Avaliação do dispositivo EpiZact
Um único centro acadêmico, coorte prospectiva com comparador retrospectivo de perda de resistência, estudo de prova de conceito para avaliar a segurança, eficácia e usabilidade do dispositivo EpiZact para parturientes submetidas à analgesia peridural do parto
The purpose of the study is to learn more about a new device, the EpiZact Epidural Device, that may be helpful for reducing the likelihood of wet tap when a labor epidural is placed. When epidurals are placed correctly the needle stops just in front of the space in the body that contains cerebrospinal fluid. When the epidural is advanced a short distance further, leakage of cerebrospinal fluid can occur. This is known as a wet tap. A wet tap can result in a severe headache and at times other complications. The goal of this clinical trial is to learn if the use of the EpiZact device prevents wet taps in pregnant women receiving epidurals for relief of labor pain. The main questions this study aims to answer are:
- What the rate of wet tap with the EpiZact device?
- What is the rate of failed epidural placement with the EpiZact device?
All patients in the study will receive an epidural with the EpiZact device. The investigators will compare the results with patients not in the study that receive an epidural without the EpiZact device.
Visão geral do estudo
Status
Intervenção / Tratamento
Descrição detalhada
The purpose of the study is to learn if use of the EpiZact device decreases the number of wet taps that occur when epidurals are placed. The EpiZact device has a technology that automatically stops the epidural needle when it enters the epidural space. In theory, this should decrease the likelihood of a wet tap. The standard epidural to treat labor pain does not have this capability. The study device may cause fewer wet taps than the standard epidural. All study participants will receive an epidural to treat labor pain with the EpiZact device. Results will be compared with patients receiving epidurals around the same time without the EpiZact device.
If a participant's screening questions show that they can take part in the study, and the participant chooses to be a part of the study, medical records from OHSU or other facilities may be reviewed. Information reviewed may include treatment history, medications, surgeries, allergies, scans, and lab results. The participant will then receive an epidural with the EpiZact device. After the epidural is placed, the participant will be asked a few questions to determine how well the epidural is working and if a complication occurred. The questions are expected to take less than 5 minutes to answer. Participants will also be called 1 week after receiving an epidural to determine if a complication occurred. This call is expected to last less than 5 minutes.
Tipo de estudo
Inscrição (Real)
Estágio
- Fase 4
Contactos e Locais
Locais de estudo
-
-
Oregon
-
Portland, Oregon, Estados Unidos, 97239
- Oregon Health and Science University
-
-
Critérios de participação
Critérios de elegibilidade
Idades elegíveis para estudo
Aceita Voluntários Saudáveis
Gêneros Elegíveis para o Estudo
Descrição
Inclusion Criteria:
- Pregnant women requesting labor epidural analgesia
- Women pregnant with a single baby (not twins or triplets)
Exclusion Criteria:
- Women requesting not to have a labor epidural
- Height less than 150 cm
- A bleeding disorder, nerve problem, or skin infection that contraindicates receiving an epidural
- A women receiving a combined spinal epidural (CSE) or a dural puncture epidural (DPE)
- A history of spine surgery
- Allergy to local anesthetics (numbing medicine) or skin cleaning solutions
- Decisionally impaired individuals
Plano de estudo
Como o estudo é projetado?
Detalhes do projeto
- Finalidade Principal: Prevenção
- Alocação: N / D
- Modelo Intervencional: Atribuição de grupo único
- Mascaramento: Nenhum (rótulo aberto)
Armas e Intervenções
Grupo de Participantes / Braço |
Intervenção / Tratamento |
|---|---|
|
Experimental: EpiZact assisted epidural placement
All patients enrolled in the study will receive an epidural to treat labor pain with the assistance of the EpiZact device. |
The standard epidural kit used with the EpiZact device to place a midline epidural using the landmark technique between the L2 and L5 vertebral bodies. The patient's back will be sterilized with a 2% chlorhexidine/70% isopropyl alcohol applicator. The epidural will be placed with a 17-gauge 3.5-inch Tuohy needle. Once the Tuohy needle is appreciated to be engaged in ligament the EpiZact device will be filled with sterile and attached to the Tuohy via a Luer connector. The green button on the EpiZact device will then be pressed to prime the device. The EpiZact device and Tuohy needle will be advanced as a unit in continuous fashion or with intermittent advancement. Upon sensing a loss of resistance, the EpiZact device will provide a visual and tactile signal, halting needle advancement and signaling that the epidural catheter should be threaded through the Tuohy needle. |
O que o estudo está medindo?
Medidas de resultados primários
Medida de resultado |
Descrição da medida |
Prazo |
|---|---|---|
|
Proportion of Participants with an Unintentional Dural Puncture
Prazo: 1 week
|
Unintentional dural puncture will be a composite outcome diagnosed by any of the following: (1) free flow of cerebrospinal fluid from the Tuohy needle at any time during the epidural technique, (2) requirement of blood patch within one week of neuraxial procedure, or (3) postural headache that worsens with upright posture and improves with supine position diagnosed within one week of neuraxial placement. |
1 week
|
Medidas de resultados secundários
Medida de resultado |
Descrição da medida |
Prazo |
|---|---|---|
|
Proportion of Participants with Failed Labor Epidural Placement
Prazo: 2 hours
|
Failed epidural placement will be defined as the inability to place a catheter in the epidural place or the need to replace the epidural within 2 hours of insertion. Decision to replace the epidural will be according to standard practice on labor and delivery at OHSU and will typically involve assessment of loss of cold to ice sensation in the lower thoracic and lumbar dermatomes, 50% decrease in labor contraction pain, and new lower extremity paresthesia. |
2 hours
|
Colaboradores e Investigadores
Patrocinador
Colaboradores
Investigadores
- Investigador principal: Brandon M Togioka, MD, Oregon Health and Science University
Datas de registro do estudo
Datas Principais do Estudo
Início do estudo (Real)
Conclusão Primária (Real)
Conclusão do estudo (Real)
Datas de inscrição no estudo
Enviado pela primeira vez
Enviado pela primeira vez que atendeu aos critérios de CQ
Primeira postagem (Real)
Atualizações de registro de estudo
Última Atualização Postada (Real)
Última atualização enviada que atendeu aos critérios de controle de qualidade
Última verificação
Mais Informações
Termos relacionados a este estudo
Palavras-chave
Termos MeSH relevantes adicionais
Outros números de identificação do estudo
- STUDY00027678
Plano para dados de participantes individuais (IPD)
Planeja compartilhar dados de participantes individuais (IPD)?
Descrição do plano IPD
There is currently no plan to make IPD available to other researchers.
Informações sobre medicamentos e dispositivos, documentos de estudo
Estuda um medicamento regulamentado pela FDA dos EUA
Estuda um produto de dispositivo regulamentado pela FDA dos EUA
produto fabricado e exportado dos EUA
Essas informações foram obtidas diretamente do site clinicaltrials.gov sem nenhuma alteração. Se você tiver alguma solicitação para alterar, remover ou atualizar os detalhes do seu estudo, entre em contato com register@clinicaltrials.gov. Assim que uma alteração for implementada em clinicaltrials.gov, ela também será atualizada automaticamente em nosso site .