POC Analysis of IO Blood Samples Within Critically Ill Patients
Point-of-care Analyses of Intraosseous Blood Samples Within Critically Ill Patients
Studie Overzicht
Toestand
Conditie
Interventie / Behandeling
Gedetailleerde beschrijving
Intravenous access is sometimes difficult to achieve due to vasoconstriction or centralization of circulation. Feasible IO-access with power-driven device is a standard alternative method of vascular access for critically ill patients. It is still unclear whether results of IO samples agree with venous or arterial blood samples so that they could be used in clinical decision-making.
After intraosseous access has been established for fluid resuscitation or medication, a common practice is to aspirate a small amount of blood to verify the proper location of the IO-needle.
POC-testing can provide emergency physicians or paramedics with important information about the patients. Could bone marrow samples be used for POC-testing instead of arterial or venous blood for emergency treatment decision-making or patient allocation?
Several animal and human studies have been published to investigate the correlation between IO versus venous or arterial blood values. Investigators of this study have performed a study in healthy volunteers and discovered that there is a good agreement for some laboratory parameters (pH, gluc, lact), but for K the agreement is poor. Our study group has as well performed a study in critically ill animals investigating the agreement during cardiac arrest and resuscitation. Te evidence about critically ill people is still week.
Aim of this study is to analyze the reliability of analysis of IO samples compared to arterial blood samples within critically ill adult emergency patients.
The hypotheses of this study are:
- Principally, it is possible to analyze IO blood samples with a POC device.
- The results are reliable enough to guide emergency treatment.
If both hypotheses prove to be true, the method can be immediately used in emergency medical situations, e.g. in searching for reversible causes of cardiac arrest.
Setting
Prospective, observational study with 35 prehospital emergency patients. The IO-samples are collected from the small amount of blood, which is aspirated from intraosseous space to confirm the correct needle placement. The IO needles are inserted to the patients for emergency treatment, not because of the study. Critically ill patients, to whom emergency doctors are inserting an intraosseous needle for emergency fluid or drug treatment, are included in the study.
Samples will be analyzed by using an i-STAT point of care analyzer (i-STAT® handheld, Abbot Point of Care Inc. U.S.A) using CG8+ cartridges to analyse Hb, Na, K, pH, pCO2, pO2, TCO2, HCO3, BE, and SpO2. The reliability of the results comparing the IO and arterial samples will be tested with Bland-Altman method by calculating the bias with 95% confidence intervals.
Ethical issues
The ethical committee of Medical Faculty of University of Helsinki has approved the research. Informed consent is waived.
Time plan
The samples will be collected during years 2017 - 2019. The analysis will be performed in the end of the year 2019.
Studietype
Inschrijving (Werkelijk)
Contacten en locaties
Studie Locaties
-
-
-
Vantaa, Finland
- FinnHEMS10, Emergency Medicine and Services, Helsinki University Hospital
-
-
Deelname Criteria
Geschiktheidscriteria
Leeftijden die in aanmerking komen voor studie
Accepteert gezonde vrijwilligers
Geslachten die in aanmerking komen voor studie
Bemonsteringsmethode
Studie Bevolking
Beschrijving
Inclusion Criteria:
- need for point-of-care analysis
- need for intraosseous access
Exclusion Criteria:
- cardiac arrest
- ability to understand the consent issues in Finnish, Swedish or English
Studie plan
Hoe is de studie opgezet?
Ontwerpdetails
Cohorten en interventies
Groep / Cohort |
Interventie / Behandeling |
---|---|
pre-hospital emergency patients
Patient in a need of an IO-access.
Patient in a need of point-of-care laboratory analysis.
Over 18 years.
Alive (no Cardiac arrest.)
|
POC analysis with i-STAT POC analyser from the intraosseous and arterial blood
|
Wat meet het onderzoek?
Primaire uitkomstmaten
Uitkomstmaat |
Maatregel Beschrijving |
Tijdsspanne |
---|---|---|
agreement between intraosseous (IO) and arterial POC analyses
Tijdsspanne: Samples are analysed immediately after the sample has been taken. Statistical analysis about the agreement performed when the whole data has been collected.
|
agreement analysed with Bland-Altman method from IO and arterial blood samples, which have been analysed with iSTAT point-of care analyser for following parameters: Hb, pH, BE, HCO3, pO2, pCO2, glucose, Na, K, iCa
|
Samples are analysed immediately after the sample has been taken. Statistical analysis about the agreement performed when the whole data has been collected.
|
Medewerkers en onderzoekers
Medewerkers
Onderzoekers
- Hoofdonderzoeker: Jouni Nurmi, M.D. Ph D, University of Helsinki, Helsinki University hospital
Publicaties en nuttige links
Algemene publicaties
- Jousi M, Laukkanen-Nevala P, Nurmi J. Analysing blood from intraosseous access: a systematic review. Eur J Emerg Med. 2019 Apr;26(2):77-85. doi: 10.1097/MEJ.0000000000000569.
- Jousi M, Saikko S, Nurmi J. Intraosseous blood samples for point-of-care analysis: agreement between intraosseous and arterial analyses. Scand J Trauma Resusc Emerg Med. 2017 Sep 11;25(1):92. doi: 10.1186/s13049-017-0435-4.
- Backman S, Angerman-Haasmaa S, Jousi M, Siitonen S, Salmela K. ABO and D typing and alloantibody screening in marrow samples: relevance to intraosseous blood transfusion. Transfusion. 2018 Jun;58(6):1372-1376. doi: 10.1111/trf.14557. Epub 2018 Mar 1.
- Tallman CI, Darracq M, Young M. Analysis of intraosseous blood samples using an EPOC point of care analyzer during resuscitation. Am J Emerg Med. 2017 Mar;35(3):499-501. doi: 10.1016/j.ajem.2016.12.005. Epub 2016 Dec 12.
- Veldhoen ES, de Vooght KM, Slieker MG, Versluys AB, Turner NM. Analysis of bloodgas, electrolytes and glucose from intraosseous samples using an i-STAT((R)) point-of-care analyser. Resuscitation. 2014 Mar;85(3):359-63. doi: 10.1016/j.resuscitation.2013.12.002. Epub 2013 Dec 31.
Studie record data
Bestudeer belangrijke data
Studie start (Werkelijk)
Primaire voltooiing (Werkelijk)
Studie voltooiing (Werkelijk)
Studieregistratiedata
Eerst ingediend
Eerst ingediend dat voldeed aan de QC-criteria
Eerst geplaatst (Werkelijk)
Updates van studierecords
Laatste update geplaatst (Werkelijk)
Laatste update ingediend die voldeed aan QC-criteria
Laatst geverifieerd
Meer informatie
Termen gerelateerd aan deze studie
Aanvullende relevante MeSH-voorwaarden
Andere studie-ID-nummers
- 1234567890
Plan Individuele Deelnemersgegevens (IPD)
Bent u van plan om gegevens van individuele deelnemers (IPD) te delen?
Informatie over medicijnen en apparaten, studiedocumenten
Bestudeert een door de Amerikaanse FDA gereguleerd geneesmiddel
Bestudeert een door de Amerikaanse FDA gereguleerd apparaatproduct
Deze informatie is zonder wijzigingen rechtstreeks van de website clinicaltrials.gov gehaald. Als u verzoeken heeft om uw onderzoeksgegevens te wijzigen, te verwijderen of bij te werken, neem dan contact op met register@clinicaltrials.gov. Zodra er een wijziging wordt doorgevoerd op clinicaltrials.gov, wordt deze ook automatisch bijgewerkt op onze website .