Visuell ytelse med Clareon® Vivity® intraokulære linser (IOL)
Evaluering av innvirkningen av gjenværende astigmatisme på den visuelle ytelsen med Clareon Vivity
Studieoversikt
Status
Status
Forhold
Forhold
Intervensjon / Behandling
Intervensjon / Behandling
Detaljert beskrivelse
Studietype
Studietype
Registrering (Faktiske)
Registrering
Fase
Fase
- Ikke aktuelt
Kontakter og plasseringer
Studiekontakt
Studiekontakt
- Navn: Alcon Call Center
- Telefonnummer: 1-888-451-3937
- E-post: alcon.medinfo@alcon.com
Studiesteder
-
-
Michigan
-
Bloomfield Hills, Michigan, Forente stater, 48302
- Grosinger, Spigelman & Grey Eye Surgeons, P.C
-
-
Texas
-
Dallas, Texas, Forente stater, 75243
- Key-Whitman Eye Center
-
Sugar Land, Texas, Forente stater, 77478
- Berkeley Eye Center
-
-
Deltakelseskriterier
Kvalifikasjonskriterier
Kvalifikasjonskriterier
Alder som er kvalifisert for studier
- Voksen
- Eldre voksen
Tar imot friske frivillige
Beskrivelse
Inkluderingskriterier:
- Kunne forstå og signere et skjema for informert samtykke som er godkjent av institusjonsvurderingsstyret;
- Villig og i stand til å delta på alle planlagte studiebesøk som kreves av protokollen;
- Implantert med IOL-er i begge øyne minst 4 måneder før studiedeltakelse;
- BCDVA på 0 logMAR eller bedre i hvert øye.
Ekskluderingskriterier:
- Resterende astigmatisme større enn eller lik 1,00 D;
- Klinisk signifikante okulære tilstander som spesifisert i protokollen;
- Anamnese med andre øyeoperasjoner enn limbalavslappende snitt
- Andre protokollspesifiserte eksklusjonskriterier kan gjelde.
Studieplan
Hvordan er studiet utformet?
Designdetaljer
- Primært formål: Behandling
- Tildeling: N/A
- Intervensjonsmodell: Enkeltgruppeoppdrag
- Masking: Ingen (Open Label)
Antall våpen
Våpen og intervensjoner
Deltakergruppe / ArmDeltakergruppe / Arm |
Intervensjon / BehandlingIntervensjon / Behandling |
|---|---|
|
Eksperimentell: Clareon Vivity/Vivity Toric
Tidligere implantasjon i begge øyne med Clareon Vivity/Vivity Toric Extended Vision IOL, med tidligere implantasjon definert som minst 4 måneder før studiedeltakelse.
|
Intraokulær linse med utvidet syn implantert i kapselposen i øyet under kataraktkirurgi.
IOL er beregnet for langtidsbruk over levetiden til kataraktpasienten.
Andre navn:
|
Hva måler studien?
Primære resultatmål
Primære resultatmål
Resultatmål |
Tiltaksbeskrivelse |
Tidsramme |
|---|---|---|
|
Mean Difference in Monocular Best Corrected Distance Visual Acuity (BCDVA) (4 Meters) Before and After Inducing 1.0 Diopter (D) Astigmatism - WTR
Tidsramme: Day 0 before induced astigmatism; Day 0 after induced astigmatism
|
BCDVA was assessed with best refractive correction in place at a distance of 4 meters (m) using Early Treatment Diabetic Retinopathy Study (ETDRS) letter charts.
Trial lenses were used to induce 1.0 diopter (D) of astigmatism with-the-rule (WTR), defined as -0.50/+1.00
D at approximately 180 degrees on the optical axis.
VA was measured in the first implanted eye in logarithm minimum angle of resolution (logMAR), with a lower logMAR value indicating better visual acuity.
The change in VA before and after inducing astigmatism is reported, with a positive value representing a decline in VA.
No hypothesis testing was prespecified for this co-primary endpoint.
|
Day 0 before induced astigmatism; Day 0 after induced astigmatism
|
|
Mean Difference in Monocular Distance Corrected Intermediate Visual Acuity (DCIVA) (66 Centimeters) Before and After Inducing 1.0 D Astigmatism - WTR
Tidsramme: Day 0 before induced astigmatism; Day 0 after induced astigmatism
|
DCIVA was assessed with distance correction in place (plus or minus power) at a distance of 66 centimeters (cm) from the spectacle plane using ETDRS letter charts.
Trial lenses were used to induce 1.0 D of astigmatism with-the-rule (WTR), defined as -0.50/+1.00
D at approximately 180 degrees on the optical axis.
VA was measured in the first implanted eye in logarithm minimum angle of resolution (logMAR), with a lower logMAR value indicating better visual acuity.
The change in VA before and after inducing astigmatism is reported, with a positive value representing a decline in VA.
No hypothesis testing was prespecified for this co-primary endpoint.
|
Day 0 before induced astigmatism; Day 0 after induced astigmatism
|
|
Mean Difference in Monocular Distance Corrected Near Visual Acuity (DCNVA) (40 cm) Before and After Inducing 1.0 D Astigmatism - WTR
Tidsramme: Day 0 before induced astigmatism; Day 0 after induced astigmatism
|
DCNVA was assessed with distance correction in place (plus or minus power) at a distance of 40 cm from the spectacle plane using ETDRS letter charts.
