Visuele prestaties met Clareon® Vivity® intraoculaire lenzen (IOL's)
Evaluatie van de impact van restastigmatisme op de visuele prestaties met Clareon Vivity
Studie Overzicht
Toestand
Toestand
Conditie
Conditie
Interventie / Behandeling
Interventie / Behandeling
Gedetailleerde beschrijving
Studietype
Studietype
Inschrijving (Werkelijk)
Inschrijving
Fase
Fase
- Niet toepasbaar
Contacten en locaties
Studiecontact
Studiecontact
- Naam: Alcon Call Center
- Telefoonnummer: 1-888-451-3937
- E-mail: alcon.medinfo@alcon.com
Studie Locaties
-
-
Michigan
-
Bloomfield Hills, Michigan, Verenigde Staten, 48302
- Grosinger, Spigelman & Grey Eye Surgeons, P.C
-
-
Texas
-
Dallas, Texas, Verenigde Staten, 75243
- Key-Whitman Eye Center
-
Sugar Land, Texas, Verenigde Staten, 77478
- Berkeley Eye Center
-
-
Deelname Criteria
Geschiktheidscriteria
Geschiktheidscriteria
Leeftijden die in aanmerking komen voor studie
- Volwassen
- Oudere volwassene
Accepteert gezonde vrijwilligers
Beschrijving
Inclusiecriteria:
- In staat zijn om een door de Institution Review Board goedgekeurd formulier voor geïnformeerde toestemming te begrijpen en te ondertekenen;
- Bereid en in staat om alle geplande studiebezoeken bij te wonen, zoals vereist door het protocol;
- Geïmplanteerd met onderzoeks-IOL's in beide ogen, minimaal 4 maanden vóór deelname aan het onderzoek;
- BCDVA van 0 logMAR of beter in elk oog.
Uitsluitingscriteria:
- Resterend astigmatisme groter dan of gelijk aan 1,00 D;
- Klinisch significante oogaandoeningen zoals gespecificeerd in het protocol;
- Geschiedenis van oogoperaties anders dan limbale ontspannende incisies
- Andere protocolgespecificeerde uitsluitingscriteria kunnen van toepassing zijn.
Studie plan
Hoe is de studie opgezet?
Ontwerpdetails
- Primair doel: Behandeling
- Toewijzing: NVT
- Interventioneel model: Opdracht voor een enkele groep
- Masker: Geen (open label)
Aantal wapens
Wapens en interventies
Deelnemersgroep / ArmDeelnemersgroep / Arm |
Interventie / BehandelingInterventie / Behandeling |
|---|---|
|
Experimenteel: Clareon Vivity/Vivity Torisch
Eerdere implantatie in beide ogen met Clareon Vivity/Vivity Toric Extended Vision IOL's, waarbij eerdere implantatie is gedefinieerd als minimaal 4 maanden vóór deelname aan het onderzoek.
|
Intraoculaire lens met verlengd zicht, geïmplanteerd in de kapselzak van het oog tijdens een cataractoperatie.
IOL's zijn bedoeld voor langdurig gebruik gedurende de levensduur van de cataractpatiënt.
Andere namen:
|
Wat meet het onderzoek?
Primaire uitkomstmaten
Primaire uitkomstmaten
Uitkomstmaat |
Maatregel Beschrijving |
Tijdsspanne |
|---|---|---|
|
Mean Difference in Monocular Best Corrected Distance Visual Acuity (BCDVA) (4 Meters) Before and After Inducing 1.0 Diopter (D) Astigmatism - WTR
Tijdsspanne: Day 0 before induced astigmatism; Day 0 after induced astigmatism
|
BCDVA was assessed with best refractive correction in place at a distance of 4 meters (m) using Early Treatment Diabetic Retinopathy Study (ETDRS) letter charts.
Trial lenses were used to induce 1.0 diopter (D) of astigmatism with-the-rule (WTR), defined as -0.50/+1.00
D at approximately 180 degrees on the optical axis.
VA was measured in the first implanted eye in logarithm minimum angle of resolution (logMAR), with a lower logMAR value indicating better visual acuity.
The change in VA before and after inducing astigmatism is reported, with a positive value representing a decline in VA.
No hypothesis testing was prespecified for this co-primary endpoint.
|
Day 0 before induced astigmatism; Day 0 after induced astigmatism
|
|
Mean Difference in Monocular Distance Corrected Intermediate Visual Acuity (DCIVA) (66 Centimeters) Before and After Inducing 1.0 D Astigmatism - WTR
Tijdsspanne: Day 0 before induced astigmatism; Day 0 after induced astigmatism
|
DCIVA was assessed with distance correction in place (plus or minus power) at a distance of 66 centimeters (cm) from the spectacle plane using ETDRS letter charts.
Trial lenses were used to induce 1.0 D of astigmatism with-the-rule (WTR), defined as -0.50/+1.00
D at approximately 180 degrees on the optical axis.
VA was measured in the first implanted eye in logarithm minimum angle of resolution (logMAR), with a lower logMAR value indicating better visual acuity.
The change in VA before and after inducing astigmatism is reported, with a positive value representing a decline in VA.
No hypothesis testing was prespecified for this co-primary endpoint.
|
Day 0 before induced astigmatism; Day 0 after induced astigmatism
|
|
Mean Difference in Monocular Distance Corrected Near Visual Acuity (DCNVA) (40 cm) Before and After Inducing 1.0 D Astigmatism - WTR
Tijdsspanne: Day 0 before induced astigmatism; Day 0 after induced astigmatism
|
DCNVA was assessed with distance correction in place (plus or minus power) at a distance of 40 cm from the spectacle plane using ETDRS letter charts.
