Denne side blev automatisk oversat, og nøjagtigheden af ​​oversættelsen er ikke garanteret. Der henvises til engelsk version for en kildetekst.

Evaluering af EpiZact-enheden

10. november 2026 opdateret af: Brandon M Togioka, Oregon Health and Science University

Et enkelt akademisk center, prospektiv kohorte med retrospektiv resistenstab-komparator, bevis på koncept-undersøgelse for at evaluere sikkerheden, effektiviteten og anvendeligheden af ​​EpiZact-anordningen til fødende, der gennemgår epidural arbejdsanalgesi

The purpose of the study is to learn more about a new device, the EpiZact Epidural Device, that may be helpful for reducing the likelihood of wet tap when a labor epidural is placed. When epidurals are placed correctly the needle stops just in front of the space in the body that contains cerebrospinal fluid. When the epidural is advanced a short distance further, leakage of cerebrospinal fluid can occur. This is known as a wet tap. A wet tap can result in a severe headache and at times other complications. The goal of this clinical trial is to learn if the use of the EpiZact device prevents wet taps in pregnant women receiving epidurals for relief of labor pain. The main questions this study aims to answer are:

  1. What the rate of wet tap with the EpiZact device?
  2. What is the rate of failed epidural placement with the EpiZact device?

All patients in the study will receive an epidural with the EpiZact device. The investigators will compare the results with patients not in the study that receive an epidural without the EpiZact device.

Studieoversigt

Detaljeret beskrivelse

The purpose of the study is to learn if use of the EpiZact device decreases the number of wet taps that occur when epidurals are placed. The EpiZact device has a technology that automatically stops the epidural needle when it enters the epidural space. In theory, this should decrease the likelihood of a wet tap. The standard epidural to treat labor pain does not have this capability. The study device may cause fewer wet taps than the standard epidural. All study participants will receive an epidural to treat labor pain with the EpiZact device. Results will be compared with patients receiving epidurals around the same time without the EpiZact device.

If a participant's screening questions show that they can take part in the study, and the participant chooses to be a part of the study, medical records from OHSU or other facilities may be reviewed. Information reviewed may include treatment history, medications, surgeries, allergies, scans, and lab results. The participant will then receive an epidural with the EpiZact device. After the epidural is placed, the participant will be asked a few questions to determine how well the epidural is working and if a complication occurred. The questions are expected to take less than 5 minutes to answer. Participants will also be called 1 week after receiving an epidural to determine if a complication occurred. This call is expected to last less than 5 minutes.

Undersøgelsestype

Interventionel

Tilmelding (Faktiske)

1

Fase

  • Fase 4

Kontakter og lokationer

Dette afsnit indeholder kontaktoplysninger for dem, der udfører undersøgelsen, og oplysninger om, hvor denne undersøgelse udføres.

Studiesteder

    • Oregon
      • Portland, Oregon, Forenede Stater, 97239
        • Oregon Health and Science University

Deltagelseskriterier

Forskere leder efter personer, der passer til en bestemt beskrivelse, kaldet berettigelseskriterier. Nogle eksempler på disse kriterier er en persons generelle helbredstilstand eller tidligere behandlinger.

Berettigelseskriterier

Aldre berettiget til at studere

15 år til 50 år (Barn, Voksen)

Tager imod sunde frivillige

Ingen

Køn, der er berettiget til at studere

Kvinde

Beskrivelse

Inclusion Criteria:

  • Pregnant women requesting labor epidural analgesia
  • Women pregnant with a single baby (not twins or triplets)

Exclusion Criteria:

  • Women requesting not to have a labor epidural
  • Height less than 150 cm
  • A bleeding disorder, nerve problem, or skin infection that contraindicates receiving an epidural
  • A women receiving a combined spinal epidural (CSE) or a dural puncture epidural (DPE)
  • A history of spine surgery
  • Allergy to local anesthetics (numbing medicine) or skin cleaning solutions
  • Decisionally impaired individuals

Studieplan

Dette afsnit indeholder detaljer om studieplanen, herunder hvordan undersøgelsen er designet, og hvad undersøgelsen måler.

Hvordan er undersøgelsen tilrettelagt?

Design detaljer

  • Primært formål: Forebyggelse
  • Tildeling: N/A
  • Interventionel model: Enkelt gruppeopgave
  • Maskning: Ingen (Åben etiket)

Våben og indgreb

Deltagergruppe / Arm
Intervention / Behandling
Eksperimentel: EpiZact assisted epidural placement

All patients enrolled in the study will receive an epidural to treat labor pain with the assistance of the EpiZact device.

