Evaluación del dispositivo EpiZact

10 de noviembre de 2026 actualizado por: Brandon M Togioka, Oregon Health and Science University

Un único centro académico, cohorte prospectiva con comparador retrospectivo de pérdida de resistencia, estudio de prueba de concepto para evaluar la seguridad, eficacia y usabilidad del dispositivo EpiZact para parturientas sometidas a analgesia epidural del parto

The purpose of the study is to learn more about a new device, the EpiZact Epidural Device, that may be helpful for reducing the likelihood of wet tap when a labor epidural is placed. When epidurals are placed correctly the needle stops just in front of the space in the body that contains cerebrospinal fluid. When the epidural is advanced a short distance further, leakage of cerebrospinal fluid can occur. This is known as a wet tap. A wet tap can result in a severe headache and at times other complications. The goal of this clinical trial is to learn if the use of the EpiZact device prevents wet taps in pregnant women receiving epidurals for relief of labor pain. The main questions this study aims to answer are:

  1. What the rate of wet tap with the EpiZact device?
  2. What is the rate of failed epidural placement with the EpiZact device?

All patients in the study will receive an epidural with the EpiZact device. The investigators will compare the results with patients not in the study that receive an epidural without the EpiZact device.

Descripción general del estudio

Descripción detallada

The purpose of the study is to learn if use of the EpiZact device decreases the number of wet taps that occur when epidurals are placed. The EpiZact device has a technology that automatically stops the epidural needle when it enters the epidural space. In theory, this should decrease the likelihood of a wet tap. The standard epidural to treat labor pain does not have this capability. The study device may cause fewer wet taps than the standard epidural. All study participants will receive an epidural to treat labor pain with the EpiZact device. Results will be compared with patients receiving epidurals around the same time without the EpiZact device.

If a participant's screening questions show that they can take part in the study, and the participant chooses to be a part of the study, medical records from OHSU or other facilities may be reviewed. Information reviewed may include treatment history, medications, surgeries, allergies, scans, and lab results. The participant will then receive an epidural with the EpiZact device. After the epidural is placed, the participant will be asked a few questions to determine how well the epidural is working and if a complication occurred. The questions are expected to take less than 5 minutes to answer. Participants will also be called 1 week after receiving an epidural to determine if a complication occurred. This call is expected to last less than 5 minutes.

Tipo de estudio

Intervencionista

Inscripción (Actual)

1

Fase

  • Fase 4

Contactos y Ubicaciones

Esta sección proporciona los datos de contacto de quienes realizan el estudio e información sobre dónde se lleva a cabo este estudio.

Ubicaciones de estudio

    • Oregon
      • Portland, Oregon, Estados Unidos, 97239
        • Oregon Health and Science University

Criterios de participación

Los investigadores buscan personas que se ajusten a una determinada descripción, denominada criterio de elegibilidad. Algunos ejemplos de estos criterios son el estado de salud general de una persona o tratamientos previos.

Criterio de elegibilidad

Edades elegibles para estudiar

15 años a 50 años (Niño, Adulto)

Acepta Voluntarios Saludables

No

Géneros elegibles para el estudio

Femenino

Descripción

Inclusion Criteria:

  • Pregnant women requesting labor epidural analgesia
  • Women pregnant with a single baby (not twins or triplets)

Exclusion Criteria:

  • Women requesting not to have a labor epidural
  • Height less than 150 cm
  • A bleeding disorder, nerve problem, or skin infection that contraindicates receiving an epidural
  • A women receiving a combined spinal epidural (CSE) or a dural puncture epidural (DPE)
  • A history of spine surgery
  • Allergy to local anesthetics (numbing medicine) or skin cleaning solutions
  • Decisionally impaired individuals

Plan de estudios

Esta sección proporciona detalles del plan de estudio, incluido cómo está diseñado el estudio y qué mide el estudio.

¿Cómo está diseñado el estudio?

Detalles de diseño

  • Propósito principal: Prevención
  • Asignación: N / A
  • Modelo Intervencionista: Asignación de un solo grupo
  • Enmascaramiento: Ninguno (etiqueta abierta)

Armas e Intervenciones

Grupo de participantes/brazo
Intervención / Tratamiento
Experimental: EpiZact assisted epidural placement

All patients enrolled in the study will receive an epidural to treat labor pain with the assistance of the EpiZact device.