Trial lenses were used to induce 1.0 D of astigmatism WTR, defined as -0.50/+1.00
D at approximately 180 degrees on the optical axis.VA was measured in the first implanted eye in logarithm minimum angle of resolution (logMAR), with a lower logMAR value indicating better visual acuity.
The change in VA before and after inducing astigmatism is reported, with a positive value representing a decline in VA.
No hypothesis testing was prespecified for this co-primary endpoint.
|
Day 0 before induced astigmatism; Day 0 after induced astigmatism
|
|
Mean Difference in Monocular Best Corrected Distance Visual Acuity (BCDVA) (4 Meters) Before and After Inducing 1.0 Diopter (D) Astigmatism - ATR
Tidsramme: Day 0 before induced astigmatism; Day 0 after induced astigmatism
|
BCDVA was assessed with best refractive correction in place at a distance of 4 m using ETDRS letter charts.
Trial lenses were used to induce 1.0 D of astigmatism against the rule (ATR), defined as -0.50/+1.00
D at approximately 90 degrees on the optical axis.
VA was measured in the first implanted eye in logarithm minimum angle of resolution (logMAR), with a lower logMAR value indicating better visual acuity.
The change in VA before and after inducing astigmatism is reported, with a positive value representing a decline in VA.
No hypothesis testing was prespecified for this co-primary endpoint.
|
Day 0 before induced astigmatism; Day 0 after induced astigmatism
|
|
Mean Difference in Monocular Distance Corrected Intermediate Visual Acuity (DCIVA) (66 Centimeters) Before and After Inducing 1.0 D Astigmatism.- ATR
Tidsramme: Day 0 before induced astigmatism; Day 0 after induced astigmatism
|
DCIVA was assessed with distance correction in place (plus or minus power) at a distance of 66 cm from the spectacle plane using ETDRS letter charts.
Trial lenses were used to induce 1.0 D of astigmatism ATR, defined as -0.50/+1.00
D at approximately 90 degrees on the optical axis.
VA was measured in the first implanted eye in logarithm minimum angle of resolution (logMAR), with a lower logMAR value indicating better visual acuity.
The change in VA before and after inducing astigmatism is reported, with a positive value representing a decline in VA.
No hypothesis testing was prespecified for this co-primary endpoint.
|
Day 0 before induced astigmatism; Day 0 after induced astigmatism
|
|
Mean Difference in Monocular Distance Corrected Near Visual Acuity (DCNVA) (40 cm) Before and After Inducing 1.0 D Astigmatism - ATR
Tidsramme: Day 0 before induced astigmatism; Day 0 after induced astigmatism
|
DCNVA was assessed with distance correction in place (plus or minus power) at a distance of 40 cm from the spectacle plane using ETDRS letter charts.
Trial lenses were used to induce 1.0 D of astigmatism ATR, defined as -0.50/+1.00
D at approximately 90 degrees on the optical axis.
VA was measured in the first implanted eye in logarithm minimum angle of resolution (logMAR), with a lower logMAR value indicating better visual acuity.
The change in VA before and after inducing astigmatism is reported, with a positive value representing a decline in VA.
No hypothesis testing was prespecified for this co-primary endpoint.
|
Day 0 before induced astigmatism; Day 0 after induced astigmatism
|
Samarbeidspartnere og etterforskere
Sponsor
Sponsor
Etterforskere
Etterforskere
- Studieleder: Clinical Trial Lead, Surgical CRD, Alcon Research, LLC
Publikasjoner og nyttige lenker
Hjelpsomme linker
Studierekorddatoer
Studer hoveddatoer
Studiestart (Faktiske)
Studiestart
Primær fullføring (Faktiske)
Primær fullføring
Studiet fullført (Faktiske)
Studiet fullført
Datoer for studieregistrering
Først innsendt
Først innsendt
Først innsendt som oppfylte QC-kriteriene
Først innsendt som oppfylte QC-kriteriene
Først lagt ut (Faktiske)
Først lagt ut
Oppdateringer av studieposter
Sist oppdatering lagt ut (Faktiske)
Sist oppdatering lagt ut
Siste oppdatering sendt inn som oppfylte QC-kriteriene
Siste oppdatering sendt inn som oppfylte QC-kriteriene
Sist bekreftet
Sist bekreftet
Mer informasjon
Begreper knyttet til denne studien
Ytterligere relevante MeSH-vilkår
Andre studie-ID-numre
Andre studie-ID-numre
- ILE632-C003
Plan for individuelle deltakerdata (IPD)
Planlegger du å dele individuelle deltakerdata (IPD)?
Legemiddel- og utstyrsinformasjon, studiedokumenter
Studerer et amerikansk FDA-regulert medikamentprodukt
Studerer et amerikansk FDA-regulert enhetsprodukt
produkt produsert i og eksportert fra USA
Denne informasjonen ble hentet direkte fra nettstedet clinicaltrials.gov uten noen endringer. Hvis du har noen forespørsler om å endre, fjerne eller oppdatere studiedetaljene dine, vennligst kontakt register@clinicaltrials.gov. Så snart en endring er implementert på clinicaltrials.gov, vil denne også bli oppdatert automatisk på nettstedet vårt. .