Trial lenses were used to induce 1.0 D of astigmatism WTR, defined as -0.50/+1.00
D at approximately 180 degrees on the optical axis.VA was measured in the first implanted eye in logarithm minimum angle of resolution (logMAR), with a lower logMAR value indicating better visual acuity.
The change in VA before and after inducing astigmatism is reported, with a positive value representing a decline in VA.
No hypothesis testing was prespecified for this co-primary endpoint.
|
Day 0 before induced astigmatism; Day 0 after induced astigmatism
|
|
Mean Difference in Monocular Best Corrected Distance Visual Acuity (BCDVA) (4 Meters) Before and After Inducing 1.0 Diopter (D) Astigmatism - ATR
Tijdsspanne: Day 0 before induced astigmatism; Day 0 after induced astigmatism
|
BCDVA was assessed with best refractive correction in place at a distance of 4 m using ETDRS letter charts.
Trial lenses were used to induce 1.0 D of astigmatism against the rule (ATR), defined as -0.50/+1.00
D at approximately 90 degrees on the optical axis.
VA was measured in the first implanted eye in logarithm minimum angle of resolution (logMAR), with a lower logMAR value indicating better visual acuity.
The change in VA before and after inducing astigmatism is reported, with a positive value representing a decline in VA.
No hypothesis testing was prespecified for this co-primary endpoint.
|
Day 0 before induced astigmatism; Day 0 after induced astigmatism
|
|
Mean Difference in Monocular Distance Corrected Intermediate Visual Acuity (DCIVA) (66 Centimeters) Before and After Inducing 1.0 D Astigmatism.- ATR
Tijdsspanne: Day 0 before induced astigmatism; Day 0 after induced astigmatism
|
DCIVA was assessed with distance correction in place (plus or minus power) at a distance of 66 cm from the spectacle plane using ETDRS letter charts.
Trial lenses were used to induce 1.0 D of astigmatism ATR, defined as -0.50/+1.00
D at approximately 90 degrees on the optical axis.
VA was measured in the first implanted eye in logarithm minimum angle of resolution (logMAR), with a lower logMAR value indicating better visual acuity.
The change in VA before and after inducing astigmatism is reported, with a positive value representing a decline in VA.
No hypothesis testing was prespecified for this co-primary endpoint.
|
Day 0 before induced astigmatism; Day 0 after induced astigmatism
|
|
Mean Difference in Monocular Distance Corrected Near Visual Acuity (DCNVA) (40 cm) Before and After Inducing 1.0 D Astigmatism - ATR
Tijdsspanne: Day 0 before induced astigmatism; Day 0 after induced astigmatism
|
DCNVA was assessed with distance correction in place (plus or minus power) at a distance of 40 cm from the spectacle plane using ETDRS letter charts.
Trial lenses were used to induce 1.0 D of astigmatism ATR, defined as -0.50/+1.00
D at approximately 90 degrees on the optical axis.
VA was measured in the first implanted eye in logarithm minimum angle of resolution (logMAR), with a lower logMAR value indicating better visual acuity.
The change in VA before and after inducing astigmatism is reported, with a positive value representing a decline in VA.
No hypothesis testing was prespecified for this co-primary endpoint.
|
Day 0 before induced astigmatism; Day 0 after induced astigmatism
|
Medewerkers en onderzoekers
Sponsor
Sponsor
Onderzoekers
Onderzoekers
- Studie directeur: Clinical Trial Lead, Surgical CRD, Alcon Research, LLC
Publicaties en nuttige links
Nuttige links
Studie record data
Bestudeer belangrijke data
Studie start (Werkelijk)
Studie start
Primaire voltooiing (Werkelijk)
Primaire voltooiing
Studie voltooiing (Werkelijk)
Studie voltooiing
Studieregistratiedata
Eerst ingediend
Eerst ingediend
Eerst ingediend dat voldeed aan de QC-criteria
Eerst ingediend dat voldeed aan de QC-criteria
Eerst geplaatst (Werkelijk)
Eerst geplaatst
Updates van studierecords
Laatste update geplaatst (Werkelijk)
Laatste update geplaatst
Laatste update ingediend die voldeed aan QC-criteria
Laatste update ingediend die voldeed aan QC-criteria
Laatst geverifieerd
Laatst geverifieerd
Meer informatie
Termen gerelateerd aan deze studie
Aanvullende relevante MeSH-voorwaarden
Andere studie-ID-nummers
Andere studie-ID-nummers
- ILE632-C003
Plan Individuele Deelnemersgegevens (IPD)
Bent u van plan om gegevens van individuele deelnemers (IPD) te delen?
Informatie over medicijnen en apparaten, studiedocumenten
Bestudeert een door de Amerikaanse FDA gereguleerd geneesmiddel
Bestudeert een door de Amerikaanse FDA gereguleerd apparaatproduct
product vervaardigd in en geëxporteerd uit de V.S.
Deze informatie is zonder wijzigingen rechtstreeks van de website clinicaltrials.gov gehaald. Als u verzoeken heeft om uw onderzoeksgegevens te wijzigen, te verwijderen of bij te werken, neem dan contact op met register@clinicaltrials.gov. Zodra er een wijziging wordt doorgevoerd op clinicaltrials.gov, wordt deze ook automatisch bijgewerkt op onze website .