Comparative Evaluation of Optrell and Octaray Mapping Catheters for Conduction Block Confirmation and Voltage Assessment During Atrial Ablation

31 maggio 2026 aggiornato da: Maxime Zabern

This prospective, single-center observational study aims to compare the clinical performance, workflow efficiency, and electrophysiological metrics of two commercially available, high-density multielectrode mapping catheters (Octaray and Optrell) in patients undergoing catheter ablation for complex atrial arrhythmias.

The primary focus is to determine if the real-time vector propagation capabilities of the Optrell catheter combined with V8 software can accurately confirm conduction block across linear ablation lesions without the need for full-chamber activation remapping, which is the current standard of care. A secondary focus will pair-compare the bipolar voltage amplitude distributions recorded by both catheters, given their differing electrode sizes and spacing.

Panoramica dello studio

Tipo di studio

Interventistico

Iscrizione (Stimato)

30

Fase

  • Non applicabile

Contatti e Sedi

Questa sezione fornisce i recapiti di coloro che conducono lo studio e informazioni su dove viene condotto lo studio.

Contatto studio

Backup dei contatti dello studio

Luoghi di studio

Criteri di partecipazione

I ricercatori cercano persone che corrispondano a una certa descrizione, chiamata criteri di ammissibilità. Alcuni esempi di questi criteri sono le condizioni generali di salute di una persona o trattamenti precedenti.

Criteri di ammissibilità

Età idonea allo studio

  • Adulto
  • Adulto più anziano

Accetta volontari sani

No

Descrizione

Inclusion Criteria:

  • • Age ≥18 years.
  • • Documented persistent atrial fibrillation and/or atrial flutter.
  • • Undergoing a clinically indicated first-time (de novo) or redo catheter ablation where linear ablation lesion sets (e.g., posterior wall, roof line, lateral mitral line, CTI) are planned or anticipated.
  • • Ability to provide written informed consent.

Exclusion Criteria:

  • • Paroxysmal or persistent AF where only standard PVI is planned.
  • • Mechanical mitral valve.
  • • Presence of intracardiac thrombus or contraindication to systemic anticoagulation.
  • • Any medical condition or anatomical contraindication that, in the investigator's judgment, precludes safe participation.
  • • Pregnant women.

Piano di studio

Questa sezione fornisce i dettagli del piano di studio, compreso il modo in cui lo studio è progettato e ciò che lo studio sta misurando.

Come è strutturato lo studio?

Dettagli di progettazione

  • Scopo principale: Altro
  • Assegnazione: N / A
  • Modello interventistico: Assegnazione di gruppo singolo
  • Mascheramento: Nessuno (etichetta aperta)

Armi e interventi

Gruppo di partecipanti / Arm
Intervento / Trattamento
Altro: Comparison between two mapping catheters, Optrell vs Octaray, in the same patient
Comparison between two mapping catheters, Optrell vs Octaray, in the same patient

Cosa sta misurando lo studio?

Misure di risultato primarie

Misura del risultato
Misura Descrizione
Lasso di tempo
Concordance
Lasso di tempo: During ablation procedure
Percentage of agreement (presence vs. absence of bidirectional block) between the Optrell real-time vector assessment and standard activation remapping on a per-line basis.
During ablation procedure
Workflow Efficiency
Lasso di tempo: During ablation procedure
Difference in time (in minutes) required to confirm conduction block status using the Optrell vector method versus full activation mapping.
During ablation procedure

Misure di risultato secondarie

Misura del risultato
Misura Descrizione
Lasso di tempo
Bipolar Voltage Correlation
Lasso di tempo: During ablation procedure
Statistical correlation and evaluation of systematic bias (limits of agreement) between the bipolar voltage amplitudes recorded by the two catheters in matched anatomical regions.
During ablation procedure
Multipolar Voltage Correlation
Lasso di tempo: During ablation procedure
Statistical correlation and evaluation of systematic bias (limits of agreement) between the multipolar voltage amplitudes recorded by the Optrell catheter and the bipolar voltage amplitudes recorded by the Octaray catheter in matched anatomical regions.
During ablation procedure
Clinical Efficacy
Lasso di tempo: 12 month follow up period
Freedom from any documented atrial tachyarrhythmia lasting ≥30 seconds at 1, 3, 6, 9, and 12 months
12 month follow up period

Collaboratori e investigatori

Qui è dove troverai le persone e le organizzazioni coinvolte in questo studio.

Sponsor

Investigatori

  • Investigatore principale: Maxime Zabern, MD, Assaf-Harofeh Medical Center

Studiare le date dei record

Queste date tengono traccia dell'avanzamento della registrazione dello studio e dell'invio dei risultati di sintesi a ClinicalTrials.gov. I record degli studi e i risultati riportati vengono esaminati dalla National Library of Medicine (NLM) per assicurarsi che soddisfino specifici standard di controllo della qualità prima di essere pubblicati sul sito Web pubblico.

Studia le date principali

Inizio studio (Effettivo)

12 maggio 2026

Completamento primario (Stimato)

1 maggio 2027

Completamento dello studio (Stimato)

1 maggio 2028

Date di iscrizione allo studio

Primo inviato

26 aprile 2026

Primo inviato che soddisfa i criteri di controllo qualità

26 aprile 2026

Primo Inserito (Effettivo)

1 maggio 2026

Aggiornamenti dei record di studio

Ultimo aggiornamento pubblicato (Effettivo)

3 giugno 2026

Ultimo aggiornamento inviato che soddisfa i criteri QC

31 maggio 2026

Ultimo verificato

1 maggio 2026

Maggiori informazioni

Termini relativi a questo studio

Altri numeri di identificazione dello studio

  • Optrell vs Octaray

Informazioni su farmaci e dispositivi, documenti di studio

Studia un prodotto farmaceutico regolamentato dalla FDA degli Stati Uniti

No

Studia un dispositivo regolamentato dalla FDA degli Stati Uniti

prodotto fabbricato ed esportato dagli Stati Uniti

Queste informazioni sono state recuperate direttamente dal sito web clinicaltrials.gov senza alcuna modifica. In caso di richieste di modifica, rimozione o aggiornamento dei dettagli dello studio, contattare register@clinicaltrials.gov. Non appena verrà implementata una modifica su clinicaltrials.gov, questa verrà aggiornata automaticamente anche sul nostro sito web .

Prove cliniche su Fibrillazione atriale (FA)

Sottoscrivi