Rendimiento visual con lentes intraoculares (LIO) Clareon® Vivity®
Evaluación del impacto del astigmatismo residual en el rendimiento visual con Clareon Vivity
Descripción general del estudio
Estado
Estado
Condiciones
Condiciones
Intervención / Tratamiento
Intervención / Tratamiento
Descripción detallada
Tipo de estudio
Tipo de estudio
Inscripción (Actual)
Inscripción
Fase
Fase
- No aplica
Contactos y Ubicaciones
Estudio Contacto
Estudio Contacto
- Nombre: Alcon Call Center
- Número de teléfono: 1-888-451-3937
- Correo electrónico: alcon.medinfo@alcon.com
Ubicaciones de estudio
-
-
Michigan
-
Bloomfield Hills, Michigan, Estados Unidos, 48302
- Grosinger, Spigelman & Grey Eye Surgeons, P.C
-
-
Texas
-
Dallas, Texas, Estados Unidos, 75243
- Key-Whitman Eye Center
-
Sugar Land, Texas, Estados Unidos, 77478
- Berkeley Eye Center
-
-
Criterios de participación
Criterio de elegibilidad
Criterio de elegibilidad
Edades elegibles para estudiar
- Adulto
- Adulto Mayor
Acepta Voluntarios Saludables
Descripción
Criterios de inclusión:
- Capaz de comprender y firmar un formulario de consentimiento informado aprobado por la Junta de Revisión Institucional;
- Dispuesto y capaz de asistir a todas las visitas de estudio programadas según lo exige el protocolo;
- Implantado con LIO del estudio en ambos ojos al menos 4 meses antes de la participación en el estudio;
- BCDVA de 0 logMAR o mejor en cada ojo.
Criterios de exclusión:
- Astigmatismo residual mayor o igual a 1,00 D;
- Condiciones oculares clínicamente significativas según lo especificado en el protocolo;
- Historia de cirugías oculares distintas de las incisiones relajantes limbales.
- Pueden aplicarse otros criterios de exclusión especificados en el protocolo.
Plan de estudios
¿Cómo está diseñado el estudio?
Detalles de diseño
- Propósito principal: Tratamiento
- Asignación: N / A
- Modelo Intervencionista: Asignación de un solo grupo
- Enmascaramiento: Ninguno (etiqueta abierta)
Número de brazos
Armas e Intervenciones
Grupo de participantes/brazoGrupo de participantes/brazo |
Intervención / TratamientoIntervención / Tratamiento |
|---|---|
|
Experimental: Clareon Vivity/Vivity Tórico
Implantación previa en ambos ojos de LIO Clareon Vivity/Vivity Toric Extended Vision, con implantación previa definida como al menos 4 meses antes de la participación en el estudio.
|
Lente intraocular de visión extendida implantada en la bolsa capsular del ojo durante la cirugía de cataratas.
Las LIO están diseñadas para un uso prolongado durante toda la vida del paciente con cataratas.
Otros nombres:
|
¿Qué mide el estudio?
Medidas de resultado primarias
Medidas de resultado primarias
Medida de resultado |
Medida Descripción |
Periodo de tiempo |
|---|---|---|
|
Mean Difference in Monocular Best Corrected Distance Visual Acuity (BCDVA) (4 Meters) Before and After Inducing 1.0 Diopter (D) Astigmatism - WTR
Periodo de tiempo: Day 0 before induced astigmatism; Day 0 after induced astigmatism
|
BCDVA was assessed with best refractive correction in place at a distance of 4 meters (m) using Early Treatment Diabetic Retinopathy Study (ETDRS) letter charts.
Trial lenses were used to induce 1.0 diopter (D) of astigmatism with-the-rule (WTR), defined as -0.50/+1.00
D at approximately 180 degrees on the optical axis.
VA was measured in the first implanted eye in logarithm minimum angle of resolution (logMAR), with a lower logMAR value indicating better visual acuity.
The change in VA before and after inducing astigmatism is reported, with a positive value representing a decline in VA.
No hypothesis testing was prespecified for this co-primary endpoint.
|
Day 0 before induced astigmatism; Day 0 after induced astigmatism
|
|
Mean Difference in Monocular Distance Corrected Intermediate Visual Acuity (DCIVA) (66 Centimeters) Before and After Inducing 1.0 D Astigmatism - WTR
Periodo de tiempo: Day 0 before induced astigmatism; Day 0 after induced astigmatism
|
DCIVA was assessed with distance correction in place (plus or minus power) at a distance of 66 centimeters (cm) from the spectacle plane using ETDRS letter charts.
Trial lenses were used to induce 1.0 D of astigmatism with-the-rule (WTR), defined as -0.50/+1.00
D at approximately 180 degrees on the optical axis.
VA was measured in the first implanted eye in logarithm minimum angle of resolution (logMAR), with a lower logMAR value indicating better visual acuity.
The change in VA before and after inducing astigmatism is reported, with a positive value representing a decline in VA.
No hypothesis testing was prespecified for this co-primary endpoint.
|
Day 0 before induced astigmatism; Day 0 after induced astigmatism
|
|
Mean Difference in Monocular Distance Corrected Near Visual Acuity (DCNVA) (40 cm) Before and After Inducing 1.0 D Astigmatism - WTR
Periodo de tiempo: Day 0 before induced astigmatism; Day 0 after induced astigmatism
|
DCNVA was assessed with distance correction in place (plus or minus power) at a distance of 40 cm from the spectacle plane using ETDRS letter charts.