The standard epidural kit used with the EpiZact device to place a midline epidural using the landmark technique between the L2 and L5 vertebral bodies. The patient's back will be sterilized with a 2% chlorhexidine/70% isopropyl alcohol applicator. The epidural will be placed with a 17-gauge 3.5-inch Tuohy needle. Once the Tuohy needle is appreciated to be engaged in ligament the EpiZact device will be filled with sterile and attached to the Tuohy via a Luer connector. The green button on the EpiZact device will then be pressed to prime the device. The EpiZact device and Tuohy needle will be advanced as a unit in continuous fashion or with intermittent advancement. Upon sensing a loss of resistance, the EpiZact device will provide a visual and tactile signal, halting needle advancement and signaling that the epidural catheter should be threaded through the Tuohy needle.

Hvad måler undersøgelsen?

Primære resultatmål

Resultatmål
Foranstaltningsbeskrivelse
Tidsramme
Proportion of Participants with an Unintentional Dural Puncture
Tidsramme: 1 week

Unintentional dural puncture will be a composite outcome diagnosed by any of the following: (1) free flow of cerebrospinal fluid from the Tuohy needle at any time during the epidural technique, (2) requirement of blood patch within one week of neuraxial procedure, or (3) postural headache that worsens with upright posture and improves with supine position diagnosed within one week of neuraxial placement.

1 week

Sekundære resultatmål

Resultatmål
Foranstaltningsbeskrivelse
Tidsramme
Proportion of Participants with Failed Labor Epidural Placement
Tidsramme: 2 hours

Failed epidural placement will be defined as the inability to place a catheter in the epidural place or the need to replace the epidural within 2 hours of insertion. Decision to replace the epidural will be according to standard practice on labor and delivery at OHSU and will typically involve assessment of loss of cold to ice sensation in the lower thoracic and lumbar dermatomes, 50% decrease in labor contraction pain, and new lower extremity paresthesia.

2 hours

Samarbejdspartnere og efterforskere

Det er her, du vil finde personer og organisationer, der er involveret i denne undersøgelse.

Samarbejdspartnere

Efterforskere

  • Ledende efterforsker: Brandon M Togioka, MD, Oregon Health and Science University

Datoer for undersøgelser

Disse datoer sporer fremskridtene for indsendelser af undersøgelsesrekord og resumeresultater til ClinicalTrials.gov. Studieregistreringer og rapporterede resultater gennemgås af National Library of Medicine (NLM) for at sikre, at de opfylder specifikke kvalitetskontrolstandarder, før de offentliggøres på den offentlige hjemmeside.

Studer store datoer

Studiestart (Faktiske)

27. februar 2025

Primær færdiggørelse (Faktiske)

6. marts 2025

Studieafslutning (Faktiske)

13. marts 2025

Datoer for studieregistrering

Først indsendt

7. december 2024

Først indsendt, der opfyldte QC-kriterier

7. december 2024

Først opslået (Faktiske)

11. december 2024

Opdateringer af undersøgelsesjournaler

Sidste opdatering sendt (Faktiske)

16. november 2026

Sidste opdatering indsendt, der opfyldte kvalitetskontrolkriterier

10. november 2026

Sidst verificeret

1. november 2026

Mere information

Begreber relateret til denne undersøgelse

Plan for individuelle deltagerdata (IPD)

Planlægger du at dele individuelle deltagerdata (IPD)?

INGEN

IPD-planbeskrivelse

There is currently no plan to make IPD available to other researchers.

Lægemiddel- og udstyrsoplysninger, undersøgelsesdokumenter

Studerer et amerikansk FDA-reguleret lægemiddelprodukt

Ingen

Studerer et amerikansk FDA-reguleret enhedsprodukt

Ja

produkt fremstillet i og eksporteret fra U.S.A.

Ingen

Disse oplysninger blev hentet direkte fra webstedet clinicaltrials.gov uden ændringer. Hvis du har nogen anmodninger om at ændre, fjerne eller opdatere dine undersøgelsesoplysninger, bedes du kontakte register@clinicaltrials.gov. Så snart en ændring er implementeret på clinicaltrials.gov, vil denne også blive opdateret automatisk på vores hjemmeside .