Positive and Expiratory Pressure and Hemorrhagic Stroke
Effect of Positive End Expiratory Pressure on Cerebral Perfusion Pressure in Adult Patients With Hemorrhagic Stroke
Panoramica dello studio
Stato
Stato
Condizioni
Condizioni
Intervento / Trattamento
Intervento / Trattamento
Descrizione dettagliata
All patients were from the surgical implantation of the ventricular catheter, arriving to the ICU intubated orally and manually ventilated with an Ambu bag. Were subjected to routine procedures: adjusting the mechanical ventilator (Inter5, Intermed, BR) during assisted controlled cycled pressure and facility to monitor vital signs. After thirty minutes of stable ICU patient in a supine position with head elevated 30 °, the protocol was initiated to assess the impact of PEEP on ICP. To perform the evaluation of lung mechanics ventilatory mode was changed to control volume with the following parameters: tidal volume (Vt) = 8ml/kg weight, peak flow (PF) = 6 x minute volume, fraction of inspired O2 (FiO2) = 40%, respiratory rate (RR) = 16 bpm, sensitivity = 1 cmH2O. The following variables were monitored: PIC, Blood Pressure (BP), heart rate (HR), peak pressure in the airways (pp.) and plateau pressure of the respiratory system (Ppl.), these values were monitored with PEEP = 5 cmH2O.
During the assessment protocol was changed to pressure control ventilation mode with the following values of ventilatory parameters: Pp = 30 cm H2O, inspiratory time = 1s; FiO2 = 40%, RR = 16 bpm; Sensitivity = 1 cmH2O. PEEP employed ranged from 0 to 14 cmH2O. To eliminate a possible physiological accommodation by the progressive increase of PEEP, the range of values was determined by drawing a sealed envelope for each patient, ranging from 2 to 2 cmH2O. At each value of PEEP the patient was ventilated for a period of five minutes to carry out monitoring of ICP, BP, HR, PPC and peripheral oxygen saturation (SpO2). The ICP monitoring catheter was kept closed for drainage and open for monitoring, since the arrival of the surgical block, was only open for drainage if there was an increase in ICP above 20 mmHg.
The monitoring was carried out using the multiparameter monitor (Siemens 7000). For ICP monitoring the ventricular catheter was connected to a pressure transducer and this monitor.
After the parameters evaluated with seven different PEEP values, the ventilatory mode was changed again to control volume again to evaluate pulmonary mechanics with the same initial parameters.
Tipo di studio
Tipo di studio
Iscrizione (Effettivo)
Iscrizione
Fase
Fase
- Non applicabile
Contatti e Sedi
Luoghi di studio
-
-
Pernambuco
-
Recife, Pernambuco, Brasile
- Hospital Português
-
-
Criteri di partecipazione
Criteri di ammissibilità
Criteri di ammissibilità
Età idonea allo studio
Accetta volontari sani
Sessi ammissibili allo studio
Descrizione
Inclusion Criteria:
- adult patients with acute CVA
- presence of ventricular drainage catheter for invasive monitoring of ICP
- without intracranial hypertension.
Exclusion Criteria:
- intracranial hypertension
- hemodynamic instability.
Piano di studio
Come è strutturato lo studio?
Dettagli di progettazione
- Scopo principale: Trattamento
- Assegnazione: Randomizzato
- Modello interventistico: Assegnazione incrociata
- Mascheramento: Doppio
Cosa sta misurando lo studio?
Misure di risultato primarie
Misure di risultato primarie
Misura del risultato |
Misura Descrizione |
|---|---|
|
Measurement of intracranial pressure
|
The ICP monitoring catheter was kept closed for drainage and open for monitoring, since the arrival of the surgical block, was only open for drainage if there was an increase in ICP above 20 mmHg.
|
Misure di risultato secondarie
Misure di risultato secondarie
Misura del risultato |
Misura Descrizione |
|---|---|
|
Effect of different values of PEEP on lung mechanics
|
- Lung compliance:During the assessment protocol was changed to pressure control ventilation mode with the following values of ventilatory parameters: Pp = 30 cm H2O, inspiratory time = 1s; FiO2 = 40%, RR = 16 bpm; Sensitivity = 1 cmH2O.
PEEP employed ranged from 0 to 14 cmH2O.
|
|
Effect of different PEEP levels on aspects hemodynamic
|
At each value of PEEP the patient was ventilated for a period of five minutes to carry out monitoring of heart rate and cerebral perfusion pressure(CPP).
|
Collaboratori e investigatori
Sponsor
Sponsor
Investigatori
Investigatori
- Cattedra di studio: Roberto Campelo, Universidade Federal de Pernambuco
Pubblicazioni e link utili
Pubblicazioni generali
- Pillai S, Praharaj SS, Rao GS, Kolluri VR. Cerebral perfusion pressure management of severe diffuse head injury: effect on brain compliance and intracranial pressure. Neurol India. 2004 Mar;52(1):67-71.
- Mayer SA, Copeland D, Bernardini GL, Boden-Albala B, Lennihan L, Kossoff S, Sacco RL. Cost and outcome of mechanical ventilation for life-threatening stroke. Stroke. 2000 Oct;31(10):2346-53. doi: 10.1161/01.str.31.10.2346.
Studiare le date dei record
Studia le date principali
Inizio studio
Inizio studio
Completamento primario (Effettivo)
Completamento primario
Completamento dello studio (Effettivo)
Completamento dello studio
Date di iscrizione allo studio
Primo inviato
Primo inviato
Primo inviato che soddisfa i criteri di controllo qualità
Primo inviato che soddisfa i criteri di controllo qualità
Primo Inserito (Stima)
Primo Inserito
Aggiornamenti dei record di studio
Ultimo aggiornamento pubblicato (Stima)
Ultimo aggiornamento pubblicato
Ultimo aggiornamento inviato che soddisfa i criteri QC
Ultimo aggiornamento inviato che soddisfa i criteri QC
Ultimo verificato
Ultimo verificato
Maggiori informazioni
Termini relativi a questo studio
Termini MeSH pertinenti aggiuntivi
Altri numeri di identificazione dello studio
Altri numeri di identificazione dello studio
- Berg 01
Queste informazioni sono state recuperate direttamente dal sito web clinicaltrials.gov senza alcuna modifica. In caso di richieste di modifica, rimozione o aggiornamento dei dettagli dello studio, contattare register@clinicaltrials.gov. Non appena verrà implementata una modifica su clinicaltrials.gov, questa verrà aggiornata automaticamente anche sul nostro sito web .