Trial lenses were used to induce 1.0 D of astigmatism WTR, defined as -0.50/+1.00
D at approximately 180 degrees on the optical axis.VA was measured in the first implanted eye in logarithm minimum angle of resolution (logMAR), with a lower logMAR value indicating better visual acuity.
The change in VA before and after inducing astigmatism is reported, with a positive value representing a decline in VA.
No hypothesis testing was prespecified for this co-primary endpoint.
|
Day 0 before induced astigmatism; Day 0 after induced astigmatism
|
|
Mean Difference in Monocular Best Corrected Distance Visual Acuity (BCDVA) (4 Meters) Before and After Inducing 1.0 Diopter (D) Astigmatism - ATR
Periodo de tiempo: Day 0 before induced astigmatism; Day 0 after induced astigmatism
|
BCDVA was assessed with best refractive correction in place at a distance of 4 m using ETDRS letter charts.
Trial lenses were used to induce 1.0 D of astigmatism against the rule (ATR), defined as -0.50/+1.00
D at approximately 90 degrees on the optical axis.
VA was measured in the first implanted eye in logarithm minimum angle of resolution (logMAR), with a lower logMAR value indicating better visual acuity.
The change in VA before and after inducing astigmatism is reported, with a positive value representing a decline in VA.
No hypothesis testing was prespecified for this co-primary endpoint.
|
Day 0 before induced astigmatism; Day 0 after induced astigmatism
|
|
Mean Difference in Monocular Distance Corrected Intermediate Visual Acuity (DCIVA) (66 Centimeters) Before and After Inducing 1.0 D Astigmatism.- ATR
Periodo de tiempo: Day 0 before induced astigmatism; Day 0 after induced astigmatism
|
DCIVA was assessed with distance correction in place (plus or minus power) at a distance of 66 cm from the spectacle plane using ETDRS letter charts.
Trial lenses were used to induce 1.0 D of astigmatism ATR, defined as -0.50/+1.00
D at approximately 90 degrees on the optical axis.
VA was measured in the first implanted eye in logarithm minimum angle of resolution (logMAR), with a lower logMAR value indicating better visual acuity.
The change in VA before and after inducing astigmatism is reported, with a positive value representing a decline in VA.
No hypothesis testing was prespecified for this co-primary endpoint.
|
Day 0 before induced astigmatism; Day 0 after induced astigmatism
|
|
Mean Difference in Monocular Distance Corrected Near Visual Acuity (DCNVA) (40 cm) Before and After Inducing 1.0 D Astigmatism - ATR
Periodo de tiempo: Day 0 before induced astigmatism; Day 0 after induced astigmatism
|
DCNVA was assessed with distance correction in place (plus or minus power) at a distance of 40 cm from the spectacle plane using ETDRS letter charts.
Trial lenses were used to induce 1.0 D of astigmatism ATR, defined as -0.50/+1.00
D at approximately 90 degrees on the optical axis.
VA was measured in the first implanted eye in logarithm minimum angle of resolution (logMAR), with a lower logMAR value indicating better visual acuity.
The change in VA before and after inducing astigmatism is reported, with a positive value representing a decline in VA.
No hypothesis testing was prespecified for this co-primary endpoint.
|
Day 0 before induced astigmatism; Day 0 after induced astigmatism
|
Colaboradores e Investigadores
Patrocinador
Patrocinador
Investigadores
Investigadores
- Director de estudio: Clinical Trial Lead, Surgical CRD, Alcon Research, LLC
Publicaciones y enlaces útiles
Enlaces Útiles
Fechas de registro del estudio
Fechas importantes del estudio
Inicio del estudio (Actual)
Inicio del estudio
Finalización primaria (Actual)
Finalización primaria
Finalización del estudio (Actual)
Finalización del estudio
Fechas de registro del estudio
Enviado por primera vez
Enviado por primera vez
Primero enviado que cumplió con los criterios de control de calidad
Primero enviado que cumplió con los criterios de control de calidad
Publicado por primera vez (Actual)
Publicado por primera vez
Actualizaciones de registros de estudio
Última actualización publicada (Actual)
Última actualización publicada
Última actualización enviada que cumplió con los criterios de control de calidad
Última actualización enviada que cumplió con los criterios de control de calidad
Última verificación
Última verificación
Más información
Términos relacionados con este estudio
Términos MeSH relevantes adicionales
Otros números de identificación del estudio
Otros números de identificación del estudio
- ILE632-C003
Plan de datos de participantes individuales (IPD)
¿Planea compartir datos de participantes individuales (IPD)?
Información sobre medicamentos y dispositivos, documentos del estudio
Estudia un producto farmacéutico regulado por la FDA de EE. UU.
Estudia un producto de dispositivo regulado por la FDA de EE. UU.
producto fabricado y exportado desde los EE. UU.
Esta información se obtuvo directamente del sitio web clinicaltrials.gov sin cambios. Si tiene alguna solicitud para cambiar, eliminar o actualizar los detalles de su estudio, comuníquese con register@clinicaltrials.gov. Tan pronto como se implemente un cambio en clinicaltrials.gov, también se actualizará automáticamente en nuestro sitio web. .