bedside caregiver to alert the clinical team of any glucose abnormalities. In order to view and document CGM glucose levels in real-time, clinical and research teams must be physically at the bedside viewing the CGM receiver or the patient's personal cell phone app screen. Teams do have access to the Dexcom Clarity database, an online cloud-based data viewing dashboard, but data transmission is delayed 3 hours from the Dexcom cloud database to Dexcom Clarity.
This observational study tests a companion mobile Android application that receives real-time data from the Dexcom G7 CGM app running on a hospital provided mobile device. The companion app and G7 app modifications were developed from a protocol created by patients in the diabetes community who wanted expanded access to their own data, and was adapted by BCH co-investigators to work entirely within the BCH firewall to access and share the data within BCH. Investigators will recruit individuals using their personal Dexcom G7 CGM in the hospital for clinical care, have the participants use the Dexcom G7 app on a hospital provided Android device, and will test the accuracy and efficiency of data capture by the companion app, all via secure BCH wifi or ethernet. On the hospital device, the Dexcom G7 app has been modified to send Android intra-application messages with CGM readings without identifying information to a companion app. The companion app will transmit the de-identified CGM data to a centralized hospital database that feeds into an internal BCH web dashboard for live data viewing, analysis, and integration into future clinical decision support tools. One anticipated future use of the companion app will be in an upcoming interventional clinical trial of ICU patients receiving IV insulin, in which CGM data will be used to guide insulin infusion rates.
Typ studie
Kontakty a umístění
Studijní místa
-
-
Massachusetts
-
Boston, Massachusetts, Spojené státy, 02115
- Boston Children's Hospital
-
-
Kritéria účasti
Kritéria způsobilosti
Věk způsobilý ke studiu
Přijímá zdravé dobrovolníky
Pohlaví způsobilá ke studiu
Metoda odběru vzorků
Studijní populace
Inpatients at Boston Children's Hospital (BCH) with diabetes, admitted for clinical care purposes, using their personal Dexcom G7 CGM during the inpatient admission.
Popis
Inclusion Criteria:
- Inpatients with diabetes using a Dexcom G7 CGM
- Technology/data sharing criteria: use of the Dexcom G7 phone app and data sharing with the BCH Dexcom Clarity Database
- Anticipated discharge date >24 hours after study eligibility
Exclusion Criteria:
- Use of medications that interfere with CGM accuracy
- Use of a Dexcom receiver to view data (rather than the typical mobile phone)
- Enrollment in a competing clinical trial
Studijní plán
Jak je studie koncipována?
Detaily designu
- Observační modely: COHORT
- Časové perspektivy: PROSPECTIVE
Kohorty a intervence
Skupina / kohorta |
|---|
|
Pediatric diabetic Dexcom G7 users
Diabetic pediatric patients using their personal Dexcom G7 admitted for clinical care purposes. |
Co je měření studie?
Primární výstupní opatření
Měření výsledku |
Popis opatření |
Časové okno |
|---|---|---|
|
Accuracy of CGM data capture.
Časové okno: The earlier of a maximum of 20 days or 72 hours following the first sensor exchange.
|
The accuracy of the CGM readings captured will be measured as the mean absolute relative difference (MARD) between the CGM glucose reading captured by the companion app compared to the CGM glucose reading documented in the Dexcom Clarity database within the same 5-minute interval. |
The earlier of a maximum of 20 days or 72 hours following the first sensor exchange.
|
Sekundární výstupní opatření
Měření výsledku |
Popis opatření |
Časové okno |
|---|---|---|
|
Efficiency of CGM data capture.
Časové okno: The earlier of a maximum of 20 days or 72 hours following the first sensor exchange.
|
Efficiency of CGM data capture, measured as the percent of CGM readings documented in Dexcom Clarity that were also captured by the companion app within the same 5-minute interval. |
The earlier of a maximum of 20 days or 72 hours following the first sensor exchange.
|
Spolupracovníci a vyšetřovatelé
Sponzor
Termíny studijních záznamů
Hlavní termíny studia
Začátek studia (Odhadovaný)
Primární dokončení (Odhadovaný)
Dokončení studie (Odhadovaný)
Termíny zápisu do studia
První předloženo
První předloženo, které splnilo kritéria kontroly kvality
První zveřejněno (Aktuální)
Aktualizace studijních záznamů
Poslední zveřejněná aktualizace (Aktuální)
Odeslaná poslední aktualizace, která splnila kritéria kontroly kvality
Naposledy ověřeno
Více informací
Termíny související s touto studií
Další relevantní podmínky MeSH
Další identifikační čísla studie
- IRB-P00049941
Plán pro data jednotlivých účastníků (IPD)
Plánujete sdílet data jednotlivých účastníků (IPD)?
Informace o lécích a zařízeních, studijní dokumenty
Studuje lékový produkt regulovaný americkým FDA
Studuje produkt zařízení regulovaný americkým úřadem FDA
Tyto informace byly beze změn načteny přímo z webu clinicaltrials.gov. Máte-li jakékoli požadavky na změnu, odstranění nebo aktualizaci podrobností studie, kontaktujte prosím register@clinicaltrials.gov. Jakmile bude změna implementována na clinicaltrials.gov, bude automaticky aktualizována i na našem webu .