- ICH GCP
- US-Register für klinische Studien
- Klinische Studie NCT02290600
Feasibility Study Using Run-to-Run Control to Optimize Continuous Glucose Sensor Bias
10. Juli 2018 aktualisiert von: Sansum Diabetes Research Institute
The purpose of this 12 week feasibility study is to see how much more accurate a continuous glucose monitor (CGM) might be in individuals with type 1 diabetes if the investigators personalize sensor calibrations to the user, based on previous wear.
Studienübersicht
Status
Abgeschlossen
Bedingungen
Intervention / Behandlung
Detaillierte Beschreibung
The purpose of this 12 week study is to see how much more accurate a continuous glucose monitor (CGM) might be if we personalize the sensor calibration algorithm for each individual.
The calibration algorithm is a mathematical calculation built into the CGM.
Currently, the CGM asks for a regular blood test (finger stick) measurement twice a day in order to make its calibration algorithm more accurate.
Our goal is to keep refining the calculating ability so that it will provide personalized blood sugar control throughout the day for each individual using the sensor with the CGM.
Studientyp
Beobachtungs
Einschreibung (Tatsächlich)
23
Kontakte und Standorte
Dieser Abschnitt enthält die Kontaktdaten derjenigen, die die Studie durchführen, und Informationen darüber, wo diese Studie durchgeführt wird.
Studienorte
-
-
California
-
Santa Barbara, California, Vereinigte Staaten, 93105
- Sansum Diabetes Research Institute
-
-
Teilnahmekriterien
Forscher suchen nach Personen, die einer bestimmten Beschreibung entsprechen, die als Auswahlkriterien bezeichnet werden. Einige Beispiele für diese Kriterien sind der allgemeine Gesundheitszustand einer Person oder frühere Behandlungen.
Zulassungskriterien
Studienberechtigtes Alter
18 Jahre bis 75 Jahre (Erwachsene, Älterer Erwachsener)
Akzeptiert gesunde Freiwillige
Ja
Studienberechtigte Geschlechter
Alle
Probenahmeverfahren
Nicht-Wahrscheinlichkeitsprobe
Studienpopulation
Type 1 diabetes
Beschreibung
Inclusion Criteria:
- Clinical diagnosis of type 1 diabetes for at least one year.
- The diagnosis of type 1 diabetes is based on the investigator's judgment; C peptide level and antibody determinations are not needed.
- Actively using an insulin pump (CSII therapy) for the last 6 months, and will continue to use an insulin pump throughout the trial.
- Age 18 to 75 years.
- For females, not currently known to be pregnant or nursing.
- HbA1c < 9.5%, as measured with DCA2000 or equivalent device.
- Willing to perform the calibration of the study CGMs using a finger stick only and willing to follow instructions for CGM wear.
- Able to and agrees to avoid the following medication starting 24 hours before sensor wear through completion of the study: acetaminophen, prednisone.
- An understanding of and willingness to follow the protocol and sign the informed consent.
Exclusion Criteria:
- Pregnancy (as determined by a positive blood pregnancy test performed in females of childbearing capacity during screening visit and before phase II) or nursing mother.
- Diabetic ketoacidosis in the past 6 months prior to enrollment requiring emergency room visit or hospitalization.
- Severe hypoglycemia resulting in seizure or loss of consciousness in the 6 months prior to enrollment.
- Current treatment for a seizure disorder: Subjects with a history of seizures may be included in the study if they receive written clearance from their neurologist.
- A known medical condition that in the judgment of the investigator might interfere with the completion of the protocol such as cognitive deficit.
- Mental incapacity, unwillingness or language barriers precluding adequate understanding or co-operation, including subjects not able to read or write.
- Active coronary artery disease or heart failure
- Active gastroparesis
- If on antihypertensive, thyroid, anti-depressant or lipid lowering medication, lack of stability on the medication for the past 2 months prior to enrollment in the study.
- Uncontrolled thyroid disease: Adequately treated thyroid disease and celiac disease do not exclude subjects from enrollment.
- Abuse of alcohol.
- A recent injury to body or limb, muscular disorder, use of any medication, any carcinogenic disease, or other significant medical disorder if that injury, medication or disease in the judgment of the investigator will affect the completion of the protocol.
- Current use of a beta blocker medication.
- Laboratory results:
- Hematocrit < 30% or >55%.
- A1C ≥ 9.5%
- Abnormal liver or renal function (Transaminase >2 times the upper limit of normal, Creatinine> 1.5 mg/dL)
- Labs drawn at screening visit or within two months prior to screening (for other purposes) will suffice for enrollment purposes.
- Subject has skin conditions that, in the determination of the investigator, would preclude wearing the study devices (CGM sensor), in the abdomen or upper buttocks. Examples include but are not limited to: psoriasis, burns, scaring, eczema, tattoos, and significant hypertrophy at sites of device wear; any known allergy to medical adhesives.
- Current participation in another investigational trial or has previously participated in this study. However, for this study, if the augmented calibration algorithm is modified or more data is needed to update the calibration algorithm, subjects will be allowed to re-consent for additional studies at the judgment of the investigator. Similarly, if there is missing data from the data collection due to device failure, transmission errors or other problems, the subject will be allowed to extend the time required to continue each phase as per the judgment of the investigator and will not be excluded from continuing the study.