The standard epidural kit used with the EpiZact device to place a midline epidural using the landmark technique between the L2 and L5 vertebral bodies. The patient's back will be sterilized with a 2% chlorhexidine/70% isopropyl alcohol applicator. The epidural will be placed with a 17-gauge 3.5-inch Tuohy needle. Once the Tuohy needle is appreciated to be engaged in ligament the EpiZact device will be filled with sterile and attached to the Tuohy via a Luer connector. The green button on the EpiZact device will then be pressed to prime the device. The EpiZact device and Tuohy needle will be advanced as a unit in continuous fashion or with intermittent advancement. Upon sensing a loss of resistance, the EpiZact device will provide a visual and tactile signal, halting needle advancement and signaling that the epidural catheter should be threaded through the Tuohy needle.

¿Qué mide el estudio?

Medidas de resultado primarias

Medida de resultado
Medida Descripción
Periodo de tiempo
Proportion of Participants with an Unintentional Dural Puncture
Periodo de tiempo: 1 week

Unintentional dural puncture will be a composite outcome diagnosed by any of the following: (1) free flow of cerebrospinal fluid from the Tuohy needle at any time during the epidural technique, (2) requirement of blood patch within one week of neuraxial procedure, or (3) postural headache that worsens with upright posture and improves with supine position diagnosed within one week of neuraxial placement.

1 week

Medidas de resultado secundarias

Medida de resultado
Medida Descripción
Periodo de tiempo
Proportion of Participants with Failed Labor Epidural Placement
Periodo de tiempo: 2 hours

Failed epidural placement will be defined as the inability to place a catheter in the epidural place or the need to replace the epidural within 2 hours of insertion. Decision to replace the epidural will be according to standard practice on labor and delivery at OHSU and will typically involve assessment of loss of cold to ice sensation in the lower thoracic and lumbar dermatomes, 50% decrease in labor contraction pain, and new lower extremity paresthesia.

2 hours

Colaboradores e Investigadores

Aquí es donde encontrará personas y organizaciones involucradas en este estudio.

Investigadores

  • Investigador principal: Brandon M Togioka, MD, Oregon Health and Science University

Fechas de registro del estudio

Estas fechas rastrean el progreso del registro del estudio y los envíos de resultados resumidos a ClinicalTrials.gov. Los registros del estudio y los resultados informados son revisados ​​por la Biblioteca Nacional de Medicina (NLM) para asegurarse de que cumplan con los estándares de control de calidad específicos antes de publicarlos en el sitio web público.

Fechas importantes del estudio

Inicio del estudio (Actual)

27 de febrero de 2025

Finalización primaria (Actual)

6 de marzo de 2025

Finalización del estudio (Actual)

13 de marzo de 2025

Fechas de registro del estudio

Enviado por primera vez

7 de diciembre de 2024

Primero enviado que cumplió con los criterios de control de calidad

7 de diciembre de 2024

Publicado por primera vez (Actual)

11 de diciembre de 2024

Actualizaciones de registros de estudio

Última actualización publicada (Actual)

16 de noviembre de 2026

Última actualización enviada que cumplió con los criterios de control de calidad

10 de noviembre de 2026

Última verificación

1 de noviembre de 2026

Más información

Términos relacionados con este estudio

Plan de datos de participantes individuales (IPD)

¿Planea compartir datos de participantes individuales (IPD)?

NO

Descripción del plan IPD

There is currently no plan to make IPD available to other researchers.

Información sobre medicamentos y dispositivos, documentos del estudio

Estudia un producto farmacéutico regulado por la FDA de EE. UU.

No

Estudia un producto de dispositivo regulado por la FDA de EE. UU.

Sí

producto fabricado y exportado desde los EE. UU.

No

Esta información se obtuvo directamente del sitio web clinicaltrials.gov sin cambios. Si tiene alguna solicitud para cambiar, eliminar o actualizar los detalles de su estudio, comuníquese con register@clinicaltrials.gov. Tan pronto como se implemente un cambio en clinicaltrials.gov, también se actualizará automáticamente en nuestro sitio web. .

Suscribir