Studienplan
Dieser Abschnitt enthält Einzelheiten zum Studienplan, einschließlich des Studiendesigns und der Messung der Studieninhalte.
Wie ist die Studie aufgebaut?
Designdetails
Kohorten und Interventionen
Gruppe / Kohorte |
Intervention / Behandlung |
|---|---|
|
type 1 diabetes
type 1 diabetes with continuous glucose monitor (CGM)
|
CGM sensor calibration individualization
Andere Namen:
|
Was misst die Studie?
Primäre Ergebnismessungen
Ergebnis Maßnahme |
Zeitfenster |
|---|---|
|
Accuracy of continuous glucose sensor output by individualizing sensor bias in subjects with type 1 diabetes, assessed by quantifying the improvement in CGM accuracy as measured by MARD (mean absolute relative difference).
Zeitfenster: 3 months
|
3 months
|
Sekundäre Ergebnismessungen
Ergebnis Maßnahme |
Maßnahmenbeschreibung |
Zeitfenster |
|---|---|---|
|
Continuous Glucose Monitor sensor bias
Zeitfenster: 3 months
|
continuous glucose monitor accuracy when compared to self monitoring of blood glucose (SMBG) using a standardized meter
|
3 months
|
|
Continuous Glucose Monitor sensor failure
Zeitfenster: 3 months
|
3 months
|
Mitarbeiter und Ermittler
Hier finden Sie Personen und Organisationen, die an dieser Studie beteiligt sind.
Ermittler
- Hauptermittler: Jordan Pinsker, M.D., Sansum Diabetes Research Institute
Publikationen und hilfreiche Links
Die Bereitstellung dieser Publikationen erfolgt freiwillig durch die für die Eingabe von Informationen über die Studie verantwortliche Person. Diese können sich auf alles beziehen, was mit dem Studium zu tun hat.
Studienaufzeichnungsdaten
Diese Daten verfolgen den Fortschritt der Übermittlung von Studienaufzeichnungen und zusammenfassenden Ergebnissen an ClinicalTrials.gov. Studienaufzeichnungen und gemeldete Ergebnisse werden von der National Library of Medicine (NLM) überprüft, um sicherzustellen, dass sie bestimmten Qualitätskontrollstandards entsprechen, bevor sie auf der öffentlichen Website veröffentlicht werden.
Haupttermine studieren
Studienbeginn
1. Januar 2015
Primärer Abschluss (Tatsächlich)
1. Oktober 2016
Studienabschluss (Tatsächlich)
1. Dezember 2016
Studienanmeldedaten
Zuerst eingereicht
25. August 2014
Zuerst eingereicht, das die QC-Kriterien erfüllt hat
10. November 2014
Zuerst gepostet (Schätzen)
14. November 2014
Studienaufzeichnungsaktualisierungen
Letztes Update gepostet (Tatsächlich)
12. Juli 2018
Letztes eingereichtes Update, das die QC-Kriterien erfüllt
10. Juli 2018
Zuletzt verifiziert
1. Juli 2018
Mehr Informationen
Begriffe im Zusammenhang mit dieser Studie
Andere Studien-ID-Nummern
- JDRF KK1413
Arzneimittel- und Geräteinformationen, Studienunterlagen
Studiert ein von der US-amerikanischen FDA reguliertes Arzneimittelprodukt
Nein
Studiert ein von der US-amerikanischen FDA reguliertes Geräteprodukt
Ja
Diese Informationen wurden ohne Änderungen direkt von der Website clinicaltrials.gov abgerufen. Wenn Sie Ihre Studiendaten ändern, entfernen oder aktualisieren möchten, wenden Sie sich bitte an register@clinicaltrials.gov. Sobald eine Änderung auf clinicaltrials.gov implementiert wird, wird diese automatisch auch auf unserer Website aktualisiert .
Klinische Studien zur Continuous Glucose Monitor
-
Wake Forest University Health SciencesNoch keine RekrutierungStörungen des Glukosestoffwechsels (einschließlich Diabetes mellitus)Vereinigte Staaten
-
Vanderbilt University Medical CenterRekrutierungHyperglykämie | Typ-2-Diabetes mellitus (T2DM) | Typ-1-Diabetes mellitus (T1DM)Vereinigte Staaten
-
Hamilton Health Sciences CorporationMcMaster University; Population Health Research Institute; Cloud DX Inc.Noch keine Rekrutierung
-
KOS Inc.Noch keine RekrutierungDiabetes mellitus, Typ 2 | Diabetes | Prädiabetes | Diabetes mellitus, Typ-1-DiabetesVereinigte Staaten
-
Hospital Universitario San IgnacioPontificia Universidad JaverianaNoch keine Rekrutierung
-
TC Erciyes UniversityNoch keine RekrutierungEnterale Ernährung | Magen-Darm-System - Anomalien
-
Pleural Dynamics, Inc.Medical Metrics Diagnostics, IncAktiv, nicht rekrutierend
-
Glenn GreenUniversity of Michigan; American Academy of Otolaryngology-Head and Neck Surgery...AbgeschlossenSchlafapnoe, obstruktiv | Kraniofaziale Anomalien | Pädiatrische StörungVereinigte Staaten
-
Tokat Gaziosmanpasa UniversityAbgeschlossenKaiserschnitt-Komplikationen | IsthmozeleTruthahn
-
Mayo ClinicSnap40 Ltd.Abgeschlossen